CT INSTRUMENTATION AND
TYPES OF DETECTOR
CONFIGURATION
Presenter :Sujan Karki
B.Sc. MIT 3rd year
National Academy of Medical Sciences(NAMS)
Bir Hospital
Contents
•Introduction
•Generations of CT
•System components
•References
Introduction
• Computed tomography(CT) is a
revolutionary tool of medicine,
particularly in medical imaging.
• tomos = slice, graphein = to write
• Method for acquiring and
reconstructing an image of a thin cross
section of an object
• Based on measurement of x-ray
attenuation through the section plane
using many different projection. Image courtesy :Damien Hirst Autopsy with Sliced
Human Brain 2004
Historical perspectives
• Invention of CT was made possible through the work of several
individuals, most notably Godfrey Newbold Hounsfield and
Allan MacLeod Cormack .
• In 1967, Hounsfield was investigating pattern recognition and
reconstruction techniques by using the computer.
• He concluded that if x-rays are passed through an object from all
direction and if are measured then information of internal
structure could be obtained.
• The radiation used was from an americium 241 gamma source
coupled with a crystal detector.
• Because of the low radiation output, the apparatus took about 9
days to scan the object.
• Because this procedure was too long, various modifications were
made and the gamma radiation source was replaced by a
powerful x-ray tube.
(Hounsfield, 1980).
Fig: second prototype CT
scanner
Limitation of conventional tomography and radiography
The major shortcoming of
radiography is that the
superimposition of all structures on
the radiograph (3D collapsed to 2D)
and Low soft-tissue contrast
The limitations of tomography
include persistent image blurring that
cannot be completely removed,
degradation of image contrast
because of the presence of scattered
radiation created by the open
geometry of the x-ray beam, and
other problems resulting from film-
screen combinations.
Imaging before CT
• entire body areas were
inaccessible to radiography
- brain, mediastinum,
retroperitoneum
• diagnostic procedures
showing better detail in
these areas were
potentially harmful and or
poorly tolerated by the
patient -
pneumoencephalography,
diagnostic laparotomy
Fig: ventriculography Fig: pneumoencephalography
Fig: preparing patient and positioning during pneumoencephalogram.
Computed Tomography Developmental chronology
• 1924 - mathematical theory of tomographic image reconstructions (Johann Radon)
• 1930 - conventional tomography (A. Vallebona)
• 1963 - theoretical basis of CT (A. McLeod Cormack)
• 1972 - first commercial CT (Sir Godfrey Hounsfield)
• 1979 - Nobel price (Cormack & Hounsfield)
• 1989 - single slice helical CT scanner
• 1992 – dual slice CT scanner
• 2004 – 64 slice CT scanner
• 2006 – dual source CT scanner
• 2007 – 320 slice CT scanner (per revolution)
• 2011 – Steller detector
• 2012 – 640 slice scanner
• 2015 – GE introduced 512 slice ct with 256 rows of detector
• 2018 – 360 degree rotation only in 0.24 sec (x-ray-tube)
Original EMI Scanner
• Designed specifically for brain imaging.
• Head is enclosed in a water bath between x-ray tube and
pair of detectors.
• Patient remains stationary and gantry move through
rotation and translation motion
• Linear motion was repeated 180 times and 1degree
gantry rotation after each linear motion.
• The x-ray tube was on throughout linear motion and off
during rotary motion.
• The transmitted radiation was measured 160 times in
each linear movement. (180x160=28,800)
• Each pair of tomographic sections took 5 minutes, so
scan time was 25 minutes.
Fig: EMI Mark 1
EMI contd.
• CT image was reconstructed and displayed in
80x80 matrix in 2 different formats: a paper
printout and visual image on cathode ray tube.
• The x-ray field was collimated to 3x3x13 mm.
• Data from the measurement were computer
processed and presented in 3x3 mm of patient
cross section.
• An oil cooled stationary x-ray tube was used with
focal spot size 2.25x12mm and operated at
120kVp and 33mA.
• A large difference in attenuation between adjacent
areas such as air surrounding the head makes
computation complicated so water bath was used.
Fig: pixel and voxel in
EMI scanner
First generation :Rotate/Translate, Pencil Beam
• Original EMI unit was first generation
scanner.
• Used a pinhole collimator
• NaI as detector element with PM tube
• X-ray tube and detector system translate
linearly across the 24 cm FOV acquiring
160 parallel rays
• pencil beam and a single detector.
• A five view study of head took about 25-
30 minutes. fig: Image of first generation CT
Second generation :Rotate/Translate, Narrow Fan Beam
• Narrow fan beam was used with an angle of 10-30
degree and detector 5-30 detectors.
• with 10 degree rotation increment, only 18
translation would be required for 180 degree image
acquisition.
• Produced a tomographic section between 20-90
seconds.
• Using fan beam increased radiation intensity
towards edge and was compensated by bow tie
filter.
• Acquired more data (600 rays X 540 views=32400
data point)
• Limited only for head scan .
Third generation : Rotate/Rotate, Wide Fan Beam
• Translational motion was oblirated.
• The path traced by the tube describe circle rather
than the semicircle characteristic of 1st and 2nd
generation CT scanner.
• The main goal was to cut the acquisition time to
less than 20 seconds.
• Wide aperture fan beam angle ranging between
40 to 60 degrees.
• This new array consisted of 400 to 1000 detector
elements.
• Major disadvantages: ring artefact(occasional )
• Software corrected image reconstruction
algorithms now remove such artefacts.
Fourth generation : Rotate/Stationary
• Developed specifically to alleviate the ring artifacts
produced by the third generation.
• Removing the detectors from the rotating gantry and
putting them in a stationary ring around the patient.
• This stationary 360 degree ring of detectors required
an increased number of detector elements (~5000
total).
• x-ray tube can rotate either outside or inside the
detector ring.
• If the x-ray tube rotates outside of the detector ring,
it is crucial that the detector ring is tilted so that the
x-rays only interact with the detectors once they pass
through the patient .
Fifth generation : Stationary/Stationary
• Developed for cardiac imaging, the EBCT
reduces scan time to as little as 50 ms.
• does not use an x-ray tube; instead it uses a beam
of electrons generated outside the gantry.
• Inside the gantry there are 180-degree rows of
fixed detectors on one side and 180 degrees of
tungsten arcs opposite.
• The electron beam is rapidly moved to bombard
the tungsten arcs, producing an x-ray beam.
• The x-rays then pass through the patient and
transmission information is collected by the
detectors.
• Capable of producing 17slice/second.
Sixth Generation: Helical CT
• In previous generation detectors, the gantry had to be
stopped after ever slice, so that data acquisition could
not be a continuous process.
• This problem was solved in the 1990's when slip ring
technology was introduced to the field of medical
imaging.
• Three technological developments were required: slip
ring gantry designs, very high power x-ray tubes,
and interpolation algorithms.
• Interpolation is essentially a weighted average of the
data from either side of the reconstruction plane, with
slightly different weighting factors used for each
projection angle.
Z-AXIS
Seventh Generation: Multiple Detector Array
• The most recent generation of CT scanner consists
of a multiple detector array and a cone shaped x-ray
beam.
• Unlike the pencil beam and fan beam, the cone
beam does not pass through a narrow collimator.
Therefore, the intensity of the initial x-ray beam is
not as strongly reduced and hence can interact more
efficiently and effectively with the detector array .
• In order to use a cone beam x-ray geometry, the
linear detector array found in previous generations
of CT scanners had to be modified to multiple
detector array.
• With multiple detector array scanner, slice
thickness is determined by the detector size, not by
the collimator.
• Rows or detector along z-axis are increased.
Flat panel detector CT ….which generation?????
• Flat-panel digital detectors similar to the ones
used in digital radiography are now being
considered for use in CT
• The detector consists of a cesium iodide (CsI)
scintillator coupled to an amorphous, silicon thin-
film transistor array.
• These flat-panel detectors produce excellent
spatial resolution but lack good contrast
resolution.
• Its fluoroscopic and angiographic capabilities are
useful for intraoperative, vascular, interventional
and intraoperative applications makes it a system
of choice.
• flat-panel detectors are also being investigated for
use in CT of the breast. Fig: Photograph of a prototypic flatpanel
volume CT scanner
Fig: cadaveric temporal bone and 3D model of
auditory apparatus
System Components
• The three major components of CT
scanner are:
1)Gantry
2)Computer and operators console
3)Patient Table
Gantry
• The gantry is a circular device that houses the x-ray
tube, DAS, slip rings and detector array .
• Helical CT unit have continuous slip ring and high
voltage generator in the gantry.
• The gantry can be tilted forward and backward to 30
degree
• aperture measures about 28 inches(71.1cm ) wide to
accommodate variety of patient sizes.
High voltage/frequency Generator
• Xray tube require a high voltage generator to achieve the necessary power required of
an x-ray tube
• AC power will supply x-ray tube with sinusoidal currents, resulting in peak and trough
• A single phase high voltage generator converts this AC into a half or full wave rectified
supply with a measure in the thousands of volts
• The half wave rectification results in a peak voltage that will dip to zero reoccurring:
this will consequently have an effect on this behavior of radiation produced and hence
the name kilovoltage peak (kVp) was born
• The advancement of highvoltage generator from single phase to three phase to
constant potential generators have overcome this voltage ripple creating a
continuous, uninterrupted voltage.
• Modern x-ray units, which largely utilize constant high voltage generators have voltage
ripple of lee than 1% and consequently employ the term kV rather than kVp.
High frequency generator
• Ratings can range from 20 to 100 kilowatts (kW; Kalender, 2005). More
recently CT manufacturers have generators capable of 120 kW. An output
capacity of, say, 60 kW will provide a range of kilovolt and milliampere
settings, where 80 and 120 to 140 kV and 20 to 500 milliamperes (mA) with 1-
mA increments are typical.
Filters
• Copper or aluminum filters are used to filter the x-
ray beam.
• The typical filtration on a CT x-ray tube is ~6 mm Al.
• A bow tie filter is used to minimize the dynamic
range of exposures at the detector.
• Bow tie filters attenuate little in the center, but
attenuation increases with increasing distance from
the central ray.
• Bow tie filters are made of a low Z material such
as Teflon to minimize beam hardening differences.
• Bow tie filters also reduce scatter and patient dose.
Fig: Two types of beam-shaping filters
for
use in CT
Via: Radiology key
Collimators
• Collimation usually refers to the act of
constraining the x-ray beam.
• In the early step and shoot scanner the
beam collimation =slice thickness.
Pre patient collimation
-determines dose profile and patient dose.
Pre detector collimation
-restrict the X-ray beam viewed by
detector array, this reduces the scatter on
detector array. when properly coupled
with pre patient collimator define the
slice thickness.
-It reduces scatter radiation there by
improving contrast. Detector collimation
Beam collimation
Fig: 3D collimator
Over ranging and Over beaming in MDCT
Over-ranging
Over-beaming
•Over beaming “relates to x-ray beams being
slightly wider than the detector which means that
patients are exposed over a small area without the
signal being detected”
•Over ranging “refers to exposure of the patient
outside the imaged range which occurs for spiral CT
with multi-row detectors at the start and the end of
the scan
(Kalender, 2014)
Adaptive Section Collimation
• The problems of over ranging and
overbearing can be solved using a
technique called adaptive section
collimation.
• Parts of the x-ray beam exposing tissue
outside of the volume to be imaged are
blocked in the z-direction by
dynamically adjusted collimators at the
beginning and at the end of the CT
scan
(Deak et al., 2009).
Data Acquisition System
1)The DAS system:
• ‘reads’ the measurements from the detector
array,
• converts these analogue signals into digital
format,
• and transmits the digital signal to the
computer systems for reconstruction into
the presented images.
FIG: Data acquisition system (DAS) channels are
used to specify slice width and thickness in multi-
detector CT by electrically coupling detectors.
Slip ring Technology
• Slip rings are electromechanical conducting brushes that
transmits power to gantry components for its continuous
rotation.
• Due to the slip ring technology, helical scanning is possible.
• One surface is a smooth ring and other a ring with brushes
that sweep the smooth ring.
• In a slip ring gantry system, power and electrical signals are
transmitted through stationary rings within the gantry.
• Thus eliminating the need for electrical cables.
• Recently GE have introduced the contactless whisper slip
ring technology.(Optical fibers rely on optical radiation to
transfer the data and operate typically at infrared
wavelength between 850 and 1550 nm allowing EMI free
transmission at very high rate of several dozens of Gbps)
Slip ring contd.
• Two slip-ring designs are the disk (Fig A) or
pancake type (Fig B)
• In the disk design, the conductive rings form
concentric circles in the plane of rotation
• The cylindrical design includes conductive rings
positioned along the axis of rotation to form a
cylinder
• The brushes that transmit electrical power to the
CT components glide in contact grooves on the
stationary slip ring.
Fig: A Fig: B
Slip ring contd.
• Two common brush designs are the wire brush and the
composite brush
• Two brushes per ring are often used to increase either
communication reliability or current carrying capacity.
• The composite brush uses a block of some conductive
material (e.g., a silver graphite alloy) as a sliding contact.
• This design makes it possible to transfer electrical energy
across a rotating interface without the use of electrical
contacts.
Types of slip ring based on power supply
• low-voltage slip-ring system :AC power and x-
ray controlling signal are passed to slip ring by
low voltage brushes then slip ring provides
power to high voltage transformer then high
voltage is provided to x-ray tube. In this case,
the x-ray generator, x-ray tube, and other
controls are positioned on the orbital scan
frame.
• High voltage slip ring system: AC delivers
power to the high-voltage generator ,which
subsequently supply to x-ray tube. In this case,
the high-voltage generator does not rotate with
the x-ray tube.
CT x-ray tube
• First- and second-generation scanners used fixed anode, oil-cooled x-ray tubes.
• Rotating anode x-ray tubes have become common in CT because of the demand
for increased output.
• The disk is usually made of (RTM) alloy and other materials with a small target
angle (12 degrees) and a rotation speed of 3600 to 10,000 rpm
• The introduction of spiral/helical CT with continuous rotation scanners has
placed new demands on x-ray tubes.
• Because the tube rotates continually for a longer period compared with
conventional scanners, the tube must be able to sustain higher power levels.
Technical advancement's
• Electrical arcing results from tungsten deposits on the
glass caused by vaporization was solved by replacing
metal envelops.
• Metal envelope tubes have larger anode disks 200-mm
diameter compared with the 120- to 160-mm diameter
typical of conventional tubes.
• Heat-storage capacity is also increased with an
improvement in heat dissipation rates
• The cathode assembly consists of one or more tungsten
filaments positioned in a focusing cup.
• Working life of recent tube are 10000 to 40000 hours
compared to 1000hours of conventional tube.
• Heat storage capacity of modern tube is 5-8MHUor more
compared to 1-3MHU of early 3rd generation scanner
tube.
Upgrade
Disc design of x-ray tube in modern CT
The brazed graphite anode disk consists of a tungsten-
rhenium focal track brazed to a graphite base body. Graphite
increases the heat storage capacity because of its high
thermal capacity, which is about 10 times that of tungsten.
CVD graphite disk consists of a graphite base
body with a tungsten-rhenium layer deposited
on the focal track by a chemical vapor process.
(Fox, 1995).
Fig: The anode assembly of a modern x-ray tube used in CT
STRATON X-RAY TUBE
• Siemens introduced a new design of x-ray tube in
which entire tube body rotates, rather than just
the anode, as is the case with conventional
designs. This change allows all the bearings to be
located outside the evacuated tube & enables the
anode to be cooled more efficiently.
• Anode is in direct contact with the oil.
The Straton has a low inherent heat capacity of
0.8 MHU, but an extremely fast cooling rate of 5
MHU / min.
The heat capacity & cooling rate combine to
produce a tube which Siemens claim is ‘0 MHU’
The beam is deflected to strike the anode at two
precisely located focal spots that vary in size.
the sizes can be 0.6 mm × 0.17 mm, 0.8 mm ×
1.1 mm, and 0.7 mm × 0.7 mm.
The electron beam alternates at about 4640 times
per second to create two separate x-ray beams
that pass through the patient and fall on the
detectors.
Comparison of Straton tube with conventional tube
MAXIMUS ROTALIX CERAMIC (MRC) TUBE
• Based on the technology of spiral
groove bearing using liquid metal
alloy as lubricant.
• The excellent heat dissipation of
the bearings via the liquid metal
lubricant gave the tube higher
cooling capacity.
• Noiseless rotating anode and had
a very long lifetime.
• Replaced ball bearings which has
to run in vacuum, conventional oil
based lubricants could not be
used.
LIMAX (Liquid Metal Anode X-ray) by Philips
• Liquid metal jet of SnPb, GaInSn, turbulently
streaming through a tube close to the cathode, is
heated at the focal spot & subjected to fast
electrons. .
• While the heated material is transported through
the tubing, cold metal enters the focal spot area &
cooled effectively by circulation through a heat
exchanger.
• Liquid metal separated from the vacuum by a
diamond, tungsten or molybdenum
• The requirements on mechanical stability
(especially at elevated temperatures) and
transparency to electrons (in the energy range of
50 – 150 keV) are much more demanding.
CT Detectors
When the x-rays passes through the patient the x-rays that were not attenuated
by the patient passes and reaches the film in screen-film radiography .
A similar process occurs in CT but the film is replaced by the detectors.
In CT there are two main types of detectors
1) Xenon gas detectors
2)Scintillation detectors
Detector Characteristics
• Detectors exhibit several characteristics essential for CT image production
affecting good image quality. Some of them are as mentioned below:
1. Efficiency :refers to the ability to capture, absorb, and convert x-ray photons
to electrical signals. CT detectors must possess high capture efficiency,
absorption efficiency, and conversion efficiency.
2. Dynamic range: The dynamic range describes the range of x-ray intensities a
detector can differentiate. A high dynamic range provides the discrimination
between small differences in x-ray attenuation., the dynamic range is 1
million to 1)
3. Afterglow :refers to the persistence of the image even after the radiation has
been turned off. CT detectors should have low afterglow values, such as less
than 0.01 percentage, 100 milliseconds after the radiation has been
terminated
4. Stability :refers to the steadiness of the detector response. If the system is not
stable, frequent calibrations are required to render the signals useful.
Xenon Detector (Convert x-ray energy directly into electrical signal)
• Xenon detectors consist of two metal electrodes (an anode and a
cathode) surrounding high-pressure(25-30atm pressure) xenon gas.
• X-ray beam pass through the patient, it interacts with the xenon
atoms ionizing the gas.
• Ionization of xenon causes a build up of electrons within the
compartment.
• The electrons move toward the positively charged anode and
continue to build up so that a charge accumulates on the anode.
• Charge is then amplified into an electronic signal.
• Xenon gas is used because of its ability to remain stable under
pressure ,it has QDE(60-87%)
• It was used in third generation CT
• Xenon gas is the element of choice because of it's ability to remain
stable under extreme amounts of pressure and high atomic no.
Scintillation Detectors(Convert x-ray energy into light)
• The more modern version of the CT detector is a
luminescent detector.
• The solid-state detectors consist of a scintillator and a
photodetector that converts x-rays into light and then
photodiode converts light into electrical signal.
• This type of photodetector is referred to as a photovoltaic
detector array (PDA) and it is based on front illumination.
• Current CT scanners now make use of back-illuminated
PDA.
Major types of detector
Design innovations in CT detectors
FIG: Various types of latest detector
Coverage of 64,128 and 320 rows of
detector along z axis
Image: Robert pelberg 2015
Detectors array configuration
•Matrix Type(uniform)
•Adaptive Type(non-
uniform)
•Hybrid Type(mixed)
Fig:4 slice MDCT Scanner
UFC Detector
• Ultrafast Ceramic is a hard yellow
substances that resembles plastic
and weighs about as much as gold
• It includes rare earth elements
gadolinium , yttrium and sulfur.
• The material is formed through
the process involves mixing,
chemical reduction , sintering and
pressing.
UFC COntd
• Ultrafast Ceramic scintillator material was developed with high X-ray
absorption efficiency, a fast decay behavior and low afterglow for the CT
system with the highest rotation speed and the shortest integration time .
Stellar Detector
• Third generation CT detector developed by siemens.
• Major components of the Stellar Detector, which include the
UFC scintillator, the back-illuminated photodiode, and the metal
oxide semiconductor (CMOS) wafer that includes the ADC, and a
ceramics substrate.
• The principle of operation is as follows: the light emitted from
the UFC scintillator reaches the backside-illuminated
photodiode, a digital signal is then produced on the other side
of the wafer.
• Stellar detectors can measure smaller signals over a wide
dynamic range which reduces the noise in CT images and
enhances CT image quality.
• The electrical components like ADC are integrated together at
the back of scintillation layer so the electronic noise can be
reduced compared to the second generation detector as shown
in fig 2 Fig: 2
Fig : 1
Stellar detector contd..
Fig: Major components of Stellar Detector
Fig: photograph of the detector array
Fig: hip phantom with a resolution insert that shows the
improvement in image quality using the Stellar Detector and a
conventional CT detector
Gem Stone detector
• Transparent polycrystalline scintillator
• It has higher sensitivity to radiation and
allow faster sampling rate.
• It is used in single source ultrafast dual
energy switching , promising almost
simultaneous spatial and temporal
registration and material decomposition.
• It has primary decay time only 30nsec i.e.
100 fold faster than conventional scintillator.
• The most advantage of gemstone detector is
the improved spatial resolution.( high
definition imaging up to 230mm resolution.)
• Primary speed is only 0.03 µs, 100 times
faster than GOS and is the fastest scintillator
in CT industry, at 40ms, the afterglow is
0.001%, only 25% of GOS
Patient couch
• The table is automated device linked to the computer
and gantry and move in increment according to scan
programme.
• Indexing must be accurate and reliable, especially
when thin slices(1 or 2 mm ) are taken through the
area of interest.
• Ct table are made of up low density carbon
composite which supports the patient without
causing artefact
• All ct table have maximum patient weight limit this
limit varies by manufacturing from 136-272kg
Computer System
• Unique component of the CT system.
• Sufficient speed & memory to solve several thousands
calculations simultaneously.
• Is designed to control data acquisition, processing, display,
retrieve & storage.
• Calculates the attenuation of the individual voxels using
algorithm.
• Calculations of the CT numbers must be very fast to
produce images for immediate viewing.
Operator’s Console
• Permits control of all scan parameter
including selecting proper technical
factor, movement of the gantry and
patient table
• Commands computer to reconstruct and
transfer of image data for storage in data
file
• Pre programmed with the kV an mA
values for individual anatomic parts
Display console
• In the display and manipulation of grayscale
images for diagnosis, it is important to optimize
image fidelity (i.e., the faithfulness with which
the device can display the image)
• This is influenced by physical characteristics
such as luminance, resolution, noise, and
dynamic range.
• Resolution, however, is an important physical
parameter of the grayscale display monitor and
is related to the size of the pixel matrix, or
matrix size.
• The display matrix can range from 64 × 64 to
1024 × 1024, but high-performance monitors
can display an image with a 2048 × 2048
matrix(Dwyer et al., 1992).
Other Accessories
• Head rest/support
• Table straps for immobilization
• Automatic contrast injector
• ECG machine
• Phantoms for quality, performance and dose test
• Emergency trolly
• Radiation protective materials
Bir CT Specifications ( Philips Ingenuity -128 slice CT
• Xray tube
Anode heat storage:8MHU, Anode cooling rate: 1608KHU/min
dual focus (1x1mm)and (0.5x0.1mm),Generator rating 80 KW,
KV Range: 80-140,MA range: 20-665
gantry rotation:0.4 sec for 360 degree rotation with 70 cm gantry aperture
• Detector
scintillation detector ,gadolinium oxysulphide
no of detector rows 64 (128 slice per rotation by z-flying focal spot technique)
• Reconstruction
iterative reconstruction (idose4)
standard reconstruction matrix:512x512
standard reconstruction speed:25images/sec
temporal resolution : 0.053sec
operating monitor:2 lcd monitors with screen resolution of 1280x1024
Bibliography
• Computed Tomography (CT or CAT Scan). RadiologyInfo.org Website: http://radiologyinfo.org/en/sitemap/modal-
alias.cfm?modal=CT. Published 2013. Updated 2013. Accessed November 20, 2013.
• Bharath AA. Introductory Medical Imaging (Synthesis Lectures on Biomedical Engineering). Morgan and Claypool Publishers;
2008.
• Ulzheimer S, Flohr T. Current Technology and Future Direction. In: Multislice CT. Medical Radiology. Germany, Springer
Berlin Heidelberg. 2009; 3-23.
• Dr. EUCLID SEERAM, computed tomography ,Physical Principles, Clinical Applications, and Quality Control
• Multidetector CT in children: current concepts and dose reduction strategiesRutger A. J. Nievelstein, Ingrid M. van
Dam & Aart J. van der Molen
• Robert Pelberg Cardiac CT Angiography Manual Second Edition
• Flat-Panel Volume CT: Fundamental Principles, Technology, and Applications Rajiv Gupta, MD, PhD • Arnold C. Cheung, MD
• A Comparison of Front- and Backside-Illuminated High-Saturation Power Partially Depleted Absorber Photodetecters Xiaowei
Li, Ning Li, Stéphane Demiguel, Joe C. Campbell, Fellow, IEEE, David Tulchinsky, and Keith J. Williams,
• Liquid metal anode x-ray tube Bernd David a , Hans Barschdorfa , Volker Doormanna , Rainer Eckarta , Geoffrey Hardinga ,
JensPeter Schlomkaa , Axel Thrana
• ADVANCES IN CT IMAGING (NJ PELC, SECTION EDITOR) CT Systems Thomas Flohr second edition.
• Instrumentation and Principles of CT Douglas P. Boyd
• Computed Tomography: Ian A. Cunningham Victoria Hospital, the John P. Robarts Research Institute, and the University of
Western Ontario Philip F. Judy Brigham and Women’s Hospital and Harvard Medical School
• Radiologic Science for Technologist: physics, biology and protection tenth edition, Stewart Carlyle Bushong
Questions
1. What is the advantage of slip ring technology? What were the technological
developments of helical CT?
2. What do you understand by overranging and overbeaming? How are
scintillation detectors different from gas ionization detectors?
3. What are the type of detector array configuration ? Which type is commonly
used in modern CT scanners?
4. How is modern CT tube different from Conventional tube ?
5. What is the difference between Straton ,maximus rotalix and liquid metal
xray tube ?

Ct instrumentation and types of detector configuration

  • 1.
    CT INSTRUMENTATION AND TYPESOF DETECTOR CONFIGURATION Presenter :Sujan Karki B.Sc. MIT 3rd year National Academy of Medical Sciences(NAMS) Bir Hospital
  • 2.
  • 3.
    Introduction • Computed tomography(CT)is a revolutionary tool of medicine, particularly in medical imaging. • tomos = slice, graphein = to write • Method for acquiring and reconstructing an image of a thin cross section of an object • Based on measurement of x-ray attenuation through the section plane using many different projection. Image courtesy :Damien Hirst Autopsy with Sliced Human Brain 2004
  • 4.
    Historical perspectives • Inventionof CT was made possible through the work of several individuals, most notably Godfrey Newbold Hounsfield and Allan MacLeod Cormack . • In 1967, Hounsfield was investigating pattern recognition and reconstruction techniques by using the computer. • He concluded that if x-rays are passed through an object from all direction and if are measured then information of internal structure could be obtained. • The radiation used was from an americium 241 gamma source coupled with a crystal detector. • Because of the low radiation output, the apparatus took about 9 days to scan the object. • Because this procedure was too long, various modifications were made and the gamma radiation source was replaced by a powerful x-ray tube. (Hounsfield, 1980). Fig: second prototype CT scanner
  • 5.
    Limitation of conventionaltomography and radiography The major shortcoming of radiography is that the superimposition of all structures on the radiograph (3D collapsed to 2D) and Low soft-tissue contrast The limitations of tomography include persistent image blurring that cannot be completely removed, degradation of image contrast because of the presence of scattered radiation created by the open geometry of the x-ray beam, and other problems resulting from film- screen combinations.
  • 6.
    Imaging before CT •entire body areas were inaccessible to radiography - brain, mediastinum, retroperitoneum • diagnostic procedures showing better detail in these areas were potentially harmful and or poorly tolerated by the patient - pneumoencephalography, diagnostic laparotomy Fig: ventriculography Fig: pneumoencephalography Fig: preparing patient and positioning during pneumoencephalogram.
  • 7.
    Computed Tomography Developmentalchronology • 1924 - mathematical theory of tomographic image reconstructions (Johann Radon) • 1930 - conventional tomography (A. Vallebona) • 1963 - theoretical basis of CT (A. McLeod Cormack) • 1972 - first commercial CT (Sir Godfrey Hounsfield) • 1979 - Nobel price (Cormack & Hounsfield) • 1989 - single slice helical CT scanner • 1992 – dual slice CT scanner • 2004 – 64 slice CT scanner • 2006 – dual source CT scanner • 2007 – 320 slice CT scanner (per revolution) • 2011 – Steller detector • 2012 – 640 slice scanner • 2015 – GE introduced 512 slice ct with 256 rows of detector • 2018 – 360 degree rotation only in 0.24 sec (x-ray-tube)
  • 8.
    Original EMI Scanner •Designed specifically for brain imaging. • Head is enclosed in a water bath between x-ray tube and pair of detectors. • Patient remains stationary and gantry move through rotation and translation motion • Linear motion was repeated 180 times and 1degree gantry rotation after each linear motion. • The x-ray tube was on throughout linear motion and off during rotary motion. • The transmitted radiation was measured 160 times in each linear movement. (180x160=28,800) • Each pair of tomographic sections took 5 minutes, so scan time was 25 minutes. Fig: EMI Mark 1
  • 9.
    EMI contd. • CTimage was reconstructed and displayed in 80x80 matrix in 2 different formats: a paper printout and visual image on cathode ray tube. • The x-ray field was collimated to 3x3x13 mm. • Data from the measurement were computer processed and presented in 3x3 mm of patient cross section. • An oil cooled stationary x-ray tube was used with focal spot size 2.25x12mm and operated at 120kVp and 33mA. • A large difference in attenuation between adjacent areas such as air surrounding the head makes computation complicated so water bath was used. Fig: pixel and voxel in EMI scanner
  • 10.
    First generation :Rotate/Translate,Pencil Beam • Original EMI unit was first generation scanner. • Used a pinhole collimator • NaI as detector element with PM tube • X-ray tube and detector system translate linearly across the 24 cm FOV acquiring 160 parallel rays • pencil beam and a single detector. • A five view study of head took about 25- 30 minutes. fig: Image of first generation CT
  • 11.
    Second generation :Rotate/Translate,Narrow Fan Beam • Narrow fan beam was used with an angle of 10-30 degree and detector 5-30 detectors. • with 10 degree rotation increment, only 18 translation would be required for 180 degree image acquisition. • Produced a tomographic section between 20-90 seconds. • Using fan beam increased radiation intensity towards edge and was compensated by bow tie filter. • Acquired more data (600 rays X 540 views=32400 data point) • Limited only for head scan .
  • 12.
    Third generation :Rotate/Rotate, Wide Fan Beam • Translational motion was oblirated. • The path traced by the tube describe circle rather than the semicircle characteristic of 1st and 2nd generation CT scanner. • The main goal was to cut the acquisition time to less than 20 seconds. • Wide aperture fan beam angle ranging between 40 to 60 degrees. • This new array consisted of 400 to 1000 detector elements. • Major disadvantages: ring artefact(occasional ) • Software corrected image reconstruction algorithms now remove such artefacts.
  • 13.
    Fourth generation :Rotate/Stationary • Developed specifically to alleviate the ring artifacts produced by the third generation. • Removing the detectors from the rotating gantry and putting them in a stationary ring around the patient. • This stationary 360 degree ring of detectors required an increased number of detector elements (~5000 total). • x-ray tube can rotate either outside or inside the detector ring. • If the x-ray tube rotates outside of the detector ring, it is crucial that the detector ring is tilted so that the x-rays only interact with the detectors once they pass through the patient .
  • 14.
    Fifth generation :Stationary/Stationary • Developed for cardiac imaging, the EBCT reduces scan time to as little as 50 ms. • does not use an x-ray tube; instead it uses a beam of electrons generated outside the gantry. • Inside the gantry there are 180-degree rows of fixed detectors on one side and 180 degrees of tungsten arcs opposite. • The electron beam is rapidly moved to bombard the tungsten arcs, producing an x-ray beam. • The x-rays then pass through the patient and transmission information is collected by the detectors. • Capable of producing 17slice/second.
  • 15.
    Sixth Generation: HelicalCT • In previous generation detectors, the gantry had to be stopped after ever slice, so that data acquisition could not be a continuous process. • This problem was solved in the 1990's when slip ring technology was introduced to the field of medical imaging. • Three technological developments were required: slip ring gantry designs, very high power x-ray tubes, and interpolation algorithms. • Interpolation is essentially a weighted average of the data from either side of the reconstruction plane, with slightly different weighting factors used for each projection angle. Z-AXIS
  • 16.
    Seventh Generation: MultipleDetector Array • The most recent generation of CT scanner consists of a multiple detector array and a cone shaped x-ray beam. • Unlike the pencil beam and fan beam, the cone beam does not pass through a narrow collimator. Therefore, the intensity of the initial x-ray beam is not as strongly reduced and hence can interact more efficiently and effectively with the detector array . • In order to use a cone beam x-ray geometry, the linear detector array found in previous generations of CT scanners had to be modified to multiple detector array. • With multiple detector array scanner, slice thickness is determined by the detector size, not by the collimator. • Rows or detector along z-axis are increased.
  • 17.
    Flat panel detectorCT ….which generation????? • Flat-panel digital detectors similar to the ones used in digital radiography are now being considered for use in CT • The detector consists of a cesium iodide (CsI) scintillator coupled to an amorphous, silicon thin- film transistor array. • These flat-panel detectors produce excellent spatial resolution but lack good contrast resolution. • Its fluoroscopic and angiographic capabilities are useful for intraoperative, vascular, interventional and intraoperative applications makes it a system of choice. • flat-panel detectors are also being investigated for use in CT of the breast. Fig: Photograph of a prototypic flatpanel volume CT scanner Fig: cadaveric temporal bone and 3D model of auditory apparatus
  • 18.
    System Components • Thethree major components of CT scanner are: 1)Gantry 2)Computer and operators console 3)Patient Table
  • 19.
    Gantry • The gantryis a circular device that houses the x-ray tube, DAS, slip rings and detector array . • Helical CT unit have continuous slip ring and high voltage generator in the gantry. • The gantry can be tilted forward and backward to 30 degree • aperture measures about 28 inches(71.1cm ) wide to accommodate variety of patient sizes.
  • 20.
    High voltage/frequency Generator •Xray tube require a high voltage generator to achieve the necessary power required of an x-ray tube • AC power will supply x-ray tube with sinusoidal currents, resulting in peak and trough • A single phase high voltage generator converts this AC into a half or full wave rectified supply with a measure in the thousands of volts • The half wave rectification results in a peak voltage that will dip to zero reoccurring: this will consequently have an effect on this behavior of radiation produced and hence the name kilovoltage peak (kVp) was born • The advancement of highvoltage generator from single phase to three phase to constant potential generators have overcome this voltage ripple creating a continuous, uninterrupted voltage. • Modern x-ray units, which largely utilize constant high voltage generators have voltage ripple of lee than 1% and consequently employ the term kV rather than kVp.
  • 21.
    High frequency generator •Ratings can range from 20 to 100 kilowatts (kW; Kalender, 2005). More recently CT manufacturers have generators capable of 120 kW. An output capacity of, say, 60 kW will provide a range of kilovolt and milliampere settings, where 80 and 120 to 140 kV and 20 to 500 milliamperes (mA) with 1- mA increments are typical.
  • 22.
    Filters • Copper oraluminum filters are used to filter the x- ray beam. • The typical filtration on a CT x-ray tube is ~6 mm Al. • A bow tie filter is used to minimize the dynamic range of exposures at the detector. • Bow tie filters attenuate little in the center, but attenuation increases with increasing distance from the central ray. • Bow tie filters are made of a low Z material such as Teflon to minimize beam hardening differences. • Bow tie filters also reduce scatter and patient dose. Fig: Two types of beam-shaping filters for use in CT Via: Radiology key
  • 23.
    Collimators • Collimation usuallyrefers to the act of constraining the x-ray beam. • In the early step and shoot scanner the beam collimation =slice thickness. Pre patient collimation -determines dose profile and patient dose. Pre detector collimation -restrict the X-ray beam viewed by detector array, this reduces the scatter on detector array. when properly coupled with pre patient collimator define the slice thickness. -It reduces scatter radiation there by improving contrast. Detector collimation Beam collimation Fig: 3D collimator
  • 24.
    Over ranging andOver beaming in MDCT Over-ranging Over-beaming •Over beaming “relates to x-ray beams being slightly wider than the detector which means that patients are exposed over a small area without the signal being detected” •Over ranging “refers to exposure of the patient outside the imaged range which occurs for spiral CT with multi-row detectors at the start and the end of the scan (Kalender, 2014)
  • 25.
    Adaptive Section Collimation •The problems of over ranging and overbearing can be solved using a technique called adaptive section collimation. • Parts of the x-ray beam exposing tissue outside of the volume to be imaged are blocked in the z-direction by dynamically adjusted collimators at the beginning and at the end of the CT scan (Deak et al., 2009).
  • 26.
    Data Acquisition System 1)TheDAS system: • ‘reads’ the measurements from the detector array, • converts these analogue signals into digital format, • and transmits the digital signal to the computer systems for reconstruction into the presented images. FIG: Data acquisition system (DAS) channels are used to specify slice width and thickness in multi- detector CT by electrically coupling detectors.
  • 27.
    Slip ring Technology •Slip rings are electromechanical conducting brushes that transmits power to gantry components for its continuous rotation. • Due to the slip ring technology, helical scanning is possible. • One surface is a smooth ring and other a ring with brushes that sweep the smooth ring. • In a slip ring gantry system, power and electrical signals are transmitted through stationary rings within the gantry. • Thus eliminating the need for electrical cables. • Recently GE have introduced the contactless whisper slip ring technology.(Optical fibers rely on optical radiation to transfer the data and operate typically at infrared wavelength between 850 and 1550 nm allowing EMI free transmission at very high rate of several dozens of Gbps)
  • 28.
    Slip ring contd. •Two slip-ring designs are the disk (Fig A) or pancake type (Fig B) • In the disk design, the conductive rings form concentric circles in the plane of rotation • The cylindrical design includes conductive rings positioned along the axis of rotation to form a cylinder • The brushes that transmit electrical power to the CT components glide in contact grooves on the stationary slip ring. Fig: A Fig: B
  • 29.
    Slip ring contd. •Two common brush designs are the wire brush and the composite brush • Two brushes per ring are often used to increase either communication reliability or current carrying capacity. • The composite brush uses a block of some conductive material (e.g., a silver graphite alloy) as a sliding contact. • This design makes it possible to transfer electrical energy across a rotating interface without the use of electrical contacts.
  • 30.
    Types of slipring based on power supply • low-voltage slip-ring system :AC power and x- ray controlling signal are passed to slip ring by low voltage brushes then slip ring provides power to high voltage transformer then high voltage is provided to x-ray tube. In this case, the x-ray generator, x-ray tube, and other controls are positioned on the orbital scan frame. • High voltage slip ring system: AC delivers power to the high-voltage generator ,which subsequently supply to x-ray tube. In this case, the high-voltage generator does not rotate with the x-ray tube.
  • 31.
    CT x-ray tube •First- and second-generation scanners used fixed anode, oil-cooled x-ray tubes. • Rotating anode x-ray tubes have become common in CT because of the demand for increased output. • The disk is usually made of (RTM) alloy and other materials with a small target angle (12 degrees) and a rotation speed of 3600 to 10,000 rpm • The introduction of spiral/helical CT with continuous rotation scanners has placed new demands on x-ray tubes. • Because the tube rotates continually for a longer period compared with conventional scanners, the tube must be able to sustain higher power levels.
  • 32.
    Technical advancement's • Electricalarcing results from tungsten deposits on the glass caused by vaporization was solved by replacing metal envelops. • Metal envelope tubes have larger anode disks 200-mm diameter compared with the 120- to 160-mm diameter typical of conventional tubes. • Heat-storage capacity is also increased with an improvement in heat dissipation rates • The cathode assembly consists of one or more tungsten filaments positioned in a focusing cup. • Working life of recent tube are 10000 to 40000 hours compared to 1000hours of conventional tube. • Heat storage capacity of modern tube is 5-8MHUor more compared to 1-3MHU of early 3rd generation scanner tube. Upgrade
  • 33.
    Disc design ofx-ray tube in modern CT The brazed graphite anode disk consists of a tungsten- rhenium focal track brazed to a graphite base body. Graphite increases the heat storage capacity because of its high thermal capacity, which is about 10 times that of tungsten. CVD graphite disk consists of a graphite base body with a tungsten-rhenium layer deposited on the focal track by a chemical vapor process. (Fox, 1995). Fig: The anode assembly of a modern x-ray tube used in CT
  • 34.
    STRATON X-RAY TUBE •Siemens introduced a new design of x-ray tube in which entire tube body rotates, rather than just the anode, as is the case with conventional designs. This change allows all the bearings to be located outside the evacuated tube & enables the anode to be cooled more efficiently. • Anode is in direct contact with the oil. The Straton has a low inherent heat capacity of 0.8 MHU, but an extremely fast cooling rate of 5 MHU / min. The heat capacity & cooling rate combine to produce a tube which Siemens claim is ‘0 MHU’ The beam is deflected to strike the anode at two precisely located focal spots that vary in size. the sizes can be 0.6 mm × 0.17 mm, 0.8 mm × 1.1 mm, and 0.7 mm × 0.7 mm. The electron beam alternates at about 4640 times per second to create two separate x-ray beams that pass through the patient and fall on the detectors.
  • 35.
    Comparison of Stratontube with conventional tube
  • 36.
    MAXIMUS ROTALIX CERAMIC(MRC) TUBE • Based on the technology of spiral groove bearing using liquid metal alloy as lubricant. • The excellent heat dissipation of the bearings via the liquid metal lubricant gave the tube higher cooling capacity. • Noiseless rotating anode and had a very long lifetime. • Replaced ball bearings which has to run in vacuum, conventional oil based lubricants could not be used.
  • 37.
    LIMAX (Liquid MetalAnode X-ray) by Philips • Liquid metal jet of SnPb, GaInSn, turbulently streaming through a tube close to the cathode, is heated at the focal spot & subjected to fast electrons. . • While the heated material is transported through the tubing, cold metal enters the focal spot area & cooled effectively by circulation through a heat exchanger. • Liquid metal separated from the vacuum by a diamond, tungsten or molybdenum • The requirements on mechanical stability (especially at elevated temperatures) and transparency to electrons (in the energy range of 50 – 150 keV) are much more demanding.
  • 38.
    CT Detectors When thex-rays passes through the patient the x-rays that were not attenuated by the patient passes and reaches the film in screen-film radiography . A similar process occurs in CT but the film is replaced by the detectors. In CT there are two main types of detectors 1) Xenon gas detectors 2)Scintillation detectors
  • 39.
    Detector Characteristics • Detectorsexhibit several characteristics essential for CT image production affecting good image quality. Some of them are as mentioned below: 1. Efficiency :refers to the ability to capture, absorb, and convert x-ray photons to electrical signals. CT detectors must possess high capture efficiency, absorption efficiency, and conversion efficiency. 2. Dynamic range: The dynamic range describes the range of x-ray intensities a detector can differentiate. A high dynamic range provides the discrimination between small differences in x-ray attenuation., the dynamic range is 1 million to 1) 3. Afterglow :refers to the persistence of the image even after the radiation has been turned off. CT detectors should have low afterglow values, such as less than 0.01 percentage, 100 milliseconds after the radiation has been terminated 4. Stability :refers to the steadiness of the detector response. If the system is not stable, frequent calibrations are required to render the signals useful.
  • 40.
    Xenon Detector (Convertx-ray energy directly into electrical signal) • Xenon detectors consist of two metal electrodes (an anode and a cathode) surrounding high-pressure(25-30atm pressure) xenon gas. • X-ray beam pass through the patient, it interacts with the xenon atoms ionizing the gas. • Ionization of xenon causes a build up of electrons within the compartment. • The electrons move toward the positively charged anode and continue to build up so that a charge accumulates on the anode. • Charge is then amplified into an electronic signal. • Xenon gas is used because of its ability to remain stable under pressure ,it has QDE(60-87%) • It was used in third generation CT • Xenon gas is the element of choice because of it's ability to remain stable under extreme amounts of pressure and high atomic no.
  • 41.
    Scintillation Detectors(Convert x-rayenergy into light) • The more modern version of the CT detector is a luminescent detector. • The solid-state detectors consist of a scintillator and a photodetector that converts x-rays into light and then photodiode converts light into electrical signal. • This type of photodetector is referred to as a photovoltaic detector array (PDA) and it is based on front illumination. • Current CT scanners now make use of back-illuminated PDA.
  • 42.
  • 43.
    Design innovations inCT detectors FIG: Various types of latest detector Coverage of 64,128 and 320 rows of detector along z axis Image: Robert pelberg 2015
  • 44.
    Detectors array configuration •MatrixType(uniform) •Adaptive Type(non- uniform) •Hybrid Type(mixed) Fig:4 slice MDCT Scanner
  • 45.
    UFC Detector • UltrafastCeramic is a hard yellow substances that resembles plastic and weighs about as much as gold • It includes rare earth elements gadolinium , yttrium and sulfur. • The material is formed through the process involves mixing, chemical reduction , sintering and pressing.
  • 46.
    UFC COntd • UltrafastCeramic scintillator material was developed with high X-ray absorption efficiency, a fast decay behavior and low afterglow for the CT system with the highest rotation speed and the shortest integration time .
  • 47.
    Stellar Detector • Thirdgeneration CT detector developed by siemens. • Major components of the Stellar Detector, which include the UFC scintillator, the back-illuminated photodiode, and the metal oxide semiconductor (CMOS) wafer that includes the ADC, and a ceramics substrate. • The principle of operation is as follows: the light emitted from the UFC scintillator reaches the backside-illuminated photodiode, a digital signal is then produced on the other side of the wafer. • Stellar detectors can measure smaller signals over a wide dynamic range which reduces the noise in CT images and enhances CT image quality. • The electrical components like ADC are integrated together at the back of scintillation layer so the electronic noise can be reduced compared to the second generation detector as shown in fig 2 Fig: 2 Fig : 1
  • 48.
    Stellar detector contd.. Fig:Major components of Stellar Detector Fig: photograph of the detector array Fig: hip phantom with a resolution insert that shows the improvement in image quality using the Stellar Detector and a conventional CT detector
  • 49.
    Gem Stone detector •Transparent polycrystalline scintillator • It has higher sensitivity to radiation and allow faster sampling rate. • It is used in single source ultrafast dual energy switching , promising almost simultaneous spatial and temporal registration and material decomposition. • It has primary decay time only 30nsec i.e. 100 fold faster than conventional scintillator. • The most advantage of gemstone detector is the improved spatial resolution.( high definition imaging up to 230mm resolution.) • Primary speed is only 0.03 µs, 100 times faster than GOS and is the fastest scintillator in CT industry, at 40ms, the afterglow is 0.001%, only 25% of GOS
  • 50.
    Patient couch • Thetable is automated device linked to the computer and gantry and move in increment according to scan programme. • Indexing must be accurate and reliable, especially when thin slices(1 or 2 mm ) are taken through the area of interest. • Ct table are made of up low density carbon composite which supports the patient without causing artefact • All ct table have maximum patient weight limit this limit varies by manufacturing from 136-272kg
  • 51.
    Computer System • Uniquecomponent of the CT system. • Sufficient speed & memory to solve several thousands calculations simultaneously. • Is designed to control data acquisition, processing, display, retrieve & storage. • Calculates the attenuation of the individual voxels using algorithm. • Calculations of the CT numbers must be very fast to produce images for immediate viewing.
  • 52.
    Operator’s Console • Permitscontrol of all scan parameter including selecting proper technical factor, movement of the gantry and patient table • Commands computer to reconstruct and transfer of image data for storage in data file • Pre programmed with the kV an mA values for individual anatomic parts
  • 53.
    Display console • Inthe display and manipulation of grayscale images for diagnosis, it is important to optimize image fidelity (i.e., the faithfulness with which the device can display the image) • This is influenced by physical characteristics such as luminance, resolution, noise, and dynamic range. • Resolution, however, is an important physical parameter of the grayscale display monitor and is related to the size of the pixel matrix, or matrix size. • The display matrix can range from 64 × 64 to 1024 × 1024, but high-performance monitors can display an image with a 2048 × 2048 matrix(Dwyer et al., 1992).
  • 54.
    Other Accessories • Headrest/support • Table straps for immobilization • Automatic contrast injector • ECG machine • Phantoms for quality, performance and dose test • Emergency trolly • Radiation protective materials
  • 55.
    Bir CT Specifications( Philips Ingenuity -128 slice CT • Xray tube Anode heat storage:8MHU, Anode cooling rate: 1608KHU/min dual focus (1x1mm)and (0.5x0.1mm),Generator rating 80 KW, KV Range: 80-140,MA range: 20-665 gantry rotation:0.4 sec for 360 degree rotation with 70 cm gantry aperture • Detector scintillation detector ,gadolinium oxysulphide no of detector rows 64 (128 slice per rotation by z-flying focal spot technique) • Reconstruction iterative reconstruction (idose4) standard reconstruction matrix:512x512 standard reconstruction speed:25images/sec temporal resolution : 0.053sec operating monitor:2 lcd monitors with screen resolution of 1280x1024
  • 56.
    Bibliography • Computed Tomography(CT or CAT Scan). RadiologyInfo.org Website: http://radiologyinfo.org/en/sitemap/modal- alias.cfm?modal=CT. Published 2013. Updated 2013. Accessed November 20, 2013. • Bharath AA. Introductory Medical Imaging (Synthesis Lectures on Biomedical Engineering). Morgan and Claypool Publishers; 2008. • Ulzheimer S, Flohr T. Current Technology and Future Direction. In: Multislice CT. Medical Radiology. Germany, Springer Berlin Heidelberg. 2009; 3-23. • Dr. EUCLID SEERAM, computed tomography ,Physical Principles, Clinical Applications, and Quality Control • Multidetector CT in children: current concepts and dose reduction strategiesRutger A. J. Nievelstein, Ingrid M. van Dam & Aart J. van der Molen • Robert Pelberg Cardiac CT Angiography Manual Second Edition • Flat-Panel Volume CT: Fundamental Principles, Technology, and Applications Rajiv Gupta, MD, PhD • Arnold C. Cheung, MD • A Comparison of Front- and Backside-Illuminated High-Saturation Power Partially Depleted Absorber Photodetecters Xiaowei Li, Ning Li, Stéphane Demiguel, Joe C. Campbell, Fellow, IEEE, David Tulchinsky, and Keith J. Williams, • Liquid metal anode x-ray tube Bernd David a , Hans Barschdorfa , Volker Doormanna , Rainer Eckarta , Geoffrey Hardinga , JensPeter Schlomkaa , Axel Thrana • ADVANCES IN CT IMAGING (NJ PELC, SECTION EDITOR) CT Systems Thomas Flohr second edition. • Instrumentation and Principles of CT Douglas P. Boyd • Computed Tomography: Ian A. Cunningham Victoria Hospital, the John P. Robarts Research Institute, and the University of Western Ontario Philip F. Judy Brigham and Women’s Hospital and Harvard Medical School • Radiologic Science for Technologist: physics, biology and protection tenth edition, Stewart Carlyle Bushong
  • 57.
    Questions 1. What isthe advantage of slip ring technology? What were the technological developments of helical CT? 2. What do you understand by overranging and overbeaming? How are scintillation detectors different from gas ionization detectors? 3. What are the type of detector array configuration ? Which type is commonly used in modern CT scanners? 4. How is modern CT tube different from Conventional tube ? 5. What is the difference between Straton ,maximus rotalix and liquid metal xray tube ?

Editor's Notes