Agilent 6850 Series II Network GC System
Specifications
Description • 100 V (+10%, –10%), 15 amps, - CISPR 11/EN 55011:
47–63 Hz Group 1 Class B
The single-channel Agilent 6850
Series II Network GC system has • 120 V (+10%, –10%), 12 amps, - IEC/EN 61326-1:2005 Electri-
similar state-of-the-art performance 47– 63 Hz cal equipment for measure-
as the Agilent 6890N GC, but is only ment, control, and laboratory
• 230 V (+10%, –10%), 9 amps,
half as wide. The six-button, two- use - EMC requirements
47–63 Hz
line local interface provides run - IEC68-2-27 Mechanical Shock
control and status information. The Fast Oven Test Standard
GC is network-ready with a built-in
• 120 V (+10%, –10%), 20 amps, - Canada ISEC/NMB-001
LAN communications interface. The
47–63 Hz
GC is also automatic liquid sampler - AS/NZC CISPR 11
(ALS)-ready with built-in controller • 200/208 V (+10%, –10%),
for the G2880 autosampler or the 12 amps, 47–63 Hz • Designed and manufactured
Agilent 7683/7693 autoinjector. under a quality system registered
• 230 V (+10%, –10%), 11 amps, to ISO 9001
47–63 Hz
• Declaration of Conformity
Environmental Conditions available
• Ambient operating temperature: Certifications
15 °C to 35 °C
• Conforms to the following safety Column Oven
• Ambient operating humidity: standards:
5% to 95% • Dimensions: 202 × 200 ×
- Canadian Standards Associa- 105 mm (HWD)
• Storage extremes: tion (CSA): C22.2 No. 1010
–40 °C to 70 °C • Oven power is turned off auto-
- CSA/Nationally Recognized matically when the lid is opened
• Heat dissipation: 3,000 Btu/h, Test Laboratory (NRTL):
0.88 kW typical (at 100–120 V) UL 3101 • Column basket diameter:
130 mm
- International Electrotechnical
• Standard: 5 °C above ambient to
Power Requirement Commission (IEC): 61010-1
350 °C. With CO2 cryo:
- EuroNorm (EN): 61010-1 –20 °C to 350 °C.
Standard Oven
• Conforms to the following • Temperature setpoint resolution:
•Approximately 1,440 VA (max) at
regulations on Electromagnetic 1 °C
100–120 V, 2,000 VA at 230 V
Compatibility (EMC) and Radio
Frequency Interference (RFI): • Column bleed compensation
• Maximum run time: 999.99 min
• Temperature programming: Split/Splitless Capillary Inlet • Maximum temperature ramp
six ramps with seven plateaus rate: 350 °C/min
• Maximum temperature: 375 °C
• Ambient rejection: <0.01 °C per • Pressure setting range: 0–100 psi
1 °C ambient change • Pressure range: 0–100 psi
(0–150 psi optional) • Total flow setting range:
• Automatic H2 carrier safety turn- 0 to 100 mL/min
off when inlet gas pressure drops • Accepts columns from 50-µm id
to 530-µm id • Available with LCO2 cyrogenic
• Typical cool-down rate: cooling only
350 °C to 50 °C: 5.2 min • Pressure pulse mode
• Lower temperature limit without
• Cryogenic gas consumption • Gas saver mode LCO2 cyrogenic cooling:
is typically half that of • Total flow ranges through the Oven temperature plus 3 °C with
dual-channel ovens inlet: He, 20 to 1,000; H2, 26 to oven track on, –20 °C with oven
1,000; N2, 20 to 200 mL/min track off
Table 1. Typical Oven Ramp Rates • Split ratio may be adjusted elec- • Lower temperature with LCO2
(at nominal line voltage) cyrogenic cooling:
tronically without affecting
Standard Fast column flow or head pressure -17 °C with oven track on
Temperature oven rates oven rates -20 °C with oven track off
range in °C (°C/min) (°C/min) Purged-Packed Injection Port
50 to 75 120 240 Detectors
• Maximum temperature: 375 °C
75 to 115 95 190
115 to 175 65 130 • Accepts 530-µm capillary columns Detectors include EPC.
175 to 300 45 90 or 1/8-inch metal-packed columns
300 to 350 35 70 up to 20 feet in length Flame Ionization Detector (FID)
• Total flow setting range: • Maximum temperature: 375 °C
Inlets 0–100 mL/min
• Automatic ignition
Inlets Available:
PTV • Flame-out detection and
• Split/Splitless capillary inlet reignition
(S/SL) • Maximum temperature: 375 °C
• Minimum detectable: <5 pg
• Purged packed injection port • Three temperature program carbon/s as propane using N2
(PPIP) ramps carrier and a 0.29-mm id jet
• Temperature–programmable • Temperature ramp rates 0.1 to • Linear dynamic range:
vaporizer (PTV) 720 °C/min 107 (±10%)
• Temperature–programmable • Pressure setting range: 0–100 psi • Full range auto scaling
cool-on-column (PCOC) • Total flow setting range: • Data acquisition rate: up to 200 Hz
• Electronic pneumatics control 0 to 200 mL/min N2
0 to 1,000 mL/Min H2 or He • Flows: air, 0 to 800; hydrogen,
(EPC)
0 to 100; make-up (He or N2),
• When capillary column dimen- • Available with LCO2 cyrogenic 0 to 100 mL/min
sions are entered, the actual flow cooling only
and pressure are shown as cali- • Lower temperature limit without Thermal Conductivity Detector (TCD)
brated digital readouts CO2: ambient plus 10 °C
• Maximum temperature: 375 °C
• Carrier gas selection provided • Lower temperature with LCO2
for: He, H2, N2 and Ar-CH4 • Minimum detectable: <400 pg
cyrogenic cooling: –30 °C (10 °C
propane/mL helium carrier (MDL
• Pressure setting range: less than oven)
can be affected by laboratory
0 to 100 psi • Available with Gerstel septumless environment.)
• Constant pressure, constant flow head or Merlin MicroSeal septum
• Linear dynamic range: 105 (± 5%)
modes head
• Flows: reference gas, 0 to 100;
• Three ramps for pressure or flow make-up (He, H2, N2),
PCOC
• Optional Merlin Microseal 1 to 12 mL/min
septum guarantees at least • Maximum temperature: 375 °C
2,000 injections • Three temperature program
• Septum purge set automatically ramps or oven track mode
2
FPD General Valving (optional)
• Available in single wavelength • Clock-time programming (24 h) One 6-port gas sampling valve or one
only 4-port liquid sampling valve can be
• Up to five methods stored, built- mounted in a heated compartment.
• 250 °C maximum operating in SERVICE method
temperature The valve can be connected to the
• One sequence column either directly or via the
• MDL: <20 pg S/s, <0.9 pg P/s with inlet. Valves may be run-time pro-
dodecanethiol/tributylphosphate • Run deviation log grammed. Actuating a sampling
mixture valve will START a run.
• Contact closure (48 VAC/VDC
• Selectivity: 105 gS/gC, 106 gP/gC 250 mA) with BCD input to con- Both the traditional actuator/rotary
3 4
trol external multiposition valve valves as well as piston/diaphragm
• Dynamic range: >10 S, 10 P (up to 16 positions).
with dodecanethiol/tributylphos- valves are available. The rotary
phate mixture • Internal valve control; 24 VDC, valves are suitable for most applica-
200 mA tions. The air-actuated diaphragm
valves have a much longer life and
Micro-ECD • Built-in power/control for the are easier to maintain; however,
Agilent 7683 autoinjector they are not suitable for ammonia,
• Equipped with hidden anode and
primary and secondary amines,
high velocity flows for
hydrazines, strong oxidizers, and
contamination resistance Data Communications alkaline solutions (pH >10).
• 400 °C maximum operating • RS-232-C: maximum Baud rate is The recommended operating tem-
temperature 57,600 perature and pressure ranges for the
• Makeup gas types: argon/5% • Analog outputs (1 mV, 1 V, 10 V) miniature VICI® diaphragm design
methane or nitrogen are: 50 to 150 °C and 20–300 psi.
• Remote start/stop The valves are actuated pneumati-
• Radioactive source: <15 mCi63 Ni
• LAN interface cally and require 40 to 50 psi air
• MDL: <0.008 pg/s lindane pressure.
• Dynamic range: >5 × 105 with The recommended operating tem-
lindane EPC perature range for the rotary 6-port
gas sampling valve is 50 to 200 °C
• Data acquisition rate: up to and the maximum sample pressure
50 Hz Standard
is 400 psi.
• Pressure setpoint increments:
0.01 psi The operating temperature range for
the rotary 4-port liquid sampling
• Temperature and pressure sen- valves is 70 °C to 175 °C. At 5000 psi,
sors compensate for ambient the maximum temperature is 75 °C
variation and altitude while at 1000 psi the maximum tem-
• All EPC setpoints are included in perature is 175 °C.
Local User Interface the method The gas sampling valve, rotary or
diaphragm, is supplied with a
• Six-button interface with two-line
Auxiliary (Aux) EPC 0.25-cc loop. Other loops (0.5, 1, and
display:
2 cc) are available separately.
- Status is on the top line; lists • Maximum number of modules:
one Liquid sample valves have a built-in
and messages are on the
sample loop (0.5 or 1.0-µL size).
bottom line.
• Maximum number of channels:
- ] ] buttons scroll through a three
list.
• Pressure increments: 0.01 psi
- LOAD button loads a method.
• Maximum pressure: 100 psig
- PREP RUN button prepares the
• Maximum pressure ramp rate
unit for manual injection.
(when configured as Inlet):
• START/STOP buttons control the 150 psi/min
sequence or method.
• Keyboard can be locked.
3
www.agilent.com/chem
Maintenance and Support
• Service method sets tempera-
tures and flows for routine main-
tenance (such as septum or liner
changing)
• Operating and service informa-
tion on CD-ROM. Includes multi-
media maintenance procedures
• Diagnostics built into the
(optional) handheld controller
for use by the operator:
- Inlet (leak) test
- Inlet vent trap (restriction) test
- FID jet (restriction) test
• Additional diagnostic tests built
in for hardware fault detection
Dimensions and Weight
• Height: 490 mm to FID cover top,
and 505 mm to top of valve box
• Width: 283 mm, or 333 mm if the
cryogenic option is installed
• Depth: 568 mm
• Weight: 29 kg (maximum)
Agilent shall not be liable for errors contained herein or
for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
Information, descriptions, and specifications in this
publication are subject to change without notice.
VICI® is a registered trademark of Valco Instruments
Co. Inc. and VICI AG.
© Agilent Technologies, Inc. 2010