0% found this document useful (0 votes)
251 views117 pages

Ophthalmic Instruments Overview

The document discusses various ophthalmic instruments including the slit lamp biomicroscope, direct ophthalmoscope, indirect ophthalmoscope, and retinoscope. It provides details on the design, components, illumination techniques, and clinical uses of the slit lamp biomicroscope, which is a high-power binocular microscope used to examine the eye with a slit-shaped light source. The summary focuses on the slit lamp as it is covered in most depth in the document.

Uploaded by

maf man
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
251 views117 pages

Ophthalmic Instruments Overview

The document discusses various ophthalmic instruments including the slit lamp biomicroscope, direct ophthalmoscope, indirect ophthalmoscope, and retinoscope. It provides details on the design, components, illumination techniques, and clinical uses of the slit lamp biomicroscope, which is a high-power binocular microscope used to examine the eye with a slit-shaped light source. The summary focuses on the slit lamp as it is covered in most depth in the document.

Uploaded by

maf man
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 117

Ophthalmic instruments part I

MODERATOR –MEKETE (MSC OPTOMETRIST)


PRESENTED BY- DR. MINTESNOT A (R1)
SEP,2019
HU-CHS
Outline

 Slit lamp biomicroscopy


 Direct ophthalmoscope
 Indirect ophthalmoscope
 Retinsoscope
SLIT LAMP BIOMICROSCOPE
Introduction

 Slit lamp biomicroscope is a high-power binocular microscope with a


slit-shaped illumination source
 Speciallydesigned for viewing the different optically transparent or
translucent tissues of the eye and adnexa

 Slit
Lamp: It is a misnomer since slit is only one of the various other
diaphragmatic opening present in the instrument
Cont…
 Allvar Gullstrand: An ophthalmologist and 1911 Nobel laureate
introduced the illumination system which had for the first time a slit
diaphragm, therefore Gullstrand is credited with the invention of the slit
lamp.
Basic design of slit lamp biomicroscope – 3 parts

illumination unit

Observation
unit

Mechanical
unit
Koehler illumination principle

 It create an even illumination of sample and ensures the image of the


illumination source isn’t visible in the resulting image

 The light filament (F) is imaged on to the objective lens (O) but the
mechanical slit (MS) is imaged on to the patient’s eye (S
Types – based on - how illumination unit aim light?

 Zeiss type the illumination comes  Haag Streit type the illumination
from below comes from above
Illumination unit – consist of :

 Light source ( the halogen bulb)


 Condensing lens system
 Consist of 2 planconvex lenses with
their convexities in apposition
 Slit and other diaphragm
 Slit
width and diameter can be
changed
 Filters
 Reflective mirror or prisms
Filters

 Cobalt
blue filter - enhances fluorescein
observations
 Tear film assessment
 Cornea staining to exclude abnormality
 Goldmann applanation tonometry
 Green or Red free filter
 Increase visibility of rose Bengal staining
 Normal density – ND filter
 Decrease beam brightness
Observation or stereomicroscope unit

 Eye piece
complex
 Inverting prism
 Magnification
changer
 Objective lens
 Binocular
viewing system
Eye piece lens complex

 Astronomical telescope
a system of 2 convex lenses
 Result in image which is
 More magnified and inverted
 Freer from optical aberrations than
a single convex lens
Cont.…

 Inverting prism - Porro-Abbe prism The objective lens


 2 triangular prisms arranged to
reflect light several times  Object is located in the
 Resulting in an optically sharp, objective lens focal point
inverted image with no
magnification and little loss of light  Magnifies the object image
projecting it virtually to infinity
 Compensate for the inverted image
produced by eyepiece  Move the working distance
 Shorten tube length from infinity to approximately
10 cm in front of the
microscope
Magnification changer-
Galilean telescope type

 Positioned between objective


lens and tube
 Single convex lens and a single
concave lens, separated by
the difference of their focal
lengths
 The image produced is upright,
virtual image and at infinity
 Most commonly used are 10×,
16×, and 25×
The Grenough type magnification changer

Flip lever to change


magnification
Change eyepieces or objective

 Eyepieces
 Often two sets provided with slit lamp
 Typical values 10x, 12.5x, 15x or 20x
 Objective
 Flip arrangement for rapid change
 Usually only two options due to space confinements
 Typical values are 1x and 2x
Binocular system
Total angular magnification

Objective lens, O - f1
Tube lenses ,T - f2
Telescopic system W - magnification factor g
Ocular, k –f3

M=5x -50x ,Resolution is limited with greater M


Mechanical unit

 It is mainly concerned with:


 Positioning of patient.
 Adjustment for observer and patient.
 Adjustment of illumination and observations system.
 Consist of viewing arm, illumination arm and patient
positioning frame
Cont…

 Viewing arm
 Pairof eyepieces
(oculars)
 Housing of
magnification
elements
 Magnification knob
Cont..

 Illumination arm
 Calibrated scale
that indicates the
length of beam
being used; also
used to measure
lesions
Cont.…

 The point at which the


microscope is focused
corresponds to the point on
which the light is focused, this
coupling effect is called
parfocality
 Allowing the examiner to
direct the light beam
anywhere
Cont.…

 Patient
positioning
frame
 Base and
table
Clinical procedure

 Switch on power & unlock


base screw
 Cleaning the forehead band
 Changing paper strip from
chin rest
 Proper positioning of the pt.
 Adjust eyepieces to correct for
examiner’s refractive error
 Adjust interpupillary distance Correct positioning
Cont.…

 The examination should be commenced using the X10 eyepieces or


the lower powered objective to locate the pathology
 Higher magnification should then be used to get more detail.

 Select the longest slit-length by means of the appropriate lever.


 The angulation between the observation arm and the illumination
arm is adjusted.
Slit lamp illumination types

 The slit lamp offers 6 main illuminating options, each with


its own special properties and particular uses:

 Direct illumination  Indirect illumination


 Transillumination, or
 Diffuse illumination
retroillumination
 Direct focal illumination  Indirect lateral illumination
 Specular reflection  Sclerotic scatter
Diffuse illumination
 Illumination of the eye with a broad,
unfocused light beam
 Directed obliquely between 30-450
 Magnification 6x to 10x
 Neutral density filter

 Used mainly for obtaining an overview of


ocular surface tissues (e.g., bulbar and
palpebral conjunctiva, cornea)
 To pick gross abnormalities of the eyelids,
conjunctiva, sclera, cornea, and iris
 Diffuse illumination with the cobalt-  Diffuse illumination with red-
blue filter (to enhance the free(green) filter is here used to
fluorescence of fluorescein dye) is enhance visibility of rose Bengal
used here to demonstrate the red dye, which has stained
dendritic corneal ulceration of keratin in intraepithelial
herpes simplex epithelial keratitis (squamous) neoplasia.
Direct focal illumination

 Used mainly to show corneal cross


section
 Variation in the application of this
illumination technique can be made just
by changing the width and or/height of
the light source - resulting in:  Technique
 Illumination and observation
 Optic section
are focused in the same
 Parallelepiped plane
 Conical beam.  Slit width narrow to broad
 Illumination angle 45° to 60°
 Magnification 10x-40x
Optical section

 Technique
 Slit width is less than 1 mm
 High intensity of light is used
 High illumination and magnification
 Assess cornea from one side to  Corneal cross –section
other (temporal limbus to nasal  Surface of tear film -brighter
limbus)  Epithelium layer - darker
 Gray wider granular area is
 Facilitates a cross-sectional view the stroma
of the cornea and lens  Endothelium - brighter
Cont.…

 Application
 Todetermine the depth or elevation of a
defect or scar of the cornea or
conjunctiva
 Locating the depth of an opacity within
the lens of the eye - grading of cataract
 To assess iridocorneal angle - Van Herick
method – beam 60 degree on temporal
side
Parallelepiped

 Parallelepiped is basically an optic


section, with the exception of the slit
width can vary from 1mm to its full
size
 Providing a more three
dimensional view of the cornea or
crystalline lens
 Application :
 To examine corneal surface,
stroma
 To ascertain depth (FB)
 Corneal nerves
 Pigment on the back of the Corneal scar
 Blood vessels
cornea - Krukenbergs spindle
 Ghost vessels
-diagnostic of iris atrophy and
pigmentary glaucoma
Conical beam

 Conical beam: is a small circular beam


used to examine the presence of cells and
flare
 Narrow,short & bright slit of light also
can be used - 1 × 3 mm in size
 Light Source - 45-60’ Temporally and
directed in to the pupil
 Bright illumination and high magnification
(16X–25X)
 Beam is focused between cornea and
anterior lens surface.
Cont.…

 Application
 Inflammatory cells
 Flare

 Pigmented cells
 Tyndall phenomenon – scattering of light by
particles in the aqueous humor
Specular Reflection
 It is used mainly for examination of
the corneal endothelium
 It can also be used for examining
the cornea epithelium, tear film
and lens
 Techniques
 Observation and illumination
system have same angle with
perpendicular axis to each other
 Parallelepiped beam of light is  When the slit-lamp’s illumination
used - Slit width < 4mm system and the biomicroscope are at
 Low illumination is needed with equal angles of incidence and
higher magnification – 25X–40X reflection the cornea’s endothelium is
viewable
Cont.…

 From axial point, slowly move slit beam


across the cornea temporal until a glittering
reflection of the filament is seen

 Then focus is moved toward the endothelial


cells
 Mosaics of hexagonal cells are seen –  Bright reflection is obtained
require high magnification from the surface of the
 Endothelial reflection is observed for cornea and a less bright
continuity and uniform intensity reflection from the posterior
surface of the cornea
Cont.
 Compromised corneal surface - broken
or rough reflections may indicate
 Inadequate tear film
 the presence of foreign material
 Corneal edema or bullae
 The regular endothelial mosaic pattern
can be disrupted by
 Largeand small cells of corneal
endothelium
 Deep corneal guttae (orange
 Irregularities in Descemet’s peel-like, dark indentations of the
membrane endothelium caused by focal
 Pigment deposits/ keratic excrescences of Descemet
precipitates - may reflect light membrane) in early Fuchs corneal
dystrophy
Transillumination or
Retroillumination
 Techniques
 Moderate wide slit beam -
projected onto a part of the eye
that lies deeper than the area to
be studied
 Retroillumination - 2 types that
based on the alignment of the
reflected light beam with the area
under observation:
 Direct
Indirect - observe adjacent to area of
 Indirect
illuminated by the reflected light.
 Direct retroillumination
 Findings are made visible with high
Cont..
contrast
 Illumination angle 45-60°
 Indirect illumination
 Illumination angle greatly reduced or
increased
 Feature on the cornea is viewed
against a dark background
 Application
 Infiltrations,small scars, corneal vessels,
micro cysts, vacuoles
Retroillumination from
the fundus

 This technique is used to observe


media clarities and opacities
 With a diffuse red glow (red reflex)
from the fundus
 Dilated pupil
 Viewing and illuminating systems of the
slit lamp not parfocal
 Slit
beam decentered to the edge of
the pupil and nearly straight ahead
Indirect lateral or proximal illumination

 Technique
 Placeillumination light source at
about 45°
 Use a parallelepiped beam
 Dark area just lateral or proximal to
the parallelepiped is the area of the
cornea which one examined
through the biomicroscope.
Cont.…

 Proximal illumination is generally used to study the corneal


epithelium and tears.
 Detection of microcytic edema, faint corneal infiltrates and
other types of irregularities of the epithelium and tears
 Facilitates
to locate and determine of size and shape of an
imbedded foreign body
Sclerotic Scatter

 Technique
 Moderate width and high intensity
beam at the corneoscleral junction
while observer view center of cornea
 The light travels the breadth and width
of the cornea by total internal reflection
 Normal cornea – limbus glows along its
entire circumference
 The normal cornea itself will appear
unilluminated.
Glowing halo
Cont.…

 Sclerotic scatter is especially useful


for detecting subtle corneal
changes over a large area of
distribution
 Identifycertain disease through
detection of a characteristic,
overall pattern
Cornea verticillata (whorl-like
changes) secondary to epithelial
deposition of the oral drug
amiodarone
Associated instruments
The Slit Lamp as a Measuring Device

 By matching the length of the slit-lamp


to to the horizontal and vertical extents
of the subject of interest

 Another method for measuring lesions


is the use of an ocular that contains a
micrometer scale
Fundus Examination With the Slit Lamp

 Possible with the slit lamp and an accessory lens


A Hruby lens is a high-minus (- 55 D) Plano-concave lens that is
often attached to the slit lamp
 High-plus condensing fundus lenses - handheld lenses
 +90 D and +78 D lenses commonly but available 60-132 D
 Contact lens - Goldmann 3 mirror lens
DIRECT OPHTHALMOSCOPE
Direct Ophthalmoscope

 Ophthalmoscope is an
instrument which
commonly used for
examination of the fundus
of the eye
 It was introduced by
Hermann von Helmholtz in
1850
A German physician
and physicist
Cont…

 It is called direct ophthalmoscope :


 Because image of the subject’s retina is formed directly on the
observer’s retina
 Unlike this, indirect ophthalmoscope use intermediate inverted
image hence the name indirect
Basic principle of Direct ophthalmoscope

 If patient and observer are both emmetropic, rays emanating from


a point in the patient's fundus will emerge as a parallel beam and
will be focused on the observer's retina
Cont…
Field of view of Direct Ophthalmoscope

 The field of view is limited by the most oblique pencil of light that
can still pass from the patient's pupil to the observer's pupil

Angle α is the field of views


Increased in dilate pupil
or when the eyes are
brought more closely
together
Refractive error of the
patient also can affect field
of ophthalmoscope
 Limited field of view in the direct method because peripheral pencils of
light do not reach the observer's pupil.

Upright image

 Extended field of view in the indirect method because the


ophthalmoscopy lens redirects peripheral pencils of light toward the
observer.

Inverted image
Magnification in Direct Ophthalmoscope

 Conventional defining of magnification is to compare the


observer's view of a given object with the view that would be
obtained when looking at the same object from a standard
distance.
 The usual standard for comparison is 25 cm
 The magnification M under which the image is seen through
the ophthalmoscope is given by:
M= P P is the refractive power of the eye -
4 Standard reduced eye, P= 60 and M = 15
Cont….
 Myopic patients have extra
plus power - ophthalmoscope
must carry a negative lens
 Galileantelescope effect -
fundus details are seen
larger
 Reduce field of vision
 In hyperopic or aphakic
patients - ophthalmoscope
must carry positive lens
 Reverse Galilean telescope
effect – fundus details
appear smaller
 Larger field of vision
Cont…

 Refractory error can be compensated by:


 By having patient and physician wear their respective spectacle
(or contact Lens) correction
 For patients with high refractive errors and especially in the case
of marked astigmatism
 For small refractive errors,
 Single lens in the ophthalmoscope must replace the
mathematical sum of the patient's and the observer's correction
Parts of direct ophthalmoscope
Illumination system
Clinical procedure

 Performed with the eye that corresponds to the eye being examined
 Focused by twirling the dial for the Rekoss disk
 With a dilated pupil patient is instructed to stare into the distance
 The focusing lens is set at 0 (or the examiner's refractive error)
 The patient's red reflex is checked from a distance of 2 feet
 Check opacity in the optical media – dark shadow
 The light beam must remain centered within the pupil – to avoid
troublesome corneal light reflexes
 While holding the patient's eyelids open, the examiner dials the focusing
lenses to clarify the fund us image
 As the patient star at a distance target, the ophthalmoscope is angled
about 15° temporal to fixation
Cont.…

 The retinal nerve fiber layer bundles are seen as fine, bright
striations fanning off the optic disc
 The green (i.e., red-free) filter enhances the visibility of the
retinal nerve fiber layer
 Exmination begin at the inferotemporal region close to the optic
disc then proceeds to the superotemporal region, followed by
the superonasal and the inferonasal parts
The Fundus Record

 The retinal drawing is made inside a circle centered on the


fovea
 Shows the relative positions of the optic disc, major retinal
blood vessels, and ora serrata
 Record of the examination is kept by Standard preprinted
fund us charts , or vitreoretinal charts
Vitreoretinal drawing chart
Optic disc drawing chart
 Bulbous retinal detachment with peripheral exudates
(yellow) and large and small horseshoe tears.
INDIRECT OPHTHALMOSCOPE
Introduction
Parts - head set

 gggg
Condensing lens

 The Examiner positions a variety of convex,


handheld magnifying ("condensing")
diagnostic lenses close to the patient's eye
 Usual powers +13 D, +20 D and +28 D.
 The observer holds the condensing lens at
arm's length
Illumination

 The illumination is provided by an electric lamp mounted on the


observer's head which passes through the condensing lens into
the observed eye.
 The light reflected from the observed eye is refracted by the
condensing lens to form a real inverted image between the
condensing lens and the observer
 The level of the observer's eye: Is the same as that of the observed
eye.
Binocular viewing system

 Is mounted on the observer's head and consists of


 Eyepiece – a convex lenses with +2 D - help the observer to view
the fundus
 Withoutusing his accommodation - especially if the patient is
aphakic or highly Hyperopic
 If the observer is presbyopic
 Two reflecting prisms
 Reflecting mirrors – 2 in number
Two reflecting prisms -Provide light to the observer's eye by total
internal reflection as the angle of incidence (45°) is greater than the
critical glass/air angle (41°).
Field of view of Indirect Ophthalmoscope

 Field of view of various lenses =


Lens diameter/Focal length = Lens diameter ×
dioptric power

 Lenses with equal power - a larger lens provides a


wider field of view
 Lenses with equal diameter , a stronger lens provides
a wider field of view
Imaging in Indirect Ophthalmoscope

Aerial image is an intermediate, inverted and reversed real


image of the patient's fundus is formed in the focal plane of the
ophthalmoscopy lens.
 The observer must accommodate on this image
Cont..

 Indirect ophthalmoscopy is that it requires a considerable distance


between the patient and the observer
 Act like an astronomical telescope
 Patient's
cornea and crystalline lens act as astronomical
telescope's objective lens
 Condensing lens acts as the astronomical telescope's eyepiece
lens
 It offers a wider field of view than does direct ophthalmoscopy, but this
advantage is at the expense of decreased magnification.
Magnification of Indirect Ophthalmoscope
2 component:

Aerial image
magnification
=Aerial
image/Fundus
detail
=60/lens power
Magnification
from aerial image
to the observer's
retinal image
Refractory error compensation
Cont…

An interesting case exists for a patient with 20-D myopia.


The myopic eye forms its own aerial image without the help of the
ophthalmoscopy lens.
Examination overview

 Headset Adjustment
 Should be positioned comfortably on the examiner's head
 Use the frontalis muscle to raise and lower the headset
slightly.
Eyepiece adjustment

 The eyepieces should be situated as close as possible to the


examiner's eyes perpendicular to the pupillary plane, without
touching the bridge of the nose
 Knob to adjust the angle of the eyepiece-light housing
 Refractory error – near and distant – of the observer if there are
greater the ocular lens power it should be corrected
 Adjust the interpupillary distance of the oculars
Light beam adjustment

 Mirror angle or tilt control knob – used to aligned the light beam
vertically until the light occupies the upper half of the field of
view for an arm's length working distance
 Brightness control knob
 Prolonged examination should use a reduced power setting of
the headset light and a condensing lens with an ultraviolet filter
or yellow coating
Choosing and Positioning the Condensing Lens
 +20 D aspheric lens – used for routine examination
 +30 D lens – when wider field is needed – diffuse retinal anomalies
 +14 D lens - reserved for examining the optic nerve lesion
Working distance

 Examination is performed at arm 's length, normally with about 40-


50 cm
 Difficulty seeing through a small pupil
 Withdrawing to a greater examination distance
 Use a higher-power lens.
 Distance between the examiner and the lens does not have to be
fixed
Sequence of the Examination

 Begin indirect ophthalmoscopy without the condensing lens – to


a get red reflex
 Condensing lens is then brought into position
 Image is completely inverted and reversed.
 Identifying the optic disc
 Retinal vessels from the optic disc to the equator.
 Superior or nasal before the inferior and temporal fundus
 To reduce photophobia
 Finally, scleral depression
Scleral depression

 To examine the area between the equator of the fundus (14 mm from
the limbus) and the ora serrata (8 mm from the limbus)
RETINOSCOPE
Introduction

 Retinoscope – an instrument that we use to illuminate inside of eye


and to observe light reflected from the fundus
 Retinoscopy is the name given to the objective method of
determining the refractive errors by using retinoscope
 Provide a starting point of subjective refraction
 Only option for people who are unable to communicate
Non-refractive use of retinoscopy

 To see optical media opacity - on lens, cornea, iris even early


opacity by retro-illumination from retina
 Keratoconus –b/c it distort the reflex
 Retinal detachment –b/c central distort the reflecting surface and
a gray reflex may be seen
 Perform indirect ophthalmoscopy using high plus lens (monocular
indirect ophthalmoscopy
Cont.…

 Retinoscope parts
 Eyepiece
 Mirror assembly – plane and concave mirror
 Light source
 Spot or streak bulb
 Collar
 Moves up and down to change the
vergence of the light
 Rotates to change the angle of the beam
 On/off/brightness control
Types of Static retinoscopy

 Static retinoscope - use in patient with relaxed accommodative


status achieved by looking at distance or using cycloplegic drugs
 The spot retinoscope: The source of light used produces a
circular image.
 The streak retinoscope: The source of light used produces a
linear image. - More commonly used
 Adv. - Each meridian can be neutralized separately
 Easier to apply and require less time
Cont.…

 Retinoscope has 2 system


 Observation system -
the examiner looks
through a plane mirror
with central perforation
 Illumination system

 The illuminated area of


fundus is not seen by the
observer but only acts as a
source for - retinoscope
shadow or reflex at subject
pupil
Illumination
system

 Light source
 Condensing lens
 Mirror at 45°
 To give both systems
of illumination - plane
and concave mirror
effects by altering
distance between light
source and
condensing lens
Cont.….

 The illuminated fundus area moves with the movement of the plane mirror
and against the movement of the concave mirror, irrespective the
refractive state of the observed eye
Observation system

 Reflex movement under


plane mirror effect can be
 Opposite movement
with concave mirror
effect
“ with” movement

 A image of the illuminate retina is formed at the patient’s far point


 Far
point is at infinity for emmetropic, in front of the eye for
myopic and behind the eye for the hyperopic
 “ with” movement seen when the Far point is behind the examiner
 Emmetrope

 Hyperope

 Myope less than working distance (FP> WD) – less than –1.5 D if
working distance is 0.75 m
“against” movement

 “against” movement seen when the Far point is in front


the examiner
 Myopia greater than the working distance greater
than – 1.5 D if working distance is 0.75 m
 Neutral point
 Achieved when the far point of the patient’s eye is
equal to the examiners working distance
Refractive error Plane mirror effect Concave mirror
effect

Myopia With- < WD Against


No- = WD NO
Against > WD With

Hyperopic With against

Emmetropia No No
Observe the reflex optical characteristics

 The direction of the reflex


 The relative speed of the movement
 The brightness of the reflex
 The width of the reflex
 Shape of the reflex
Cont.….

 The speed of movement of the reflex is useful in determining refractive


status
 the slower the movement- higher refractive error
 the faster the movement –the less the refractive error

 The speed of movement does not depend on the speed of movement


of the retinoscope, but is related/compared to it
Cont.….

 The brightness of the red reflex


If the reflex is dull, then a high refractive error
If the reflex is brighter, then a low refractive error

The width of the reflex


The higher the refractive error, the narrow reflex width
The Lower the refractive error, the wider reflex.
Cont.….The shapes of the red reflex
Retinoscopy techniques

 The examining room light s are dimmed.


 Examiner sits facing the patient
 Patient should look past the examiner's ear at a fixation target
located at effective optical infinity (20 feet or more)
 The examiner’s eyes on the same level as the patient
 At standard distance, usually about arm's length
 Examine right eye of the patient with right eye of yours and do
the same to the left eye
Cont.…

 It is critical to know that the eye being examined is the fixating


eye
 Retinoscopic reflex should be in the visual axis ,rather in the
axis of the deviation - in order to measure the correct
refractive error in the visual axis
 Manifest strabismus
 Occlude the eye not being examined so that the eye being
examined can be fixating eye
Working distance

 The distance between the examiner and the patient 's eye
 Must be measured and converted into diopters
 Most examiners use a working distance of arm's length, usually about
66 cm - corresponding working lens of +1.50 D.
 The working distance must remain constant throughout the
examination
Cont.…

 If retinoscopy is done through a working lens


 The additional lens power that results in neutralization is the
refractive err or.
 If a working lens is not used-
 theexaminer must subtract the dioptric equivalent of the
working distance from the sphere of the refractive error
reached in retinoscopy.
Neutralization With a Retinoscope

 Neutralization refers to the achievement of the point at which a


lens placed before the patient's eye effectively "neutralizes" the
retinoscopic reflex which means :
 The patient's pupil fills with reflected light – full width of
retinoscopy reflex
 No movement of the retinoscopy reflex
 The closer one approaches to neutral:
The wider, brighter and faster the reflex.
Recording of the retinoscopic results

 This is usually done in the form of a cross (power


cross) which indicates the neutralization point of the
two main meridians and also their orientation
 Vertical streak – move nasal to temporal
 Cylindrical axis is – 900
 Power meridian or axis is – 1800
 Change the power cross into axis cross Power cross
 Lens transposition
Checking of the neutral point

 If the power of the lens in the trial frame is Increased by 0.5D -


The movement of the red reflex will be reversed.
 If the observer moves his head backwards - The red reflex
moves against the movement of the mirror.
 If the observer moves his head forwards - The red reflex moves
with the movement of the mirror
References

 Slit-lamp Biomicroscopy in Primary Eye Care


 Clinical optics - third edition
 Practical ophthamology, manual for beginning residents,
seventh edition
 Duane’s clinical ophthalmology 12 edition
 BCSC, Clinical optics, 2016-2017
 internet source – slideshare and pictures
Thank you
Questions?

You might also like