INTRODUCTION
 Computed Tomography is a well accepted imaging
modality for evaluation of the entire body.
 Computed Tomography(CT) Scan Machines Uses X-
rays, a powerful form of Electromagnetic Radiation.
 The images are obtained directly in the axial plane of
varying tissue thickness with the help of a computer.
 Some pathology can be seen in saggital or coronal
plane by reconstruction of the images by computer.
 CT has undergone several evolutions and nowadays
multi- detectors CT scanners have been evolved which
have better application in clinical field.
COMPARISION OF CT WITH
CONVIENTIONAL RADIOGRAPHY
Conventional radiography suffers
from the collapsing of 3D structures
onto a 2D image.
CT gives accurate diagnostic
information about the
distribution of structures inside
the body.
COMPARISION OF CT WITH CONVIENTIONAL RADIOGRAPHY.
A conventional X-ray image is basically a shadow.
Shadows give you an incomplete picture of an object's shape.
This is the basic idea of computer aided tomography. In a CT scan machine,
the X-ray beam moves all around the patient, scanning from hundreds of
different angles
Comparison of CT with Conventional
Radiography
Radiographic procedure is qualitative and not quantitative
ADVANTAGE OF COMPUTED TOMOGRAPHY
OVER CONVIENTIONAL RADIOGRAPHY.
 To overcome superimposition of structures.
 To improve contrast of the image.
 To measure small differences in tissue contrast.
TOMOGRAPHY
 Imaging of Layer/Slice.
Principle
Images of structures lying above and below the plane
are blurred out due to motion unsharpness while the
structures lying in plane of interest appear sharp in in
the image.
Tomography
PRINCIPLE OF COMPUTED
TOMOGRAPHY
 The internal structure of the object can be
reconstructed from multiple projections of the
object.
 Mathematically principle of CT was first developed in
1917 by Radon.
 Proved that image of unknown object could be
produced if one had several number of projections
throughout the object.
VARIOUS PARAMETERS OF CT
 SLICE
 MATRIX
 PIXEL
 VOXEL
 CT NUMBER
 WINDOWING
 WINDOW WIDTH
 WINDOW LEVEL
 PITCH
SLICE/CUT
 The cross section portion of body which is scanned for
production of CT image is called Slice.
 The slice has width and therefore volume.
 The width is determined by width of the x rays beam.
Cross Sectional Slices
Think like looking into the loaf of bread by cutting into
the thin slices and then viewing the slice individually.
MATRIX
 The CT image is represented as the Matrix of the
number.
 A two dimensional array of numbers arranged in rows
and columns is called Matrix.
 Each number represent
the value of the image at
that location.
PIXEL
 Each square in a matrix is called a pixel.
 Also known as picture element.
VOXEL
 Each individual element or number in the image
matrix represents a three dimensional volume element
in object called VOXEL.
CT NUMBER
 The numbers in the image matrix is called CT
NUMBER.
 Each pixel has a number which represents the x-ray
attenuation in the corresponding voxel of the object.
HOUNSFIELD UNITS(HU)
 Related to different composition and nature of Tissue.
 The CT NUMBER is also known as Hounsfield
units(HU).
 Represent the density of tissue.
 Different Tissue have different CT number Range in
HU.
Air - 1000
Fat -100
Pure water 0
CSF 15
White matter 45
Gray matter 40
Blood 20
Bone/calcification +1000
TISSUE AND CT NUMBER APPROXIMATE
WINDOWING is a system where the CT no. range of
interest is spread cover the full grey scale available on
the display system
WINDOW WIDTH –Means total range of CT no.
values selected for gray scale interpretation.
It corresponds to contrast of the image.
WINDOW LEVEL– represents the CT no. selected for
the centre of the range of the no. displayed on the
image. It corresponds to brightness of image .
Pitch
 The relationship between patient and tube motion is
called Pitch.
 It is defined as table movement during each
revolution of x-ray tube divided by collimation
width.
 For example: For a 5mm section, if patient moves
10mm during the time it takes for the x-ray tube to
rotate through 360˚, the pitch is 2.
 Increasing pitch reduces the scan time and patient
dose.
STEPS OF CT IMAGE FORMATION
Phase of CT scanning
1.Scanning the patient or data Acquisition
a)X-ray Generator
b)X-ray Tube
c)X-ray Filtration System
d)Detector System
2.Reconstruction
a)Simple back projection
b)Iterative method
c)Analytical method
3.Display
DATA ACQUISTION
The scanning process begins with data
acquisition
Data Acquisition refers to a method by
which the patient is systematically
scanned by the X ray tube and detectors
to collect enough information for image
reconstruction
Major components of Data Acquisition
System(DAS)
a)X-ray Generators
Generators are located on rotating scan frames within
the CT gantry to accommodate slip Ring.
Power: 50 to 80kw
Frequency: 5 to 50kHz
KVp: 80-120
mA:80-500
b) X-ray Tube
 Rotating anode x-ray tube with unique cooling.
 Small focal spot size (0.6mm) to improve spatial resolution.
 Anode heating capacity:1MHU to 7MHU
 Cooling rate:1MHU per minute.
c)X-ray Beam Filtration System
 CT employs monochromatic beam but radiation from
CT X-ray tube is polychromatic. so, X-ray beam is shaped by
compensation filter.
a)Pre patient Collimators: Reduces the patient dose.
b)Post patient Collimators: Reduces the scattered radiation detectors.
Overall Functions of
Collimators.
 To decrease scatter
radiation
 To reduce patient
dose
 To improve image
quality
 Collimator width
determines the slice
thickness
d)Detectors
The detectors gather information by measuring the x-ray
transmission through the patient.
 Two types:
Scintillation crystal detector
(Cadmium tungstate+ Si Photodiode)
Can be used in third and fourth generation scanners
Xenon gas ionisation chamber
Can be used in third generation scanners only
2)Reconstruction
 Reproduction of an image from raw data is called
Reconstruction.
A)Simple back projection
The image is created by reflecting the attenuation
profiles back in same direction they were obtained.
B)Iterative method
It start with assumption that all point in matrix have
same value and it was compared with measured value
and make correction until Values come with in
acceptable range.
It contain three correction factor
1. SIMULTANEOUS RECONSTRUCTION
2. RAY BY RAY CORRECTION
3. POINT BY POINT CORRECTION
C)Analytical Method
Today commonly used .
Two popular method used in that method are:-
1. 2-D FOURIER ANALYSIS
2.FILTERED BACK PROJECTION
2-D FOURIER ANALYSIS
In it any function of time or space can be represented by the
sum of various frequencies and amplitude of sine and
cosine waves.
For example the actual projected image of original object is
more rounded than those shown which would be slowly
simplify and corrected by Fourier transformation.
2-D FOURIER ANALYSIS
FILTERED BACK PROJECTION
Same as back projection except that the image is filtered,
or Modified to exactly counterbalance the effect of
sudden density Changes , which cause blurring(star
like pattern) in simple back projection.
Computed tomography
3)Display
 The reconstructed image is displayed on the monitor.
It is a digital image.
 It consists of 2D representation of 3D object in the
form of pixels.
 CT pixel size is determined by dividing the FOV by
matrix Size which is generally 512*512.
 PIXEL SIZE= FOV (mm)/ MATRIX SIZE
Generations of CT Scan
First Generation
 Narrow pencil beam
 Single detector
 Detector used is made up of NaI.
 Translate –Rotate movements of
Tube- detector combination
 Scan time-5mins.
 Designed only for evaluation
of brain.
First generation CT Scanner
•Head kept enclosed in a water
bath
•Paired detectors
•A reference detector
Second Generation CT Scanner
Second Generation
 Narrow fan beam
 Linear detector array(5 to30)
 Translate-Rotate movements of Tube-Detector
combination
 Fewer linear movements are needed as there are
more detectors to gather the data.
 Between linear movements, the gantry rotated 30o
 Scan time~30secs(advantage over first generation)
Third Generation
•Rotate(tube)Rotate(detectors)
Motion.
•Pulsed wide fan beam.
•Arc of detectors(600-900)
•Detectors are perfectly aligned
with the X-Ray tube
•Both Xenon and scintillation
crystal detectors can be used
•Scan time< 5secs
•Disadvantage: Ring Artifacts
due to electronic drift between
many detectors.
Fourth Generation
Fourth Generation
 Complete circular array of about 1200 to 4800
stationary detectors
 Single x-ray tube rotates with in the circular array
of detectors
 Wide fan beam to cover the entire patient
 Scan time of newer scanners is about ½ s or, <2s.
 Designed to address ring artifacts by keeping
detector assembly stationary.
 Disadvantage: High cost.
Fifth Generation
 stationary/stationary
 Developed specifically for cardiac tomographic imaging
 No conventional x-ray tube; large arc of tungsten encircles
patient and lies directly opposite to the detector ring
 Electron beam steered around the patient to strike the
annular tungsten target
 Capable of 50-msec scan times; can produce fast-frame-
rate CT movies of the beating heart
 Electron gun
 Large Arcs of tungsten
targets
 Detector ring
 17 slices per second
CT SCAN IN OUR RADIOLOGY
DEPARTMENT.(16 SLICES)
Patient Positioning and Data
Acquisition
TECHNICAL SPECIFICATION
(Hitachi ECLOS 16)
 CT scanner mode: Multislice
 Slices per rotation: 16
 Other rotation speed options: 1.0, 1.5, 2.0, 3.0 seconds
 Minimum rotation speed: 0.8 seconds
 Gantry diameter: 70 cm
 Maximum beam width (cm): 2 cm
 Table weight limit: 495 lbs
 Table movement range vertically/longitudinally: 42 to 100 cm
vertical, 186 cm horizontal
 X-ray generator kV range: 100, 120, 130 kVp
 Maximum scan range: 175 cm
 X-ray tube heat capacity: 5.0 MHU
 Power requirements: 3-phase, 200 Amp max.
Computed tomography
SPIRAL/HELICAL CT
 Spiral/Helical scanning uses third generation or fourth
generation slip ring design.
 Spiral computed tomography (or helical computed
tomography) is a computed tomography(CT) technology in
which the source and detector travel along a helical path
relative to the object. Typical implementations involve
moving the patient couch through the bore of the scanner
whilst the gantry rotates. Spiral CT can achieve improved
image resolution for a given radiation dose, compared to
individual slice acquisition. Most modern hospitals
currently use spiral CT scanners.
SLIP RING TECHNOLOGY
 In conventional CT scanning there was a paused
between each gantry rotation. But in Helical CT, Slip
Ring technology is used which allows continuous
rotation of gantry without interruption.
 Slip Rings are electrical conducting brushes and
component of gantry transferring the data or,
electrical energy to and from the stationary part of
gantry to rotating part of gantry for continuous
rotation of gantry.
Contd….
 There are usually three slip rings made up of
conduction materials(i.e. Sliver and Graphite.)
 First Slip Ring provides high voltage power to X-ray
tube.
 Second provide low voltage to control system on
rotating gantry and
 Third Slip Ring transfers digital data from rotating
detectors arrays.
Computed tomography
MULTISLICE/MULTIDETECTOR CT
 Multidetector computed tomography(MDCT) is a form of
computed tomography (CT) technology for diagnostic
imaging.
 In MDCT, a two-dimensional array of detector elements
replaces the linear array of detector elements used in
typical conventional and helical CT scanners. The two-
dimensional detector array permits CT scanners to acquire
multiple slices or sections simultaneously and greatly
increase the speed of CT image acquisition.
 Image reconstruction in MDCT is more complicated than
that in single section CT.
DUAL SOURCE CT
Dual Source CT (DSCT) is equipped
with two X-ray tubes.
Two corresponding detectors are
oriented in the gantry with an angular
offset of 90 degrees.
ADVANTAGES
1) High temporal resolution(in
response to time domain) for
cardiac imaging without β
blockers which means heart rate
is independent upon temporal
resolution.
2) Less radiation dose even for
obese patient.
3) Faster acquisition time with
shortest breathe hold.
DUAL ENERGY CT
Standard computed tomography (CT)
scanners use normal X-rays to make
cross-sectional ‘slice-like’ pictures or
images of the body.
A dual energy CT scanner is fairly new
technology that uses both the normal
X-ray and also a second less powerful X-
ray to make the images.
Two pictures are taken of the same
slice at different energies.(i.e. 80/140KV
,100/140KV Or,70/150KV)
This gives dual energy CT additional
advantages over standard CT for a wide
range of tests and procedures.
Most commonly used for CT
Angiography.
PORTABLE CT SCAN
PORTABLE CT SCAN
 As the world’s first portable, full-body, 32-slice CT (computed
tomography) scanner, BodyTom is a multi-departmental
imaging solution capable of transforming any room in the
hospital into an advanced imaging suite. The system boasts an
impressive 85cm gantry and 60cm field of view, the largest field
of view available in a portable CT scanner.
 The battery-powered BodyTom with an innovative internal drive
system can easily be transported from room to room and is
compatible with PACS, planning systems, surgical and robotic
navigation systems.
 Uniquely designed to accommodate patients of all sizes,
BodyTom provides point-of-care CT imaging wherever high-
quality CT images are needed, including the operating room,
intensive care unit, radiation oncology suites, and the emergency
department.
RADIATION DOSE IN CT
 Volume Computed Tomography Dose Index (CTDIvol) is a
standardized parameter to measure Scanner Radiation Output
 CTDIvol provides information about the amount of radiation
used to perform the study.
 CTDIvol is NOT patient dose
 CTDIvol is reported in units of mGy for either a 16-cm
(for head exams) or 32-cm (for body exams) diameter.
 AAPM (American Association of Physicts in Medicine)
introduces a parameter known as the Size Specific Dose
Estimate (SSDE) to allow estimation of patient dose based on
CTDIvol and patient size.
 For the same CTDIvol, a smaller patient will tend to have a
higher patient dose than a larger patient.
Computed tomography
EFFECTIVE DOSE (mSv)
CT
 HEAD------------ 2
 CHEST ---------- 8
 ABDOMEN --10
 PELVIS -------10
GENERAL RADIOGRAPHY
 SKULL -------- 0.07
 CHEST PA ---- 0.02
 ABDOMEN --- 1
 PELVIS -------- 0.7
Computed tomography

More Related Content

PPTX
Ct scan
Shreya Feliz
 
PPTX
Ct scan final (2)
shekhar star
 
PPT
Computer Tomography (CT Scan)
Likan Patra
 
PPTX
CT - computed tomography
@Saudi_nmc
 
PPTX
Basic principle of ct and ct generations
Tarun Goyal
 
PPTX
Computed tomography
Rituraj Mishra
 
PPTX
CT Generation (Generation of CT)
Upakar Paudel
 
Ct scan
Shreya Feliz
 
Ct scan final (2)
shekhar star
 
Computer Tomography (CT Scan)
Likan Patra
 
CT - computed tomography
@Saudi_nmc
 
Basic principle of ct and ct generations
Tarun Goyal
 
Computed tomography
Rituraj Mishra
 
CT Generation (Generation of CT)
Upakar Paudel
 

What's hot (20)

PPTX
Gamma Camera
Muhammad Musaddiq
 
PPTX
Computed Radiography and digital radiography
Durga Singh
 
PPTX
image intensifier.pptx
GajananWattamwar1
 
PPTX
Generations of CT scanner
sadhamhussain52
 
PPTX
X ray tube
Aiims New Delhi
 
PPTX
Basic principle of gamma camera
Upakar Paudel
 
PPTX
Fluoroscopy ppt
Brittany Jansante
 
PPTX
Spect technology
PRADEEP Cheekatla
 
PPTX
CT image quality and image resolution.pptx
Rohit Bansal
 
PPTX
Magnification(macro and micro radiography), distortion
parthajyotidas11
 
PPTX
CT Procedure OF Abdomen & Pelvis
Upakar Paudel
 
PPTX
Ct instrument
Riadh Al-Haidari
 
PPTX
MDCT Principles and Applications- Avinesh Shrestha
Avinesh Shrestha
 
PPTX
Usg diplay modes
Ripan Sharma
 
PPTX
Ultrasound Transducer Constriction And It’s Physics.pptx
Dr. Dheeraj Kumar
 
PPTX
Rectifiers
Madhu Sudana
 
PPTX
Digital radiography
haribudke
 
PPTX
GENERATIONS OF COMPUTED TOMOGRAPHY
MamtaSharma158
 
PPTX
CT ARTIFACTS.pptx
dypradio
 
PPT
Ct Basics
Pramod Krishnan
 
Gamma Camera
Muhammad Musaddiq
 
Computed Radiography and digital radiography
Durga Singh
 
image intensifier.pptx
GajananWattamwar1
 
Generations of CT scanner
sadhamhussain52
 
X ray tube
Aiims New Delhi
 
Basic principle of gamma camera
Upakar Paudel
 
Fluoroscopy ppt
Brittany Jansante
 
Spect technology
PRADEEP Cheekatla
 
CT image quality and image resolution.pptx
Rohit Bansal
 
Magnification(macro and micro radiography), distortion
parthajyotidas11
 
CT Procedure OF Abdomen & Pelvis
Upakar Paudel
 
Ct instrument
Riadh Al-Haidari
 
MDCT Principles and Applications- Avinesh Shrestha
Avinesh Shrestha
 
Usg diplay modes
Ripan Sharma
 
Ultrasound Transducer Constriction And It’s Physics.pptx
Dr. Dheeraj Kumar
 
Rectifiers
Madhu Sudana
 
Digital radiography
haribudke
 
GENERATIONS OF COMPUTED TOMOGRAPHY
MamtaSharma158
 
CT ARTIFACTS.pptx
dypradio
 
Ct Basics
Pramod Krishnan
 
Ad

Viewers also liked (9)

PPTX
L 4 ct physics
Shahid Younas
 
PPTX
Computer Tomography
Lovnish Thakur
 
PPTX
Slip ring
Rohan Deokar
 
PPTX
L 7 ct physics
Shahid Younas
 
PPT
CT Physics
RMLIMS
 
PPT
Physics of ct mri
Lokender Yadav
 
PPTX
L 5 ct physics
Shahid Younas
 
PPTX
5lab components of ct scanner
Khaleeque Memon
 
PPT
Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOU...
NCHANJI NKEH KENETH
 
L 4 ct physics
Shahid Younas
 
Computer Tomography
Lovnish Thakur
 
Slip ring
Rohan Deokar
 
L 7 ct physics
Shahid Younas
 
CT Physics
RMLIMS
 
Physics of ct mri
Lokender Yadav
 
L 5 ct physics
Shahid Younas
 
5lab components of ct scanner
Khaleeque Memon
 
Ct instrumentation lecture. NCHANJI NKEH KENETH, RADIOLOGY DEPARTMENT, ST LOU...
NCHANJI NKEH KENETH
 
Ad

Similar to Computed tomography (20)

PPTX
CT PHYSICS INTRODUCTION WITH BASICS.pptx
sukheswer
 
PPTX
Basics of CT & MRI
Vibhuti Kaul
 
PPT
Computed tomography - Physics and Clinical
Sambasivaselli R
 
PPTX
Physics_of_CT , CT machine and it’s parts, ct generations
NeetuMeena56
 
PPT
Ct scan by sajid ejaz
SAJIDEJAZ1
 
PPT
CT PHYSICS for first year radiology residents
drvanajabulkapuram
 
PPT
ct physics final jhlgkuyfkytfukyguktfkytfkytfiyt
bijayap1119
 
PPTX
Computed Tomography BOOk: physical processes of medical imaging
iamhadi664
 
PPTX
COMPUTED TOMOGRAPHY SCAN
Shounak Nandi
 
PPTX
Computed Tomography and Spiral Computed Tomography
JAMES JACKY
 
PPTX
COMPUTED TOMOGRAPHY upload.pptx
Preethi Natarajan
 
PPTX
Presentation on Computed tomography in english
NiloyAhmed6
 
PDF
BASICS OF WORKING OF COMPUTER TOMOGRAPHY
anishacotta2
 
PPTX
INTRODUCTION OF CT , VARIOUS GENERATIONS .pptx
PrashantVerma227710
 
PPTX
Computed Tomography.pptx
jerelubarco1
 
PPTX
Basics of CT.pptx
AntonyGodson3
 
PPTX
Computed tomography by jay&jay
Patel Jay
 
PPT
15_-_fgdgdfgdgdgdgComputed_Tomography_I.ppt
RenuJohn12
 
PPT
Computed Tomography Basic Principles and protocol
awangsapuan1
 
CT PHYSICS INTRODUCTION WITH BASICS.pptx
sukheswer
 
Basics of CT & MRI
Vibhuti Kaul
 
Computed tomography - Physics and Clinical
Sambasivaselli R
 
Physics_of_CT , CT machine and it’s parts, ct generations
NeetuMeena56
 
Ct scan by sajid ejaz
SAJIDEJAZ1
 
CT PHYSICS for first year radiology residents
drvanajabulkapuram
 
ct physics final jhlgkuyfkytfukyguktfkytfkytfiyt
bijayap1119
 
Computed Tomography BOOk: physical processes of medical imaging
iamhadi664
 
COMPUTED TOMOGRAPHY SCAN
Shounak Nandi
 
Computed Tomography and Spiral Computed Tomography
JAMES JACKY
 
COMPUTED TOMOGRAPHY upload.pptx
Preethi Natarajan
 
Presentation on Computed tomography in english
NiloyAhmed6
 
BASICS OF WORKING OF COMPUTER TOMOGRAPHY
anishacotta2
 
INTRODUCTION OF CT , VARIOUS GENERATIONS .pptx
PrashantVerma227710
 
Computed Tomography.pptx
jerelubarco1
 
Basics of CT.pptx
AntonyGodson3
 
Computed tomography by jay&jay
Patel Jay
 
15_-_fgdgdfgdgdgdgComputed_Tomography_I.ppt
RenuJohn12
 
Computed Tomography Basic Principles and protocol
awangsapuan1
 

Recently uploaded (20)

PPTX
unit2_cdunit2_cdunit2_cdunit2_cdunit2_cd.pptx
shella20221
 
PDF
LeadIAS – Best IAS Coaching in Kerala.pdf
LeadIAS
 
PPT
Gas turbine mark VIe control Monitoring IO.ppt
aliyu4ahmad
 
PDF
A Guide To Why Doing Nothing Is Powerful
Lokesh Agrawal
 
PPTX
Title The Power of Oral Communication (2).pptx
amankumar7762044
 
PPTX
PPT Lapkas helminthiasiiiiiiiiiiiiis.pptx
ratnaernawati4
 
PDF
Invincible season 2 storyboard revisions seq2 by Mark G
MarkGalez
 
PPTX
PRESENTATION OF SEPSIS, SEPTIC SHOCK.pptx
ericklouiseopio
 
PPTX
beforjkkkvbjkklkccghjjjkjjjjjje after.pptx
JayeshTaneja4
 
PPTX
锡根大学文凭办理|办理Uni Siegen毕业证电子版成绩单在线办理办成绩单
xxxihn4u
 
PPTX
FARZ ACADEMY MRCP EXAM PREPARATION-GUIDE & TIPS.pptx
dawnmarketingmaveric
 
PDF
Professor Dr. Nazrul Islam - Curriculum Vitae.pdf
Dr. Nazrul Islam
 
PDF
Left Holding the Bag sequence 2 Storyboard by Mark G
MarkGalez
 
PDF
【2nd】Explanatory material of DTU(230207).pdf
kewalsinghpuriya
 
PPTX
opportunities in biophysics for Bsc.pptx
MukeshPimpliskar2
 
PDF
Applying Lean Six Sigma in Pre-Sales & Pre-Development: Setting the Stage for...
alekhyamandadi1
 
PPTX
How To Write A ResumeCV - Resume Writing Tips
yeasinArafath6
 
PPTX
ASP MVC asderfewerwrwerwrefeewwfdewfewfdsfsd
faresslaam82
 
PPT
Cryptographyhsjckhyhbghvdsnbfgnhgvhnnbfrrnb
mannamsarath224
 
PDF
Meatball of Canyon Valley sequence 3 storyboard by Mark G.
MarkGalez
 
unit2_cdunit2_cdunit2_cdunit2_cdunit2_cd.pptx
shella20221
 
LeadIAS – Best IAS Coaching in Kerala.pdf
LeadIAS
 
Gas turbine mark VIe control Monitoring IO.ppt
aliyu4ahmad
 
A Guide To Why Doing Nothing Is Powerful
Lokesh Agrawal
 
Title The Power of Oral Communication (2).pptx
amankumar7762044
 
PPT Lapkas helminthiasiiiiiiiiiiiiis.pptx
ratnaernawati4
 
Invincible season 2 storyboard revisions seq2 by Mark G
MarkGalez
 
PRESENTATION OF SEPSIS, SEPTIC SHOCK.pptx
ericklouiseopio
 
beforjkkkvbjkklkccghjjjkjjjjjje after.pptx
JayeshTaneja4
 
锡根大学文凭办理|办理Uni Siegen毕业证电子版成绩单在线办理办成绩单
xxxihn4u
 
FARZ ACADEMY MRCP EXAM PREPARATION-GUIDE & TIPS.pptx
dawnmarketingmaveric
 
Professor Dr. Nazrul Islam - Curriculum Vitae.pdf
Dr. Nazrul Islam
 
Left Holding the Bag sequence 2 Storyboard by Mark G
MarkGalez
 
【2nd】Explanatory material of DTU(230207).pdf
kewalsinghpuriya
 
opportunities in biophysics for Bsc.pptx
MukeshPimpliskar2
 
Applying Lean Six Sigma in Pre-Sales & Pre-Development: Setting the Stage for...
alekhyamandadi1
 
How To Write A ResumeCV - Resume Writing Tips
yeasinArafath6
 
ASP MVC asderfewerwrwerwrefeewwfdewfewfdsfsd
faresslaam82
 
Cryptographyhsjckhyhbghvdsnbfgnhgvhnnbfrrnb
mannamsarath224
 
Meatball of Canyon Valley sequence 3 storyboard by Mark G.
MarkGalez
 

Computed tomography

  • 1. Presented by :Rituraj Mishra. B.Sc.MIT, 2013 Moderator: Mr. Ranjit Kumar Jha
  • 2. INTRODUCTION  Computed Tomography is a well accepted imaging modality for evaluation of the entire body.  Computed Tomography(CT) Scan Machines Uses X- rays, a powerful form of Electromagnetic Radiation.  The images are obtained directly in the axial plane of varying tissue thickness with the help of a computer.  Some pathology can be seen in saggital or coronal plane by reconstruction of the images by computer.  CT has undergone several evolutions and nowadays multi- detectors CT scanners have been evolved which have better application in clinical field.
  • 3. COMPARISION OF CT WITH CONVIENTIONAL RADIOGRAPHY Conventional radiography suffers from the collapsing of 3D structures onto a 2D image. CT gives accurate diagnostic information about the distribution of structures inside the body.
  • 4. COMPARISION OF CT WITH CONVIENTIONAL RADIOGRAPHY. A conventional X-ray image is basically a shadow. Shadows give you an incomplete picture of an object's shape. This is the basic idea of computer aided tomography. In a CT scan machine, the X-ray beam moves all around the patient, scanning from hundreds of different angles
  • 5. Comparison of CT with Conventional Radiography Radiographic procedure is qualitative and not quantitative
  • 6. ADVANTAGE OF COMPUTED TOMOGRAPHY OVER CONVIENTIONAL RADIOGRAPHY.  To overcome superimposition of structures.  To improve contrast of the image.  To measure small differences in tissue contrast.
  • 7. TOMOGRAPHY  Imaging of Layer/Slice. Principle Images of structures lying above and below the plane are blurred out due to motion unsharpness while the structures lying in plane of interest appear sharp in in the image.
  • 9. PRINCIPLE OF COMPUTED TOMOGRAPHY  The internal structure of the object can be reconstructed from multiple projections of the object.  Mathematically principle of CT was first developed in 1917 by Radon.  Proved that image of unknown object could be produced if one had several number of projections throughout the object.
  • 10. VARIOUS PARAMETERS OF CT  SLICE  MATRIX  PIXEL  VOXEL  CT NUMBER  WINDOWING  WINDOW WIDTH  WINDOW LEVEL  PITCH
  • 11. SLICE/CUT  The cross section portion of body which is scanned for production of CT image is called Slice.  The slice has width and therefore volume.  The width is determined by width of the x rays beam.
  • 12. Cross Sectional Slices Think like looking into the loaf of bread by cutting into the thin slices and then viewing the slice individually.
  • 13. MATRIX  The CT image is represented as the Matrix of the number.  A two dimensional array of numbers arranged in rows and columns is called Matrix.  Each number represent the value of the image at that location.
  • 14. PIXEL  Each square in a matrix is called a pixel.  Also known as picture element.
  • 15. VOXEL  Each individual element or number in the image matrix represents a three dimensional volume element in object called VOXEL.
  • 16. CT NUMBER  The numbers in the image matrix is called CT NUMBER.  Each pixel has a number which represents the x-ray attenuation in the corresponding voxel of the object.
  • 17. HOUNSFIELD UNITS(HU)  Related to different composition and nature of Tissue.  The CT NUMBER is also known as Hounsfield units(HU).  Represent the density of tissue.  Different Tissue have different CT number Range in HU.
  • 18. Air - 1000 Fat -100 Pure water 0 CSF 15 White matter 45 Gray matter 40 Blood 20 Bone/calcification +1000 TISSUE AND CT NUMBER APPROXIMATE
  • 19. WINDOWING is a system where the CT no. range of interest is spread cover the full grey scale available on the display system WINDOW WIDTH –Means total range of CT no. values selected for gray scale interpretation. It corresponds to contrast of the image. WINDOW LEVEL– represents the CT no. selected for the centre of the range of the no. displayed on the image. It corresponds to brightness of image .
  • 20. Pitch  The relationship between patient and tube motion is called Pitch.  It is defined as table movement during each revolution of x-ray tube divided by collimation width.  For example: For a 5mm section, if patient moves 10mm during the time it takes for the x-ray tube to rotate through 360˚, the pitch is 2.  Increasing pitch reduces the scan time and patient dose.
  • 21. STEPS OF CT IMAGE FORMATION
  • 22. Phase of CT scanning 1.Scanning the patient or data Acquisition a)X-ray Generator b)X-ray Tube c)X-ray Filtration System d)Detector System 2.Reconstruction a)Simple back projection b)Iterative method c)Analytical method 3.Display
  • 23. DATA ACQUISTION The scanning process begins with data acquisition Data Acquisition refers to a method by which the patient is systematically scanned by the X ray tube and detectors to collect enough information for image reconstruction
  • 24. Major components of Data Acquisition System(DAS) a)X-ray Generators Generators are located on rotating scan frames within the CT gantry to accommodate slip Ring. Power: 50 to 80kw Frequency: 5 to 50kHz KVp: 80-120 mA:80-500
  • 25. b) X-ray Tube  Rotating anode x-ray tube with unique cooling.  Small focal spot size (0.6mm) to improve spatial resolution.  Anode heating capacity:1MHU to 7MHU  Cooling rate:1MHU per minute. c)X-ray Beam Filtration System  CT employs monochromatic beam but radiation from CT X-ray tube is polychromatic. so, X-ray beam is shaped by compensation filter. a)Pre patient Collimators: Reduces the patient dose. b)Post patient Collimators: Reduces the scattered radiation detectors.
  • 26. Overall Functions of Collimators.  To decrease scatter radiation  To reduce patient dose  To improve image quality  Collimator width determines the slice thickness
  • 27. d)Detectors The detectors gather information by measuring the x-ray transmission through the patient.  Two types: Scintillation crystal detector (Cadmium tungstate+ Si Photodiode) Can be used in third and fourth generation scanners Xenon gas ionisation chamber Can be used in third generation scanners only
  • 28. 2)Reconstruction  Reproduction of an image from raw data is called Reconstruction. A)Simple back projection The image is created by reflecting the attenuation profiles back in same direction they were obtained.
  • 29. B)Iterative method It start with assumption that all point in matrix have same value and it was compared with measured value and make correction until Values come with in acceptable range. It contain three correction factor 1. SIMULTANEOUS RECONSTRUCTION 2. RAY BY RAY CORRECTION 3. POINT BY POINT CORRECTION
  • 30. C)Analytical Method Today commonly used . Two popular method used in that method are:- 1. 2-D FOURIER ANALYSIS 2.FILTERED BACK PROJECTION
  • 31. 2-D FOURIER ANALYSIS In it any function of time or space can be represented by the sum of various frequencies and amplitude of sine and cosine waves. For example the actual projected image of original object is more rounded than those shown which would be slowly simplify and corrected by Fourier transformation.
  • 33. FILTERED BACK PROJECTION Same as back projection except that the image is filtered, or Modified to exactly counterbalance the effect of sudden density Changes , which cause blurring(star like pattern) in simple back projection.
  • 35. 3)Display  The reconstructed image is displayed on the monitor. It is a digital image.  It consists of 2D representation of 3D object in the form of pixels.  CT pixel size is determined by dividing the FOV by matrix Size which is generally 512*512.  PIXEL SIZE= FOV (mm)/ MATRIX SIZE
  • 36. Generations of CT Scan First Generation  Narrow pencil beam  Single detector  Detector used is made up of NaI.  Translate –Rotate movements of Tube- detector combination  Scan time-5mins.  Designed only for evaluation of brain.
  • 37. First generation CT Scanner •Head kept enclosed in a water bath •Paired detectors •A reference detector
  • 39. Second Generation  Narrow fan beam  Linear detector array(5 to30)  Translate-Rotate movements of Tube-Detector combination  Fewer linear movements are needed as there are more detectors to gather the data.  Between linear movements, the gantry rotated 30o  Scan time~30secs(advantage over first generation)
  • 40. Third Generation •Rotate(tube)Rotate(detectors) Motion. •Pulsed wide fan beam. •Arc of detectors(600-900) •Detectors are perfectly aligned with the X-Ray tube •Both Xenon and scintillation crystal detectors can be used •Scan time< 5secs •Disadvantage: Ring Artifacts due to electronic drift between many detectors.
  • 42. Fourth Generation  Complete circular array of about 1200 to 4800 stationary detectors  Single x-ray tube rotates with in the circular array of detectors  Wide fan beam to cover the entire patient  Scan time of newer scanners is about ½ s or, <2s.  Designed to address ring artifacts by keeping detector assembly stationary.  Disadvantage: High cost.
  • 43. Fifth Generation  stationary/stationary  Developed specifically for cardiac tomographic imaging  No conventional x-ray tube; large arc of tungsten encircles patient and lies directly opposite to the detector ring  Electron beam steered around the patient to strike the annular tungsten target  Capable of 50-msec scan times; can produce fast-frame- rate CT movies of the beating heart
  • 44.  Electron gun  Large Arcs of tungsten targets  Detector ring  17 slices per second
  • 45. CT SCAN IN OUR RADIOLOGY DEPARTMENT.(16 SLICES)
  • 46. Patient Positioning and Data Acquisition
  • 47. TECHNICAL SPECIFICATION (Hitachi ECLOS 16)  CT scanner mode: Multislice  Slices per rotation: 16  Other rotation speed options: 1.0, 1.5, 2.0, 3.0 seconds  Minimum rotation speed: 0.8 seconds  Gantry diameter: 70 cm  Maximum beam width (cm): 2 cm  Table weight limit: 495 lbs  Table movement range vertically/longitudinally: 42 to 100 cm vertical, 186 cm horizontal  X-ray generator kV range: 100, 120, 130 kVp  Maximum scan range: 175 cm  X-ray tube heat capacity: 5.0 MHU  Power requirements: 3-phase, 200 Amp max.
  • 49. SPIRAL/HELICAL CT  Spiral/Helical scanning uses third generation or fourth generation slip ring design.  Spiral computed tomography (or helical computed tomography) is a computed tomography(CT) technology in which the source and detector travel along a helical path relative to the object. Typical implementations involve moving the patient couch through the bore of the scanner whilst the gantry rotates. Spiral CT can achieve improved image resolution for a given radiation dose, compared to individual slice acquisition. Most modern hospitals currently use spiral CT scanners.
  • 50. SLIP RING TECHNOLOGY  In conventional CT scanning there was a paused between each gantry rotation. But in Helical CT, Slip Ring technology is used which allows continuous rotation of gantry without interruption.  Slip Rings are electrical conducting brushes and component of gantry transferring the data or, electrical energy to and from the stationary part of gantry to rotating part of gantry for continuous rotation of gantry.
  • 51. Contd….  There are usually three slip rings made up of conduction materials(i.e. Sliver and Graphite.)  First Slip Ring provides high voltage power to X-ray tube.  Second provide low voltage to control system on rotating gantry and  Third Slip Ring transfers digital data from rotating detectors arrays.
  • 53. MULTISLICE/MULTIDETECTOR CT  Multidetector computed tomography(MDCT) is a form of computed tomography (CT) technology for diagnostic imaging.  In MDCT, a two-dimensional array of detector elements replaces the linear array of detector elements used in typical conventional and helical CT scanners. The two- dimensional detector array permits CT scanners to acquire multiple slices or sections simultaneously and greatly increase the speed of CT image acquisition.  Image reconstruction in MDCT is more complicated than that in single section CT.
  • 54. DUAL SOURCE CT Dual Source CT (DSCT) is equipped with two X-ray tubes. Two corresponding detectors are oriented in the gantry with an angular offset of 90 degrees. ADVANTAGES 1) High temporal resolution(in response to time domain) for cardiac imaging without β blockers which means heart rate is independent upon temporal resolution. 2) Less radiation dose even for obese patient. 3) Faster acquisition time with shortest breathe hold.
  • 55. DUAL ENERGY CT Standard computed tomography (CT) scanners use normal X-rays to make cross-sectional ‘slice-like’ pictures or images of the body. A dual energy CT scanner is fairly new technology that uses both the normal X-ray and also a second less powerful X- ray to make the images. Two pictures are taken of the same slice at different energies.(i.e. 80/140KV ,100/140KV Or,70/150KV) This gives dual energy CT additional advantages over standard CT for a wide range of tests and procedures. Most commonly used for CT Angiography.
  • 57. PORTABLE CT SCAN  As the world’s first portable, full-body, 32-slice CT (computed tomography) scanner, BodyTom is a multi-departmental imaging solution capable of transforming any room in the hospital into an advanced imaging suite. The system boasts an impressive 85cm gantry and 60cm field of view, the largest field of view available in a portable CT scanner.  The battery-powered BodyTom with an innovative internal drive system can easily be transported from room to room and is compatible with PACS, planning systems, surgical and robotic navigation systems.  Uniquely designed to accommodate patients of all sizes, BodyTom provides point-of-care CT imaging wherever high- quality CT images are needed, including the operating room, intensive care unit, radiation oncology suites, and the emergency department.
  • 58. RADIATION DOSE IN CT  Volume Computed Tomography Dose Index (CTDIvol) is a standardized parameter to measure Scanner Radiation Output  CTDIvol provides information about the amount of radiation used to perform the study.  CTDIvol is NOT patient dose  CTDIvol is reported in units of mGy for either a 16-cm (for head exams) or 32-cm (for body exams) diameter.  AAPM (American Association of Physicts in Medicine) introduces a parameter known as the Size Specific Dose Estimate (SSDE) to allow estimation of patient dose based on CTDIvol and patient size.  For the same CTDIvol, a smaller patient will tend to have a higher patient dose than a larger patient.
  • 60. EFFECTIVE DOSE (mSv) CT  HEAD------------ 2  CHEST ---------- 8  ABDOMEN --10  PELVIS -------10 GENERAL RADIOGRAPHY  SKULL -------- 0.07  CHEST PA ---- 0.02  ABDOMEN --- 1  PELVIS -------- 0.7

Editor's Notes

  • #5: Imagine you are standing in front of a wall, holding a pineapple against your chest with your right hand and a banana out to your side with your left hand. Your friend is looking only at the wall, not at you. If there's a lamp in front of you, your friend will see the outline of you holding the banana, but not the pineapple -- the shadow of your torso blocks the pineapple. If the lamp is to your left, your friend will see the outline of the pineapple, but not the banana. In order to know that you are holding a pineapple and a banana, your friend would have to see your shadow in both positions and form a complete mental image. This is the basic idea of computer aided tomography
  • #6: Differential absorption of X ray is recorded in film so qualitative measurement and differential absorption of x rays are recorded by special detectors so quantitative.(minute differences.)