The Anaesthesia Machine- 2
Presented by: Dr. Vineet Chowdhary
Moderated by: Dr. Avnish Bharadwaj
LOWPRESSURE SYSTEM
The low-pressure system is downstream of the flow
control devices. Pressure in this section is only
slightly above atmospheric and variable.
Consists of:
1. Flowmeters
2. Hypoxia prevention safety devices
3. Unidirectional valve
4. Pressure relief devices
5. The common gas outlet
6. Vaporizers and their mounting devices
FLOWMETER
Flowmeters (flow indicators, flow tubes, rotameters)
indicate the rate of flow of a gas passing through
them.
They may be mechanical or electronic.
Electronic flowmeters usually have a representation of
a mechanical flowmeter on a screen or a number
representing the flow.
Mechanical Flowmeters
 Measuring gas flow in a mechanical flowmeter is
based on the principle that flow past a
resistance is proportional to pressure.
 Mechanical flowmeters measure the drop in
pressure that occurs when a gas passes through a
resistance.
• Flowmeters on modern anaesthetic machines consist of a
tapered glass tube containing a bobbin or ball, which
floats on stream of moving gas
• As gas flow rate increases,float is carried further up
tube, thereby indicating flow rate
• Flowmeters are specifically constructed for each gas,
since flow rate depends on both viscosity and density
of gas
• Only correct tube and bobbin or ball can be used to
repair broken flowmeters
• Since bobbin floats in gas stream, flowmeters will
only function correctly if tube is vertical
• Flowmeters will not function correctly if the tube
cracked
Physical Principle of Gas Flow
 Traditional mechanical flow indicators used in
anesthesia machines have been of the variable orifice
(variable area, Thorpe tube) type.
 A vertical glass tube is internally tapered with its
smallest diameter at the bottom. It contains an
indicator that is free to move up and down inside the
tube.
 When there is no gas flow, the indicator rests at the
bottom of the tube.
 When the flow control valve is opened, gas enters at
the bottom and flows up the tube, elevating the
indicator.
 Gas passes through the annular opening between the
indicator and the tube and on to the outlet at the
top of the tube.
Anaesthesia machine 2
The indicator floats freely at a position where the
downward force caused by gravity equals the upward
force caused by the gas pressure on the bottom of the
indicator.
Because the tube is tapered, the size of the annular
opening around the indicator increases with height,
and more gas flows.
A scale marked on or beside the tube indicates the gas
flowrate
The rate of flow through the tube will depend
on three factors:



Pressure drop across
the constriction
Physical properties of
the gas.
Size of the annular
opening
Pressure Drop across the Constriction
 As gas flows around the indicator, it encounters
frictional resistance between the indicator and
the tube wall. There is a resultant loss of
energy reflected in a pressure drop.
 This pressure drop is constant for all positions
in the tube and is equal to the weight of the
float divided by its cross-sectional area.
Size of the Annular Opening
Annular cross-sectional area varies while the pressure
drop across the indicator remains constant for all
positions in the tube.
Flowmeters are often called constant-pressure
flowmeters.
Increasing the flow does not increase the pressure
drop but causes the indicator to rise to a higher
position in the tube, thereby providing greater flow
area for the gas.
The elevation of the indicator is a measure of the
annular area for flow and, therefore, of the flow
itself
Physical Characteristics of the Gas
At (low flows) flow is laminar and is a function of
the viscosity of the gas (Hagen-Poiseuille equation).
When the constriction is shorter and wider (high
flows), flow is more turbulent and depends on gas
density (Graham's law).
Q is flow; n is viscosity; p/l
is the pressure gradient; r
is the radius of the tube
Temperature and Pressure Effects
 Flowmeters are calibrated at atmospheric pressure
(760 torr) and room temperature 20°C
 Temperature and pressure changes will affect both
the viscosity and the density of a gas and so
influence the accuracy
Variations in temperature- no significant effect
Hyperbaric( high pressure)- less gas delivered
than intended
Low pressure(increased altitude)- More gas
delivered
Anaesthesia machine 2
Anaesthesia machine 2
Anaesthesia machine 2
Anaesthesia machine 2
Flow indicator indicators. The plumb bob and skirted floats are kept
centered in the tube by constant rotation. The reading is taken at the top.
The ball indicator is kept centered by rib guides. The reading is taken at the
center. The nonrotating float does not rotate and is kept centered by gas
flow
Stop
 At the top of the flowmeter tube.
 Prevents the indicator from plugging the outlet which
could lead to damage to the tube
 Prevents the indicator from ascending to a point in
the tube where it cannot be seen.
 Flowmeter with the indicator hidden at the top looks
much like one that is turned OFF
Scale
Flow indicator scale either be marked on or
immediately adjacent to the tube
 Flowmeters are calibrated in liters per minute.
 For flows below 1 L/minute, the flow may be expressed
either in milliliters or in decimal fractions of a
liter per minute with a zero before the decimal point.
Lights
• Flow indicator lights are offered as an option on most
modern anesthesia machines. These are useful when the
machine is used in a darkened room.
Anaesthesia machine 2
Flowmeter Tube Arrangement
Flowmeter tubes for different gases are grouped side
by side. The various gas flows meet at the common
manifold (mixing chamber) at the top.
Sometimes, there are two flowmeters for the same gas:
One for low and one for high flows.
The tubes are arranged in series (tandem)
There is one flow control valve for the both flowmeter
tubes. Gas from the flow control valve first passes
through a tube calibrated up to 1 L/minute, then
passes to a second tube that is calibrated for higher
flows.
Series flowmeter tubes offer increased accuracy .
Flow indicator sequence
A,B: Potentially
dangerous arrangements,
with the oxygen flow
indicator upstream.
If a leak occurs, oxygen
will be selectively
lost.
C,D: Oxygen is
downstream from other
gases, which is a safer
situation because
anesthetic gas rather
than oxygen will be
lost.
Arrows represent flows
of gases.
Auxiliary Oxygen Flowmeter
It is a self-contained flowmeter with its own flow
control valve, flow indicator, and outlet.
It usually has a short tube with a maximum flow of 10
L/minute and a barbed fitting on the outlet.
It is usually mounted on the left side of the machine.
This can be used to supply oxygen to the patient
without turning ON the anesthesia machine.
Problems with Flowmeters
Inaccuracy
The flowmeter scale, tube, and indicator must be
regarded as an inseparable unit.
Should any one of the components need replacement, a
complete new set must be obtained.
The tube assembly calibrated for one gas cannot be
used for a different gas.
The degree of inaccuracy that will occur if flowmeters
are transposed depends on the gases for which the
flowmeters were intended
The greatest accuracy in flowmeter tubes is in the
middle half of the tube.
Having one tube in the series arrangement for low flows
(less than 1 L/minute) and one for high flows (1 to 10
L/minute) will result in greater accuracy.
Indicator Problems
Damage to the flow indicator can result from a sudden
projection to the top of the tube due to a sudden rise
in pressure
Flow indicators can become worn or distorted by
handling
The stop at the top of the flowmeter tube can become
dislodged and rest on top of the indicator.
Leaks
A leak in a flowmeter downstream of the indicator but
upstream of the common manifold will result in a lower
than expected concentration of that gas in the fresh
gas
A leak may occur if a flow control valve is left open and
there is no cylinder or yoke plug in the yoke
Using the Wrong Flowmeter
The problem is most likely to occur between air and
nitrous oxide, as the position of oxygen is generally
fixed by national custom or standards.
Anaesthesia machine 2
Care of flowmeters
includes ensuring that:
Floats spin freely,
Qualified service personnel regularly maintain gas
machines,
An oxygen analyzer used always
One never adjusts a flowmeter without looking at it,
One includes flowmeters in visual monitoring sweeps
One turns flowmeters off before opening cylinders,
connecting pipelines, or turning machine "on".
HYPOXIAPREVENTIONSAFETY DEVICE
Anaesthesia machine 2
Mandatory Minimum Oxygen
 Anesthesia machines require a minimum (50 to 250
mL/minute) flow of oxygen before other gases will
flow.
 This is preset by the manufacturer (sometimes to the
customer's specification).
 Some machines activate an alarm if the oxygen flow
goes below a certain level even if no gases are being
used.
 The minimum flow is activated when the master switch
is turned ON.
 On some machines, it is disabled when air is used.
The minimum oxygen flow does not in itself prevent a
hypoxic gas concentration from being delivered. A
hypoxic gas mixture can be delivered with only
modest anesthetic gas flows
MINIMUM OXYGEN RATIO
The anesthesia workstation standard requires that an
anesthesia machine be provided with a device to
protect against an operator-selected delivery of a
mixture of oxygen and nitrous oxide having an oxygen
concentration below 21% oxygen (V/V) in the fresh gas
or the inspiratory gas
Mechanical Linkage
• There is a 14-tooth sprocket on the nitrous oxide
flow control valve and a 28-tooth sprocket on the
oxygen flow control valve.
• If the flow control valves are adjusted so that a 25%
concentration of oxygen is reached, a pin on the
oxygen sprocket engages a pin on the oxygen flow
control knob.
• This causes the oxygen and nitrous oxide flow control
valves to turn together to maintain a minimum of 25%
oxygen.
• This minimum oxygen ratio device (proportioning
system) permits independent control of each gas as
long as the percentage of oxygen is above the
minimum.
If the operator attempts to increase the nitrous
oxide flow beyond that ratio, the oxygen flow is
automatically increased. If the operator attempts to
lower the oxygen flow too much, the nitrous oxide
flow is lowered proportionally
It should be noted that these devices only link two
gases, normally nitrous oxide and oxygen.
Administering a third gas such as helium can result
in a hypoxic mixture.
Anaesthesia machine 2
In contrast to the Datex-Ohmeda Link-
25 system, which actively increases
oxygen flow to maintain a fresh gas-
oxygen concentration greater than 25%,
the Dräger ORMC and S-ORC systems
limit nitrous oxide flow to prevent
delivery of a fresh gas mixture with
an oxygen concentration of less than
25%
Limitations of Proportioning System
Anti hypoxic device may also deliver hypoxic
mixture under following conditions-
Wrong supply of gas.
Defective pneumatics/mechanics.
Inert gas administration(3rd gas-He,N2,CO2).
Leaks downstream.
UNIDIRECTIONAL CHECK VALVES
When ventilation is controlled or assisted, positive
pressure from the breathing system can be transmitted
back into the machine.
Using the oxygen flush valve may also create a
positive back pressure.
This pressure can affect flowmeter readings and the
concentration of volatile anesthetic agents delivered
from the vaporizers on the machine.
Some machines have a unidirectional (check) valve to
minimize these effects.
This valve is located between the vaporizers and the
common gas outlet, upstream of where the oxygen flush
flow joins the fresh gas flow.
This valve will lessen the pressure increase but not
prevent it, because gas will be continually flowing
from the flowmeters.
These check valves are of great importance when
checking the machine for leaks. Testing the breathing
system for leaks will not detect a leak upstream of
the check valve in a machine equipped with a check
valve.
Anaesthesia machine 2
Common (Fresh) Gas Outlet
Anaesthesia machine 2
Some new anesthesia machines provide two common gas
outlets.
A hazard with this arrangement is that the fresh gas
may be directed to the wrong outlet
The anesthesia workstation standard states that there
shall be only one functional common gas outlet at a
time
Many new anesthesia machines have internal connections
to the breathing system. These machines may not have
the conventional common gas outlet described
previously.
CAUTION!!
The common gas outlet should not be used to administer
supplemental oxygen to a patient
This will delay use of the breathing system if an
emergency arises.
Another potential problem is that a vaporizer may be
accidentally left ON, leading to undesired
administration of inhalational agent.
Either the auxiliary oxygen flowmeter or a separate
flowmeter should be used to supply oxygen.
Alternative Oxygen Control
When using an anesthesia machine, there is always the
possibility that the electronics will fail.
Some machines provide an alternative means to
administer oxygen.
This is separate from the auxiliary (courtesy)
flowmeter. If there is a mechanical total flow
flowmeter, it can be used to measure the delivered
oxygen.
Anaesthesia machine 2
Anaesthesia machine 2
Anaesthesia machine 2
Anaesthesia machine 2

More Related Content

PPTX
Low pressure system in anaesthesia machine
PPTX
ANESTHESIA MACHINE PPT .pptx
PPTX
Safety features of anaesthesia machine
PPTX
Anesthesia workstation
PPT
Anesthetic machine for student
PPTX
Intermediate & low pressure system
PPT
Anaesthesia machine 1
PPTX
Anaesthesia gas cylinders & pipeline gas supply
Low pressure system in anaesthesia machine
ANESTHESIA MACHINE PPT .pptx
Safety features of anaesthesia machine
Anesthesia workstation
Anesthetic machine for student
Intermediate & low pressure system
Anaesthesia machine 1
Anaesthesia gas cylinders & pipeline gas supply

What's hot (20)

PPT
Scavenging system in operating room
PPTX
Physics In Anaesthesia
PPT
Vaporizers
PDF
caudal anesthesia.pdf
PPTX
Double lumen tubes
PPTX
Safety features in anesthesia machine
PPTX
Vaporizers
PPTX
High pressure system- Anaesthesia Machine
PPTX
CIRCLE SYSTEM.pptx
PPTX
The canister (the absorber)
PPT
Mapleson circuits
PPTX
Anaesthesia machine
PPTX
Low flow anaesthesia
PPTX
Neuromuscular Monitoring
PPTX
Supraglottic airway device
PPTX
Medical gas cylinders
PPTX
Circle system low flow anesthesia
PPTX
CO2 absorbents and waste gas scavenging system.pptx
PPTX
anaesthesia Vaporizers tec1 to tec5
PPTX
Medical gas cylinders
Scavenging system in operating room
Physics In Anaesthesia
Vaporizers
caudal anesthesia.pdf
Double lumen tubes
Safety features in anesthesia machine
Vaporizers
High pressure system- Anaesthesia Machine
CIRCLE SYSTEM.pptx
The canister (the absorber)
Mapleson circuits
Anaesthesia machine
Low flow anaesthesia
Neuromuscular Monitoring
Supraglottic airway device
Medical gas cylinders
Circle system low flow anesthesia
CO2 absorbents and waste gas scavenging system.pptx
anaesthesia Vaporizers tec1 to tec5
Medical gas cylinders
Ad

Viewers also liked (20)

PPT
Anesthesia Machine
PPTX
Anaesthesia machine 2014 spmc
PPTX
Anesthesia machine
PPTX
Introduction to anesthesia
PPTX
Perioperative hypertension- Definition, management
PPTX
Hyponatremia and hypernatremia
PPTX
How to Conduct a Literature Review (ISRAPM 2014)
PPTX
Slideshare chapter #9
PPT
Quit Smoking
PPTX
Slideshare chapter #9
PDF
Improving home oxygen: testing the case for change
PPTX
Why you should quit smoking
PDF
Healthy living specification
PPT
Corrosion Circuits
PPT
Vets 238 Anesthetic Equipment Final
PPT
Start exercising daily before you quit smoking
PDF
Stroke and its types
PPT
Anaesthesia machine manoj
PPT
Safety features in anesthesia machines-madras medical college
Anesthesia Machine
Anaesthesia machine 2014 spmc
Anesthesia machine
Introduction to anesthesia
Perioperative hypertension- Definition, management
Hyponatremia and hypernatremia
How to Conduct a Literature Review (ISRAPM 2014)
Slideshare chapter #9
Quit Smoking
Slideshare chapter #9
Improving home oxygen: testing the case for change
Why you should quit smoking
Healthy living specification
Corrosion Circuits
Vets 238 Anesthetic Equipment Final
Start exercising daily before you quit smoking
Stroke and its types
Anaesthesia machine manoj
Safety features in anesthesia machines-madras medical college
Ad

Similar to Anaesthesia machine 2 (20)

PPTX
lowpressuresysteminanaesthesiamachine-130521144917-phpapp02 (1).pptx
PPTX
Flowmeter colour bobbins
PPTX
Low Pressure System anaesthesia apparatus. pptx
PPTX
flowmeters.pptx
PPTX
anaesthesia machine and equipments.pptx
PPTX
anaesthesia machine-140810030802-pptx.pptx
PDF
ANAESTHESIA MACHINE.pptx.pdf different parts
PPTX
ANAESTHESIA MACHINE -vaporisers, general anesthesia.pptx
PPTX
APPROACH TO ANAESTHESIA MACHINE- Dr ASHISH.pptx
PPT
anesthetic machine for student new.ppt
PPT
Safety Anesthesia Work Station
PPTX
Anaesthesia workstation
PPTX
ANESTHESIA_MACHINE-_PRESSURE_REDUCING_VALVES,_FLOWMETER_AND[1].pptx
PPT
Anesthesiamachine
PPTX
INTERMEDIATE PRESSURE SYSTEM.pptx
PPTX
Anesthesia Machine -.pptx
PPTX
old vs new anesthesia machine comparison
PDF
Anesthesia machines and breathing systems
PDF
019 the anesthetic machine
PPTX
Máquina de Anestesia
lowpressuresysteminanaesthesiamachine-130521144917-phpapp02 (1).pptx
Flowmeter colour bobbins
Low Pressure System anaesthesia apparatus. pptx
flowmeters.pptx
anaesthesia machine and equipments.pptx
anaesthesia machine-140810030802-pptx.pptx
ANAESTHESIA MACHINE.pptx.pdf different parts
ANAESTHESIA MACHINE -vaporisers, general anesthesia.pptx
APPROACH TO ANAESTHESIA MACHINE- Dr ASHISH.pptx
anesthetic machine for student new.ppt
Safety Anesthesia Work Station
Anaesthesia workstation
ANESTHESIA_MACHINE-_PRESSURE_REDUCING_VALVES,_FLOWMETER_AND[1].pptx
Anesthesiamachine
INTERMEDIATE PRESSURE SYSTEM.pptx
Anesthesia Machine -.pptx
old vs new anesthesia machine comparison
Anesthesia machines and breathing systems
019 the anesthetic machine
Máquina de Anestesia

Recently uploaded (20)

PPTX
PSYCHIATRIC SEQUALAE OF HEAD INJURY.pptx
PPTX
المحاضرة الثالثة Urosurgery (Inflammation).pptx
PPTX
A Detailed Physiology of Endocrine System.pptx
PPTX
FORENSIC MEDICINE and branches of forensic medicine.pptx
PDF
Gonadotropin-releasing hormone agonist versus HCG for oocyte triggering in an...
PPTX
ENT-DISORDERS ( ent for nursing ). (1).p
PPTX
Approch to weakness &paralysis pateint.pptx
PPTX
Geriatrics_(0).pptxxvvbbbbbbbnnnnnnnnnnk
PPTX
critical care nursing 12.pptxhhhhhhhhjhh
PDF
heliotherapy- types and advantages procedure
PDF
The Role of Medicinal Plants in Alleviating Symptoms of Diabetes-Related Com...
PPTX
Communicating with the FDA During an Inspection -August 26, 2025 - GMP.pptx
PPTX
Genetics and health: study of genes and their roles in inheritance
PPTX
Bacteriology and purification of water supply
PPTX
RESEARCH APPROACH & DESIGN.pptx presented by preeti kulshrestha
PPTX
Nutrition needs in a Surgical Patient.pptx
PPTX
Local Anesthesia Local Anesthesia Local Anesthesia
PPTX
Acute Abdomen and its management updates.pptx
PDF
Cranial nerve palsies (I-XII) - AMBOSS.pdf
PDF
FMCG-October-2021........................
PSYCHIATRIC SEQUALAE OF HEAD INJURY.pptx
المحاضرة الثالثة Urosurgery (Inflammation).pptx
A Detailed Physiology of Endocrine System.pptx
FORENSIC MEDICINE and branches of forensic medicine.pptx
Gonadotropin-releasing hormone agonist versus HCG for oocyte triggering in an...
ENT-DISORDERS ( ent for nursing ). (1).p
Approch to weakness &paralysis pateint.pptx
Geriatrics_(0).pptxxvvbbbbbbbnnnnnnnnnnk
critical care nursing 12.pptxhhhhhhhhjhh
heliotherapy- types and advantages procedure
The Role of Medicinal Plants in Alleviating Symptoms of Diabetes-Related Com...
Communicating with the FDA During an Inspection -August 26, 2025 - GMP.pptx
Genetics and health: study of genes and their roles in inheritance
Bacteriology and purification of water supply
RESEARCH APPROACH & DESIGN.pptx presented by preeti kulshrestha
Nutrition needs in a Surgical Patient.pptx
Local Anesthesia Local Anesthesia Local Anesthesia
Acute Abdomen and its management updates.pptx
Cranial nerve palsies (I-XII) - AMBOSS.pdf
FMCG-October-2021........................

Anaesthesia machine 2

  • 1. The Anaesthesia Machine- 2 Presented by: Dr. Vineet Chowdhary Moderated by: Dr. Avnish Bharadwaj
  • 2. LOWPRESSURE SYSTEM The low-pressure system is downstream of the flow control devices. Pressure in this section is only slightly above atmospheric and variable. Consists of: 1. Flowmeters 2. Hypoxia prevention safety devices 3. Unidirectional valve 4. Pressure relief devices 5. The common gas outlet 6. Vaporizers and their mounting devices
  • 4. Flowmeters (flow indicators, flow tubes, rotameters) indicate the rate of flow of a gas passing through them. They may be mechanical or electronic. Electronic flowmeters usually have a representation of a mechanical flowmeter on a screen or a number representing the flow.
  • 5. Mechanical Flowmeters  Measuring gas flow in a mechanical flowmeter is based on the principle that flow past a resistance is proportional to pressure.  Mechanical flowmeters measure the drop in pressure that occurs when a gas passes through a resistance.
  • 6. • Flowmeters on modern anaesthetic machines consist of a tapered glass tube containing a bobbin or ball, which floats on stream of moving gas • As gas flow rate increases,float is carried further up tube, thereby indicating flow rate • Flowmeters are specifically constructed for each gas, since flow rate depends on both viscosity and density of gas • Only correct tube and bobbin or ball can be used to repair broken flowmeters • Since bobbin floats in gas stream, flowmeters will only function correctly if tube is vertical • Flowmeters will not function correctly if the tube cracked
  • 7. Physical Principle of Gas Flow  Traditional mechanical flow indicators used in anesthesia machines have been of the variable orifice (variable area, Thorpe tube) type.  A vertical glass tube is internally tapered with its smallest diameter at the bottom. It contains an indicator that is free to move up and down inside the tube.  When there is no gas flow, the indicator rests at the bottom of the tube.  When the flow control valve is opened, gas enters at the bottom and flows up the tube, elevating the indicator.  Gas passes through the annular opening between the indicator and the tube and on to the outlet at the top of the tube.
  • 9. The indicator floats freely at a position where the downward force caused by gravity equals the upward force caused by the gas pressure on the bottom of the indicator. Because the tube is tapered, the size of the annular opening around the indicator increases with height, and more gas flows. A scale marked on or beside the tube indicates the gas flowrate
  • 10. The rate of flow through the tube will depend on three factors:    Pressure drop across the constriction Physical properties of the gas. Size of the annular opening
  • 11. Pressure Drop across the Constriction  As gas flows around the indicator, it encounters frictional resistance between the indicator and the tube wall. There is a resultant loss of energy reflected in a pressure drop.  This pressure drop is constant for all positions in the tube and is equal to the weight of the float divided by its cross-sectional area.
  • 12. Size of the Annular Opening Annular cross-sectional area varies while the pressure drop across the indicator remains constant for all positions in the tube. Flowmeters are often called constant-pressure flowmeters. Increasing the flow does not increase the pressure drop but causes the indicator to rise to a higher position in the tube, thereby providing greater flow area for the gas. The elevation of the indicator is a measure of the annular area for flow and, therefore, of the flow itself
  • 13. Physical Characteristics of the Gas At (low flows) flow is laminar and is a function of the viscosity of the gas (Hagen-Poiseuille equation). When the constriction is shorter and wider (high flows), flow is more turbulent and depends on gas density (Graham's law). Q is flow; n is viscosity; p/l is the pressure gradient; r is the radius of the tube
  • 14. Temperature and Pressure Effects  Flowmeters are calibrated at atmospheric pressure (760 torr) and room temperature 20°C  Temperature and pressure changes will affect both the viscosity and the density of a gas and so influence the accuracy Variations in temperature- no significant effect Hyperbaric( high pressure)- less gas delivered than intended Low pressure(increased altitude)- More gas delivered
  • 19. Flow indicator indicators. The plumb bob and skirted floats are kept centered in the tube by constant rotation. The reading is taken at the top. The ball indicator is kept centered by rib guides. The reading is taken at the center. The nonrotating float does not rotate and is kept centered by gas flow
  • 20. Stop  At the top of the flowmeter tube.  Prevents the indicator from plugging the outlet which could lead to damage to the tube  Prevents the indicator from ascending to a point in the tube where it cannot be seen.  Flowmeter with the indicator hidden at the top looks much like one that is turned OFF Scale Flow indicator scale either be marked on or immediately adjacent to the tube  Flowmeters are calibrated in liters per minute.  For flows below 1 L/minute, the flow may be expressed either in milliliters or in decimal fractions of a liter per minute with a zero before the decimal point. Lights • Flow indicator lights are offered as an option on most modern anesthesia machines. These are useful when the machine is used in a darkened room.
  • 22. Flowmeter Tube Arrangement Flowmeter tubes for different gases are grouped side by side. The various gas flows meet at the common manifold (mixing chamber) at the top. Sometimes, there are two flowmeters for the same gas: One for low and one for high flows. The tubes are arranged in series (tandem) There is one flow control valve for the both flowmeter tubes. Gas from the flow control valve first passes through a tube calibrated up to 1 L/minute, then passes to a second tube that is calibrated for higher flows. Series flowmeter tubes offer increased accuracy .
  • 23. Flow indicator sequence A,B: Potentially dangerous arrangements, with the oxygen flow indicator upstream. If a leak occurs, oxygen will be selectively lost. C,D: Oxygen is downstream from other gases, which is a safer situation because anesthetic gas rather than oxygen will be lost. Arrows represent flows of gases.
  • 24. Auxiliary Oxygen Flowmeter It is a self-contained flowmeter with its own flow control valve, flow indicator, and outlet. It usually has a short tube with a maximum flow of 10 L/minute and a barbed fitting on the outlet. It is usually mounted on the left side of the machine. This can be used to supply oxygen to the patient without turning ON the anesthesia machine.
  • 25. Problems with Flowmeters Inaccuracy The flowmeter scale, tube, and indicator must be regarded as an inseparable unit. Should any one of the components need replacement, a complete new set must be obtained. The tube assembly calibrated for one gas cannot be used for a different gas. The degree of inaccuracy that will occur if flowmeters are transposed depends on the gases for which the flowmeters were intended The greatest accuracy in flowmeter tubes is in the middle half of the tube. Having one tube in the series arrangement for low flows (less than 1 L/minute) and one for high flows (1 to 10 L/minute) will result in greater accuracy.
  • 26. Indicator Problems Damage to the flow indicator can result from a sudden projection to the top of the tube due to a sudden rise in pressure Flow indicators can become worn or distorted by handling The stop at the top of the flowmeter tube can become dislodged and rest on top of the indicator. Leaks A leak in a flowmeter downstream of the indicator but upstream of the common manifold will result in a lower than expected concentration of that gas in the fresh gas A leak may occur if a flow control valve is left open and there is no cylinder or yoke plug in the yoke
  • 27. Using the Wrong Flowmeter The problem is most likely to occur between air and nitrous oxide, as the position of oxygen is generally fixed by national custom or standards.
  • 29. Care of flowmeters includes ensuring that: Floats spin freely, Qualified service personnel regularly maintain gas machines, An oxygen analyzer used always One never adjusts a flowmeter without looking at it, One includes flowmeters in visual monitoring sweeps One turns flowmeters off before opening cylinders, connecting pipelines, or turning machine "on".
  • 32. Mandatory Minimum Oxygen  Anesthesia machines require a minimum (50 to 250 mL/minute) flow of oxygen before other gases will flow.  This is preset by the manufacturer (sometimes to the customer's specification).  Some machines activate an alarm if the oxygen flow goes below a certain level even if no gases are being used.  The minimum flow is activated when the master switch is turned ON.  On some machines, it is disabled when air is used. The minimum oxygen flow does not in itself prevent a hypoxic gas concentration from being delivered. A hypoxic gas mixture can be delivered with only modest anesthetic gas flows
  • 33. MINIMUM OXYGEN RATIO The anesthesia workstation standard requires that an anesthesia machine be provided with a device to protect against an operator-selected delivery of a mixture of oxygen and nitrous oxide having an oxygen concentration below 21% oxygen (V/V) in the fresh gas or the inspiratory gas
  • 34. Mechanical Linkage • There is a 14-tooth sprocket on the nitrous oxide flow control valve and a 28-tooth sprocket on the oxygen flow control valve. • If the flow control valves are adjusted so that a 25% concentration of oxygen is reached, a pin on the oxygen sprocket engages a pin on the oxygen flow control knob. • This causes the oxygen and nitrous oxide flow control valves to turn together to maintain a minimum of 25% oxygen. • This minimum oxygen ratio device (proportioning system) permits independent control of each gas as long as the percentage of oxygen is above the minimum.
  • 35. If the operator attempts to increase the nitrous oxide flow beyond that ratio, the oxygen flow is automatically increased. If the operator attempts to lower the oxygen flow too much, the nitrous oxide flow is lowered proportionally It should be noted that these devices only link two gases, normally nitrous oxide and oxygen. Administering a third gas such as helium can result in a hypoxic mixture.
  • 37. In contrast to the Datex-Ohmeda Link- 25 system, which actively increases oxygen flow to maintain a fresh gas- oxygen concentration greater than 25%, the Dräger ORMC and S-ORC systems limit nitrous oxide flow to prevent delivery of a fresh gas mixture with an oxygen concentration of less than 25%
  • 38. Limitations of Proportioning System Anti hypoxic device may also deliver hypoxic mixture under following conditions- Wrong supply of gas. Defective pneumatics/mechanics. Inert gas administration(3rd gas-He,N2,CO2). Leaks downstream.
  • 39. UNIDIRECTIONAL CHECK VALVES When ventilation is controlled or assisted, positive pressure from the breathing system can be transmitted back into the machine. Using the oxygen flush valve may also create a positive back pressure. This pressure can affect flowmeter readings and the concentration of volatile anesthetic agents delivered from the vaporizers on the machine. Some machines have a unidirectional (check) valve to minimize these effects. This valve is located between the vaporizers and the common gas outlet, upstream of where the oxygen flush flow joins the fresh gas flow.
  • 40. This valve will lessen the pressure increase but not prevent it, because gas will be continually flowing from the flowmeters. These check valves are of great importance when checking the machine for leaks. Testing the breathing system for leaks will not detect a leak upstream of the check valve in a machine equipped with a check valve.
  • 44. Some new anesthesia machines provide two common gas outlets. A hazard with this arrangement is that the fresh gas may be directed to the wrong outlet The anesthesia workstation standard states that there shall be only one functional common gas outlet at a time Many new anesthesia machines have internal connections to the breathing system. These machines may not have the conventional common gas outlet described previously.
  • 45. CAUTION!! The common gas outlet should not be used to administer supplemental oxygen to a patient This will delay use of the breathing system if an emergency arises. Another potential problem is that a vaporizer may be accidentally left ON, leading to undesired administration of inhalational agent. Either the auxiliary oxygen flowmeter or a separate flowmeter should be used to supply oxygen.
  • 46. Alternative Oxygen Control When using an anesthesia machine, there is always the possibility that the electronics will fail. Some machines provide an alternative means to administer oxygen. This is separate from the auxiliary (courtesy) flowmeter. If there is a mechanical total flow flowmeter, it can be used to measure the delivered oxygen.