Coil Positioning/Safety




By
Marie Thomas
Coil Positioning is important for


                Image Quality
                Patient Safety
                Patient Comfort
Image Quality
Image Quality
   The patient needs to be positioned properly in or on a
    coil for optimal imaging
   The coil selected should be the right size for the body
    part being imaged (ex. liver flex-body NOT Q-body)
   The part of the body being imaged needs to be in the
    center of the coil, and in turn the coil needs to be as
    close to isocenter as possible.
Image Quality
       
            This Survey image is
           in good position. It
           has good signal
           coverage, the next
           slide shows the
           image received from
           this survey
Image Quality
          Good coil position
           has given this image
           nice even signal over
           the area of interest
Image Quality
          This Survey is not in
           a good position. The
           signal coverage area
           is too high. The
           following is what you
           get.
Image Quality
          An Artifact caused by
           not enough signal,
           poor uneven fat-
           saturation. In order to
           see the top of the
           tibia the window level
           had to be increased
           dramatically, causing
           the fluid to be to
           bright and the
           articular cartilage
           unseen.
Image Quality
          This is a survey
           Image using a flex-
           body coil. There is an
           artifact on the survey
           images.
Image Quality
          The resultant Image
Image Quality
          The coil was
           repositioned and......
Image Quality
          Artifact is gone!
          Revealing some
           pathology in the area
           of the artifact.
Image Quality Conclusion
   An extra minute or two to center the patient
    properly,can make a big difference. A lot of
    coils have a line to show you the center or ends
     of its active area, have a look at yours.
   When you look at your survey, remember to
    look at YOUR SIGNAL COVERAGE as well as
    the anatomy. This will help tell you how good
    your coil positioning is.
Image Quality Conclusion
   On sequences with lots of signal some
    T1's you might “get away” with poorer coil
    position but on Fat- Sat's or low NSA
    sequences you wont.
   If the anatomy your aiming for is not in the
    center, take a minute and reposition it
    might save you ten minutes later on
Coils and Patient Burns
   There are two types of burns
   RF Burns
   Cable Heating Burns
Coils and Patient Burns
                  Cable Heating burns
   What can we do to prevent them?
   NO LOOPS (loops cause conduction causing
    heat causing burns)
   Don't let a cable touch a patients skin/QBC
    directly
   Keep cables aligned with the magnet
   Make sure any gating leads are properly
    attached, no air in between and MR compatible
   Check your cables for frayed parts
Coils and Patient Burns

                 Types of Loops!
   Cable Loops
   Loops between cables and QBC
   Loops between cables and patients
   Loops within patient
Unsafe Coil
   Un-Safe Cables go across the main magnet
    field by the q-body coil creating a RF antenna
Safe Coil
   Safe The cables are not touching the skin
    (2cm)
   The cables are straight and going under pads
Safe Coil
   Cables go under pads
EKG Burn
Pulse Oximeter Burn
        3rd Degree
Coils and Patient Burns

                        RF Burns
   What can we do to prevent?
   Keep SAR to a minimum
   Put a pause in,some machines reconstruct and
    scan at the same time giving continual RF
    exposure.
   Keep in contact with your patient, make sure
    they alert you if they start to feel warm any
    where
Shoulder Burn
Tattoo Burn
Burn--
RF-or Loop burn?
           This patient was
            having his thighs
            scanned due to there
            different sizes.
            Apparently his legs
            were only touching for
            one scan. After the
            exam, he complained
            of an itching, he
            stayed at the facility
            and 3 hours later...
RF-or Loop burn?





    These 3rd degree burns came to the surface?

    What do you think?
Remember
              Burns can happen
           Screen your patient well
   Keep in contact with them during the exam
        Keep your SAR to a minimum
            Check your equipment
       Check your coil/cable positioning

More Related Content

PPTX
Mri artifacts
PDF
Quality Control in MRI- Avinesh Shrestha
PPT
PDF
Digital Subtraction Angiography king saud unversity.pdf
PPTX
Mri diffusion
PPTX
Magnetic Resonance Perfusion
PPTX
Ct instrumentation and types of detector configuration
PPT
radiation protection in ct
Mri artifacts
Quality Control in MRI- Avinesh Shrestha
Digital Subtraction Angiography king saud unversity.pdf
Mri diffusion
Magnetic Resonance Perfusion
Ct instrumentation and types of detector configuration
radiation protection in ct

What's hot (20)

PPTX
MRI INSTRUMENTATION/ HARDWARE
PPTX
The Benefits of Microwave Ablation
PPTX
Direct digital radiography(1) (1)
PPTX
Nuclear imaging
PPTX
Basics of mri physics Dr. Muhammad Bin Zulfiqar
PPTX
Radiation hazards in CT scan
PPTX
K Space in MRI
PPTX
Computed Tomography Dose Index
PPTX
Mammographic equipment and its advancement
PPTX
MDCT Principles and Applications- Avinesh Shrestha
PPTX
mr angiography.pptx
PPTX
MRI (Magnetic resonance imaging)
PPTX
Principle of DSA
PPTX
MRI spin echo pulse sequences
PPTX
Hypotonic duodenography
PPTX
CT Image reconstruction
PPTX
Basic of mri
PPTX
Ct head protocols
PPT
Physics of Multidetector CT Scan
PPTX
magnetic resonance in angiography
MRI INSTRUMENTATION/ HARDWARE
The Benefits of Microwave Ablation
Direct digital radiography(1) (1)
Nuclear imaging
Basics of mri physics Dr. Muhammad Bin Zulfiqar
Radiation hazards in CT scan
K Space in MRI
Computed Tomography Dose Index
Mammographic equipment and its advancement
MDCT Principles and Applications- Avinesh Shrestha
mr angiography.pptx
MRI (Magnetic resonance imaging)
Principle of DSA
MRI spin echo pulse sequences
Hypotonic duodenography
CT Image reconstruction
Basic of mri
Ct head protocols
Physics of Multidetector CT Scan
magnetic resonance in angiography
Ad

Viewers also liked (20)

PPTX
MRI Procedure of Brain
PPT
Mr isafety overview_anaesthetists_2013_edit_4_online
PPT
Mr hazard protection
PPT
MRI sequences
PDF
Poster - STIR vs FAT SAT_1009_2015 (1) (1)
PPTX
Fat suppression imaging
PPT
Basics Of MRI
PPTX
Dr. Simone A Winkler - Ultra High-Field MRI Open Questions in Engineering an...
PDF
Safety of MRI Critical Medical Devices
PDF
The Future Of Mri Safety W Kainz 13 Jul2009
PPT
2nd lect mri safety7 very-short_form
PDF
Properties of coordination compounds part 2 of 3
PPTX
Mri safety – the dos and donts
PDF
Breast imaging pitffals mri p taourel
PDF
Breast imaging breast mri s taieb
PDF
C Balleyguier, Birads MRI jfim hanoi 2015
PPTX
Mri in surgery
PPTX
Mri physics
PPT
MRI physics
MRI Procedure of Brain
Mr isafety overview_anaesthetists_2013_edit_4_online
Mr hazard protection
MRI sequences
Poster - STIR vs FAT SAT_1009_2015 (1) (1)
Fat suppression imaging
Basics Of MRI
Dr. Simone A Winkler - Ultra High-Field MRI Open Questions in Engineering an...
Safety of MRI Critical Medical Devices
The Future Of Mri Safety W Kainz 13 Jul2009
2nd lect mri safety7 very-short_form
Properties of coordination compounds part 2 of 3
Mri safety – the dos and donts
Breast imaging pitffals mri p taourel
Breast imaging breast mri s taieb
C Balleyguier, Birads MRI jfim hanoi 2015
Mri in surgery
Mri physics
MRI physics
Ad

Similar to Coil Safe Positioning in MRI (20)

PPTX
Artifacts and Quality Control
PPTX
Artifacts
PPTX
MRI Artifacts (introduction, classification, types)
PPT
Artifacts in MRI
PPTX
MRI Artifacts
PPTX
mriphysics-1610311406mmmmmmmmmmmm05.pptx
PPTX
Physics of 3 Tesla MRI & Silent MRI
PPTX
artifacts in computed tomography and MRI by remi.pptx
PPT
Mri physics
PDF
Mri basics
PPTX
ppMRIARTIFACTS.pptx
PPTX
MRI PULSE SEQUENCES.pptx/////////////////
PPTX
MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
PPTX
MRI ARTEFACTS caused in the scan normal.pptx
PPTX
Artifacts part 1
PPTX
9 mri
PPTX
Magnetic Resonance Imaging Artefacts ppt.
PPTX
ANESTHESIA for MRI procedures
PPTX
MRI IMAGE QUALITY and Factors affecting it by T.R.B.
Artifacts and Quality Control
Artifacts
MRI Artifacts (introduction, classification, types)
Artifacts in MRI
MRI Artifacts
mriphysics-1610311406mmmmmmmmmmmm05.pptx
Physics of 3 Tesla MRI & Silent MRI
artifacts in computed tomography and MRI by remi.pptx
Mri physics
Mri basics
ppMRIARTIFACTS.pptx
MRI PULSE SEQUENCES.pptx/////////////////
MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
MRI ARTEFACTS caused in the scan normal.pptx
Artifacts part 1
9 mri
Magnetic Resonance Imaging Artefacts ppt.
ANESTHESIA for MRI procedures
MRI IMAGE QUALITY and Factors affecting it by T.R.B.

Recently uploaded (20)

PDF
495958952-Techno-Obstetric-sminiOSCE.pdf
PPT
fiscal planning in nursing and administration
PPTX
Biostatistics Lecture Notes_Dadason.pptx
PPTX
etomidate and ketamine action mechanism.pptx
PPTX
presentation on dengue and its management
PPTX
ENT-DISORDERS ( ent for nursing ). (1).p
PPTX
SUMMARY OF EAR, NOSE AND THROAT DISORDERS INCLUDING DEFINITION, CAUSES, CLINI...
PPTX
Applied anatomy and physiology of Esophagus .pptx
PPTX
Genetics and health: study of genes and their roles in inheritance
PDF
Tackling Intensified Climatic Civil and Meteorological Aviation Weather Chall...
PPTX
GAIT IN HUMAN AMD PATHOLOGICAL GAIT ...............
PPTX
gut microbiomes AND Type 2 diabetes.pptx
PPTX
Approach to Abdominal trauma Gemme(COMMENT).pptx
PPT
intrduction to nephrologDDDDDDDDDy lec1.ppt
PPTX
Tuberculosis : NTEP and recent updates (2024)
PPTX
Journal Article Review - Ankolysing Spondylitis - Dr Manasa.pptx
PPTX
Local Anesthesia Local Anesthesia Local Anesthesia
PPTX
presentation on causes and treatment of glomerular disorders
PDF
neonatology-for-nurses.pdfggghjjkkkkkkjhhg
PDF
Glaucoma Definition, Introduction, Etiology, Epidemiology, Clinical Presentat...
495958952-Techno-Obstetric-sminiOSCE.pdf
fiscal planning in nursing and administration
Biostatistics Lecture Notes_Dadason.pptx
etomidate and ketamine action mechanism.pptx
presentation on dengue and its management
ENT-DISORDERS ( ent for nursing ). (1).p
SUMMARY OF EAR, NOSE AND THROAT DISORDERS INCLUDING DEFINITION, CAUSES, CLINI...
Applied anatomy and physiology of Esophagus .pptx
Genetics and health: study of genes and their roles in inheritance
Tackling Intensified Climatic Civil and Meteorological Aviation Weather Chall...
GAIT IN HUMAN AMD PATHOLOGICAL GAIT ...............
gut microbiomes AND Type 2 diabetes.pptx
Approach to Abdominal trauma Gemme(COMMENT).pptx
intrduction to nephrologDDDDDDDDDy lec1.ppt
Tuberculosis : NTEP and recent updates (2024)
Journal Article Review - Ankolysing Spondylitis - Dr Manasa.pptx
Local Anesthesia Local Anesthesia Local Anesthesia
presentation on causes and treatment of glomerular disorders
neonatology-for-nurses.pdfggghjjkkkkkkjhhg
Glaucoma Definition, Introduction, Etiology, Epidemiology, Clinical Presentat...

Coil Safe Positioning in MRI

  • 2. Coil Positioning is important for  Image Quality  Patient Safety  Patient Comfort
  • 4. Image Quality  The patient needs to be positioned properly in or on a coil for optimal imaging  The coil selected should be the right size for the body part being imaged (ex. liver flex-body NOT Q-body)  The part of the body being imaged needs to be in the center of the coil, and in turn the coil needs to be as close to isocenter as possible.
  • 5. Image Quality  This Survey image is in good position. It has good signal coverage, the next slide shows the image received from this survey
  • 6. Image Quality  Good coil position has given this image nice even signal over the area of interest
  • 7. Image Quality  This Survey is not in a good position. The signal coverage area is too high. The following is what you get.
  • 8. Image Quality  An Artifact caused by not enough signal, poor uneven fat- saturation. In order to see the top of the tibia the window level had to be increased dramatically, causing the fluid to be to bright and the articular cartilage unseen.
  • 9. Image Quality  This is a survey Image using a flex- body coil. There is an artifact on the survey images.
  • 10. Image Quality  The resultant Image
  • 11. Image Quality  The coil was repositioned and......
  • 12. Image Quality  Artifact is gone!  Revealing some pathology in the area of the artifact.
  • 13. Image Quality Conclusion  An extra minute or two to center the patient properly,can make a big difference. A lot of coils have a line to show you the center or ends of its active area, have a look at yours.  When you look at your survey, remember to look at YOUR SIGNAL COVERAGE as well as the anatomy. This will help tell you how good your coil positioning is.
  • 14. Image Quality Conclusion  On sequences with lots of signal some T1's you might “get away” with poorer coil position but on Fat- Sat's or low NSA sequences you wont.  If the anatomy your aiming for is not in the center, take a minute and reposition it might save you ten minutes later on
  • 15. Coils and Patient Burns  There are two types of burns  RF Burns  Cable Heating Burns
  • 16. Coils and Patient Burns Cable Heating burns  What can we do to prevent them?  NO LOOPS (loops cause conduction causing heat causing burns)  Don't let a cable touch a patients skin/QBC directly  Keep cables aligned with the magnet  Make sure any gating leads are properly attached, no air in between and MR compatible  Check your cables for frayed parts
  • 17. Coils and Patient Burns Types of Loops!  Cable Loops  Loops between cables and QBC  Loops between cables and patients  Loops within patient
  • 18. Unsafe Coil  Un-Safe Cables go across the main magnet field by the q-body coil creating a RF antenna
  • 19. Safe Coil  Safe The cables are not touching the skin (2cm)  The cables are straight and going under pads
  • 20. Safe Coil  Cables go under pads
  • 22. Pulse Oximeter Burn  3rd Degree
  • 23. Coils and Patient Burns RF Burns  What can we do to prevent?  Keep SAR to a minimum  Put a pause in,some machines reconstruct and scan at the same time giving continual RF exposure.  Keep in contact with your patient, make sure they alert you if they start to feel warm any where
  • 27. RF-or Loop burn?  This patient was having his thighs scanned due to there different sizes. Apparently his legs were only touching for one scan. After the exam, he complained of an itching, he stayed at the facility and 3 hours later...
  • 28. RF-or Loop burn?  These 3rd degree burns came to the surface?  What do you think?
  • 29. Remember  Burns can happen  Screen your patient well  Keep in contact with them during the exam  Keep your SAR to a minimum  Check your equipment  Check your coil/cable positioning