Hematology Analyzer
D-Cell 30
D-Cell 60
D-Cell 60
D-Cell 30
Offering the
Small labs
excellence
shouldn't
compromise
either
Hematology Analyzer
Exellence in its class
D-Cell 30 3-part differentiation of WBC with improved
cyanide-free reagent system
D-Cell 60 19 parameters + 3 histograms
Automatic orifice clearing
Automatic sampling probe cleaning
Huge number of sample results, including
histograms can be stored
Built-in thermal printer
(optional external printer)
Bar code scanner (optional)
Touchpad and normal computer keyboard for
convenient data entry
Large color LCD display
Flexible and Easy To Use
Full blood picture from venous or capillary
blood
Automatic cleaning programs
Low reagent consumption, cost effective
operation
Low sample volume
High reliability
Maintenance free concept
Total Solution for QC and Calibration
Comprehensive QC programs
Control files for the commercial
control blood and XB programs
Fully automatic calibration
Dedicated reagents from Diagon®
Diagon® D-Cell 30 / 60
31101 Dia-Diluent-D 20 L
31102 Dia-Lyse-Diff D-CF 0,5 L
31103 Dia-Rinse-D 20 L
31104 Dia-EZ-Cleaner-D 0,1 L
31105 Dia-Probe-cleaner-D 0,05 L
TECHNICAL SPECIFICATIONS
D-Cell 60 D-Cell 30
PARAMETERS
WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH,
MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, histograms for WBC, RBC, PLT
PRINCIPLES
Electrical resistance for cell counting and sizing cyanide free method for hemoglobin
PERFORMANCE
Storage capacity
20.000 results with histograms 10.000 results with histograms
LINEARITY RANGE
WBC (109/L) 0.0-99.9
RBC (1012/L) 0.00-9.99
HGB (g/L) 0-300
MCV (fl) 40.0-150.0
PLT (l09/L) 10-999
PRECISION (CV%)
WBC (109/L) 2.5 (7.0-15.0) WBC (109/L) 3.0 (7.0-15.0)
RBC (1012/L) 2.0 (3.50-6.00) RBC (1012/L) 2.0 (3.50-6.50)
HGB (g/L) 1.5 (110-180) HGB (g/L) 2.0 (100-180)
MCV (fl) 0.5 (80.0-110.0) MCV (fl) 1.0 (70.0-110.0)
PLT (l09 /L) 5.0 (200-500) PLT (l09 /L) 5.0 (200-500)
SAMPLING
D-Cell 60: open sampler D-Cell 30: open sampler
SAMPLE VOLUME
Prediluted: 20 uL whole blood
Whole blood: 13 uL whole blood
APERTURE DIAMETER
WBC: 100 um
WBC/RBC/PLT: 80 um
RBC/PLT: 70 um
TEST SELECTION
RBC - PLT
All parameters WBC - 3PD - HGB
all parameters
THROUGHPUT
60 samples per hour 30 samples per hour
DISPLAY
Large, color LCD display
CARRYOVER
WBC, RBC, HGB < 0.5%, PLT < 1 %
INPUT/OUTPUT
External keyboard RS232x2, bar code scanner (optional)
PRINTOUT
Built-in thermal printer, external matrix printer (optional)
OPERATING ENVIRONMENT
Temperature: 15°C- 30°C
Humidity: 30% - 85%
POWER REQUIREMENT
AC 100V-240V, 50/60Hz
DIMENSION
460mm(H)x390mm(W)x390mm(D) 437mm(H)x322mm(W)x390mm(D)