0% found this document useful (0 votes)
18 views7 pages

PDIV Test Report

Uploaded by

Subhashis Sarkar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
18 views7 pages

PDIV Test Report

Uploaded by

Subhashis Sarkar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

PDIV Test Report Page 1

PDIV (Partial Discharge Inception Voltage) Test Report


for Motor / Stator Windings

Comprehensive test procedure, equipment, calibration, interpretation and reporting checklist

Generated: 2025-10-06 13:31:18

Prepared by: ChatGPT (PDIV test procedure summary based on IEC 60270, IEC 60034-27-1, IEC
60034-18-41 and industry guides)

1. Purpose
- Determine PDIV (partial discharge inception voltage) the voltage at which first measurable PD
pulses appear.
- Detect manufacturing defects (voids, poor impregnation, sharp edges) and provide baseline for
acceptance and condition monitoring.
PDIV Test Report Page 2

2. Standards & References


- IEC 60270: High-voltage test techniques Partial discharge measurements.
- IEC 60034-27-1: Off-line PD measurements on stator windings.
- IEC 60034-18-41: PD-free insulation systems (Type I) for converter-fed machines.
- Vendor guides: OMICRON, Doble, etc. for practical calibration and measurement tips.

3. Required Equipment
- High-voltage source (AC or impulse) with current-limiting.
- PD detector / analyzer compliant with IEC 60270.
- Coupling capacitor OR HFCT (high-frequency current transformer) for PD signal pickup.
- PD calibrator (known charge injection), screened HV leads, grounding, EMI filters.
- Thermocouples, data logger, safety interlocks, HV protective gear.
PDIV Test Report Page 3

4. Measurement Circuits & Calibration


- Coupling-capacitor measurement: Provides apparent charge (pC) via capacitive coupling; high
sensitivity.
- HFCT on earth/ground: Detects PD pulses returning through ground; robust for large machines or
field tests.
- Calibration: Inject known charge (q_cal) via calibration capacitor, record calibration constant
and noise floor.
- Target sensitivity: lab 1-5 pC; field 10-50 pC depending on noise.

5. Test Waveforms
- AC sinusoidal (power-frequency) is common for PDIV determination.
- Impulse or repetitive-pulse methods used per IEC TS 60034-27-5 for converter-related stresses.
- Always state waveform used when reporting PDIV/PDEV.
PDIV Test Report Page 4

6. Detailed Step-by-Step Procedure


Pre-test:
- Visual inspection, measure IR and winding resistances, disconnect from external networks, record
ambient T/RH.
- Install coupling device (C_couple or HFCT), shortest screened leads, single return path to earth.

Calibration:
- Use PD calibrator to inject known charge pulses, set instrument to display pC, record sensitivity
and noise floor.

PDIV Ramping:
- Start at 0 V. Increase in controlled steps (e.g., 0.1-0.2 kV steps for LV) or slow continuous
ramp.
- At each step hold for sufficient cycles to observe PD activity. Monitor apparent charge (pC),
pulse count and PRPD.
- PDIV = lowest applied voltage where pulses exceed detection threshold (document threshold and
method).
- After detection, continue or record and then reduce to find PDEV (extinction voltage on decreasing
ramp).
PDIV Test Report Page 5

7. Data to Capture
- Test circuit diagram and connections.
- Calibration constants (q_cal, C_cal) and measured sensitivity (pC).
- Ambient conditions (T, RH).
- Ramp table: voltages, dwell times, PD magnitudes, pulse counts, PRPD plots.
- Post-test IR, Hipot, visual inspection photos and notes.

8. Interpretation & Acceptance


- PDIV values are method-dependent; compare against baseline/type-test numbers or contractual
acceptance criteria.
- Type-I PD-free systems must show no PD at specified converter stress levels per IEC 60034-18-41.
- Examine Q V curve: slow onset and low pC is less severe; steep rise indicates serious defects.
PDIV Test Report Page 6

9. Common Pitfalls & Troubleshooting


- Electromagnetic noise: use screening, filters, avoid noisy periods.
- Poor grounding / multiple earth returns: causes false signals ensure single defined return path.
- Long unshielded leads and stray capacitance distort pulses shorten & screen leads.
- Failure to calibrate: cannot report pC without calibration.
- Mixing measurement circuits: do not compare results from different circuits without normalization.

10. Safety & Practical Cautions


- HV tests are hazardous: use interlocks, current limits, trained personnel, bleeder resistors.
- PD testing during temperature changes requires careful shielding and noise control.
- Do not perform destructive breakdown tests on production units; use representative samples for
such work.
PDIV Test Report Page 7

11. Test Report Checklist (Suggested Minimum)


- Identification: machine type, serial number, winding type, operator, date.
- Standards referenced.
- Test circuit diagram, instrument list and calibration data.
- Ambient conditions and test sequence (ramp steps, dwell times).
- PDIV, PDEV, Q V table, PRPD plots (attach plots as files or appendices).
- Post-test electrical checks: IR, Hipot (if performed), surge test results.
- Visual inspection findings and photos.
- Conclusion: pass/fail against quoted acceptance criteria and recommended actions.

Notes:
- All PD magnitudes must be reported with measurement sensitivity and coupling method.
- For formal acceptance, include type-test baseline comparisons or manufacturer-specified
thresholds.

You might also like