Cubemass DCI
Coriolis mass flow measuring system
TI099D/06/en/04.10
71112154
Cubemass DCI
Table of contents
2 Endress+Hauser
Cubemass DCI
The amplitude of the Coriolis force depends on the moving mass Δm, its velocity v in the system, and thus on
the mass flow. Instead of a constant angular velocity ω, oscillation occurs.
This causes the measuring tube loop through which the fluid is flowing to oscillate. The Coriolis forces
produced at the measuring tube loop cause a phase shift in the oscillations of the tube loop (see illustration):
• If there is zero flow, i.e. when the fluid stands still, the oscillation measured at points A and B has the same
phase, and thus there is no phase difference (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tube loop (2) and acceleration at the
outlet (3).
A A A
B B B
1 2 3
A0003383
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube loop
oscillations at the inlet and outlet.
Compared to two-tube systems, other design solutions are required in single-tube systems to ensure system
balance. In the case of the CNGmass DCI, an internal reference mass is provided for this purpose.
The measuring principle operates independently of temperature, pressure, viscosity, conductivity and flow
profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the
density of the oscillating system (comprising the measuring tube loop and fluid) results in a corresponding,
automatic adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The
microprocessor utilizes this relationship to obtain a density signal.
Temperature measurement
The temperature of the measuring tube loop is determined in order to calculate the compensation factor due
to temperature effects.
This signal corresponds to the process temperature and is also available as an output.
Endress+Hauser 3
Cubemass DCI
Measuring system The measuring system consists of a transmitter and a sensor. Two versions are available:
• Compact version: transmitter and sensor form a mechanical unit.
• Remote version: transmitter and sensor are mounted physically separate from one another.
Transmitter
A0003672
Sensor
A0011878
4 Endress+Hauser
Cubemass DCI
Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Volume flow (calculated using mass flow and density)
• Fluid density (proportional to the resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
1 1/24" 0 to 20 0 to 0.75
6 ¼" 0 to 1000 0 to 37
Endress+Hauser 5
Cubemass DCI
Output
Output signal Current output
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s),
full scale value selectable, temperature coefficient: typically 0.005% o.f.s. / °C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 Ω, RL ≥ 250 Ω (HART)
• Passive: 4 to 20 mA; supply voltage VS: 18 to 30 V DC; Ri ≥ 150 Ω
o.f.s. = of full scale value
Pulse/frequency output
Active/passive selectable, galvanically isolated
• Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• Passive: open collector, 30 V DC, 250 mA
• Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1,
pulse width max. 2 s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms)
MODBUS RS485
• MODBUS device type: slave
• Address range: 1 to 247
• Supported function codes: 03, 04, 06, 08, 16, 23
• Broadcast: supported with the function codes 06, 16, 23
• Physical interface:
RS485 in accordance with EIA/TIA-485 standard
• Supported baud rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
• Transmission mode:
RTU or ASCII
• Response times:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
• Possible output combinations È Operating Instructions
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
6 Endress+Hauser
Cubemass DCI
Power supply
Electrical connection
Measuring unit
A B
HART
b/c
–/b
a/a
MODBUS RS485
– 27 A (RxD/TxD-N) 27
+ 26 B (RxD/TxD-P) 26
– 25 b – 25 c
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20 b
N (L-) 2 d N (L-) 2
d
L1 (L+) 1 L1 (L+) 1
a a
A0012023
Connecting the transmitter, cable cross-section: max. 2.5 mm2 (14 AWG)
A View A (field housing)
B View B (wall-mount housing)
Endress+Hauser 7
Cubemass DCI
Order version 20 (+) / 21 (-) 22 (+) / 23 (-) 24 (+) / 25 (-) 26 (+) / 27 (–)
8CN**–**D********* Status input Relay output Frequency output Current output, HART
8CN**–**M********* Status input Frequency output 2 Frequency output 1 Current output, HART
8CN**–**1********* Relay output Frequency output 2 Frequency output 1 Current output, HART
8CN**–**2********* Relay output Current output 2 Frequency output Current output 1, HART
Electrical connection,
Remote version S1 S1 S2 S2 GND TM TM TT TT
a b + + + +
4 5 6 7 8 9 10 11 12 41 42
d
d
d
e
c
4 5 6 7 8 9 10 11 12 41 42
+ + + +
S1 S1 S2 S2 GND TM TM TT TT
A0003681
8 Endress+Hauser
Cubemass DCI
Cable specifications Any suitable cable with a temperature specification of at least 20 °C (68 °F) higher than the ambient
temperature of the application. We recommend the use of a cable with a temperature specification of +80 °C
(+176 °F).
Remote version:
• 6 × 0.38 mm2 (20 AWG) PVC cable with common shield and individually shielded cores
• Conductor resistance: ≤ 50 Ω/km (≤ 0.015 Ω/ft)
• Capacitance core/shield: ≤ 140 pF/m (≤ 42.7 pF/ft)
• Cable length: max. 20 m (65.6 ft)
• Permanent operating temperature: max. +105 °C (+221 °F)
! Note!
The cable must be installed securely, to prevent movement.
Endress+Hauser 9
Cubemass DCI
Performance characteristics
Reference operating • Error limits following ISO/DIS 11631
conditions • Water, typically 20 to 30 °C (68 to 86 °F); 2 to 4 bar (30 to 60 psi)
• Data as per the calibration report ±5 °C (±9 °F) and ±2 bar (±30 psi)
• Data on the measured error based on accredited calibration rigs traced back to ISO 17025
Maximum measured error The values indicated refer to the pulse/frequency output.
The additional measured error for the current output is typically ±5 μA. Basis for calculations → ä 11.
o.r. = of reading
Density (liquids)
• ±0.001 g/cc (after field density calibration or under reference conditions)
• ±0.002 g/cc (special density calibration (optional)
Calibration range: 0.0 to 2.0 g/cc, 5 to 80 °C (41 to 176 °F)
Application range: 0.0 to 5.0 g/cc, –50 to 200 °C (–58 to 392 °F)
• ±0.02 g/cc (standard calibration)
Temperature
±0.5 °C ± 0.005 · T °C
(±1.0 °F ± 0.003 · (T – 32) °F)
T = Fluid temperature
[%]
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0 5 10 15 20 25 kg/h
A0011691
10 Endress+Hauser
Cubemass DCI
15 0.5513 0.1
Density (liquids)
• ±0.0005 g/cc
1 g/cc = 1 kg/l
Temperature
±0.25 °C ± 0.0025 · T °C
(±0.5 °F ± 0.0015 · (T – 32) °F)
T = Fluid temperature
Influence of medium When there is a difference between the temperature for zero point adjustment and the process temperature,
temperature the typical measured error is ±0.0002% of the full scale value / °C (±0.0001% of the full scale value / °F).
Influence of medium pressure The tables below shows the effect on accuracy of mass flow due to a difference between calibration pressure
and process pressure.
DN Medium pressure
[mm] [inch] [% o.r./bar] [% o.r./psi]
1 1/24" −0.001 −0.00007
2 1/12" 0 0
4 1/8" −0.005 −0.0004
6 ¼" −0.003 −0.0002
o.r. = of reading
Endress+Hauser 11
Cubemass DCI
2 4
1 3
6
A0004302
A0003601
Inlet and outlet runs There are no installation requirements regarding inlet and outlet runs. If possible, mount the sensor upstream
of fittings such as valves, T-pieces, elbows etc.
12 Endress+Hauser
Cubemass DCI
System pressure It is important to ensure that cavitation does not occur as it could influence the oscillation of the measuring
tube. No special measures need to be taken for fluids which have properties similar to water under normal
conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is
important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to
boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects
can be prevented when system pressure is sufficiently high.
Consequently, it is generally best to install the sensor:
• On the pump pressure side (no risk of vacuum)
• At the lowest point in a riser
Endress+Hauser 13
Cubemass DCI
! Note!
• Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F), the readability of the display may be impaired.
Storage temperature –40 to +80 °C (–40 to +175 °F), preferably at +20 °C (+68 °F)
14 Endress+Hauser
Cubemass DCI
Seals
• Only for mounting kits with threaded connections:
– Viton: –15 to 200 °C (–5 to +392 °F)
– EPDM: –40 to +160 °C (–40 to +320 °F)
– Silicone: –60 to +200 °C (–76 to +392 °F)
– Kalrez: –20 to +275 °C (–4 to +527 °F)
2 1/12"
6 ¼"
Pressure loss (SI units) Pressure loss depends on the fluid properties and on the flow rate.
The following formulas can be used to approximately calculate the pressure loss:
4·g
Reynolds number Re =
p·d·n·r
A0003381
Re < 2300 Dp = K1 · n · g
A0003379
Endress+Hauser 15
Cubemass DCI
DN d K K1
[mbar]
10000
DN 44
DN
DN 2 DN 6
DN 1
1000
100
10
1
0.1 1 10 100 1000 [kg/h]
A0011692
Pressure loss (US units) Pressure loss depends on the nominal diameter and the fluid properties.
The "Applicator" PC software is available from Endress+Hauser and can be used to calculate the pressure loss
in US units. The "Applicator" program contains all the important device data which allows the measuring
system arrangement to be optimized.
The software is used for the following calculations:
• Nominal diameter of the sensor with fluid properties such as viscosity, density etc.
• Pressure loss downstream from the measuring point
• Conversion of mass flow to volume flow etc.
• Simultaneous display of variables determined by different measuring devices
• Determining measuring ranges
The Applicator program runs on any IBM-compatible PC with Windows.
16 Endress+Hauser
Cubemass DCI
Mechanical construction
Design/dimensions Field housing compact version (non-hazardous area II2G / zone 1)
A J
A* K
Esc
- + E
G
F
di
L
C
D M
N
B
H
E
O
A0011849
Dimensions in SI units
DN A A* B C D E F G H J K L M N O di
1 1.3
2 2
227 207 10 40 90 120 291.2 269.3 22 168 100 30 120 175 187.5
4 3.9
6 5.35
Dimensions in US units
DN A A* B C D E F G H J K L M N O di
1/24" 0.05
1/12" 0.08
8.94 8.15 0.39 1.57 3.54 4.72 11.5 10.6 0.87 6.61 3.94 1.18 4.72 6.89 7.38
1/8" 0.15
¼" 0.21
Endress+Hauser 17
Cubemass DCI
A
A*
B C
B* D
e r Sp annu
ht unt ng
Nic öffnen
Nicht-eigensichere
ove r tight wh
e
c
IP40-Abdeckung geschützt
p cover tight w
Non-intrinsically safe
circuits Ip40 protected
E
Esc
ep c
Boucles de courant – +
c
E
ve
sans sécurité intrinsèque
protégées par Ip40
sou s nsion l Ne
pas o t e pparei
uvrir l’a
M
L
J
K
F
G
A0006999
Dimensions in SI units
A A* B B* C D E F G H J K L M
∅ 8.6
265 242 240 217 206 186 178 100 130 100 144 170 355
(M8)
Dimensions in US units
A A* B B* C D E F G H J K L M
∅ 8.6
10.4 9.53 9.45 8.54 8.11 7.32 7.01 3.94 5.12 3.94 5.67 6.69 13.9
(M8)
18 Endress+Hauser
Cubemass DCI
D
Esc
- + E
B
C
F G
A E
H J K J
N
S S
M
R
L
O
P Q P
A0001150
Dimensions in SI units
A B C D E F G H J K
L M N O P Q R S T 1)
Dimensions in US units
A B C D E F G H J K
8.46 9.84 3.56 6.27 5.31 3.54 1.77 > 1.97 3.18 2.08
1)
L M N O P Q R S T
3.74 2.08 4.01 3.20 0.45 7.55 8 × M5 0.79 2 × ∅ 0.26
1)
Securing screw for wall mounting: M6 (screw head max. 0.41")
All dimensions in [inch]
Endress+Hauser 19
Cubemass DCI
A
B G
F
E
C H
A0012360
Dimensions in SI units
DN A B C D E F G H
Dimensions in US units
DN A B C D E F G H
1/24 to ¼" 6.42 5.63 6.89 4.02 5.24 9.25 5.08 4.72
All dimensions in [inch]
20 Endress+Hauser
Cubemass DCI
Esc
Q
- + E
P
L ±1.0 (±0.04)
mm (inch)
A0012329
Dimensions in SI units
DN L P Q
6 175 AF 1" 20
Dimensions in US units
DN L P Q
Endress+Hauser 21
Cubemass DCI
R
LK
Q
Esc
- + E
P
S
L ±1.5 (±0.06)
mm (inch)
A0012330
Dimensions in SI units
DN PN L P Q R S LK
1 to 6 40 278 95 17.3 4 × Ø 14 28 65
DN JIS L P Q R S LK
1 to 6 10K 278 95 15 4 × Ø 15 28 70
DN ASME L P Q R S LK
1 to 6 Cl 150 278 88.9 15.7 4 × Ø 15.7 17.7 60.5
Dimensions in US units
DN JIS L P Q R S LK
DN ASME L P Q R S LK
1/24 to ¼" Cl 150 11 3.50 0.62 4 × Ø 0.62 0.70 2.38
22 Endress+Hauser
Cubemass DCI
Esc
Q
- + E
P
+1.5 (+0.06)
L –2.0 (–0.08)
mm (inch)
A0012332
Dimensions in SI units
DN L P Q
Dimensions in US units
Endress+Hauser 23
Cubemass DCI
" Caution!
The secondary containment is filled with dry nitrogen (N2). Do not open the purge connections unless the
containment can be filled immediately with a dry inert gas. Use only low overpressure to purge.
Maximum pressure 5 bar (73 psi).
C
B D
A0012335
Dimensions in SI units
DN A B C D
1 to 6 ½" NPT 30 37 33
All dimensions in [mm]
Dimensions in US units
DN A B C D
24 Endress+Hauser
Cubemass DCI
Rupture element/disk
The sensor housing is optionally available with an integrated rupture element.
# Warning!
• Make sure that the function of the rupture element is not impeded by the installation. The triggering pressure
in the housing is indicated on the information notice. Take suitable measures to ensure that no damage can
occur if the rupture disk is tripped and personal injury is ruled out.
Triggering pressure in the housing 10 to 15 bar (145 to 217 psi)
• Please note that if a rupture disk is used, the housing can no longer assume a secondary containment
function.
• It is not permitted to open the connections or remove the rupture disk.
" Caution!
The existing connection nozzles are not designed for a purge or pressure monitoring function.
! Note!
• The transportation guard on the rupture disk must be removed prior to commissioning.
• Comply with the information on information notices.
A0012344
i i
Additional sign regarding the position of the rupture disk (RUPTURE DISK)
A Sensor housing with rupture element (incl. rupture disk) È defined fluid exit
A Sensor housing with rupture disk È undefined fluid exit
! Note!
The location of the rupture disk with undefined fluid exit is covered over by an adhesive label. If the rupture
disk is tripped, the adhesive label is destroyed, meaning that a visual check is possible.
Endress+Hauser 25
Cubemass DCI
A B
C
D
1 2
A0012349
Dimensions in SI units
DN A B C D
Dimensions in US units
DN A B C D
26 Endress+Hauser
Cubemass DCI
# Warning!
The following load curves relate to the entire measuring device and not just to the process connection.
[psi] [bar]
700 50
600 40
500 PN 40
30
400
300 20
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0012140
[psi] [bar]
700 50
600 40 Class 300
500
30
400
300 20
200 Class 150
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0012141
Endress+Hauser 27
Cubemass DCI
[psi] [bar]
300 20
200
10
100 10K
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0012143
[psi] [bar]
440
6200
5800 400
DN 1
5400
360
5000
4600 320
4200 280
3800
3400 240
3000 200
2600
2200 160
DN 2, DN 4, DN 6
1800
120
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0011882
28 Endress+Hauser
Cubemass DCI
Human interface
Display elements • Liquid crystal display: illuminated, four lines with 16 characters per line
• Selectable display of different measured values and status variables
• 3 totalizers
• At ambient temperatures below –20 °C (–4 °F), the readability of the display may be impaired.
Operating elements • Onsite operation with three optical sensor keys (S/O/ F)
• Application specific Quick Setup menus for straightforward commissioning
! Note!
You can change the language group via the operating program FieldCare.
Endress+Hauser 29
Cubemass DCI
C-Tick symbol The measuring system meets the EMC requirements of the Australian Communications and Media Authority
(ACMA).
Ex approval Information about currently available Ex versions (ATEX, NEC/CEC etc.) can be supplied by your
Endress+Hauser sales office on request. All information relevant to explosion protection is available in separate
Ex documents that you can order as necessary.
Pressure Equipment Directive Measuring devices with a nominal diameter smaller than or equal to DN 25 (1") correspond to Article 3(3
) of the EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and manufactured
according to good engineering practice.
30 Endress+Hauser
Cubemass DCI
Ordering information
The Endress +Hauser service organization can provide ordering information and detailed information on the
order code.
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor.
! Note!
The Endress +Hauser service organization can provide detailed information on the relevant order codes.
Documentation
❑ Flow measurement (FA005D/06)
❑ System Information Promass (SI032D/06)
❑ Operating Instructions (BA139D/06)
❑ Operating Instructions MODBUS RS485 (BA141D/06)
❑ Description of Device Parameters (GP002D/06)
❑ Description of Device Parameters MODBUS RS485 (GP004D/06)
❑ Ex-Supplementary documentation ATEX (II2G): (XA139D/06)
❑ Ex-Supplementary documentation NEC/CEC (Div. 1): (XA141D/06)
❑ Ex-Supplementary documentation NEPSI (Zone 1, Zone 21): (XA142D/06)
Registered trademarks
KALREZ® and VITON®
Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA
HART®
Registered trademark of HART Communication Foundation, Austin, USA
MODBUS®
Registered trademark of the MODBUS Organization
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, FieldCare®, Fieldcheck®, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
Endress+Hauser 31