Measuring Device
Measuring Device
Technical Information
Proline Promag L 400
Electromagnetic flowmeter
Table of contents
2 Endress+Hauser
Proline Promag L 400
Supplementary documentation . . . . . . . . . . . . . . . 86
Standard documentation . . . . . . . . . . . . . . . . . . . . . . . . 87
Supplementary device-dependent documentation . . . . . . . 87
Registered trademarks . . . . . . . . . . . . . . . . . . . . . . 88
Endress+Hauser 3
Proline Promag L 400
Document information
Symbols used Electrical symbols
Symbol Meaning
Direct current
Alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Equipotential connection
A connection that has to be connected to the plant grounding system: This may be a
potential equalization line or a star grounding system depending on national or
company codes of practice.
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
A Reference to page
Reference to graphic
Visual inspection
Symbols in graphics
Symbol Meaning
1. , 2. , 3. … Series of steps
A, B, C, ... Views
Hazardous area
-
Safe area (non-hazardous area)
.
Flow direction
4 Endress+Hauser
Proline Promag L 400
Ue
I
v
B
I
L
A0028962
Ue Induced voltage
B Magnetic induction (magnetic field)
L Electrode spacing
I Current
v Flow velocity
In the electromagnetic measuring principle, the flowing medium is the moving conductor. The
voltage induced (Ue) is proportional to the flow velocity (v) and is supplied to the amplifier by means
of two measuring electrodes. The flow volume (Q) is calculated via the pipe cross-section (A). The DC
magnetic field is created through a switched direct current of alternating polarity.
Formulae for calculation
• Induced voltage Ue = B · L · v
• Volume flow Q = A · v
Endress+Hauser 5
Proline Promag L 400
Transmitter
Sensor
A0017041
6 Endress+Hauser
Proline Promag L 400
Equipment architecture
1
2 3 4 5
6
A0021560
Safety IT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional
protection for the device and device data transfer must be implemented by the operators themselves.
Device-specific IT security
The device offers a range of specific functions to support protective measures on the operator's side.
These functions can be configured by the user and guarantee greater in-operation safety if used
correctly. An overview of the most important functions is provided in the following section.
Endress+Hauser 7
Proline Promag L 400
WLAN passphrase
A connection between an operating unit (e.g. notebook or tablet) and the device via the WLAN
interface which can be ordered as an option is protected by the network key. The WLAN
authentication of the network key complies with the IEEE 802.11 standard.
When the device is delivered, the network key is pre-defined depending on the device. It can be
changed via the WLAN settings submenu in the WLAN passphrase parameter.
Input
Measured variable Direct measured variables
• Volume flow (proportional to induced voltage)
• Electrical conductivity
Measuring range Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy
Electrical conductivity: ≥ 5 μS/cm for liquids in general
8 Endress+Hauser
Proline Promag L 400
Endress+Hauser 9
Proline Promag L 400
– 65 16 to 500 130 1 2
10 Endress+Hauser
Proline Promag L 400
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Digital communication
The measured values can be written from the automation system to the measuring via:
• PROFIBUS DP
• Modbus RS485
• EtherNet/IP
Status input
Output
Output signal Current output
Load 0 to 700 Ω
Resolution 0.5 µA
Endress+Hauser 11
Proline Promag L 400
Pulse/frequency/switch output
Function • With the order code for "Output; Input", option H: output 2 can be set as a pulse
or frequency output
• With the order code for "Output; Input", option I: output 2 and 3 can be set as a
pulse, frequency or switch output
Pulse output
Frequency output
Switch output
PROFIBUS DP
12 Endress+Hauser
Proline Promag L 400
Modbus RS485
Terminating resistor Integrated, can be activated via DIP switch on the transmitter electronics module
EtherNet/IP
Current output 4 to 20 mA
4 to 20 mA
Failure mode Choose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
0 to 20 mA
Failure mode Choose from:
• Maximum alarm: 22 mA
• Freely definable value between: 0 to 22.5 mA
Device diagnostics Device condition can be read out via HART Command 48
Pulse/frequency/switch output
Pulse output
Frequency output
Switch output
PROFIBUS DP
Endress+Hauser 13
Proline Promag L 400
Modbus RS485
EtherNet/IP
Local display
Interface/protocol
• Via digital communication:
– HART protocol
– PROFIBUS DP
– Modbus RS485
– EtherNet/IP
• Via service interface
Web server
Low flow cut off The switch points for low flow cut off are user-selectable.
Galvanic isolation The following connections are galvanically isolated from each other:
• Inputs
• Outputs
• Power supply
Manufacturer ID 0x11
14 Endress+Hauser
Proline Promag L 400
PROFIBUS DP
Manufacturer ID 0x11
Endress+Hauser 15
Proline Promag L 400
Modbus RS485
Data access Each device parameter can be accessed via Modbus RS485.
16 Endress+Hauser
Proline Promag L 400
EtherNet/IP
Manufacturer ID 0x49E
Baud rates Automatic ¹⁰⁄₁₀₀ Mbit with half-duplex and full-duplex detection
Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs
Configuration options for • DIP switches on the electronics module for IP addressing
measuring device • Manufacturer-specific software (FieldCare)
• Custom Add-on Profile for Rockwell Automation control systems
• Web browser
• Electronic Data Sheet (EDS) integrated in the measuring device
Configuration of the EtherNet • Speed: 10 MBit, 100 MBit, auto (factory setting)
interface • Duplex: half-duplex, full-duplex, auto (factory setting)
Configuration of the device • DIP switches on the electronics module for IP addressing (last octet)
address • DHCP
• Manufacturer-specific software (FieldCare)
• Custom Add-on Profile for Rockwell Automation control systems
• Web browser
• EtherNet/IP tools, e.g. RSLinx (Rockwell Automation)
Fix Input
O → T configuration: 0x66 56
T → O configuration: 0x64 32
O → T configuration: 0x66 56
T → O configuration: 0x64 32
O → T configuration: 0xC7 -
T → O configuration: 0x64 32
O → T configuration: 0xC7 -
T → O configuration: 0x64 32
Endress+Hauser 17
Proline Promag L 400
Configurable Input
O → T configuration: 0x66 56
T → O configuration: 0x65 88
O → T configuration: 0x66 56
T → O configuration: 0x65 88
O → T configuration: 0xC7 -
T → O configuration: 0x65 88
O → T configuration: 0xC7 -
T → O configuration: 0x65 88
Fix Output
18 Endress+Hauser
Proline Promag L 400
Configuration
Configuration Assembly Only the most common configurations are listed below.
• Software write protection
• Mass flow unit
• Mass unit
• Volume flow unit
• Volume unit
• Density unit
• Conductivity
• Temperature unit
• Totalizer 1-3:
– Assignment
– Unit
– Operating mode
– Failure mode
• Alarm delay
Power supply
Terminal assignment Transmitter: 0-20 mA/4-20 mA HART
The sensor can be ordered with terminals.
Supply voltage
Order code for "Power supply" Terminal numbers
1 (L+/L) 2 (L-/N)
Signal transmission 0-20 mA/4-20 mA HART and additional outputs and inputs
Order code for Terminal numbers
"Output" and
Output 1 Output 2 Output 3 Input
"Input"
26 (+) 27 (-) 24 (+) 25 (-) 22 (+) 23 (-) 20 (+) 21 (-)
Endress+Hauser 19
Proline Promag L 400
Transmitter: PROFIBUS DP
The sensor can be ordered with terminals.
Supply voltage
Order code for "Power supply" Terminal numbers
1 (L+/L) 2 (L-/N)
26 (RxD/TxD-P) 27 (RxD/TxD-N)
Option L B A
Supply voltage
Order code for "Power supply" Terminal numbers
1 (L+/L) 2 (L-/N)
26 (+) 27 (-)
Option M B A
20 Endress+Hauser
Proline Promag L 400
Transmitter: EtherNet/IP
The sensor can be ordered with terminals or a device plug.
Supply voltage
Order code for "Power supply" Terminal numbers
1 (L+/L) 2 (L-/N)
Remote version
1 2
GND
E1
E2
S1
S2
A
E
S
6 5 7 8 4 37 36 42 41
1
2 n.c. n.c. n.c.
5 7 4 37 42 41
B
E2
E1
GND
E
A0032059
Terminal No. and cable colors: 6/5 = brown; 7/8 = white; 4 = green; 36/37 = yellow
Pin assignment, device plug Order codes for the M12x1 connectors, see the "Order code for electrical connection" column:
EtherNet/IP → 21
Endress+Hauser 21
Proline Promag L 400
EtherNet/IP
1 + Tx D Socket
2 + Rx
1 3
3 - Tx
4 - Rx
4
A0032047
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
22 Endress+Hauser
Proline Promag L 400
A B
1 2 3 1 2 3
A0032041
Connecting cable
A B
1 2 1 2
A0032042
Endress+Hauser 23
Proline Promag L 400
Connection examples
1 2 3
4...20 mA
4 5
A0029055
1 2 3 4
4...20 mA
A0028762
0/4...20 mA
+
+
– 3
_
2
A0017162
7 Connection example for 0-20 mA current output (active) and 4-20 mA current output (active)
1 Automation system with current input (e.g. PLC)
2 Analog display unit: observe maximum load
3 Transmitter
24 Endress+Hauser
Proline Promag L 400
Pulse/frequency output
1 2
12345
A0028761
Switch output
1 2
A0028760
Endress+Hauser 25
Proline Promag L 400
PROFIBUS DP
1 2 3
A A
4
B
B
A
4
B
A0028765
10 Connection example for PROFIBUS DP, non-hazardous area and Zone 2/Div. 2
1 Control system (e.g. PLC)
2 Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3 Transmitter
If baud rates > 1.5 MBaud an EMC cable entry must be used and the cable shield must continue
as far as the terminal wherever possible.
Modbus RS485
1 2 3
A A
4
B
B
A
4
B
A0028765
11 Connection example for Modbus RS485, non-hazardous area and Zone 2/Div. 2
1 Control system (e.g. PLC)
2 Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3 Distribution box
4 Transmitter
26 Endress+Hauser
Proline Promag L 400
EtherNet/IP
1 2 3 4
A0028767
Status input
1 2
A0028764
A0016315
Endress+Hauser 27
Proline Promag L 400
DN ≤ 300 DN ≥ 350
A0029338
A0029339
28 Endress+Hauser
Proline Promag L 400
3
+
2 –
1
A0030377
Terminals Transmitter
• Supply voltage cable: plug-in spring terminals for wire cross-sections
0.5 to 2.5 mm2 (20 to 14 AWG)
• Signal cable: plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Electrode cable: spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Coil current cable: spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
Sensor connection housing
Spring terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Signal cable
Endress+Hauser 29
Proline Promag L 400
Pulse/frequency/switch output
Standard installation cable is sufficient.
Status input
Standard installation cable is sufficient.
PROFIBUS DP
The IEC 61158 standard specifies two types of cable (A and B) for the bus line which can be used for
every transmission rate. Cable type A is recommended.
Cable type A
Signal damping Max. 9 dB over the entire length of the cable cross-section
Shield Copper braided shielding or braided shielding with foil shield. When grounding
the cable shield, observe the grounding concept of the plant.
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used
for every transmission rate. Cable type A is recommended.
Cable type A
Signal damping Max. 9 dB over the entire length of the cable cross-section
Shield Copper braided shielding or braided shielding with foil shield. When grounding
the cable shield, observe the grounding concept of the plant.
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for a cable
used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
Electrode cable
Standard cable 3 ×0.38 mm2 (20 AWG) with common, braided copper shield ( ~9.5 mm
(0.37 in)) and individual shielded cores
Cable for empty pipe 4 ×0.38 mm2 (20 AWG) with common, braided copper shield ( ~9.5 mm
detection (EPD) (0.37 in)) and individual shielded cores
30 Endress+Hauser
Proline Promag L 400
Standard cable 3 ×0.75 mm2 (18 AWG) with common, braided copper shield ( ~9 mm
(0.35 in))
1
2
3
4
5
6
7
a b
A0029151
17 Cable cross-section
a Electrode cable
b Coil current cable
1 Core
2 Core insulation
3 Core shield
4 Core jacket
5 Core reinforcement
6 Cable shield
7 Outer jacket
Endress+Hauser 31
Proline Promag L 400
Performance characteristics
Reference operating • Error limits following DIN EN 29104, in future ISO 20456
conditions • Water, typically +15 to +45 °C (+59 to +113 °F); 0.5 to 7 bar (73 to 101 psi)
• Data as indicated in the calibration protocol
• Accuracy based on accredited calibration rigs according to ISO 17025
[%]
2.5
2.0
0.5 %
1.5
0.2 %
1.0
0.5
0
0 1 2 4 6 8 10 [m/s]
v
0 5 10 15 20 25 30 32 [ft/s]
A0032069
Electrical conductivity
Max. measured error not specified.
Accuracy of outputs
The outputs have the following base accuracy specifications.
Current output
Accuracy Max. ±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (across the entire ambient temperature range)
32 Endress+Hauser
Proline Promag L 400
Pulse/frequency output
Installation
No special measures such as supports etc. are necessary. External forces are absorbed by the
construction of the device.
Mounting location
A0029343
Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe
elbow: h ≥ 2 × DN
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
1
2
h
A0028981
Endress+Hauser 33
Proline Promag L 400
N
xD
³2
N
xD
³5
A0029257
A0016276
Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the
flow direction (direction of medium flow through the piping).
An optimum orientation position helps avoid gas and air accumulations and deposits in the
measuring tube.
The measuring device also offers the empty pipe detection function to detect partially filled
measuring pipes in the event of outgassing fluids or variable process pressures.
Vertical
A0015591
Optimum for self-emptying pipe systems and for use in conjunction with empty pipe detection.
34 Endress+Hauser
Proline Promag L 400
Horizontal
2 2
A0029344
• Ideally, the measuring electrode plane should be horizontal. This prevents brief insulation of
the two measuring electrodes by entrained air bubbles.
• Empty pipe detection only works if the transmitter housing is pointing upwards as otherwise
there is no guarantee that the empty pipe detection function will actually respond to a
partially filled or empty measuring tube.
Inlet and outlet runs If possible, install the sensor upstream from fittings such as valves, T-pieces or elbows.
Observe the following inlet and outlet runs to comply with accuracy specifications:
≥ 5 x DN ≥ 2 x DN
A0028997
Adapters Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-
diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very
slow-moving fluids.
The nomogram shown here can be used to calculate the pressure loss caused by reducers and
expanders:
• Calculate the ratio of the diameters d/D.
• From the nomogram read off the pressure loss as a function of flow velocity (downstream from
the reduction) and the d/D ratio.
The nomogram only applies to liquids with a viscosity similar to that of water.
Endress+Hauser 35
Proline Promag L 400
100 [mbar]
8 m/s
7 m/s
6 m/s
5 m/s
4 m/s
3 m/s 10
max. 8° 2 m/s
d D
1 m/s 1
Length of connecting cable To ensure correct measuring results when using the remote version,
observe the maximum permitted length of the connecting cable Lmax. This length is determined by
the conductivity of the fluid.
If measuring liquids in general: 5 µS/cm
[µS/cm]
200
100
L max
5
[m]
10 100 200 L max
[ft]
0 200 400 600
A0016539
36 Endress+Hauser
Proline Promag L 400
14 (0.55)
5.8 (0.23)
210.5 (8.29)
5.8 (0.23)
149 (5.85)
A0020523
Post mounting
ø 20…70 3x TX 25
(ø 0.79…2.75)
4x SW 10
.0)
~4
0 2(
~1
A0029051
Endress+Hauser 37
Proline Promag L 400
≤ 3 (≤ 10)
A0029320
Environment
Ambient temperature range Transmitter –40 to +60 °C (–40 to +140 °F)
Local display –20 to +60 °C (–4 to +140 °F), the readability of the display may be
impaired at temperatures outside the temperature range.
Liner Do not exceed or fall below the permitted temperature range of the liner .
If operating outdoors:
• Install the measuring device in a shady location.
• Avoid direct sunlight, particularly in warm climatic regions.
• Avoid direct exposure to weather conditions.
• If the compact version of the device is insulated at low temperatures, the insulation must also
include the device neck.
• Protect the display against impact.
• Protect the display from abrasion by sand in desert areas.
You can order a display guard from Endress+Hauser : → 84
Temperature tables
Observe the interdependencies between the permitted ambient and fluid temperatures when
operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document entitled "Safety
Instructions" (XA) for the device.
Storage temperature The storage temperature corresponds to the operating temperature range of the measuring
transmitter and the appropriate measuring sensors.→ 38
• Protect the measuring device against direct sunlight during storage in order to avoid unacceptably
high surface temperatures.
• Select a storage location where moisture cannot collect in the measuring device as fungus or
bacteria infestation can damage the liner.
• If protection caps or protective covers are mounted these should never be removed before
installing the measuring device.
38 Endress+Hauser
Proline Promag L 400
Atmosphere If a plastic transmitter housing is permanently exposed to certain steam and air mixtures, this can
damage the housing.
If you are unsure, please contact your Endress+Hauser Sales Center for clarification.
Mechanical load • Protect the transmitter housing against mechanical effects, such as shock or impact; the use of the
remote version is sometimes preferable.
• Never use the transmitter housing as a ladder or climbing aid.
Process
Medium temperature range • 0 to +80 °C (+32 to +176 °F) for hard rubber, DN 350 to 2400 (14 to 90")
• –20 to +50 °C (–4 to +122 °F) for polyurethane, DN 25 to 1200 (1 to 48")
• –20 to +90 °C (–4 to +194 °F) for PTFE, DN 25 to 300 (1 to 12")
Endress+Hauser 39
Proline Promag L 400
Conductivity ≥ 5 μS/cm for liquids in general. Stronger filter damping is required for very low conductivity values.
Note that in the case of the remote version, the requisite minimum conductivity also depends
on the cable length → 36.
Pressure-temperature The following pressure/temperature diagrams apply to all pressure-bearing parts of the device and
ratings not just the process connection.
Process connection: lap joint flange/lap joint flange, stamped plate according to EN 1092-1
(DIN 2501) and ASME B16.5; DN 25 to 300 (1 to 12")
[psi] [bar] 1 2
300 20
200 15
10
100
5
0 0
-60 -40 -20 0 20 40 60 80 100120140 [°C]
24 Process connection material: stainless steel (min. -40 °C (-40 °F)), carbon steel (min. -10 °C (+14 °F))
1 Lap joint flange PN16/ Class150
2 Lap joint flange, stamped plate PN10, lap joint flange PN10
[psi] [bar]
300 20
200 15
PN 10
10
100
5 PN 6
0 0
-40 -20 0 20 40 60 80 100120140 [°C]
[psi] [bar]
40
500
400 30
300 20
200 Class 150
10
100
0 0
-40 -20 0 20 40 60 80 100120140 [°C]
26 Process connection material: stainless steel, F316L similar to 1.4404; carbon steel, A105/A515(70)
40 Endress+Hauser
Proline Promag L 400
[psi] [bar]
160 11
145 10 Class D
130 9
116 8
0 0
-40 -20 0 20 40 60 80 100120140 [°C]
A0032072-EN
[psi] [bar]
300 20
PN 16
200 15
10
100
5
0 0
-40 -20 0 20 40 60 80 100120140 [°C]
[mm] [in] +25 °C (+77 °F) +50 °C (+122 °F) +80 °C (+176 °F)
Endress+Hauser 41
Proline Promag L 400
Liner: PTFE
Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
25 1 0 (0) 0 (0)
40 2 0 (0) 0 (0)
50 2 0 (0) 0 (0)
65 2½ 0 (0) 40 (0.58)
80 3 0 (0) 40 (0.58)
Flow limit The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum velocity of flow is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow
(v) to the physical properties of the fluid:
• v < 2 m/s (6.56 ft/s): for abrasive fluids (e.g. potter's clay, lime milk, ore slurry)
• v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. wastewater sludge)
A necessary increase in the flow velocity can be achieved by reducing the sensor nominal
diameter.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section → 8
Pressure loss • No pressure loss occurs if the sensor is installed in a pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545 → 35
System pressure
A0028777
Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and thus
damage to the liner.
Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.
Vibrations In the event of very strong vibrations, the pipe and sensor must be supported and fixed.
It is also advisable to mount the sensor and transmitter separately.
• For information on the shock resistance of the measuring system
• For information on the vibration resistance of the measuring system
42 Endress+Hauser
Proline Promag L 400
L
A0029004
Mechanical construction
Dimensions in SI units Compact version
Order code for "Housing", option M "Compact, polycarbonate" or option A "Compact, aluminum, coated"
A D
B C
E
G
F
J
H K
A0020352
DN A B C D E F G H J K 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
125 193 103 90 167 287 150 437 260 140 250
150 193 103 90 167 287 150 437 260 140 300
200 193 103 90 167 312 180 492 324 156 350
250 193 103 90 167 337 205 542 400 166 450
300 193 103 90 167 362 230 592 460 166 500
1) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
Endress+Hauser 43
Proline Promag L 400
A D
B C
E
F
H
G J
A0017153
DN A B C D E H J
44 Endress+Hauser
Proline Promag L 400
Dimension F Dimension G
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
350 631 638 702 653 648 490 505 520 533 525
400 682 694 760 710 702 540 565 580 597 580
450 737 747 823 757 760 595 615 640 635 640
500 787 800 926 814 817 645 670 715 699 705
600 883 896 1026 912 918 755 780 840 813 825
700 1001 1018 1145 1034 1026 860 895 910 927 910
800 1115 1135 1240 1157 1157 975 1015 1025 1060 1060
900 1215 1235 1240 1261 1265 1075 1115 1125 1168 1175
1000 1315 1342 1355 1372 1355 1175 1230 1255 1289 1255
1200 1544 1569 1584 1597 1586 1405 1455 1485 1511 1490
Order code for "Housing", option Q "Compact, polycarbonate, inclined" or option R "Compact, aluminum,
coated, inclined"
A D
B C
22.5°
E
G
F
J
H K
A0020353
Endress+Hauser 45
Proline Promag L 400
DN A B C D E F G H J K 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
125 199 119 80 167 332 150 482 260 140 250
150 199 119 80 167 332 150 482 260 140 300
200 199 119 80 167 357 180 537 324 156 350
250 199 119 80 167 382 205 587 400 166 450
300 199 119 80 167 407 230 637 460 166 500
1) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
A D
B C
22.5°
E
F
G J
A0020393
DN A B C D E H J
46 Endress+Hauser
Proline Promag L 400
DN A B C D E H J
Dimension F Dimension G
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
350 676 683 618 698 693 490 505 520 533 525
400 727 739 672 755 747 540 565 580 597 580
450 782 792 732 802 805 595 615 640 635 640
500 832 845 795 859 862 645 670 715 699 705
600 928 941 898 957 963 755 780 840 813 825
700 1046 1063 1008 1079 1071 860 895 910 927 910
800 1160 1180 1112 1202 1202 975 1015 1025 1060 1060
900 1260 1280 1212 1306 1310 1075 1115 1125 1168 1175
1000 1360 1387 1327 1417 1400 1175 1230 1225 1289 1255
1200 1589 1614 1556 1642 1631 1405 1455 1255 1511 1490
Endress+Hauser 47
Proline Promag L 400
Dimension F Dimension G
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
Order code for "Housing", option N "Remote, polycarbonate" or option P "Remote, aluminum coated"
A F
D
C
E
B
A0020522
A B C D E F
[mm] [mm] [mm] [mm] [mm] [mm]
A
B
D
C
F
E G
A0017282
48 Endress+Hauser
Proline Promag L 400
DN A B C D E F G 1)
1) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
B
C
E
D F
A0017284
DN A B E F
Endress+Hauser 49
Proline Promag L 400
DN A B E F
Dimension C Dimension D
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
350 603 610 616 625 620 490 505 520 533 525
400 654 666 672 682 674 540 565 580 597 580
450 709 719 729 729 732 595 615 640 635 640
500 759 772 791 786 789 645 670 715 699 705
600 855 868 903 884 890 755 780 840 813 825
700 973 990 1009 1006 998 860 895 910 927 910
800 1087 1107 1123 1129 1129 975 1015 1025 1060 1060
900 1187 1207 1223 1233 1237 1075 1115 1125 1168 1175
1000 1287 1314 1338 1344 1327 1175 1230 1225 1289 1255
1200 1516 1541 1567 1569 1558 1405 1455 1255 1511 1490
50 Endress+Hauser
Proline Promag L 400
Dimension C Dimension D
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
Order code for "Sensor option", option CK "IP68, Type 6P, waterproof"
B
C
E
D F
A0020436
DN A B C D E F
Endress+Hauser 51
Proline Promag L 400
DN A B C D E F
Accessories
H
H
H
!B A
!B ! !D !A !D
!A !D
DN Pressure rating A B D H
52 Endress+Hauser
Proline Promag L 400
DN Pressure rating A B D H
1) Ground disks can be used for all the flange standards/pressure ratings which can be supplied in the
standard version.
Endress+Hauser 53
Proline Promag L 400
Order code for "Housing", option M "Compact, polycarbonate" or option A "Compact, aluminum, coated"
A D
B C
E
G
F
J
H K
A0020352
DN A B C D E F G H J K 1)
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
1 7.60 4.06 3.54 6.57 8.74 3.31 12.1 4.72 3.70 7.87
1½ 7.60 4.06 3.54 6.57 8.74 3.31 12.1 4.72 3.70 7.87
2 7.60 4.06 3.54 6.57 8.74 3.31 12.1 4.72 3.70 7.87
3 7.60 4.06 3.54 6.57 9.72 4.29 14.0 7.09 3.70 7.87
4 7.60 4.06 3.54 6.57 9.72 4.29 14.0 7.09 3.70 9.84
6 7.60 4.06 3.54 6.57 11.3 5.91 17.2 10.2 5.51 11.8
8 7.60 4.06 3.54 6.57 12.3 7.09 19.4 12.8 6.14 13.8
10 7.60 4.06 3.54 6.57 13.3 8.07 21.3 15.8 6.54 17.7
12 7.60 4.06 3.54 6.57 14.3 9.06 23.3 18.1 6.54 19.7
1) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
A D
B C
E
F
G J
A0017153
54 Endress+Hauser
Proline Promag L 400
DN A B C D E H J
Dimension F Dimension G
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
14 24.8 25.1 27.6 25.7 25.5 19.3 19.9 20.5 21.0 20.7
15 – – – – 27 – – – – 21.7
16 26.8 27.3 30.0 27.0 27.6 21.3 22.2 22.8 23.5 22.8
18 29.0 29.4 32.4 29.8 29.9 23.4 24.2 25.2 25.0 25.2
20 31.0 31.5 36.5 32.0 32.1 25.4 26.4 28.1 27.5 27.8
24 34.7 35.3 40.4 35.9 36.1 29.7 30.7 33.1 32.0 32.5
28 39.4 40.1 45.1 40.7 40.4 33.9 35.2 35.8 36.5 35.8
32 43.9 44.7 48.8 45.5 45.5 38.4 40.0 40.4 41.7 41.7
36 47.8 48.6 48.8 49.6 49.8 42.3 43.9 44.3 46.0 46.3
40 51.7 52.8 53.4 54.0 53.3 46.3 48.4 49.4 50.7 49.4
42 – – – 56.5 – – – – 53.0 –
48 60.8 61.7 62.4 62.9 62.4 55.3 57.3 58.5 59.5 58.7
54 – – – 70.6 – – – – 66.3 –
60 – – – 77.9 – – – – 73.0 –
66 – – – 83.4 – – – – 80.0 –
Endress+Hauser 55
Proline Promag L 400
Dimension F Dimension G
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
84 – – – 104.0 – – – – 99.8 –
90 – – – 111.0 – – – – – –
Order code for "Housing", option Q "Compact, polycarbonate, inclined" or option R "Compact, aluminum,
coated, inclined"
A D
B C
22.5°
E
G
F
J
H K
A0020353
DN A B C D E F G H J K 1)
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
1 7.83 4.69 3.15 6.57 10.5 3.31 13.8 4.72 3.70 7.87
1½ 7.83 4.69 3.15 6.57 10.5 3.31 13.8 4.72 3.70 7.87
2 7.83 4.69 3.15 6.57 10.5 3.31 13.8 4.72 3.70 7.87
3 7.83 4.69 3.15 6.57 11.5 4.29 15.8 7.09 3.70 7.87
4 7.83 4.69 3.15 6.57 11.5 4.29 15.8 7.09 3.70 9.84
6 7.83 4.69 3.15 6.57 13.1 5.91 19.0 10.2 5.51 11.8
8 7.83 4.69 3.15 6.57 14.1 7.09 21.1 12.8 6.14 13.8
10 7.83 4.69 3.15 6.57 15.0 8.07 23.1 15.8 6.54 17.7
12 7.83 4.69 3.15 6.57 16.0 9.06 25.1 18.1 6.54 19.7
1) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
56 Endress+Hauser
Proline Promag L 400
A D
B C
22.5°
E
F
H
G J
A0020393
DN A B C D E H J
Endress+Hauser 57
Proline Promag L 400
Dimension F Dimension G
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
14 26.6 26.9 24.3 27.5 27.3 19.3 19.9 20.5 21.0 20.7
15 – – – – 28.8 – – – – 21.7
16 28.6 29.1 26.5 28.8 29.4 21.3 22.2 22.8 23.5 22.8
18 30.8 31.2 28.8 31.6 31.7 23.4 24.2 25.2 25.0 25.2
20 32.8 33.3 31.3 33.8 33.9 25.4 26.4 28.1 27.5 27.8
24 36.5 37.1 35.4 37.7 37.9 29.7 30.7 33.1 32.0 32.5
28 41.2 41.9 39.7 42.5 42.2 33.9 35.2 35.8 36.5 35.8
32 45.7 46.5 43.8 47.3 47.3 38.4 40.0 40.4 41.7 41.7
36 49.6 50.4 47.7 51.4 49.8 42.3 43.9 44.3 46.0 46.3
40 53.5 54.6 52.2 55.8 55.1 46.3 48.4 48.2 50.7 49.4
42 – – – 58.3 – – – – 53.0 –
48 62.6 63.5 61.3 64.7 64.2 55.3 57.3 49.4 59.5 58.7
54 – – – 72.4 – – – – 66.3 –
60 – – – 79.7 – – – – 73.0 –
66 – – – 85.2 – – – – 80.0 –
84 – – – 105.8 – – – – 99.8 –
90 – – – 112.8 – – – – – –
Order code for "Housing", option N "Remote, polycarbonate" or option P "Remote, aluminum coated"
A F
D
C
E
B
A0020522
A B C D E F
[in] [in] [in] [in] [in] [in]
58 Endress+Hauser
Proline Promag L 400
B
D
C
F
E G
A0017282
DN A B C D E F G 1)
1) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
A
B
C
D F
A0017284
Endress+Hauser 59
Proline Promag L 400
DN A B E F
Dimension C Dimension D
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
14 23.7 24.0 24.3 24.6 24.4 19.3 19.9 20.5 21.0 20.7
15 – – – – 25.9 – – – – 21.7
16 25.8 26.2 26.5 26.9 26.5 21.3 22.2 22.8 23.5 22.8
18 27.9 28.3 28.7 28.7 28.8 23.4 24.2 25.2 25.0 25.2
20 29.9 30.4 31.1 30.9 31.1 25.4 26.4 28.1 27.5 27.8
24 33.7 34.2 35.6 34.8 35.0 29.7 30.7 33.1 32.0 32.5
28 38.5 39.2 39.7 39.8 39.5 33.9 35.2 35.8 36.5 35.8
32 43.0 43.8 44.2 44.6 44.6 38.4 40.0 40.4 41.7 41.7
36 46.9 47.7 48.2 48.7 48.9 42.3 43.9 44.3 46.0 46.3
40 50.8 51.9 52.7 53.1 52.4 46.3 48.4 48.2 50.7 49.4
42 – – – 55.6 – – – – 53.0 –
48 59.9 60.8 61.7 62.0 61.5 55.3 57.3 49.4 59.5 58.7
54 – – – 69.6 – – – – 66.3 –
60 – – – 76.9 – – – – 73.0 –
66 – – – 82.4 – – – – 80.0 –
60 Endress+Hauser
Proline Promag L 400
Dimension C Dimension D
DN PN 6 PN 10 PN 16 AWWA PN 6 PN 10 PN 16 AWWA
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
84 – – – 102.8 – – – – 99.8 –
90 – – – 110.3 – – – – 106.5 –
Order code for "Sensor option", option CK "IP68, Type 6P, waterproof"
B
C
E
D F
A0020436
DN A B C D E F
Endress+Hauser 61
Proline Promag L 400
DN A B C D E F
Accessories
H
H
H
!B A
!B ! !D !A !D
!A !D
DN Pressure rating A B D H
62 Endress+Hauser
Proline Promag L 400
DN Pressure rating A B D H
1) Ground disks can be used for all the flange standards/pressure ratings which can be supplied in the
standard version.
Endress+Hauser 63
Proline Promag L 400
Weight in SI units
Weight [kg]
PN 6 PN 10 PN 16
25 – – 6.8
32 – – 7.5
40 – – 8.5
50 – – 9
65 – – 10
80 – – 12
100 – – 14
125 – – 20
150 – – 24
200 – 43 44.4
250 – 63 70.2
300 – 68 85.3
350 77 88 103
AS 2129, Table E
DN Weight [kg]
[mm]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
350 99 99.7
64 Endress+Hauser
Proline Promag L 400
AS 2129, Table E
DN Weight [kg]
[mm]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
AS 4087, PN 16
DN Weight [kg]
[mm]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
350 99 99.7
DN Weight [kg]
[mm]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
25 5.3 6.0
32 5.1 5.8
40 5.8 6.5
50 5 5.7
65 6 6.7
80 7 7.7
100 9 9.7
125 13 13.7
Endress+Hauser 65
Proline Promag L 400
DN Weight [kg]
[mm]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
150 17 17.7
200 35 35.7
250 54 54.7
300 55 55.7
Weight in US units
DN Weight [lbs]
[in]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
1 11.6 13.2
1½ 12.8 14.3
2 20 21.5
3 26 27.5
4 31 32.5
6 53 54.5
8 95 96.5
10 139 140.5
12 150 151.5
14 302 303.5
16 370 371.5
18 421 422.5
20 503 504.5
24 666 667.5
DN Weight [lbs]
[in]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
28 586 587.5
30 701 702.5
32 844 845.5
36 1 036 1 037.5
40 1 294 1 295.5
42 1 477 1 478.5
48 1 987 1 988.5
54 2 807 2 808.5
60 3 515 3 516.5
66 4 699 4 700.5
66 Endress+Hauser
Proline Promag L 400
DN Weight [lbs]
[in]
Order code for "Housing", option M, Q Order code for "Housing", option A, R
Polycarbonate plastic Aluminum, AlSi10Mg, coated
72 5 662 5 663.5
78 6 864 6 865.5
84 8 280 8 281.5
90 10 577 10 578.5
Wall-mount housing
Depends on the material of the wall-mount housing:
• Polycarbonate plastic: 1.3 kg (2.9 lb)
• Aluminum, AlSi10Mg, coated: 2.0 kg (4.4 lb)
Weight in SI units
DN Weight [kg]
[mm]
PN 6 PN 10 PN 16
25 – – 6.8
32 – – 7.5
40 – – 8.5
50 – – 6
65 – – 7
80 – – 9
100 – – 11
125 – – 16
150 – – 20
200 – 40 44.4
250 – 60 70.2
300 – 65 85.3
350 73 84 101
Endress+Hauser 67
Proline Promag L 400
DN Weight [kg]
[mm]
PN 6 PN 10 PN 16
AS 2129, Table E
DN Weight
[mm] [kg]
350 95
400 116
450 139
500 178
600 256
700 343
750 430
800 490
900 687
1 000 758
1 200 1 234
AS 4087, PN 16
DN Weight
[mm] [kg]
350 95
375 101
400 116
450 129
500 178
600 256
700 364
750 442
800 500
900 699
1 000 756
1 200 1 216
68 Endress+Hauser
Proline Promag L 400
DN [kg]
[mm]
25 6.0
32 5.8
40 6.5
50 3
65 4
80 5
100 7
125 11
150 15
200 33
250 52
300 53
Weight in US units
DN Weight
[in] [lbs]
1 13.2
1½ 14.3
2 13
3 20
4 24
6 44
8 88
10 132
12 143
14 293
15 –
16 361
18 412
20 494
24 657
DN Weight
[in] [lbs]
28 580
30 695
Endress+Hauser 69
Proline Promag L 400
DN Weight
[in] [lbs]
32 838
36 1 030
40 1 288
42 1 471
48 1 980
54 2 800
60 3 508
66 4 692
72 5 656
78 6 858
84 8 273
90 10 571
Measuring tube specification Nominal diameter Pressure rating Measuring tube internal diameter
AWWA AS 4087
70 Endress+Hauser
Proline Promag L 400
AWWA AS 4087
1 050 42 PN 6 – – – – – – – –
1 050 42 PN 10 – – – – – – – –
1 050 42 – – Table E, PN 16 – – – – – –
Endress+Hauser 71
Proline Promag L 400
AWWA AS 4087
72 Endress+Hauser
Proline Promag L 400
3
A0020640
Adapter for cable entry with internal thread G ½" or Nickel-plated brass
NPT ½"
Device plug
Sensor housing
• DN 25 to 300 (1 to 12"): aluminum, AlSi10Mg, coated
• DN 350 to 2400 (14 to 90"): carbon steel with protective varnish
Measuring tubes
• DN 25 to 300 (1 to 12"): stainless steel, 1.4301/1.4306/304L
• DN 350 to 1200 (14 to 48"): stainless steel, 1.4301/1.4307/202/304
• DN 1350 to 2400 (54 to 90"): stainless steel, 1.4301/1.4307
Endress+Hauser 73
Proline Promag L 400
Liner
• DN 25 to 300 (1 to 12"): PTFE
• DN 25 to 1200 (1 to 48"): polyurethane
• DN 350 to 2400 (14 to 90"): hard rubber
Electrodes
• Stainless steel, 1.4435 (316L)
• Alloy C22, 2.4602 (UNS N06022)
Process connections
ASME B16.5
DN 25 to 300 (1 to 12"):
Fixed flange:
– Stainless steel, F316L similar to 1.4404
– Carbon steel, A105 similar to 1.0432
DN 350 to 600 (14 to 24"):
Carbon steel, A105/A515 Grade 70
AWWA C207
• DN 48":
Carbon steel, A105/A181/FE410WB/P265GH/S275JR
• DN 54 to 90":
Carbon steel, A105/A181/P265GH/S275JR
AS 2129
Carbon steel, A105/FE410WB/P235GH/P265GH/S235JRG2
AS 4087
Carbon steel, A105/P265GH/S275JRG2
Seals
In accordance with DIN EN 1514-1
Accessories
Display protection
Stainless steel, 1.4301 (304L)
Ground disks
• Stainless steel, 1.4435 (316L)
• Alloy C22, 2.4602 (UNS N06022)
74 Endress+Hauser
Proline Promag L 400
Fitted electrodes Measurement, reference and empty pipe detection electrodes available as standard with:
• 1.4435 (316L)
• Alloy C22, 2.4602 (UNS N06022)
Surface roughness Electrodes with 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022):
≤ 0.3 to 0.5 µm (11.8 to 19.7 µin)
(All data relate to parts in contact with fluid)
Operability
Operating concept Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Fast and safe commissioning
• Guided menus ("Make-it-run" wizards) for applications
• Menu guidance with brief explanations of the individual parameter functions
• Device access via Web server
• Optional: WLAN access to device via mobile handheld terminal
Reliable operation
• Operation in local language
• Uniform operating philosophy applied to device and operating tools
• If replacing electronic modules, transfer the device configuration via the integrated memory
(integrated HistoROM) which contains the process and measuring device data and the event
logbook. No need to reconfigure.
Efficient diagnostics increase measurement availability
• Troubleshooting measures can be called up via the device and in the operating tools
• Diverse simulation options, logbook for events that occur and optional line recorder functions
Endress+Hauser 75
Proline Promag L 400
A0032074
Display elements
• 4-line, illuminated, graphic display
• White background lighting; switches to red in event of device errors
• Format for displaying measured variables and status variables can be individually configured
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the temperature range.
Operating elements
• External operation via touch control (3 optical keys) without opening the housing: , ,
• Operating elements also accessible in various hazardous areas
4 6
1 2 3 5
A0028747
76 Endress+Hauser
Proline Promag L 400
1 2
4 4 4
A0020903
1 2
A0029437
Endress+Hauser 77
Proline Promag L 400
1 2 3
5 5 5
A0032078
1
open
press 3
A0029163
78 Endress+Hauser
Proline Promag L 400
1 2 3
4 5
A0032079
Configurable channels 1 to 11
Supported operating tools Different operating tools can be used for local or remote access to the measuring device. Depending
on the operating tool used, access is possible with different operating units and via a variety of
interfaces.
Web browser Notebook, PC or tablet • CDI-RJ45 service Special Documentation for the
with Web browser interface device
• WLAN interface
• Ethernet-based
fieldbus (EtherNet/IP)
Endress+Hauser 79
Proline Promag L 400
Other operating tools based on FDT technology with a device driver such as DTM/iDTM or
DD/EDD can be used for device operation. These operating tools are available from the
individual manufacturers. Integration into the following operating tools, among others, is
supported:
• Process Device Manager (PDM) by Siemens → www.siemens.com
• Asset Management Solutions (AMS) by Emerson → www.emersonprocess.com
• FieldCommunicator 375/475 by Emerson → www.emersonprocess.com
• Field Device Manager (FDM) by Honeywell → www.honeywellprocess.com
• FieldMate by Yokogawa → www.yokogawa.com
• PACTWare → www.pactware.com
The associated device description files are available at: www.endress.com → Downloads
Web server
Thanks to the integrated Web server, the device can be operated and configured via a Web browser
and via a service interface (CDI-RJ45) or via a WLAN interface. The structure of the operating menu
is the same as for the local display. In addition to the measured values, status information on the
device is also displayed and allows the user to monitor the status of the device. Furthermore the
device data can be managed and the network parameters can be configured.
A device that has a WLAN interface (can be ordered as an option) is required for the WLAN
connection: order code for "Display", option W1 "WLAN display": 4-line, illuminated; touch control +
WLAN. The device acts as an Access Point and enables communication by computer or a mobile
handheld terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the measuring
device:
– Uploading the configuration from the measuring device (XML format, configuration backup)
– Save the configuration to the measuring device (XML format, restore configuration)
– Export event list (.csv file)
– Export parameter settings (.csv file, create documentation of the measuring point configuration)
– Export the Heartbeat verification log (PDF file, only available with the "Heartbeat Verification"
application package)
– Flash firmware version for device firmware upgrade, for instance
– Download driver for system integration
HistoROM data management The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making operation and
servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored as a
backup in the device memory. This memory can be overwritten with an updated data record, for
example after commissioning.
80 Endress+Hauser
Proline Promag L 400
There are different types of data storage units in which device data are stored and used by the device:
Device memory T-DAT S-DAT
Available data • Device firmware package • Event history, such as diagnostic events • Sensor data: diameter etc.
• Driver for system integration e.g.: • Measured value memory ("Extended • Serial number
– DD for HART HistoROM" order option) • User-specific access code (to use the
– GSD for PROFIBUS DP • Current parameter data record (used by "Maintenance" user role)
– EDS for EtherNet/IP firmware at run time) • Calibration data
• Maximum indicators (min/max values) • Device configuration (e.g. SW options,
• Totalizer values fixed I/O or multi I/O)
Storage location Fixed on the user interface board in Can be plugged into the user interface board In the sensor plug in the transmitter neck
the connection compartment in the connection compartment part
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the DAT
modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the previous device
data has been exchanged, the new measuring device is ready for operation again immediately
without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are transferred from
the S-DAT in the measuring device and the measuring device is ready for operation again
immediately without any errors
Data transfer
Manual
Transfer of a device configuration to another device using the export function of the specific
operating tool, e.g. with FieldCare, DeviceCare or Web server: to duplicate the configuration or to
store in an archive (e.g. for backup purposes)
Event list
Automatic
• Chronological display of up to 20 event messages in the events list
• If the Extended HistoROM application package (order option) is enabled: up to 100 event
messages are displayed in the events list along with a time stamp, plain text description and
remedial measures
• The events list can be exported and displayed via a variety of interfaces and operating tools e.g.
DeviceCare, FieldCare or Web server
Data logging
Manual
If the Extended HistoROM application package (order option) is enabled:
• Record up to 1 000 measured values via 1 to 4 channels
• User configurable recording interval
• Record up to 250 measured values via each of the 4 memory channels
• Export the measured value log via a variety of interfaces and operating tools e.g. FieldCare,
DeviceCare or Web server
• Use the recorded measured value data in the integrated device simulation function in the
Diagnostics submenu.
Endress+Hauser 81
Proline Promag L 400
C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
Ex approval The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Endress+Hauser sales center.
ATEX, IECEx
Currently, the following versions for use in hazardous areas are available:
EtherNet/IP certification The measuring device is certified and registered by the ODVA (Open Device Vendor Association).
The measuring system meets all the requirements of the following specifications:
• Certified in accordance with the ODVA Conformance Test
• EtherNet/IP Performance Test
• EtherNet/IP PlugFest compliance
• The device can also be operated with certified devices of other manufacturers (interoperability)
82 Endress+Hauser
Proline Promag L 400
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Application packages
Many different application packages are available to enhance the functionality of the device. Such
packages might be needed to address safety aspects or specific application requirements.
The application packages can be ordered with the device or subsequently from Endress+Hauser.
Detailed information on the order code in question is available from your local Endress+Hauser sales
center or on the product page of the Endress+Hauser website: www.endress.com.
Endress+Hauser 83
Proline Promag L 400
Electrode cleaning circuit The electrode cleaning circuit (ECC) function has been developed to have a solution
(ECC) for applications where magnetite (Fe3O4) deposits frequently occur (e.g. hot
water). Since magnetite is highly conductive this build up leads to measuring errors
and ultimately to the loss of signal. The application package is designed to AVOID
build up of highly conductive matter and thin layers (typical of magnetite).
Extended HistoROM Comprises extended functions concerning the event log and the activation of the
measured value memory.
Event log:
Memory volume is extended from 20 message entries (standard version) to up to
100 entries.
Data logging (line recorder):
• Memory capacity for up to 1000 measured values is activated.
• 250 measured values can be output via each of the 4 memory channels. The
recording interval can be defined and configured by the user.
• Measured value logs can be accessed via the local display or operating tool e.g.
FieldCare, DeviceCare or Web server.
Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Accessories Description
Display protection Is used to protect the display against impact or scoring from sand in desert areas.
Ground cable Set, consisting of two ground cables for potential equalization.
84 Endress+Hauser
Proline Promag L 400
Compact → Remote For converting a compact device version to a remote device version.
conversion kit
Promag 50/53 → Promag For converting a Promag with transmitter 50/53 to a Promag 400.
400 conversion kit
Accessories Description
Ground disks Are used to ground the fluid in lined measuring tubes to ensure proper
measurement.
Wireless HART adapter Is used for the wireless connection of field devices.
SWA70 The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
Fieldgate FXA520 Gateway for the remote diagnostics and remote configuration of connected HART
measuring devices via a Web browser.
Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices in the non-Ex area.
Endress+Hauser 85
Proline Promag L 400
Memograph M graphic The Memograph M graphic display recorder provides information on all relevant
display recorder measured variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on a SD card or USB stick.
Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.
86 Endress+Hauser
Proline Promag L 400
Part 1 of 2: Sensor
Part 2 of 2: Transmitter
HART
Operating Instructions
Webserver SD01811D
Webserver SD01458D
Endress+Hauser 87
Proline Promag L 400
Installation Instructions
Installation Instructions for spare part sets Specified for each individual accessory
Registered trademarks
HART®
Registered trademark of the FieldComm Group, Austin, Texas, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
Modbus®
Registered trademark of SCHNEIDER AUTOMATION, INC.
EtherNet/IPTM
Trademark of ODVA, Inc.
Microsoft®
Registered trademark of the Microsoft Corporation, Redmond, Washington, USA
Applicator®, FieldCare®, DeviceCare ®, Field XpertTM, HistoROM®, Heartbeat TechnologyTM
Registered or registration-pending trademarks of the Endress+Hauser Group
www.addresses.endress.com