ELECTROPHORESIS
BASIC CONCEPTS,
INSTRUMENTATION, TECHNIQUE,
TYPES AND APPLICATIONS
By
Dr. Basil, B – MBBS (Nigeria),
Department of Chemical Pathology/Metabolic Medicine,
Benue State University Teaching Hospital, Makurdi.
February, 2015.
INTRODUCTION
• ELECTROPHORESIS – migration of charged solutes
or particles in a liquid medium under the influence
of an electric field
• The aim of carrying out electrophoresis include:
– To determine the number, amount and mobility of
components in a given sample or to separate them.
– To obtain information about the electrical double layers
surrounding the particles.
– Determination of molecular weight of proteins and DNA
sequencing.
BASIC CONCEPTS
• Iontophoresis – migration of small ions
• Ionized species move towards the anode or cathode
depending on their charges
• Ampholytes become +vely charged in a soln more
acidic than its isoelectric point (pI), and –vely charged
in a more alkaline soln
• Rate of migration depends on:
– Net electrical charge of the molecule
– Size and shape of the molecule
– Electrical field strength
– Properties of the supporting medium
– Temperature of operation
• Electrophoretic mobility (µ) – rate of migration (cm/s)
per unit field strength (volts/cm)
µ = Q/6∏rȠ
Where:
µ = electrophoretic mobility in cm2/(V)(s)
Q = net charge on the ion
r = ionic radius of the solute
Ƞ = viscosity of the buffer soln
• Other factors affecting mobility:
– Endosmotic flow: This is the preferential movement of
water in one direction through an electrophoresis medium
due to selective binding of one type of charge on the
surface of the medium
– Wick flow: Movement of water from the buffer reservoir
towards the center of an electrophoresis gel or strip to
replace water lost by evaporation
INSTRUMENTATION/REAGENTS
Power Supply:
• Electrophoresis is done under conditions of
constant voltage, current or power.
• Current (I) flow through the medium > heat
production due to resistance (R) > increase in
thermal agitation of dissolved ions > decrease in
resistance and increase in current flow >> more
heat and evaporation of water from the buffer >
increase in ionic concentration of the buffer
• Migration rate is kept constant by using a constant-
current power supply.
• Since, EMF = IR, and increase in R will result in and
increase in EMF (V) at constant current (I), thus, no
appreciable heat production
Buffers:
• Function include:
– Carries the applied current
– Establishes the pH of the system
– Determines the electrical charge on the solute
• The ionic strength influences:
– Conductance of the support
– Thickness of the ionic cloud (buffer and non-buffer ions)
surrounding a charged molecule
– Rate of its migration
– Sharpness of the electrophoretic zone
• Features of an ideal buffer:
– Does not interfere with the ability to detect the analytes of
interest
– Maintains solubility of the analytes
– Maintains buffering capacity through the analysis
– Produce the desired separation
• Commonly used buffers include:
– Low pH (acidic): phosphate, acetate, formate, citrate
– High pH (basic): tris, tricine, borate, CAPS (N-cyclohexyl-3-
aminopropanesulfonic acid)
• Higher the ionic strength (and concentration) > higher
size of ionic cloud > lower mobility of the particle
• Also, higher ionic strength > sharper protein-band
separation and increased heat production > denaturation
of heat-labile proteins
• Buffers used are made of monovalent ions because their
valencies (ionic strength) and molality are equal.
• They are good culture media for microorganisms and
should be refrigerated when not in use,
• Also cold buffer reduces evaporation and improves
resolution
• Small volumes should be discarded after use, but large
volumes can be reused up to 4 times
Support Media:
• Insoluble gels e.g sheets, slabs or columns of starch,
agarose or polyacrylamide
• Membranes (paper) of cellulose acetate
• Pure buffer soln in a capillary
Starch Gel:
• Separate macromolecules on basis of surface charge and
molecular size
• Obsolete because preparation of reproducible starch gel
is difficult
Cellulose Acetate:
• made by treating cellulose with acetic anhydride
• They need to be soaked in buffer to soften them before
use
• Require clearing before densitometry
• Hardly used in routine clinical application
Agarose:
• Used in AGE for the separation of Serum, Urine or
CSF proteins, Hb variants, Isoenzymes,
Lipoproteins etc.
• Separation is based only on charge-to-mass ratio –
pore size is large enough for all proteins to pass
• Separate proteins into only 5 fractions (zones):
albumin, alpha-1, alpha-2, beta- and gamma-
globulins
• Advantages:
– Permits excellent densitometry – lower affinity for
proteins (migration is not affected) and native clarity
after drying
– Little Endosmosis – free of ionizable groups (neutral)
• Disadvantage:
– DNA recovery is affected by inhibitors
Polyacrylamide:
• thermostable, transparent, durable and relatively
chemically inert
• No endosmosis – uncharged
• Pore size does not allow larger proteins like fibrinogen, B1-
lipoprotiens, Y-globulins etc, to migrate
• Separation is based on both charge-to-mass ratio and
molecular mass – molecular sieving
• Carcinogenic – CAUTION when handling
• Accommodates a large amount of sample in a single
sample slot
• DNA recovered is pure with no inhibitors unlike Agarose.
• Separates proteins into 20 or more fractions
• Used to study individual proteins (e.g Isoenzymes)
Equipment used for Gel Electrophoresis:
TECHNIQUE
Separation:
• Blot the hydrated support medium to remove
excess buffer
• The sample is added to the support for about 5min
• The support is then placed into the electrophoretic
chamber in contact with the buffer
• Apply a constant-current or constant-voltage
power for a specified time
• The support is then removed and placed in a
fixative or rapidly dried to prevent diffusion of the
sample
Staining:
• The support is then stained and dried after
washing out the excess dye
• Amido Black B or members of the Coomassie
Brillant Blue series are the commonest dyes
• The amount of dye taken up is dependent on the
type of protein, degree of denaturation by the
fixing agent and quality of the dye
• To visualize isoenzymes, incubate the gel in contact
with a solution of substrate, which is linked
structurally or chemically to a dye, before fixing.
• Typical stain solution may be used several times
• When the stain solution becomes faulty, protein
zones will appear too lightly stained
• It must be stored tightly covered to prevent
evaporation
Detection and Quantification:
• Detection can be achieved using UV light, but
quantification is by Densitometry
• Densitometers integrate the area under a peak, and
the result is printed as percentage of the total
• Reliable quantification of stained zones using
densitometry requires:
– Light of an appropriate wavelength
– Linear response from the instrument
– Transparent background in the strip being scanned
• Mass spectrometers – determine the molecular
weights of proteins and their cleavage products, and
for peptide sequencing.
Blotting Techniques:
• Southern Blotting – widely used in molecular biology:
– DNA or DNA fragments separated by AGE
– Strip of nitrocellulose or a nylon membrane is laid over the
gel blotting the DNA or DNA fragments onto it by capillary,
electro-, or vacuum blotting
– Detection and identification is by hybridization with a
labeled complementary nucleic acid probe.
• Northern Blotting:
– Separates and detects RNAs
– Uses labeled RNA probe for hybridization
• Western Blotting:
– Separates and detects proteins
– The membrane is reacted with a reagent containing an
antibody raised against the protein of interest
Common Problems Encountered:
TYPES OF ELECTROPHORESIS
• Zone Electrophoresis
• Slab Gel Electrophoresis
• Disc Electrophoresis
• Isoelectric Focusing Electrophoresis
• 2-Dimensional Electrophoresis
• Capillary Electrophoresis
• Microchip Electrophoresis
Zone Electrophoresis:
• Produce zones of proteins that are heterogeneous
and physically separated from one another
• Classified according to type and structure of the
support material e.g AGE, CAE, PAGE etc
Slab Gel Electrophoresis:
• Use of a rectangular gel regardless of the thickness
• Main advantage – ability to simultaneously separate
several samples in one run
• Primary method used in clinical chemistry lab
• Gels (usually agarose) may be cast on sheet of plastic
backing or completely encased within a plastic walled
cell allowing horizontal or vertical electrophoresis
and submersion for cooling, if needed.
• May be cast with additives like:
– Ampholytes which create a pH gradient, or
– Sodium dodecyl sulfate (SDS) that denatures protiens
Schematic representation of Slab gel electrophoresis
• Useful in the separation of serum proteins, isoenzymes,
lipoproteins, hemoglobins, and fragments of DNA and RNA
Common problems encountered in SGE include:
• Discontinuities in sample application – dirty applicator;
cleaned by agitating in water
• Unequal migration of samples across the width of the gel
– dirty electrodes (uneven application of electric field),
uneven wetting of the gel
• Distorted protein zones – bent applicator, incorporation of
air bubble during sample application, over application of
sample, excessive drying of support
• Unusual bands – artefacts, other causes e.g hemolysed
samples > increased b-globulin band
• Atypical bands – denatured protein from deteriorated
serum, also rule out a true paraprotein
Disc Electrophoresis:
• 3-gel system:
– small-pore separating gel (running gel),
– a larger-pore spacer gel (stacking gel),
– and a thin layer of large-pore monomer solution (sample
gel) – containing about 3 µL of serum
• The different composition cause discontinuities in
the electrophoresis matrix
• During electrophoresis, all proteins migrate easily
through the large-pore gels and stack up on the
separation gel in a very thin zone
• This improves resolution and concentrates protein
components at the border (or starting zone)
• Separation occurs at the bottom separation gel by
the molecular sieve phenomenon
Schematic representation of Disc Electrophoresis
Isoelectric Focusing Electrophoresis:
• Separates amphoteric compounds with increased
resolution in a medium possessing a stable pH gradient
• The protein migrate to a zone in the medium where the pH
of the gel matches its pI
• At this point, the charge of the protein becomes zero and
its migration ceases – it become “focused”.
• Regions associated with a given pH are very narrow –
enough to separate proteins that differ in their pI values by
only 0.02pH units
• A high voltage power source is needed because carrier
ampholytes are used in relatively high concentrations
• Thus, the electrophoretic matrix must be cooled
• IEF is used in neonatal screening programs to test for
variant Hb
2-Dimensional Electrophoresis:
• Uses charge-dependent IEF (first dimension) and molecular
weight-dependent electrophoresis (in the second)
• 1st dimension – carried out in a large-pore medium like
agarose or large-pore PAG; to which ampholytes are added
to yield a pH gradient
• 2nd dimension is often polyacrylamide in a linear or
gradient format
• It achieves the highest resolving power for the separation
of DNA fragments
– 1st dimension – normal AGE
– 2nd dimension – ethidium bromide is added to the gel to open
up the fragments and cause changes in their mobility
• Method of choice when complex samples need to be
arrayed for characterization, as in proteomics.
Schematic representation of 2D Electrophoresis
Capillary Electrophoresis:
• Separation in narrow-bore fused silica capillaries (inner
diameter 25 – 75µm) filled with buffer – gel media can also
be used
• Sample is loaded after filling capillary with buffer, and
electric field applied
• Borate is a classic CE buffer that generates relatively low
current and heat, even with a high ionic strength
• Electro-osmotic flow (EOF) controls the amount of time
solutes remain in the capillary (also Electroendosmostic
flow or endosmosis)
– EOF = bulk flow of liquid towards the cathode upon application
of an electric field and it is superimposed on electrophoretic
migration
• Cations migrate fastest due to EOF and electrophoretic
attraction towards the cathode
• Neutral molecules are all carried by EOF but not separated
from each other
• Anions move slowest because EOF is slightly greater than
the attraction twds the anode and repulsion from cathode
• Used for determination of MWts of proteins and peptides,
analysis of PCR products, inorganic ions, organic acids,
pharmaceuticals, optic isomers, and drugs of abuse in
serum and urine.
• When using a new capillary or changing buffer, the capillary
must be equilibrated with the buffer – conditioning
– Particularly important when phosphate-containing buffer is
involved
– This should be done for at least 4hrs before electrophoresis is
started
• The capillary surface must be regenerated or reconditioned
to remove any material adsorbed onto the wall
– Done by following each run with 3- to 5-column volume rinse
using NaOH, and flushing with 5- to 8-column volumes of fresh
buffer.
Microchip Electrophoresis:
• Similar in principle to CE, but differs in that
the separation channels, sample injection
channels and reservoirs are all fabricated into
the same planar substrate using
photolithographic processes.
• More so, sample preparationand/or
precolumn or postcolumn reactors, detectors,
and excitation sources are intergrated into the
chip
QUALITY CONTROL
• A Clear SOP must be adhered to during each
run to ensure reproducibility
• Control serum must be included in each
electrophoretic run to evaluate its quality and
that of the densitometer.
APPLICATIONS
• DNA Analysis:
– specific DNA sequences can be analyzed, isolated and cloned.
– The analyzed DNA may be used in forensic investigations and
paternity tests.
• Protein Analysis:
– In the diagnosis of conditions where levels of specific proteins
or total protein is low or higher than normal e.g monoclonal
gammopathy etc.
• Antibiotic Analysis:
– Synthesis of new antibiotics
– Analysis of bacteria response to antibiotics and determining
antibiotic-resistance
• Vaccine Analysis:
– Purification, processing and analysis of vaccines e.g influenza
vaccine, hepatitis vaccine, polio vaccine
REFERENCES:
• Tietz Textbook of Clinical Chemistry and Molecular
Diagnostics, fifth ed., by Burtis et al.
• Clinical Chemistry; Principles, Techniques and
Correlations, 7th ed., by Bishop et al.
• http://www.slideshare.net/jyots23/electrophoresis-
presentation
• http://www.bio-rad.com/en-us/applications-
technologies/protein-electrophoresis-methods
• Landers JP. Molecular diagnostics on electrophoretic
microchips. Anal Chem 2003;75:2919-27
• St. claire iii RL. Capillary electrophoresis. Anal Chem
1996;379-423
THANK YOU FOR
LISTENING

Electrophoresis

  • 1.
    ELECTROPHORESIS BASIC CONCEPTS, INSTRUMENTATION, TECHNIQUE, TYPESAND APPLICATIONS By Dr. Basil, B – MBBS (Nigeria), Department of Chemical Pathology/Metabolic Medicine, Benue State University Teaching Hospital, Makurdi. February, 2015.
  • 2.
    INTRODUCTION • ELECTROPHORESIS –migration of charged solutes or particles in a liquid medium under the influence of an electric field • The aim of carrying out electrophoresis include: – To determine the number, amount and mobility of components in a given sample or to separate them. – To obtain information about the electrical double layers surrounding the particles. – Determination of molecular weight of proteins and DNA sequencing.
  • 3.
    BASIC CONCEPTS • Iontophoresis– migration of small ions • Ionized species move towards the anode or cathode depending on their charges • Ampholytes become +vely charged in a soln more acidic than its isoelectric point (pI), and –vely charged in a more alkaline soln • Rate of migration depends on: – Net electrical charge of the molecule – Size and shape of the molecule – Electrical field strength – Properties of the supporting medium – Temperature of operation
  • 4.
    • Electrophoretic mobility(µ) – rate of migration (cm/s) per unit field strength (volts/cm) µ = Q/6∏rȠ Where: µ = electrophoretic mobility in cm2/(V)(s) Q = net charge on the ion r = ionic radius of the solute Ƞ = viscosity of the buffer soln • Other factors affecting mobility: – Endosmotic flow: This is the preferential movement of water in one direction through an electrophoresis medium due to selective binding of one type of charge on the surface of the medium – Wick flow: Movement of water from the buffer reservoir towards the center of an electrophoresis gel or strip to replace water lost by evaporation
  • 5.
  • 6.
    Power Supply: • Electrophoresisis done under conditions of constant voltage, current or power. • Current (I) flow through the medium > heat production due to resistance (R) > increase in thermal agitation of dissolved ions > decrease in resistance and increase in current flow >> more heat and evaporation of water from the buffer > increase in ionic concentration of the buffer • Migration rate is kept constant by using a constant- current power supply. • Since, EMF = IR, and increase in R will result in and increase in EMF (V) at constant current (I), thus, no appreciable heat production
  • 7.
    Buffers: • Function include: –Carries the applied current – Establishes the pH of the system – Determines the electrical charge on the solute • The ionic strength influences: – Conductance of the support – Thickness of the ionic cloud (buffer and non-buffer ions) surrounding a charged molecule – Rate of its migration – Sharpness of the electrophoretic zone • Features of an ideal buffer: – Does not interfere with the ability to detect the analytes of interest – Maintains solubility of the analytes – Maintains buffering capacity through the analysis – Produce the desired separation
  • 8.
    • Commonly usedbuffers include: – Low pH (acidic): phosphate, acetate, formate, citrate – High pH (basic): tris, tricine, borate, CAPS (N-cyclohexyl-3- aminopropanesulfonic acid) • Higher the ionic strength (and concentration) > higher size of ionic cloud > lower mobility of the particle • Also, higher ionic strength > sharper protein-band separation and increased heat production > denaturation of heat-labile proteins • Buffers used are made of monovalent ions because their valencies (ionic strength) and molality are equal. • They are good culture media for microorganisms and should be refrigerated when not in use, • Also cold buffer reduces evaporation and improves resolution • Small volumes should be discarded after use, but large volumes can be reused up to 4 times
  • 9.
    Support Media: • Insolublegels e.g sheets, slabs or columns of starch, agarose or polyacrylamide • Membranes (paper) of cellulose acetate • Pure buffer soln in a capillary Starch Gel: • Separate macromolecules on basis of surface charge and molecular size • Obsolete because preparation of reproducible starch gel is difficult Cellulose Acetate: • made by treating cellulose with acetic anhydride • They need to be soaked in buffer to soften them before use • Require clearing before densitometry • Hardly used in routine clinical application
  • 10.
    Agarose: • Used inAGE for the separation of Serum, Urine or CSF proteins, Hb variants, Isoenzymes, Lipoproteins etc. • Separation is based only on charge-to-mass ratio – pore size is large enough for all proteins to pass • Separate proteins into only 5 fractions (zones): albumin, alpha-1, alpha-2, beta- and gamma- globulins • Advantages: – Permits excellent densitometry – lower affinity for proteins (migration is not affected) and native clarity after drying – Little Endosmosis – free of ionizable groups (neutral) • Disadvantage: – DNA recovery is affected by inhibitors
  • 11.
    Polyacrylamide: • thermostable, transparent,durable and relatively chemically inert • No endosmosis – uncharged • Pore size does not allow larger proteins like fibrinogen, B1- lipoprotiens, Y-globulins etc, to migrate • Separation is based on both charge-to-mass ratio and molecular mass – molecular sieving • Carcinogenic – CAUTION when handling • Accommodates a large amount of sample in a single sample slot • DNA recovered is pure with no inhibitors unlike Agarose. • Separates proteins into 20 or more fractions • Used to study individual proteins (e.g Isoenzymes)
  • 12.
    Equipment used forGel Electrophoresis:
  • 13.
    TECHNIQUE Separation: • Blot thehydrated support medium to remove excess buffer • The sample is added to the support for about 5min • The support is then placed into the electrophoretic chamber in contact with the buffer • Apply a constant-current or constant-voltage power for a specified time • The support is then removed and placed in a fixative or rapidly dried to prevent diffusion of the sample
  • 14.
    Staining: • The supportis then stained and dried after washing out the excess dye • Amido Black B or members of the Coomassie Brillant Blue series are the commonest dyes • The amount of dye taken up is dependent on the type of protein, degree of denaturation by the fixing agent and quality of the dye • To visualize isoenzymes, incubate the gel in contact with a solution of substrate, which is linked structurally or chemically to a dye, before fixing. • Typical stain solution may be used several times • When the stain solution becomes faulty, protein zones will appear too lightly stained • It must be stored tightly covered to prevent evaporation
  • 15.
    Detection and Quantification: •Detection can be achieved using UV light, but quantification is by Densitometry • Densitometers integrate the area under a peak, and the result is printed as percentage of the total • Reliable quantification of stained zones using densitometry requires: – Light of an appropriate wavelength – Linear response from the instrument – Transparent background in the strip being scanned • Mass spectrometers – determine the molecular weights of proteins and their cleavage products, and for peptide sequencing.
  • 16.
    Blotting Techniques: • SouthernBlotting – widely used in molecular biology: – DNA or DNA fragments separated by AGE – Strip of nitrocellulose or a nylon membrane is laid over the gel blotting the DNA or DNA fragments onto it by capillary, electro-, or vacuum blotting – Detection and identification is by hybridization with a labeled complementary nucleic acid probe. • Northern Blotting: – Separates and detects RNAs – Uses labeled RNA probe for hybridization • Western Blotting: – Separates and detects proteins – The membrane is reacted with a reagent containing an antibody raised against the protein of interest
  • 17.
  • 18.
    TYPES OF ELECTROPHORESIS •Zone Electrophoresis • Slab Gel Electrophoresis • Disc Electrophoresis • Isoelectric Focusing Electrophoresis • 2-Dimensional Electrophoresis • Capillary Electrophoresis • Microchip Electrophoresis
  • 19.
    Zone Electrophoresis: • Producezones of proteins that are heterogeneous and physically separated from one another • Classified according to type and structure of the support material e.g AGE, CAE, PAGE etc
  • 20.
    Slab Gel Electrophoresis: •Use of a rectangular gel regardless of the thickness • Main advantage – ability to simultaneously separate several samples in one run • Primary method used in clinical chemistry lab • Gels (usually agarose) may be cast on sheet of plastic backing or completely encased within a plastic walled cell allowing horizontal or vertical electrophoresis and submersion for cooling, if needed. • May be cast with additives like: – Ampholytes which create a pH gradient, or – Sodium dodecyl sulfate (SDS) that denatures protiens
  • 21.
    Schematic representation ofSlab gel electrophoresis
  • 22.
    • Useful inthe separation of serum proteins, isoenzymes, lipoproteins, hemoglobins, and fragments of DNA and RNA Common problems encountered in SGE include: • Discontinuities in sample application – dirty applicator; cleaned by agitating in water • Unequal migration of samples across the width of the gel – dirty electrodes (uneven application of electric field), uneven wetting of the gel • Distorted protein zones – bent applicator, incorporation of air bubble during sample application, over application of sample, excessive drying of support • Unusual bands – artefacts, other causes e.g hemolysed samples > increased b-globulin band • Atypical bands – denatured protein from deteriorated serum, also rule out a true paraprotein
  • 23.
    Disc Electrophoresis: • 3-gelsystem: – small-pore separating gel (running gel), – a larger-pore spacer gel (stacking gel), – and a thin layer of large-pore monomer solution (sample gel) – containing about 3 µL of serum • The different composition cause discontinuities in the electrophoresis matrix • During electrophoresis, all proteins migrate easily through the large-pore gels and stack up on the separation gel in a very thin zone • This improves resolution and concentrates protein components at the border (or starting zone) • Separation occurs at the bottom separation gel by the molecular sieve phenomenon
  • 24.
    Schematic representation ofDisc Electrophoresis
  • 25.
    Isoelectric Focusing Electrophoresis: •Separates amphoteric compounds with increased resolution in a medium possessing a stable pH gradient • The protein migrate to a zone in the medium where the pH of the gel matches its pI • At this point, the charge of the protein becomes zero and its migration ceases – it become “focused”. • Regions associated with a given pH are very narrow – enough to separate proteins that differ in their pI values by only 0.02pH units • A high voltage power source is needed because carrier ampholytes are used in relatively high concentrations • Thus, the electrophoretic matrix must be cooled • IEF is used in neonatal screening programs to test for variant Hb
  • 27.
    2-Dimensional Electrophoresis: • Usescharge-dependent IEF (first dimension) and molecular weight-dependent electrophoresis (in the second) • 1st dimension – carried out in a large-pore medium like agarose or large-pore PAG; to which ampholytes are added to yield a pH gradient • 2nd dimension is often polyacrylamide in a linear or gradient format • It achieves the highest resolving power for the separation of DNA fragments – 1st dimension – normal AGE – 2nd dimension – ethidium bromide is added to the gel to open up the fragments and cause changes in their mobility • Method of choice when complex samples need to be arrayed for characterization, as in proteomics.
  • 28.
    Schematic representation of2D Electrophoresis
  • 29.
    Capillary Electrophoresis: • Separationin narrow-bore fused silica capillaries (inner diameter 25 – 75µm) filled with buffer – gel media can also be used • Sample is loaded after filling capillary with buffer, and electric field applied • Borate is a classic CE buffer that generates relatively low current and heat, even with a high ionic strength • Electro-osmotic flow (EOF) controls the amount of time solutes remain in the capillary (also Electroendosmostic flow or endosmosis) – EOF = bulk flow of liquid towards the cathode upon application of an electric field and it is superimposed on electrophoretic migration • Cations migrate fastest due to EOF and electrophoretic attraction towards the cathode
  • 31.
    • Neutral moleculesare all carried by EOF but not separated from each other • Anions move slowest because EOF is slightly greater than the attraction twds the anode and repulsion from cathode • Used for determination of MWts of proteins and peptides, analysis of PCR products, inorganic ions, organic acids, pharmaceuticals, optic isomers, and drugs of abuse in serum and urine. • When using a new capillary or changing buffer, the capillary must be equilibrated with the buffer – conditioning – Particularly important when phosphate-containing buffer is involved – This should be done for at least 4hrs before electrophoresis is started • The capillary surface must be regenerated or reconditioned to remove any material adsorbed onto the wall – Done by following each run with 3- to 5-column volume rinse using NaOH, and flushing with 5- to 8-column volumes of fresh buffer.
  • 32.
    Microchip Electrophoresis: • Similarin principle to CE, but differs in that the separation channels, sample injection channels and reservoirs are all fabricated into the same planar substrate using photolithographic processes. • More so, sample preparationand/or precolumn or postcolumn reactors, detectors, and excitation sources are intergrated into the chip
  • 34.
    QUALITY CONTROL • AClear SOP must be adhered to during each run to ensure reproducibility • Control serum must be included in each electrophoretic run to evaluate its quality and that of the densitometer.
  • 35.
    APPLICATIONS • DNA Analysis: –specific DNA sequences can be analyzed, isolated and cloned. – The analyzed DNA may be used in forensic investigations and paternity tests. • Protein Analysis: – In the diagnosis of conditions where levels of specific proteins or total protein is low or higher than normal e.g monoclonal gammopathy etc. • Antibiotic Analysis: – Synthesis of new antibiotics – Analysis of bacteria response to antibiotics and determining antibiotic-resistance • Vaccine Analysis: – Purification, processing and analysis of vaccines e.g influenza vaccine, hepatitis vaccine, polio vaccine
  • 36.
    REFERENCES: • Tietz Textbookof Clinical Chemistry and Molecular Diagnostics, fifth ed., by Burtis et al. • Clinical Chemistry; Principles, Techniques and Correlations, 7th ed., by Bishop et al. • http://www.slideshare.net/jyots23/electrophoresis- presentation • http://www.bio-rad.com/en-us/applications- technologies/protein-electrophoresis-methods • Landers JP. Molecular diagnostics on electrophoretic microchips. Anal Chem 2003;75:2919-27 • St. claire iii RL. Capillary electrophoresis. Anal Chem 1996;379-423
  • 37.