XS1 2
XS1 2
(Please read through the instruction manual before the installation and operation
of the equipment)
Characteristics of the Equipment
The manufacturer shall be liable for the safety, liability and performance
problems of the equipment in the following conditions:
A. Installation, remodeling, adjustment, change or maintenance are
conducted by the staff specified by the manufacturer.
B. All relevant appliance in the room are in line with relevant
requirements.
C. Operating the equipment in accordance with the instruction manual
Please note that the manufacturer will not give any notice if there are
changes in the appearance and specifications of the equipment.
CONTENT
1.Outline…………………………………………………………………………………………1
The following three kinds of beam limiters are available to satisfy the
requirements of different X-ray machines:
a. XS5-1 type close control automatic beam limiter;
b. XS3-2 type automatic beam limiter;
c. XS1-2 type manual beam limiter.
While the ways of operation and control as well as related components of the
four beam limiters are somewhat different, they have lead gates of the same
construe. The former two beam limiters have a function of automatic operation.
The position sensors for lead gates installed on them cooperate with the
control circuit, film size sensors and focus position sensors etc. on the
corresponding bed to realize remote operation of the equipment. According to
the size of film, the radiation field can be set automatically and the function of
distance compensation can be executed. In addition, there are also knobs for
manual operation and display for light field on XS5-1 type beam limiter,
enabling manual operation of the bed closely and performing compensation
function.
XS5-1 type beam limiter is suitable for working with X-ray tube assembly
installed above the tomographic table or remote control table with a function of
tomographic etc.
XS3-2 type beam limiter is suitable for working with X-ray tube assembly
installed on all kinds of remote control table (both above and below) or C-arm
frames etc.
XS1-2 type beam limiter is suitable for working with X-ray tube assembly
supported by all kinds of stand and opening and closing of the lead gate is
controlled manually.
The ways of installation and operation of the above three kinds of beam
limiters are basically identical.
1
2.Main technical specifications
Maximum tube
150kV 150kV 150kV
voltage
Radiation field Maximum 350mm×350mm
(SID65cm) Minimum 0mm×0mm
Average
illuminance of the
≥100lx ≥100lx
light field
(SID100cm)
Edge contrast of
Not less than 4 Not less than 4
the light field
Time limit of the Not more than 60s Not more than 60s
light source
Inherent filtration Less than 2 Less than 1.5 Less than 2
(mmAl) (70kV) (70kV) (70kV)
Filtration of
organic-glass <0.2mmAl(70kv) <0.2mmAl(70kv) <0.2mmAl(70kv)
output window
Controlling ways of Remote control/
Remote control Remote control
lead gate close control
Distance between
the focus and 56mm 56mm 56mm
mounting surface
Weight (kg) 9.5 8 9
External
dimensions (length
250×224×250 224×224×250 250×224×250
×width×
height)mm
2
3.Structure and components
The structure and main components of the lead gate of XS5-1 type close
control beam limiter are shown in the following diagram.
①Outer lead gate ②Mid lead gate ③Shielding tube ④Perifocal lead gate
⑤Light reflector ⑥Small light reflector ⑦Lamp ⑧Close control box ⑨..Tapeline
The field limiting and shielding functions of the beam limiter are achieved by
vertical and horizontal outer lead gate①, mid lead gate②, perifocal lead gate
④ and shielding tube③. The opening and closing movements of the vertical
and horizontal lead gates are driven by the vertical and horizontal AC servomotors,
steel cable, pulley and sliding block and rectangular radiation fields of any size can
be achieved. There are also sensors for measuring opening of the vertical and
horizontal lead gates to realize automatic control of the radiation field.
There are knobs on the close control box⑧ for manual adjusting of the radiation field.
⑨Vertical drag tapeline bracket of Beam limiting device to diagnostic tabletop
can measure and display SID.
The displaying device for ray field is composed of light reflector⑤ and lamp⑦.
The radiation field of X-rays are displayed with visible light to help positioning
in manual operation. The rasters on small light reflector⑥ and close control
box ⑧ can display the vertical position of the center of X-rays on the
corresponding table in terms of light gap.
There is not close control box⑧ on XS3-2 type automatic beam limiter
XS1-2 type manual beam limiter is without automatic functions, XS1-2 type
3
manual beam limiter is different from XS3-2 type beam limiter in that the
former is not equipped with a motor, sensor and related transmission system
and the opening and closing of the lead gate are achieved through steel cable,
pulley and sliding block by turning the knobs on the close control box
manually.
4.Operation methods
4.1 XS5-1 type close control automatic beam limiter (working with remote
control table)
4.1.1 Remote control (automatic)
Choose the film size on the remote control operation desk and the beam
limiter will adjust the radiation field automatically to the size limited by the film
chosen at the time of photographing.
4.1.2 Close control (manual)
The layout of the panel on the close control box is shown as the following
diagram.
4
Indication of opening the field limiting plate up and down
Press time limiting switch③ and the lamp of light source in the displaying
device for ray field lights. The maximum time limit is 60s and the lamp goes
out when the time is out.
When the light field lamp lights, a light gap will be formed on the ray filter tray
of the table, which indicates the longitudinal position of the center of X-ray
beam. This is very convenient at the time of photographing.
5
between the focus and the image-receiving surface is changed, a
compensation function will be performed automatically to keep the size of
transmission field constant.
The functions and operation ways of ①horizontal knob, ②vertical knob and
③time limiting switch are totally the same as those of XS5-1 type beam
limiter.
6
Film distance 1m 1.5m 1.8m
Size of the cassettes
20cm(8in) 65° 42° 34°
23cm(9in) 75° 48° 41°
25cm(10in) 84° 55° 46°
28cm(11in) 93° 60° 52°
30cm(12in) 102° 66° 57°
36cm(14in) 121° 78° 67°
43cm(17in) 144° 94° 80°
a. The distance between the focus of the X-ray tube assembly and the
mounting surface should be 56mm.
b. The space inside the X-ray tube assembly should not be less than that
shown in the following diagram.
7
5.1.2 Installation methods
The structure of the junction connecting the series of beam limiter and X-ray
tube assembly is shown in the following diagram.
8
packaging and transportation).
c. Fix the connecting plate on the X-ray tube assembly⑤ with bolts.
d. Mount the blocking plate and screw down the bolts.
5.2 Adjustment
5.2.1 Adjustment of the center of the beam limiter
The center line of the beam limiter should pass through the focus of X-ray
tube, as shown above figure.
At the time of adjusting, unscrew the clamping bolt⑥, move the beam limiter
and align it’s center line with the radial and axial centers of the sleeve of X-ray
tube approximately, then screw down the bolt⑥. Observe the radiation field of
X-ray under the condition of transmission (make use of sensitivity enhancing
paper or fluorescent screen), and the light and shade inside the radiation field
should be even.
5.2.2 The procedures for calibrating and adjusting the centers of radiation field
and image-receiving surface are as follows:
a. Mark the center of image-receiving surface (fluorescent screen, image
intensifier, film magazine etc.).
b. Adjust the X-ray radiation field as small as possible.
c. Unscrew the clamping bolt of supporting shaft for the housing and the
blocking knob of sliding ring on the housing.
d. Adjust the position of the housing, observe or take a photo under the
condition of transmission to check the radiation field and make the center of
the radiation field overlap the center of image-receiving surface
approximately.
e. Screw down the clamping bolt and blocking knob.
5.2.3 For the sake of choosing, combining and adjusting the beam limiter
effectively, there are sizes of X-ray field on the panel of beam limiter
corresponding to all the distances between the focus and the image-receiving
surface. The operator can decide an appropriate size of X-ray field according
to the distance by adjusting vertical and horizontal knobs before a load is put
on.
There are knobs for manual operation on XS5-1 type automatic beam limiter,
which enable the operator to control closely and adjust manually. On the
9
surface that is 1m apart from the focus and vertical to the datum axis, both the
length and width of X-ray field should be less than 5cm.
Notice: The lamp and the metal components around the lamp are of
high temperature and do not touch them with your hands. Use
tools while adjusting or changing the lamp, or wait until the
high-temperature objects are cooled down completely.
After the lamp is changed, the overlapping of the center of light irradiation field
with that of X-ray radiation field needs to be checked by transmission or photo
taking. The lamp stand plate can be loosened to adjust its position if
necessary. The maximum allowable deviation between the position of light
field and that of X-ray radiation field is shown in the below figure:
Warp sketch map of light irradiation field and X-ray radiation field
Given the film distance of 1m, the maximum deviation should be less than the
10
following equation:
│a│+│b│≤ 2% of SID(film distance) and
│c│+│d│≤ 2% of SID
Remark: The solid line indicates X-ray radiation field
The dotted line indicates light field.
In the above diagram, MIV and MIH are vertical and horizontal servomotors
respectively, while CIV and CIH are phase splitting capacitors. The power
supply for the motor is AC110V/12.5V with a frequency of 50Hz/60Hz. AJ2 is
an option switch.
11
RIV and RIH are vertical and horizontal potentiometers for position sensing
respectively. When the beam limiter is fully closed, the residual resistance
between potentiometer 1and 2 should be adjusted to 35±5Ω. H1 is a light
field lamp of AC24V.
When switch AJ2 is disconnected, the beam limiter can be operated manually.
S1 is a time limiting switch for the light field lamp.
In the above diagram, MIV and MIH are vertical and horizontal servomotors
respectively, while CIV and CIH are phase splitting capacitors. The power
supply for the motor is AC110V/12.5V with a frequency of 50Hz/60Hz.
12
RIV and RIH are vertical and horizontal potentiometers for position sensing
respectively. When the beam limiter is fully closed, the residual resistance
between potentiometer 1and 2 should be adjusted to 35±5Ω.
NOTICE: If the beam limiters are matched with the products of other
company, the low-voltage power supply must be isolated from
the electricity grid. Never use the power supplied by
anauto-transformer to ensure safety.
The series of beam limiters are precision instruments and should be handled
carefully to avoid damage.
13
7.2 Failure analysis
14
Appendix 1 :Circuit diagram
D0V 7812 V4
V5 N1
R5 S1 1N4007
1 Vin 3 J1
V1 Vout
GND
100 R2
1N4740
120K
V2
C1 2 R4 C2 R1 S9014
R3 V3
470 5.1K 27K S9014
1A 5.1K 470
J1-1
J1-2
AJ2-1
V3 C2-1
K12V
C1 C2
R2 R1 0.1μ/400V
220 5.1K 5.1K
1A C2-2
K0V D2 D1
AJ2-2 -
15
Appendix 2 Element list
16