Agilent 7200 User Manual
Agilent 7200 User Manual
7200 Accurate-Mass
Quadrupole
Time-of-Flight GC/MS
System
Operation Manual
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2012, 2014 Warranty Safety Notices
No part of this manual may be reproduced in
The material contained in this docu-
any form or by any means (including elec-
ment is provided “as is,” and is sub-
tronic storage and retrieval or translation
ject to being changed, without notice,
CAUTION
into a foreign language) without prior agree-
ment and written consent from Agilent in future editions. Further, to the max-
imum extent permitted by applicable A CAUTION notice denotes a haz-
Technologies, Inc. as governed by United
law, Agilent disclaims all warranties, ard. It calls attention to an operat-
States and international copyright laws.
either express or implied, with regard ing procedure, practice, or the like
Manual Part Number to this manual and any information that, if not correctly performed or
contained herein, including but not adhered to, could result in damage
G3850-90008 limited to the implied warranties of
merchantability and fitness for a par- to the product or loss of important
Edition ticular purpose. Agilent shall not be data. Do not proceed beyond a
First Edition, August 2014 liable for errors or for incidental or CAUTION notice until the indicated
consequential damages in connec- conditions are fully understood and
Printed in USA
tion with the furnishing, use, or per- met.
Agilent Technologies, Inc. formance of this document or of any
5301 Stevens Creek Blvd. information contained herein. Should
Santa Clara, CA 95051 Agilent and the user have a separate
written agreement with warranty WA R N I N G
terms covering the material in this
document that conflict with these A WARNING notice denotes a
terms, the warranty terms in the sep-
hazard. It calls attention to an
arate agreement shall control.
operating procedure, practice, or
Technology Licenses the like that, if not correctly per-
The hardware and/or software described in formed or adhered to, could result
this document are furnished under a license in personal injury or death. Do not
and may be used or copied only in accor- proceed beyond a WARNING
dance with the terms of such license.
notice until the indicated condi-
Restricted Rights Legend tions are fully understood and
met.
U.S. Government Restricted Rights. Soft-
ware and technical data rights granted to
the federal government include only those
rights customarily provided to end user cus-
tomers. Agilent provides this customary
commercial license in Software and
technical data pursuant to FAR 12.211
(Technical Data) and 12.212 (Computer Soft-
ware) and, for the Department of Defense,
DFARS 252.227-7015 (Technical Data -Com-
mercial Items) and DFARS 227.7202-3
(Rights in Commercial Computer Software
or Computer Software Documentation).
About This Manual
This manual contains information for operating and maintaining the Agilent
7200 Accurate-Mass Quadrupole Time-of-Flight GC/MS system.
1 “Introduction”
Chapter 1 describes general information about the 7200 Q-TOF GC/MS,
including a hardware description, and general safety warnings.
2 “Installing GC Columns”
Chapter 2 shows you how to prepare a capillary column for use with the MS,
install it in the GC oven, and connect it to the MS using the GC/MS interface.
6 “General Maintenance”
Chapter 6 describes general maintenance procedures for the 7200 Q-TOF
GC/MS.
Version B.01.05
January, 2013
Agilent Technologies
GC/GCMS Hardware User Manuals and Videos
Agilent Technologies
Disk 2/3
Insert Disk 1 into your DVD drive and follow the prompts.
GC/GCMS Hardware User
To install: Manuals and Videos
1. Insert disk into DVD drive
2. Follow the instructions on the screen.
3. If autoplay is not enabled, double-click
"index_xx.html" on the DVD
where xx is:
Version B.01.05 en for English Disk 2/3
ch for 中文
January, 2013 jp for 日本語
To install:
1. Insert disk into DVD drive
2. Follow the instructions on the screen.
© 2013 Agilent Technologies, Inc.3. If autoplay is not enabled, double-click
All rights reserved. "index_xx.html" on the DVD
S1
ch for 中文
jp for 日本語
S1
information onto the receiving computer.
See the Agilent 7200 Accurate Mass Q-TOF GC/MS System Quick Start
document (G6845-90014) for more details on how to install this information
on your computer.
2 Installing GC Columns
Columns 24
To Install a Capillary Column in a Split/Splitless Inlet 26
To Condition a Capillary Column 29
To Install a Capillary Column in the GC/MS Interface 30
6 General Maintenance
Before Starting 90
Maintaining the Vacuum System 94
Maintaining the Analyzer 95
To Disassemble the EI Ion Source 97
To Assemble the EI Ion Source 99
To Disassemble the CI Ion Source 101
To Assemble the CI Ion Source 103
To Clean the Ion Source 105
To Remove a Filament 109
To Install a Filament 111
1
Introduction
Abbreviations Used 10
The 7200 Accurate-Mass Quadrupole Time-of-Flight GC/MS System 12
Hardware Description 14
Important Safety Warnings 15
Safety and Regulatory Certifications 18
Intended Use 21
Cleaning/Recycling the Product 21
Liquid Spills 21
Moving or Storing the MS 21
Agilent Technologies 9
1 Introduction
Abbreviations Used
The abbreviations in Table 1 are used in discussing this product. They are
collected here for convenience.
Table 1 Abbreviations
Abbreviation Definition
AC Alternating current
CC Collision cell
CI Chemical ionization
DC Direct current
EI Electron impact
eV Electron volt
GC Gas chromatograph
id Inside diameter
MS Mass spectrometer
Abbreviation Definition
RF Radio frequency
TOF Time-of-flight
Physical description
The 7200 Q-TOF GC/MS is approximately 48 cm high, 71 cm wide, and 89 cm
deep excluding the flight tube and RIS handle. The flight tube extends 84 cm
over the top of the instrument. The RIS handle, when attached, extends 48 cm
from the front of the instrument.
The weight of the instrument mainframe is 152 kg. The attached foreline
(roughing) pump weighs an additional 21.5 kg.
The basic components of the instrument are: the frame/cover assemblies, the
vacuum system, the GC/MS interface, the removable ion source, the flight tube
electronics, the quadrupole mass filter, the collision cell, and the detector.
Vacuum gauge
The 7200 Q-TOF GC/MS is equipped with four ion vacuum gauges. The
MassHunter Workstation can be used to read the pressure (high vacuum) in
the vacuum manifold, the turbomolecular vacuum pump discharge, and the
flight tube.
Ionization modes
The G3851BA 7200 Accurate-Mass Q-TOF GC/MS comes standard with both an
EI source and a CI source.
For the CI system, a methane/isobutane gas purifier is provided and is
required. It removes oxygen, water, hydrocarbons, and sulfur compounds.
The MS CI system has been optimized to achieve the relatively high ion source
pressure required for CI while still maintaining high vacuum in the collision
cell, quadrupole, and TOF tube. Special seals along the flow path of the
reagent gas and very small openings in the ion source keep the ion source
gases in the ionization volume long enough for the appropriate reactions to
occur. The interface has special plumbing for reagent gas. A retractable
insulating seal fits onto the tip of the interface and is used for both CI and EI.
Switching back and forth between CI and EI ion sources takes less than
30 minutes with the new removable ion source (RIS). The RIS allows the
instrument to remain under vacuum, and provides a cooling chamber with N2
purge for rapid ion source cooling without venting the machine. This saves
hours in cycle time over the traditional unit.
Hardware Description
Figure 1 is an overview of a typical Agilent 7200 Accurate-Mass Q-TOF GC/MS
system.
Flight tube
ALS
7890 GC
Q-TOF
All these parts are shielded by covers. With the covers in place, it should be difficult
WA R N I N G
to accidentally make contact with dangerous voltages. Unless specifically
instructed to, never remove a cover unless the detector, inlet, and oven are turned
off.
If the power cord insulation is frayed or worn, the cord must be replaced. Contact
WA R N I N G
your Agilent service representative.
You MUST ensure the top thumbscrew on the front analyzer side plate and the top
WA R N I N G
thumbscrew on the rear analyzer side plate are both fastened finger-tight. Do not
over-tighten the thumbscrews; this can cause air leaks.
You MUST leave the collision cell chamber top plate shipping brackets fastened.
Do not remove the shipping brackets from the top plate for normal operation; they
secure the top plate in the event of an explosion.
Be careful when working behind the instrument. During cooldown cycles, the GC
WA R N I N G
emits hot exhaust that can cause burns.
The insulation around the inlets, detectors, valve box, and the insulation cups is
WA R N I N G
made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend
the following safety procedures: ventilate your work area; wear long sleeves,
gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation
in a sealed plastic bag in accordance with local regulations; wash your hands with
mild soap and cold water after handling the insulation.
The oil pan under the standard foreline pump can be a fire hazard
Oily rags, paper towels, and similar absorbents in the oil pan could ignite and
damage the pump and other parts of the MS.
The 7200 Q-TOF GC/MS is designed and manufactured under a quality system
registered to ISO 9001.
Information
The Agilent Technologies 7200 Accurate-Mass Q-TOF GC/MS meets the
following IEC (International Electrotechnical Commission) classifications:
Equipment Class I, Laboratory Equipment, Installation Category II, and
Pollution Degree 2.
This unit has been designed and tested in accordance with recognized safety
standards, and is designed for use indoors. If the instrument is used in a
manner not specified by the manufacturer, the protection provided by the
instrument may be impaired. Whenever the safety protection of the MS has
been compromised, disconnect the unit from all power sources and secure the
unit against unintended operation.
Refer servicing to qualified service personnel. Substituting parts or
performing any unauthorized modification to the instrument may result in a
safety hazard.
Symbols
Warnings in the manual or on the instrument must be observed during all
phases of operation, service, and repair of this instrument. Failure to comply
with these precautions violates safety standards of design and the intended
use of the instrument. Agilent Technologies assumes no liability for the
customer’s failure to comply with these requirements.
Electromagnetic compatibility
This device complies with the requirements of CISPR 11. Operation is subject
to the following two conditions:
• This device may not cause harmful interference.
• This device must accept any interference received, including interference
that may cause undesired operation.
If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try one or more of the following measures:
• Relocate the radio or antenna.
• Move the device away from the radio or television.
• Plug the device into a different electrical outlet, so that the device and the
radio or television are on separate electrical circuits.
• Ensure that all peripheral devices are also certified.
• Ensure that appropriate cables are used to connect the device to peripheral
equipment.
• Consult your equipment dealer, Agilent Technologies, or an experienced
technician for assistance.
Changes or modifications not expressly approved by Agilent Technologies
could void the user’s authority to operate the equipment.
Sound pressure
Sound pressure Lp < 70 dB according to EN 27779:1991 and
EN ISO 3744:1995.
Schalldruckpegel
Schalldruckpegel LP < 70 dB nach EN 27779:1991 und EN ISO 3744:1995.
Intended Use
Agilent products must only be used in the manner described in the Agilent
product user guides. Any other use may result in damage to the product or
personal injury. Agilent is not responsible for any damages caused, in whole or
in part, by improper use of the products, unauthorized alterations,
adjustments or modifications to the products, failure to comply with
procedures in Agilent product user guides, or use of the products in violation
of applicable laws, rules or regulations.
Liquid Spills
Do not spill liquids on the MS.
2
Installing GC Columns
Columns 24
To Install a Capillary Column in a Split/Splitless Inlet 26
To Condition a Capillary Column 29
To Install a Capillary Column in the GC/MS Interface 30
Before you can operate your GC/MS system, you must select, install, and
condition a GC column. This chapter shows you how to install and condition a
column.
Agilent Technologies 23
2 Installing GC Columns
Columns
Many types of GC columns can be used with the MS, but there are some
restrictions.
During tuning or data acquisition the rate of column flow into the MS should
not exceed the maximum recommended flow. Therefore, there are limits to
column length and flow. Exceeding recommended flow will result in
degradation of mass spectral and sensitivity performance.
Remember that column flows vary greatly with oven temperature. See “To
Calibrate the Column” on page 46 for instructions on how to measure actual
flow in your column. Use the flow calculator in the Agilent Instrument
Utilities software, and Table 2 to determine an acceptable column flow. For
expected column outlet flow pressures, use the values shown in Table 3 for EI
mode and Table 6 for CI mode.
* Total gas flow into the MS = column flow + collision cell gas flow + reagent gas flow (if applicable)
+ Agilent Quick Swap flow (if applicable)
† Expect degradation of spectral performance and sensitivity.
Conditioning columns
Conditioning a column before it is connected to the GC/MS interface is
essential.
A small portion of the capillary column stationary phase is often carried away
by the carrier gas. This is called column bleed. Column bleed deposits traces of
the stationary phase in the MS ion source. This decreases MS sensitivity and
makes cleaning the ion source necessary.
Column bleed is most common in new or poorly cross-linked columns. It is
much worse if there are traces of oxygen in the carrier gas when the column is
heated. To minimize column bleed, all capillary columns should be
conditioned before they are installed in the GC/MS interface.
Conditioning ferrules
Heating ferrules to their maximum expected operating temperature a few
times before they are installed can reduce chemical bleed from the ferrules.
Always wear safety glasses when handling capillary columns. Use care to avoid
WA R N I N G
puncturing your skin with the end of the column.
General Materials
• Gloves, clean
• Large (8650-0030)
• Small (8650-0029)
• Metric ruler
• Septum (may be old, used inlet septum)
• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Self Tightening column nut materials (Recommended)
• Self Tightening column nut (5190-6194)
• Ferrules, Vespel for Self Tightening column nut
• 5181-3323 Ferrule, 0.4mm VG 0.1-0.25 col 10 pk
• 5062-3514 Ferrule, 0.5mm VG 0.32 col 10 pk
Traditional Universal column nut materials
• 5181-8830 Universal column nut, 2 pk
• Ferrules, Vespel for universal column nut
• 5181-3323 Ferrule, 0.4mm VG 0.1-0.25 col 10 pk
• 5062-3514 Ferrule, 0.5mm VG 0.32 col 10 pk
To install columns in other types of inlets, refer to your gas chromatograph
user information.
Procedure
The GC operates at high temperatures. In order to avoid burns, do not touch any
WA R N I N G
parts of the GC until you are sure they are cool.
Always wear clean gloves while handling any parts that go inside the GC or analyzer
CAUTION
chambers.
2 Wearing clean gloves, slide a septum, column nut, and conditioned ferrule
onto the free end of the column. Figure 2 shows a traditional inlet column
nut but the nut is similarly positioned for a self tightening column nut.
The tapered end of the ferrule should point away from the column nut for
a column attachment to an inlet.
Capillary column
Column cutter
Ferrule
Septum
Figure 2 Preparing a capillary column for installation (standard inlet column nut shown)
3 Use the column cutter to score the column 2 cm from the end.
4 While holding the column against the column cutter with your thumb,
break the column against the edge of the column cutter.
5 Inspect the end for jagged edges or burrs. If the break is not clean and
even, repeat steps 3 and 4.
6 Wipe the outside of the free end of the column with a lint-free cloth
moistened with methanol.
7 Position the septum under the column nut so that the column extends
4 to 6 mm past the end of the ferrule (Figure 3).
Never use a wrench to tighten a self tightening nut. This may result in damage to the
CAUTION
nut or instrument.
11 For the Self Tightening column nut, finger-tighten until it begins to grip
the column then turn an additional 30 to 60 degrees until tight.
For the traditional column nut tighten the nut an additional 1/4 to
1/2 turn. The column should not slide with a gentle tug.
12 Start carrier gas flow.
13 Verify flow by submerging the free end of the column in isopropanol. Look
for bubbles.
Insulation cup
Reducing nut
Capillary column 4 to 6 mm
Septum
Materials needed
• Carrier gas, (99.9995% pure or better)
• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
The GC operates under high temperatures. To avoid burns, do not touch any GC parts
WA R N I N G
unless you are certain they are cool.
Procedure
1 Install the column in the GC inlet. (See “To Install a Capillary Column in a
Split/Splitless Inlet” on page 26.)
2 Set a minimum velocity of 30 cm/s, or as recommended by the column
manufacturer. Allow gas to flow through the column at room temperature
for 15 to 30 minutes to remove air.
3 Program the oven from room temperature to the maximum temperature
limit for the column.
4 Increase the temperature at a rate of 10 to 15 °C/min.
5 Hold at the maximum temperature for 30 minutes.
Never exceed the maximum column temperature, either in the GC/MS interface, the
CAUTION
GC oven, or the inlet.
This procedure is for the installation of a capillary column directly into the
analyzer using the Agilent recommended self-tightening column nut.
Materials needed
• Column cutter, ceramic (5181-8836) or diamond (5183-4620)
• Flashlight
• Magnifying loupe
• Gloves, clean
• Large (8650-0030)
• Small (8650-0029)
• Septum (may be old, used inlet septum)
• Safety glasses
• Column install tool (G3850-60014)
Self Tightening column nut materials (Recommended)
• Self Tightening column nut (5190-5233)
• Ferrules, Vespel for Self Tightening column nut
• 5181-3323 Ferrule, 0.4 mm VG 0.1-0.25 col 10 pk
• 5062-3514 Ferrule, 0.5 mm VG 0.32 col 10 pk
Traditional MS interface nut materials
• MS Interface column nut (05988-20066)
• Ferrules, Vespel for universal column nut
• 5181-3323 Ferrule, 0.4 mm VG 0.1-0.25 col 10 pk
• 5062-3514 Ferrule, 0.5 mm VG 0.32 col 10 pk
Procedure
Always wear clean gloves while handling any parts that go inside the GC or the
CAUTION
analyzer chambers.
The analyzer, GC/MS interface, and other components in the analyzer chamber
WA R N I N G
operate at very high temperatures. Do not touch any part until you are sure it is cool.
Dangerous voltages exist inside the analyzer chamber, which can result in fatal
WA R N I N G
injury. Do not open the analyzer chamber door for any reason. If access is ever
required, trained service personnel must first disconnect the instrument from the
building power source.
2 Vent the MS. To vent the MS, see “To Vent the MS” on page 56.
The GC operates under high temperatures. To avoid burns, do not touch any GC parts
WA R N I N G
unless you are certain they are cool.
8 Wipe the outside of the free end of the column with a lint-free cloth
moistened with methanol.
9 Slide the column so that the end projects 4–5 mm past the end of the
column install pre-swaging tool.
10 For the Self Tightening column nut, finger-tighten until it begins to grip
the column then turn an additional 30 to 60 degrees until tight.
For the traditional column nut, tighten the nut an additional 1/4 to
1/2 turn. The column should not slide with a gentle tug.
11 Loosen the column nut and remove the column from the tool.
12 Slide the column into the GC/MS interface.
13 Hand-tighten the nut. Ensure the position of the column does not change
as you tighten the nut.
14 Check the GC oven to be sure that the column does not touch the oven
walls.
15 Check the nut’s tightness after one or two heat cycles; retighten as
appropriate.
3
Operating in Electron Impact (EI) Mode
Operating the MS from the Data System 34
The GC/MS Interface 35
Before You Turn On the MS 37
Pumping Down 38
Controlling Temperatures 38
Controlling Column Flow 38
Controlling Collision Cell Flow 39
Venting the MS 39
Typical Vacuum Pressures in EI Mode 40
To Set Monitors for Temperature and Vacuum Status 41
To Set the MS Analyzer Temperature 43
To Set the GC/MS Interface Temperature from the MassHunter
Workstation 45
To Calibrate the Column 46
To Configure the Collision Cell Gas 48
To Set the Collision Cell Gas Flow Rate 49
To Autotune the MS for EI Mode 50
To Remove the Upper RIS Cover 52
To Open the Analyzer Cover for Access to the Analyzer Sideplate 53
To Pump Down the MS 54
To Vent the MS 56
This chapter will explain how to perform some routine operating procedures
for the 7200 Accurate-Mass Q-TOF GC/MS in EI mode.
Agilent Technologies 33
3 Operating in Electron Impact (EI) Mode
See also
“To Install a Capillary Column in the GC/MS Interface” on page 30.
The GC/MS interface operates at high temperatures. If you touch it when it is hot, it
WA R N I N G
will burn you.
Spring-loaded tip
Column nut
seal entrance to
ion source
MS GC oven
IRM calibrant in
The exhaust from the foreline pump contains solvents and the chemicals you are
WA R N I N G
analyzing. The standard foreline pump, also exhausts traces of pump oil. If you are
using toxic solvents or analyzing toxic chemicals, attach a hose (11-mm id) to the
oil mist filter to take the foreline pump exhaust outside or to a fume (exhaust) hood.
Be sure to comply with local regulations. The oil mist filter supplied with the
standard pump stops only pump oil. It does not trap or filter out toxic chemicals.
Pumping Down
Pumping down is only required if the analyzer was vented. It is not required
for changing the ion source.
The data system helps you pump down the MS. The process is mostly
automated. When you turn on the main power switch (while pressing on the
analyzer sideplate), the MS pumps down by itself. The data system software
monitors and displays system status during pumpdown. When the pressure is
low enough, the program turns on the ion source and mass filter heaters and
prompts you to turn on the GC/MS interface heater. The MS will shut down if
it cannot pump down correctly.
Using the MS monitors, the data system can display:
• Motor speed for each MS turbo pump
• Analyzer chamber pressure (vacuum)
• Quadrupole and ion source temperatures
Controlling Temperatures
MS temperatures are controlled through the data system. The MS has
independent heaters and temperature sensors for the ion source and the
quadrupole mass filter. You can adjust the setpoints and view these
temperatures from the data system
The GC/MS interface heater is powered and controlled by the Thermal Aux #2
heated zone of the GC. The GC/MS interface temperature can be set and
monitored from the data system or from the GC keypad.
The MS can be used to measure actual column flow. You inject a small amount
of air or other unretained chemical and time how long it takes to reach the MS.
With this time measurement, you can calculate the column flow. See “To
Calibrate the Column” on page 46.
Venting the MS
A program in the data system automates the venting process. It turns off the
GC and MS heaters and the turbo pump at the correct time.
The Fast Vent feature loads a user-defined GC method that sets an oven
temperature, turns off the transfer line heater and sets the column flow to a
user predefined level.
The MS will be damaged by incorrect venting. A turbo pump will be damaged if
it is vented while spinning at more than 50% of its normal operating speed.
Never vent the MS by allowing air in through either end of the foreline hose. Use the
CAUTION
vent valve or remove the column nut and column.
Do not exceed the maximum recommended total gas flow. See Table 2.
Column Flow
(mL/min) Rough Vac Quad Vac TOF Vac Rough Vac Quad Vac TOF Vac
0.5 1.16 * 10-1 3.41 * 10-5 4.20* 10-7 8.29 * 10-2 3.17 * 10-7 2.15 * 10-7
0.7 1.18 * 10-1 3.41 * 10-5 4.20 * 10-7 8.67 * 10-2 3.45 * 10-7 2.15 * 10-7
1 1.22 * 10-1 3.43 * 10-5 4.22 * 10-7 9.17 * 10-2 3.79 * 10-7 2.16 * 10-7
1.2 1.24 * 10-1 3.43 * 10-5 4.22 * 10-7 9.49 * 10-2 4.06 * 10-7 2.16 * 10-7
2 1.32 * 10-1 3.44 * 10-5 4.22 * 10-7 1.07 * 10-1 5.17 * 10-7 2.17 * 10-7
3 1.42 * 10-1 3.44 * 10-5 4.25 * 10-7 1.19 * 10-1 6.51 * 10-7 2.20 * 10-7
4 1.50 * 10-1 3.46 * 10-5 4.27 * 10-7 1.29 * 10-1 7.84 * 10-7 2.21 * 10-7
If the pressure is consistently higher than those listed, refer to the online help
in the MassHunter Workstation software for information on troubleshooting
air leaks and other vacuum problems.
Procedure
1 Select Instrument > Edit Monitors to display the Select Monitors dialog box.
See Figure 5.
2 In the Available Monitors column, select a monitor and click the Add button
to move the selected monitor to the Selected Monitors column.
3 Select any other monitors you want and add them to the Selected Monitors
column.
4 Click OK. The new monitors will be stacked on top of each other in the
lower right corner of the Instrument Control window.
5 Select Window > Arrange Monitors, or click and drag each monitor to the
desired position. See Figure 6 for one way of arranging the monitors.
6 To make the new settings part of the method, select Save from the Method
menu.
Procedure
1 In Instrument Control panel, select the MS Tune icon to display the Tune dialog
box. Select the Manual Tune tab then select the Ion Source tab to display the
ion source parameters.
2 Enter the temperature setpoint in the Source Temp field. See Table 4 for
recommended setpoints.
3 Select the Quadrupole tab to display the MS parameters.
4 Enter the temperature setpoint in the Quad Temp field. See Table 4 for
recommended setpoints.
5 Select the Files and Reports tab then click the Save button to save the tune
file with these changes.
Zone EI operation
MS Source 230 °C
MS Quad 150 °C
The GC/MS interface, ion source, and the MS quadrupole heated zones
interact. The analyzer heater may not be able to accurately control
temperature if the setpoint for one zone is much different from that of an
adjacent zone.
The software will not allow you to exceed 200 °C for the quadrupole, or 350 °C for
WA R N I N G
the ion source.
Procedure
1 Select Instrument > GC Parameters from the Instrument Control panel.
2 Click the Aux Heaters icon to edit the interface temperature (Figure 8). This
example has the GC/MS interface temperature configured as Thermal Aux 2.
Ensure that the carrier gas is turned on and the column has been purged of air before
CAUTION
heating the GC/MS interface or the GC oven.
When setting the GC/MS interface temperature, never exceed the maximum for your
CAUTION
column.
3 Click the heater On check box and enter the setpoint in the Value °C
column. The typical setpoint is 280 °C. The limits are 0 °C to 400 °C. A
setpoint below ambient temperature turns off the interface heater.
4 Click Apply to download setpoints or click OK to download setpoints and
close the window.
5 To make the new settings part of the method, select Save from the Method
menu.
Procedure
1 Set Data Acquisition for splitless manual injection and set up a real time
plot to monitor m/z 28.
2 Press [Prep Run] on the GC keypad.
3 Inject 1 µL of air into the GC inlet and press [Start Run]
4 Wait until a peak elutes at m/z 28. Note the retention time.
11 When the new column length appears, click OK to save the changes.
12 Click OK on the Calibrate Columns screen to save the calibration.
select .
2 Click the Configuration icon and select the Modules tab to display the screen.
See Figure 10.
3 From the Collision Cell EPC drop-down menu, select N2 as the collision cell
gas.
4 Click OK to save the configuration.
Procedure
1 Set the system to the same conditions (GC oven temperature and column
flow, and MS analyzer temperature) that will be used for data acquisition.
2 In the Instrument Control panel, click the MS Tune icon to display the
GC/Q-TOF Tune dialog box.
3 The current tune file is displayed in the upper left corner of the
GC/Q-TOF Tune dialog box. Verify that the correct tune file is loaded.
4 If necessary, load a new tune file by clicking on the Files and Report tab then
click on the Load button in the Tune File area. Select a tune file and click the
OK button.
The tune file must match the type of ion source in the analyzer. If you are
using an EI ion source, select a tune file created for an EI ion source.
5 Click the Autotune tab and select the Tune from default settings check box if
you are restarting the system after a system vent, major servicing, or a
power outage. If you clear the Tune from default settings box, the autotune
process starts using the previous tune values.
6 Select Save tune file when done check box to save the new tune parameters
generated by the autotune. Do not select this item if you want to review
the autotune report before saving the newly generated tune parameters.
7 Select the Print autotune report check box to automatically print a tune
report.
8 Click the Autotune button to start the autotune. The Status line displays the
current step in the autotune process and the plot of the tuned parameter
for that step is shown in the top graph. If specified above, at the
completion of the autotune, a Tune Report is printed.
To stop the autotune before it completes the automatic parameter
selection, click the Abort Tune button. The parameters from the last
successful autotune are used.
9 Review the Tune Report. If the results are acceptable and you did not
select the Save tune file when done check box, save the autotune by clicking
the Files and Report tab, then click the Save button.
See the manuals or online help provided with your MassHunter Data
Acquisition Workstation software for additional information about tuning.
Figure 12 EI Autotune
The upper RIS cover is in front of the instrument (Figure 13). Remove it to
access the latches on the analyzer cover, attach the RIS probe, or access the EI
calibration vial. Do not remove the upper RIS cover for any other reasons.
Procedure
1 Grasp the cover by the outside corners.
2 Pull the cover straight up, and remove it from the instrument.
Open the analyzer cover (Figure 14) to access the analyzer sideplate only.
This is necessary during pump down or to access the analyzer sideplate
thumbscrews. You should not open the analyzer cover for any other reason.
Procedure
1 Remove the upper RIS cover to expose the analyzer cover latches.
See “To Remove the Upper RIS Cover” on page 52.
2 Open the latches on the analyzer cover at the front of the instrument.
3 Swing the analyzer cover open.
Analyzer cover
Figure 14 Covers
Ensure your MS meets all the conditions listed in the introduction to this chapter
WA R N I N G
before starting up and pumping down the MS. Failure to do so can result in personal
injury.
Dangerous voltages exist inside the analyzer chamber, which can result in fatal
WA R N I N G
injury. Do not open the analyzer chamber door for any reason. If access is ever
required, trained service personnel must first disconnect the instrument from the
building power source.
Procedure
1 Open the analyzer cover to access the analyzer quad driver board. See “To
Open the Analyzer Cover for Access to the Analyzer Sideplate” on page 53.
2 Plug the power cord into a grounded electrical outlet.
3 Turn on the Q-TOF power switch (Figure 14).
4 Press lightly on the metal box on the quad driver board to ensure a correct
seal.
Do not push on the filament board safety cover while pressing on the analyzer board.
CAUTION
This cover was not designed to withstand this type of pressure.
The foreline pump will make a gurgling noise. This noise should stop
within a minute. If the noise continues, there is a large air leak in your
system, probably at the sideplate seal or the interface column nut.
5 Start the MassHunter Data Acquisition program. If the Q-TOF was
configured for both an EI and a CI ion source, you are prompted for the
ion source type that is currently installed. Click on an EI or CI ion source
type if prompted.
6 Select the MS Tune icon from the Instrument Control panel.
7 Select the Vacuum Control tab.
Do not turn on any GC heated zones until carrier gas flow is on. Heating a column with
CAUTION
no carrier gas flow will damage the column.
To Vent the MS
Procedure
Be sure the GC oven and the GC/MS interface are cool before turning off carrier gas
CAUTION
flow.
6 Turn off the MS by pressing the power switch. (See Figure 1.)
7 Unplug the MS power cord.
Allow the analyzer to cool to near room temperature before touching it.
WA R N I N G
When the MS is vented, do not put the Workstation into Instrument Control view.
CAUTION
Doing so will turn on the interface heater, which can damage the column.
4
Operating in Chemical Ionization (CI)
Mode
Setting Up Your MS to Operate in CI Mode 58
The GC/MS Interface 59
Operating the CI MS 61
High Vacuum Pressure in CI Mode 62
Other Reagent Gases 63
CI Autotune 65
The Flow Control Module 66
To Switch from the EI Mode to CI Mode 67
To Operate the Reagent Gas Flow Control Module 69
To Set a Reagent Gas Flow 71
To Perform a CI Autotune 72
This chapter provides information and instructions for operating the 7200
Q-TOF GC/MS system in Chemical Ionization (CI) mode. Most of the
information in the preceding chapter is also relevant.
Most of the material is related to methane chemical ionization but one section
discusses the use of other reagent gases.
The software contains instructions for setting the reagent gas flow and for
performing CI autotunes. Autotunes are provided for positive CI (PCI) with
methane reagent gas and for negative CI (NCI) with any reagent gas.
Agilent Technologies 57
4 Operating in Chemical Ionization (CI) Mode
See Also
“To Install a Capillary Column in the GC/MS Interface” on page 30.
Never exceed the maximum column temperature, either in the GC/MS interface, the
CAUTION
GC oven, or the inlet.
The GC/MS interface operates at high temperatures. If you touch it when it is hot, it
WA R N I N G
will burn you.
Column nut
Spring-loaded tip
seal entrance to
ion source
MS GC oven
Operating the CI MS
Operating your GC/MS in the CI mode is slightly more complicated than
operating in the EI mode. After tuning, gas flow, temperatures (Table 5), and
electron energy may need to be optimized for your specific analyte.
MFC (%) Rough Pump Quadrupole Flight Tube Rough Pump Quadrupole Flight Tube
(mTorr) (Torr) (Torr) (mTorr) (Torr) (Torr)
Do not use nitrous oxide as a reagent gas. It radically shortens the life span of the
CAUTION
filament.
Isobutane CI
Isobutane (C4H10) is commonly used for chemical ionization when less
fragmentation is desired in the chemical ionization spectrum. This is because
the proton affinity of isobutane is higher than that of methane; hence less
energy is transferred in the ionization reaction.
Addition and proton transfer are the ionization mechanisms most often
associated with isobutane. The sample itself influences which mechanism
dominates.
Ammonia CI
Ammonia (NH3) is commonly used for chemical ionization when less
fragmentation is desired in the chemical ionization spectrum. This is because
the proton affinity of ammonia is higher than that of methane; hence less
energy is transferred in the ionization reaction.
Because many compounds of interest have insufficient proton affinities,
ammonia chemical-ionization spectra often result from the addition of NH4+
and then, in some cases, from the subsequent loss of water. Ammonia reagent
ion spectra have principal ions at m/z 18, 35, and 52, corresponding to NH4+,
NH4(NH3)+, and NH4(NH3)2+.
Use of ammonia affects the maintenance requirements of the MS. See Chapter 6,
CAUTION
“General Maintenance”” for more information.
The pressure of the ammonia supply must be less than 5 psig. Higher pressures can
CAUTION
result in ammonia condensing from a gas to a liquid.
Always keep the ammonia tank in an upright position, below the level of the flow
module. Coil the ammonia supply tubing into several vertical loops by wrapping the
tubing around a can or bottle. This will help keep any liquid ammonia out of the flow
module.
Ammonia tends to break down vacuum pump fluids and seals. Ammonia CI
makes more frequent vacuum system maintenance necessary. (See the Agilent
7200 Q-TOF GC/MS Troubleshooting and Maintenance Manual.)
When running ammonia for 5 or more hours a day, the foreline pump must be ballasted
CAUTION
(flushed with air) for at least 1 hour a day to minimize damage to pump seals. Always
purge the MS with methane after flowing ammonia.
Carbon dioxide CI
Carbon dioxide is often used as a reagent gas for CI. It has obvious advantages
of availability and safety.
CI Autotune
After the reagent gas flow is set, the lenses and electronics of the MS should be
tuned which is a good starting point for developing your application settings.
Typical CI application settings are shown inTable 7.
Perfluoro-5,8-dimethyl-3,6,9-trioxidodecane (PFDTD) is used as the calibrant.
Instead of flooding the entire vacuum chamber, the PFDTD is introduced
directly into the ionization chamber through the GC/MS interface by means of
the gas flow control module. There are no tune performance criteria. If CI
autotune completes, it passes.
Always verify MS performance in EI before switching to CI operation.
CAUTION
Always wear clean gloves while touching the analyzer or any other parts that go inside
CAUTION
the analyzer chamber.
Procedure
1 Change the source. See “To Change the Ion Source” on page 78.
2 Load a suitable PCI or NCI method for use with the CI source.
3 Click the MS Tune icon in the Instrument Control panel to display the Tune
dialog box and select the Autotune tab.
The method selects PCI or NCI autotune and the correct reagent gas.
4 Select the Tune from default settings check box because you have changed the
ion source.
5 Select Print autotune report check box to automatically print a tune report.
6 Click the Autotune button to start the autotune. At the completion of the
autotune, a Tune Report is printed.
7 Review the Tune Report. If the results are acceptable, save the autotune by
clicking the Files and Report tab, then click the Save button.
Procedure
1 In Instrument Control panel, select the MS Tune icon to display the GC/Q-TOF
Tune dialog box. Select the Manual Tune tab then select the Ion Source tab to
display the ion source parameters.
2 Use the parameters in the CI Reagent Gas Control area to control reagent gas
flow.
CI Reagent - Selects the reagent gas from the dropdown menu.
CI Gas Flow - Enter percent of maximum volumetric flow for the selected
reagent gas. The actual % transmitted by the flow controller is shown next
to this entry. 20% is a good value for PCI and 40% is a good value for NCI.
Set button - Opens the selected reagent gas supply valve and controls the
reagent gas flow to the entered setpoint.
Off button - Turns off the reagent gas flow.
Purge button - Opens the selected reagent gas valve for 6 minutes to clear
the system of unwanted compounds.
Pumpout button - Closes the reagent gas valve for 4 minutes and evacuates
the system of reagent gases. At the end of the pumpout time the selected
reagent gas valve opens.
After the system has been switched from EI to CI mode, or vented for any other reason,
CAUTION
the MS must be baked out for at least 2 hours before tuning.
Continuing with CI autotune if the MS has an air leak or large amounts of water will
CAUTION
result in severe ion source contamination. If this happens, you will need to vent the
MS and clean the ion source.
Procedure
1 In Instrument Control panel, select the MS Tune icon to display the GC/Q-TOF
Tune dialog box. Select the Manual Tune tab then select the Ion Source tab to
display the ion source parameters.
2 In the CI Reagent Gas Control area, select the CI reagent gas you are using.
3 Enter the reagent gas flow setpoint in the CI Gas Flow field. This value is
entered as a percentage of maximum flow rate. The recommended flow is
20% for a PCI source and 40% for an NCI source.
4 Click the Set button. The Flow Set indication is displayed.
The reagent gas is flowing into the ion source at the rate displayed next to
the setpoint.
5 Click the Files and Reports tab, then click the Save button to save your
changes to the currently loaded tune file.
To Perform a CI Autotune
Avoid tuning more often than is absolutely necessary; this will minimize PFDTD
CAUTION
background noise and help prevent ion source contamination.
Procedure
1 Verify that the MS performs correctly in EI mode first.
2 Click the MS Tune icon in the Instrument Control panel to display the
GC/Q-TOF Tune dialog box.
3 If necessary, load a new tune file by clicking on the Files and Report tab then
click the Load button in the Tune File area. Select a tune file and click the OK
button.
The tune file must match the type of ion source in the analyzer. For a CI
ion source select a tune file created for a positive or negative CI source.
4 Click the Autotune tab and select PCI source for a positive CI source or
NCI source for a negative CI source.
5 Select the reagent gas you are using from the dropdown menu in the
CI Reagent Gas area.
6 If a log file and associated data files of the tune are required, click the Files
and Reports tab and in the Log Files section, click on the Browse button to
create a directory and files for the logs. Click on the required log and data
files check boxes.
7 Click on the Manual Tune tab and select the Ion source tab. In the CI Reagent
Gas section, select the reagent gas and then enter a CI Gas Flow rate of 20%
for a PCI source or 40% for an NCI source. Click on the Autotune tab to
return to Autotune.
8 Select the Tune from default settings check box if you are restarting the
system after a system vent, major servicing, or a power outage. If you clear
the Tune from default settings box, the Autotune process starts using
previous tune values.
9 Select Save tune file when done check box to save the new tune parameters
generated by the autotune. Do not select this item if you want to review
the autotune report before saving the newly generated tune parameters.
10 Select the Print autotune report check box to automatically print a tune
report.
11 Click the Autotune button to start the autotune. The Status line displays the
current step in the autotune process and the plot of the tuned parameter
for that step is shown in the top graph. If specified above, at the
completion of the autotune a Tune Report is printed.
To stop the autotune before it completes the autotune parameter selection,
click the Abort Autotune button. The parameters from the last successful
autotune are used.
12 Review the Tune Report. If the results are acceptable and you did not
select the Save tune file when done check box, save the autotune by clicking
the Files and Report tab, then click the Save button.
5
Using the Removable Ion Source
The Removable Ion Source 76
To Change the Ion Source 78
To Install the RIS Probe Extraction Tool 80
To Remove the Source from the Analyzer Chamber 82
To Change the Ion Source on the RIS Probe Extraction Tool 84
To Install the Ion Source in the Analyzer Chamber 86
To Remove the RIS Probe Extraction Tool from the Instrument 87
This chapter will explain how to use the features of the removable ion source.
Agilent Technologies 75
5 Using the Removable Ion Source
The software and firmware are revised periodically. If the steps in these procedures do
CAUTION
not match your MassHunter Workstation software, refer to the manuals and online
help supplied with the software for more information.
RIS handle
RIS bayonet
Procedure
1 From MassHunter Data Acquisition software control view, click the MS
Tune icon to display the GC/Q-TOF Tune dialog box.
2 Click the Removable Ion Source tab in the GC/Q-TOF Tune screen. (Figure 19)
If the ion source is not fully installed, it may interfere with the gate valve as it closes.
CAUTION
Damage to the RIS, the valve, or the transfer line may occur.
14 Click Next to close the gate valve and vent the RIS cooling chamber.
15 Remove the RIS probe extraction tool and replace the RIS chamber cover.
See “To Remove the RIS Probe Extraction Tool from the Instrument” on
page 87.
16 Click Next to pump down the chamber behind the RIS chamber cover.
17 When prompted, click Finish.
Procedure
1 When prompted by MassHunter, remove the upper RIS cover (see “To
Remove the Upper RIS Cover” on page 52).
2 Open the latch on the right side of the RIS cover door.
3 Swing the door open.
4 Lift the RIS cover door straight up to free it from the hinges, and remove.
Do not leave the instrument with the RIS cover off. If you remove the cover, you must
CAUTION
install the RIS probe as soon as possible to minimize contamination in the instrument.
5 Holding the RIS probe extraction tool by the handle, position the RIS
cooling chamber up against the instrument RIS chamber. Hold the handle
perpendicular to the front of the instrument so that the outside edge of the
cooling chamber is flush with the RIS chamber opening on the front
instrument.
6 Line up the pins on the probe hinge with the slots on the instrument
hinge.
7 Lower the pins into the hinge.
8 Close the cooling chamber door, and secure the door with the latch on the
right side.
9 Return to step 5 in the “To Change the Ion Source” on page 78.
Procedure
1 Holding the RIS handle, turn the alignment guide, which is the white line
on the RIS handle, to the 2 o’clock position from the top of the handle.
This ensures that the RIS bayonet is aligned properly to attach to the
source.
RIS
Alignment
guide
(white line on
RIS handle)
While the RIS handle is fully extended, do not bump or push it from the side. Doing so
CAUTION
will break the seal and allow air into the cooling chamber.
Do not try to guide the ion source in manually. Push the RIS handle gently and allow
CAUTION
the RIS bayonet to slide on its own. The RIS bayonet will slide in straight if it is allowed
to move naturally.
2 Keeping the RIS handle in position, push the RIS handle gently and allow
the RIS bayonet to extend into the analyzer chamber. Ensure the RIS
handle extends fully, to ensure that the RIS bayonet seats in the holes on
the ion source body properly.
3 When the ion source is properly seated, turn the RIS handle counter
clockwise until the alignment guide is in the 12 o’clock position, to lock
the ion source in place.
4 Pull the RIS handle toward you, keeping the alignment guide at the
12 o’clock position, to retract the ion source into the RIS cooling chamber.
5 Return to step 8 in the “To Change the Ion Source” on page 78.
If not done correctly, source maintenance can introduce contaminants into the MS.
CAUTION
Always wear clean rubber gloves when handling the source.
1
2
Undo the latch on the right edge of the RIS cooling chamber.
Swing open the RIS cooling chamber until the RIS handle is parallel with
the front of the instrument.
3 Lock the RIS cooling chamber in the open position by opening it enough to
engage the locking pin on the hinge.
The 7200 Q-TOF GC/MS operates at high temperatures. Ensure the source is cooled
WA R N I N G
to a safe temperature before attempting to remove it from the RIS cooling chamber.
4 Grasp the RIS handle with one hand, and the ion source body with the
other hand.
5 While holding the ion source steady, turn the RIS handle counter
clockwise so that the alignment guide is in the 2 o’clock position. This
guides the pin on the RIS bayonet out of the track on the ion source body.
6 Gently pull the ion source free from the RIS bayonet and set aside.
7 Remove the new ion source from its storage container. The ion source
remains connected to the lid. Twist the lid clockwise to separate the ion
source from the lid.
8 Turn the RIS handle so that the alignment guide is in the 2 o’clock
position.
9 Guide the pin on the RIS bayonet into the tracks on the end of the ion
source body.
10 While holding the ion source securely, turn the RIS handle counter
clockwise so that the alignment guide is in the 12 o’clock position, to lock
the ion source securely on the bayonet.
11 Retract the RIS handle fully so that the ion source is located completely in
the cooling chamber.
12 Raise the locking pin on the RIS cooling chamber hinge and swing it shut.
13 Secure the RIS cooling chamber to the instrument with the latch on the
right side of the door.
14 Return to step 10 in the “To Change the Ion Source” on page 78.
Take care when walking past the front of the instrument when the RIS handle is fully
CAUTION
extended, to avoid damage to the instrument. When closed, the RIS handle extends 48
cm (19 inches) out from the front of the instrument.
Do not try to guide the ion source in manually. Push the RIS handle gently and allow
CAUTION
the RIS bayonet to slide on its own. The RIS bayonet will slide in straight if it is allowed
to move naturally.
Procedure
1 Ensure the alignment guide on the RIS handle is in the 12 o’clock position.
Push the RIS handle gently towards the analyzer chamber and allow it to
extend into the analyzer chamber. This will seat the ion source securely
into its socket.
2 Once the RIS handle is fully extended, turn the RIS handle clockwise until
the alignment guide is in the 2 o’clock position to release the ion source
body from the RIS bayonet.
3 Pull the RIS handle toward you until it is fully retracted into the RIS
cooling chamber.
4 Return to step 13 in the “To Change the Ion Source” on page 78.
The 7200 Q-TOF GC/MS operates at high temperatures. Ensure the RIS cooling
WA R N I N G
chamber is at a safe temperature before attempting to remove it from the
instrument.
1
2
Undo the latch on the right side of the RIS probe extraction tool and open
the door.
Lift the RIS probe extraction tool up to release it from its hinges.
3 Replace the RIS probe extraction tool with the RIS chamber door.
4 Line up the hinge pins on the RIS chamber door with the slots on the
instrument hinge, and lower the pins into the hinge.
5 Swing the RIS chamber door closed and secure the door with the latch on
the right side.
6 Return to step 16 in the “To Change the Ion Source” on page 78.
6
General Maintenance
Before Starting 90
Maintaining the Vacuum System 94
Maintaining the Analyzer 95
To Disassemble the EI Ion Source 97
To Assemble the EI Ion Source 99
To Disassemble the CI Ion Source 101
To Assemble the CI Ion Source 103
To Clean the Ion Source 105
To Remove a Filament 109
To Install a Filament 111
Agilent Technologies 89
6 General Maintenance
Before Starting
You can perform much of the maintenance required by your MS. For your
safety, read all of the information in this introduction before performing any
maintenance tasks.
Scheduled maintenance
Common maintenance tasks are listed in Table 9. Performing these tasks
when scheduled can reduce operating problems, prolong system life, and
reduce overall operating costs.
Keep a record of system performance (tune reports) and maintenance
operations performed. This makes it easier to identify variations from normal
operation and to take corrective action.
Tune the MS X
* Or as needed.
Some procedures in this chapter require access to the inside of the MS while
the power switch is on. Do not remove any of the electronics safety covers in
any of these procedures. To reduce the risk of electric shock, follow the
procedures carefully.
Dangerous temperatures
Many parts in the MS operate at, or reach, temperatures high enough to cause
serious burns. These parts include, but are not limited to:
• GC/MS interface
• Analyzer parts
• Vacuum pumps
The foreline pump can cause burns if touched when operating.
WA R N I N G
Never touch these parts while your MS is on. After the MS is turned off, give these
WA R N I N G
parts enough time to cool before handling them.
The GC/MS interface heater is powered by a thermal zone on the GC. The interface
WA R N I N G
heater can be on, and at a dangerously high temperature, even though the MS is off.
The GC/MS interface is well insulated. Even after it is turned off, it cools very
slowly.
The GC inlets and GC oven also operate at very high temperatures. Use the
same caution around these parts. See the documentation supplied with your
GC for more information.
Chemical residue
Only a small portion of your sample is ionized by the ion source. The majority
of any sample passes through the ion source without being ionized. It is
pumped away by the vacuum system. As a result, the exhaust from the foreline
pump will contain traces of the carrier gas and your samples. Exhaust from
the standard foreline pump also contains tiny droplets of foreline pump oil.
An oil mist filter is supplied with the standard foreline pump. This filter stops
only pump oil droplets. It does not trap any other chemicals. If you are using
toxic solvents or analyzing toxic chemicals, install a hose from the mist filter
outlet to the outdoors or into a fume hood vented to the outdoors. Be sure to
comply with your local air quality regulations.
The oil trap supplied with the standard foreline pump stops only foreline pump oil. It
WA R N I N G
does not trap or filter out toxic chemicals. If you are using toxic solvents or
analyzing toxic chemicals, use a hose to vent to a safe location.
The fluids in the foreline pump also collect traces of the samples being
analyzed. All used pump fluid should be considered hazardous and handled
accordingly. Dispose of used fluid as specified by your local regulations.
When replacing pump fluid, use appropriate chemical-resistant gloves and safety
WA R N I N G
glasses. Avoid all contact with the fluid.
Ammonia
Ammonia, used as a reagent gas, increases the need for foreline pump
maintenance. Ammonia causes foreline pump oil to break down more quickly.
Therefore, the oil in the standard foreline vacuum pump must be checked and
replaced more frequently.
Always purge the MS with methane after using ammonia.
Be sure to install the ammonia so the tank is in an upright position. This will
help prevent liquid ammonia from getting into the flow module.
Periodic maintenance
As listed in Table 9, some maintenance tasks for the vacuum system must be
performed periodically. These include:
• Checking the foreline pump fluid color and level (every week)
• Checking the calibration vial(s) (every 6 months)
• Replacing the foreline pump oil and oil mist filter (every 6 months or as
needed)
• Replacing the RIS probe seals (yearly or as needed)
Failure to perform these tasks as scheduled can result in decreased
instrument performance. It can also result in damage to your instrument.
Other procedures
Tasks such as replacing an ion vacuum gauge should be performed only when
needed by a certified Agilent service personnel. See the Agilent 7200
Accurate-Mass Q-TOF GC/MS Troubleshooting and Maintenance Manual and
see the online help in the MassHunter WorkStation software for symptoms
that indicate this type of maintenance is required.
Scheduling
None of the analyzer components requires periodic maintenance. Some tasks,
however, must be performed when MS behavior indicates they are necessary.
These tasks include:
• Cleaning the ion sources
• Replacing filaments
The Agilent 7200 Accurate-Mass Q-TOF GC/MS Troubleshooting and
Maintenance Manual provides information about symptoms that indicate the
need for analyzer maintenance. The troubleshooting material in the online
help in the MassHunter Workstation software provides more extensive
information.
Precautions
Keep components clean during analyzer maintenance. Never open the analyzer
chamber. There are no customer serviceable parts accessible from this door.
If not done correctly, analyzer maintenance can introduce contaminants into the MS.
CAUTION
The analyzers operate at high temperatures. Do not touch any part until you are sure
WA R N I N G
it is cool.
Electrostatic discharge
The wires, contacts, and cables connected to the analyzer components can
carry electrostatic discharges (ESDs) to the electronics boards to which they
are connected. This is especially true of the mass filter (quadrupole) and
collision cell contact wires which can conduct ESD to sensitive components on
the quad driver board. ESD damage may not cause immediate failure but will
gradually degrade performance and stability.
Incorrect handling or cleaning of the mass filter can damage it and have a serious,
CAUTION
negative effect on instrument performance. Never open the analyzer door.
Materials needed
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
• Nut driver, 5.5 mm (8710-1220)
• Tweezers (8710-2460)
Procedure
1 Remove the ion source. See “To Change the Ion Source” on page 78.
2 Remove the filaments. See “To Remove a Filament” on page 109.
3 Separate the ion focus and insulator assembly from the source body. (See
Figure 23.)
4 Remove the two screws from the ion focus insulator and separate it from
the ion focus lens.
5 Remove the extractor lens and its insulator from the source body.
6 Remove the two screws from the RIS bayonet mount on the bottom of the
source body, and remove the RIS bayonet mount.
7 Remove the repeller assembly from the source body.
8 Disassemble the repeller assembly by separating the repeller, the repeller
cap, ring heater and sensor assembly. (See Figure 23.)
9 Remove the clocking button from the repeller cap by removing the screw
that holds it in place. (See Figure 23.)
Screws
Washers
Bayonet mount
Repeller insulator
Clocking button
Repeller cap
Repeller
Filaments
Ion focus
Screws
Materials needed
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
Procedure
Always wear clean gloves when working in the analyzer chamber to avoid
CAUTION
contamination.
While installing, do not overtighten the repeller nut or the ceramic repeller insulators
CAUTION
will break when the source heats up. The nut should only be finger-tight.
Screws
Washers
Bayonet mount
Repeller insulator
Clocking button
Repeller cap
Repeller
Filaments
Ion focus
Screws
Materials needed
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
• Nut driver, 5.5 mm (8710-1220)
• Tweezers (8710-2460)
Procedure
1 Remove the CI ion source. See “To Change the Ion Source” on page 78.
2 Remove the filament. See “To Remove a Filament” on page 109.
3 Remove the ion focus lens and insulator. (See Figure 25.)
4 Remove the two screws from the ion focus lens insulator and separate it
from the ion focus lens.
5 Remove the two screws from the RIS bayonet mount and remove the
repeller assembly.
6 Disassemble the repeller assembly by removing the ceramic insulator and
heater assembly from the repeller. (See Figure 25.)
7 Take the repeller off the source body.
Bayonet mount
Clocking button
Repeller
Filament
Materials needed
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
Procedure
Always wear clean gloves when working in the analyzer chamber to avoid
CAUTION
contamination.
1 Place the ceramic heater assembly on top of the repeller, aligning the
grooves on the repeller with the slot on the heater assembly. (Figure 26).
2 Place the repeller on the source body. (Figure 26).
3 Place the assembled parts into the RIS bayonet mount, heater side first.
4 Align the interface socket in the source body with the alignment groove in
the RIS bayonet mount. Attach with the two screws.
5 Place the ion focus and ion focus insulator together and attach with the
two screws.
While installing, do not overtighten the repeller nut or the ceramic repeller insulator
CAUTION
will break when the source heats up. The nut should only be finger-tight.
Bayonet mount
Clocking button
Repeller
Filament
Frequency of cleaning
Because the CI ion source operates at much higher pressures than the EI ion
source, it will probably require more frequent cleaning than the EI ion source.
Cleaning of the source is not a scheduled maintenance procedure. The source
should be cleaned whenever there are performance anomalies that are
associated with a dirty ion source. See the 7200 Q-TOF GC/MS
Troubleshooting and Maintenance Manual for symptoms that indicate a dirty
ion source.
Visual appearance is not an accurate guide to cleanliness of the CI ion
source. The CI ion source can show little or no discoloration yet still need
cleaning. Let analytical performance be your guide.
Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Solvents
• Acetone, reagent grade
• Methanol, reagent grade
• Methylene chloride, reagent grade
• Ultrasonic bath
Preparation
1 Disassemble the ion source. See “To Disassemble the EI Ion Source” on
page 97 or “To Disassemble the CI Ion Source” on page 101.
2 Collect the following parts to be cleaned if you are cleaning a high
sensitivity extractor EI ion source: (Figure 27)
• Repeller
• Source body
• Extractor lens
• Ion focus lens
Collect the following parts to be cleaned if you are cleaning a CI source:
• Repeller
• Source body
• Ion focus lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
If insulators are dirty, clean them with a cotton swab dampened with reagent-grade
CAUTION
methanol. If that does not clean the insulators, replace them. Do not abrasively or
ultrasonically clean the insulators.
Cleaning the CI ion source is very similar to cleaning the EI ion source, with
the following exceptions:
• The CI ion source may not look dirty but deposits left by chemical
ionization are very difficult to remove. Clean the CI ion source thoroughly.
• Use a round wooden toothpick to gently clean out the electron entrance
hole in the source body.
• Do not use halogenated solvents. Use hexane for the final rinse.
4 Abrasively clean the surfaces that contact the sample or ion beam.
Use an abrasive slurry of alumina powder and reagent-grade methanol on
a cotton swab. Use enough force to remove all discolorations. Polishing the
parts is not necessary; small scratches will not harm performance. Also
abrasively clean the discolorations where electrons from the filaments
enter the source body.
5 Rinse away all abrasive residue with reagent-grade methanol.
Ensure all abrasive residue is rinsed away before ultrasonic cleaning. If
the methanol becomes cloudy or contains visible particles, rinse again.
6 Separate the parts that were abrasively cleaned from the parts that were
not abrasively cleaned.
Always wear clean gloves to prevent contamination when working in the analyzer
CAUTION
chamber.
All of these solvents are hazardous. Work in a fume hood and take all appropriate
WA R N I N G
precautions.
8 Place the parts in a clean beaker. Loosely cover the beaker with clean
aluminum foil (dull side down).
9 Dry the cleaned parts in an oven at 100 °C for 5–6 minutes.
To Remove a Filament
Materials needed
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Hex ball driver, 1.5 mm (8710-1570)
• Tweezers (8710-2460)
Procedure
Always wear clean gloves to prevent contamination when working in the analyzer
CAUTION
chamber.
1 Remove the ion source. See “To Change the Ion Source” on page 78.
2 Remove the screw holding the filament to the ion source body. (See
Figure 29.)
3 Slide the filament off the ion source assembly. (See Figure 29.)
The analyzer, GC/MS interface, and other components in the analyzer chamber
WA R N I N G
operate at very high temperatures. Do not touch any part until you are sure it is cool.
Filaments Filament
EI Source CI Source
To Install a Filament
Materials needed
• Filament assembly, EI (G7005-60061)
• Filament assembly, CI (G7005-60072)
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Tweezers (8710-2460)
Procedure
1 Remove the old filament. (See “To Remove a Filament” on page 109.)
2 Place the new filament into its position on the ion source body. (See
Figure 29.)
3 Secure the filament to the ion source body with the screw.
4 After installing the filament, verify that it is not grounded to source body.
5 Reinstall the ion source. (See “To Change the Ion Source” on page 78.)
6 Autotune the MS.
Do not overtighten the thumbscrew; it can cause air leaks or prevent successful
CAUTION
pumpdown. Do not use a screwdriver to tighten the thumbscrew.
*G3850-90008*
G3850-90008