2. Introduction
• Pacemakers are small electronic devices that senses intrinsic heart
rhythm and transmit electrical impulses if indicated to stimulate the
heart.
• It replace the defective natural pacemaker and/or conduction pathway.
• One of the miracles of modern science
• It improved the life expectancy of patients with arrhythmias.
• Its discovery was timely to the expansion of open-heart surgery
3. Epidemiology
• Globally, about 1 million pacemakers are implanted yearly (ESC 2021).
• 76 pacemaker related surgeries in 2024 at the NCTC Ghana.
• 48 - Permanent pacemaker implantation
• 20 - Pulse generator change
• 4 – Repositioning of dislodged lead
• 2-Implantable cardioverter defibrillator (ICD)
• 1 - Cardiac resynchronization therapy (CRT)
• 1- Debridement of infected PG pocket
18. Bradycardia- Atrioventricular Block
• Aetiology
• Ischaemia od the AV node or bundle e.g. Ischaemic heart disease
• Compression of AV bundle e.g. calcification of AV or MV
• Inflammatory conditions e.g. diphtheria or rheumatic fever
• Extreme vagal stimulation as part of carotid sinus syndrome
• Degeneration of AV conduction system especially older patients
• Iatrogenic (surgical): VSD repairs, MV and AV replacements
• Medications: beta blocker and digoxin
19. Bradycardia- Atrioventricular Block
• 1st
degree: Increase in PR interval by > 0.2s
• 2nd degree: Action Potential is sometimes not strong enough to pass
through the bundle
• Mobitz type I (Wenckebach periodicity)
• Mobitz type II
21. Bradycardia- Atrioventricular Block
• 3rd
degree: Complete Heart Block (complete AV dissociation)
• No relationship between p wave and QRS-T complexes
• Ventricular escape rate is called the idioventricular rate.
22. Basic Electrophysiology
• Voltage (V) is the difference in electrical potential between two points
in a circuit. It is the force that drives electrons through the circuit.
• It is measure in Volts (V)
• Voltage = Current (I) x Resistance (R)
• Current (I) is the flow of electrons through a complete circuits.
• It is measured in Ampere (A)
23. Basic Electrophysiology
• Impedance (Z) is the opposition to alternating current. It is presented by the
combined effect of resistance, inductance and capacitance in a circuit.
• It is measured in Ohm (Ω).
• Inductance (L) is the tendency of an electrical conductor to oppose a change in
electric current flowing through it.
• It is measured in Henry (H)
• Capacitance (C) is the ability of an object to store electric charge. It is measured by
the change in charge in response to a difference in electric potential i.e.
Charge/voltage
• It is measured in farad
26. Battery of Pulse Generator
• Modern PG have a lithium iodide battery (Older gen - mercury-zinc cells)
• Avg duration (single chamber – 7 to 12yrs; Dual chamber – 6 to 10 yrs)
• When replacement indicator light is triggered, it has 3 - 6mo before it
reaches its useful
• Conservation strategy
• Dual chamber to single chamber
• Rate response to rate non-response
• Change to magnetic rate behaviours
27. Pacing lead
• Pacing lead is an insulated wire connecting an electrode to a PG
• It delivers the depolarizing pulse and senses intrinsic cardiac activity.
• Parts of the pacing lead
• Connecting pin
• Conducting cables surrounded by an insulator
• Lead-tip electrode
• Fixation
31. The Superiority of Bipolar Leads
• Bipolar leads possess significant advantages over the unipolar leads:
• no sensing of musculoskeletal potentials that may inhibit pacing
• less chance of “cross-talk” between the two leads of dual-chamber devices,
• less chance of interference with an ICD
• less chance of pacing skeletal muscle.
32. Passive Fixation
• Tines, Talons and Fins
• Entrapped myocardial trabeculations.
• Advantages
• Less risk of cardiac perforation
• Little change to threshold after fixation
• Easily retrievable
• Disadvantages
• Easily dislodged
• Less control over implantation site
33. Active Fixation
• Helical screw
• Lower risk of dislodgement
• Improved choice of implantation site
• Increased risk of perforation
• Increase in threshold after
implantation
35. Basic Concepts of Pacing
• Sensing is the ability of the pacemaker to detect the intrinsic electrical
cardiac activity
• This function is aided by an amplifier and a band pass filter
• Safety margin is set to 2 to 3-fold the sensing threshold to avoid oversensing.
36. Basic Concepts of Pacing
• Stimulating threshold is the minimum amount of energy required to
produce a successful myocardial depolarization and pacing.
• It is a function of the amplitude (volts) and pulse width (ms).
• The safety margin is programmed at 2- to 3-fold that of the min threshold
• Ideal threshold is < 1.5V and pulse width is 0.5ms.
• Impedance is the opposition to the flow of current.
• Typical pacing impedance varies from about 300 to 1500 Ω.
• Abrupt changes (>30%) may suggest lead malfunction.
37. Pacing Modes
• In 1974, the Association for the Advancement of Medical Instrumentation (AAMI)
adopted the 3-lettered generic pacemaker code that allows for uniform
functional classification of all pacing systems.
• 1st
letter - chamber paced [ A – atrium; V-ventricle; D – both and O – none ]
• 2nd
letter - chamber sensed [ A – atrium; V-ventricle; D – both and O – none ]
• 3rd
letter - type of response [ T – triggered; I – inhibited; D- both and O none ]
38. Pacing Modes
• In 1981, the 3-letter code was extended to 5 to reflect advances in pacemakers
• 4th letter – programmability or rate modulation
• [P – Simple programming]
• [M – Multiprogrammable]
• [C – communication]
• [R – Rate modulation]
• [O – none]
• 5th letter – Anti-tachycardia therapy
• [P – antitachycardia pacing]
• [S – shock]
• [D – antitachycardia pacing and shock]
• [O – none]
40. Classification of Pacemaker
• Number of chamber: single, dual or CRT
• Number of pacing lead electrodes: Unipolar or Bipolar
• Fixation mechanism: Passive or Active
• Location of leads: epicardial or endocardial
• Time related: Temporary or permanent.
• Functional classification: Based on pacing mode
41. Indications for Permanent Pacing
• 3rd
or advanced 2nd
degree AV block
• Asymptomatic 3rd
degree with HR < 40bpm with cardiomegaly or ventricular dysfunction
• Symptomatic bradycardia or ventricular arrhythmia
• Asystole ≥ 3sec or an escape rhythm < 40bpm or below AV node.
• Post-op AV block that is not expected to resolve after cardiac surgery.
• NM disease (myotonic muscular dystrophy, Kearns-Sayre syndrome, Erb dystrophy etc.)
• Sinus Node Dysfunction
• SND with symptomatic bradycardia (including symptomatic frequent pauses)
• SND with HR < 40bpm, even when no clear ass with symptom is established
42. Indications for Permanent Pacing
• Chronic Bi-fascicular Block
• Syncope not due to AV block when other likely causes have been excluded.
• NM diseases with bifascicular block or any fascicular block +/- symptoms
• After the Acute Phase of MI
• Persistent 2nd
degree block with alternating bundle-branch block or 3rd
degree
• Transient advanced 2nd
or 3rd
infra-nodal AV block and ass bundle branch block.
• Persistent symptomatic or asymptomatic 2 or 3rd
degree block
43. Indications for Permanent Pacing
• Other indications
• Hypersensitive Carotid Sinus Syndrome and Neuro-cardiogenic Syncope.
• Inappropriate or symptomatic bradycardia after cardiac transplantation
• Hypertrophic obstructive cardiomyopathy
• Symptomatic recurrent SVT that is recalcitrant to medical and ablation therapy.
• High-risk patients with congenital long-QT syndrome
• Brugada syndrome (sudden unexplained nocturnal death syndrome)
44. Contraindications to Permanent PPI
• These contraindications are relative and controversial
• In patients with an active infectious illnesses e.g., endocarditis
• Reversible causes of the arrhythmia
• Terminal illness with low life expectancy
• Severe psychiatric disorders – it may preclude proper care or follow-up
45. Temporary Pacemaker
• It may be indicated for emergency or elective setting
• It may serve as a bridge to permanent pacemaker implantation.
• Indications
• Acute MI with asystole, bilateral bundle block or AV block
• Bradycardia in patients on thrombolytic therapy.
• Prophylaxis after open heart surgery
• Active infection in a pacemaker dependent patient
• Electrophysiological studies
48. Implantation of Transvenous Pacemakers
• Pre-op Clinical Considerations
• History of injury or disease within planned region for implantation.
• Recognition and treatment of an active infection.
• Assessment of respiratory status.
• Allergies to medications – antibiotics, LA, opioids, IV contrasts.
• Management of anticoagulation therapy
• Prophylactic antibiotics immediately before and for 72hrs after implantation.
• Venous access
• Temporary – internal jugular, subclavian, femoral or antecubital veins
• Permanent - Cephalic, axillary or subclavian vein
51. Implantation of Pacemakers
• Technique
• Advancement of guidewire under fluoroscopic guidance into the IVC
• A sheath (or introducer) is passed over the guidewire into the subclavian vein.
• Ventricular lead with its stylet is introduced into the vein and advanced into the PA
• The guidewire in the lead is replaced with a straight guidewire.
• The lead is withdrawn from the PA and implanted at the infero-apical part of RV
54. Videos
• Creating the Three Dimensional Stylet for Right Ventricular Lead
• https://www.howtopace.com/atrial-lead/
59. Implantation of Pacemakers
• Post-implantation Care.
• Patients should be admitted for 24 – 48hrs
• Bed rest with little movement of the limbs adjacent the insertion site.
• Regular inspection of pulse generator site
• Resumption of anticoagulation therapy after 8hrs
• A post-op X-ray to assess lead and PG positions.
• Interrogation of device on day after implantation
• Follow up appointment within 7 – 10 days
74. Future of Pacemaker
• Smart pacemakers with AI
• remote monitoring
• Improved battery longevity
• Self maintenance
• Biological pacemakers
• created from stem cells or gene therapy
75. Conclusion
• The invention of the pacemaker is due to the collaboration of cardiac
surgeon and electrical engineers.
• It has saved countless life and improved the quality of million of
patients with arrhythmias that are recalcitrant to medical therapy.
• The future prospects are exciting and advances are being made to
improve its longevity and expand its utility.
76. References
• (ACC/AHA/HRS) guidelines for device-based therapy of cardiac rhythm
abnormalities 2012
• Guyton and Hall Textbook of Medical Physiology 14th
Edition
• Clinically Oriented Anatomy 8TH
Edition
• Kirklin/Barratt-Boyes Cardiac Surgery – 4th
Edition
• Sabiston and Spencer: Surgery of the Chest – 9th
Edition
#4:In 1804, Aldini successfully stimulated the hearts of decapitated criminals.
He was In the nephew of Luigi Aldini
The works in reanimation inspired the story of Frankenstein
#5:Australian anaesthesiologist
Used it resuscitate a stillborn neonate
One pole was applied to a skin pad soaked in strong salt solution,” while the other pole “consisted of a needle insulated except at its point, and was plunged into the appropriate cardiac chamber.
#6:An American physiologist
An electro-mechanical instrument.
Hyman coined the term ‘artificial pacemaker’,
Pace the heart and lead it like an artificial maestro!
#7:External cutaneous electrodes and a large PG
Used to treat CHB patients
It was the only type available when OPS started in the 1950.
It kept some cardiac surgical patients alive until sinus rhythm returned.
Painful, excoriations and skeletal contractions worsened as days passed.
Surgeons would provoke Stokes-Adams to see if sinus rhythm or an adequate idioventricular rhythm had returned
#8:Clarence Lillehei enlisted Earl Bakken
A television engineer and later founder of Medtronics Corporation.
Developing a small portable pacemaker
Epicardial leads - Reduced the discomfort of patients
Increased risk of ascending infection
#9:Abou the size of a Walkman
Increased risk of ascending infection.
#10:It became possible by invention of transistors during the 1950s.
Elmqvist was a medical graduate who became an engineer.
He designed a portable ECG machine and ink jet printer.
A transistor – semiconductor > gen, amplify or control flow of current
.
#12:Designing pacemaker that could sense – mid 1960’s
Change of battery from mercury-zinc to lithium cell
Programming of rate and stimulus duration
Casing/housing of pacemakers
Non-invasive assessments of pacemaker function and adjustment of pacemakers via radio or telephonic transmission
#13:The heart is a muscle that contract in response to electrical impulse.
That impulse is generated and transmitted by the conduction system.
Conduction system – nodal tissue and conducting fibres
Nodal tissue – generates impulses
Conducting fibres - transmits
Impulses propagated by the cardiac muscle cells.
SA - deep to the epicardium (junction of SVC and RA)
It is supplied by a SA node artery (RCA in 60% and LCA in 40% )
A-V node post wall of the RA (just behind TV)
AV – junction between membranous and muscular part of IVS
AV node artery of posterior IV artery of the RCA (80%)
#14:The His bundle - only bridge between the atrium and ventricle.
AV node through the fibrous skeleton of the heart
Right and left bundle ramify into Purkinje fibres
Pacemaker system
SA node – 70 to 80bpm
AV node - 40 to 60 bpm
Purkinje - 15 to 40 bpm
Delays
SA to AV is 0.03sec
Delay within AV 0.09sec
Delay along His-Purkinje – 0.04sec
#15:PR interval – delay in transmission
QT interval – Contraction of vent
#16:Abnormal rhythmicity of the pacemaker
Shift of the pacemaker from the sinus node
Blocks along normal conduction pathway
Abnormal pathways of impulse transmission through the heart
Spontaneous generation of impulses in almost any part of the heart
#17:Junctional rhythm – absent p wave.
Causes
Congenital
Iatrogenic (surgery) – node or artery (Fontan and Atrial switch)
Myocardial infarction
Myocarditis
Electrolyte imbalances
Well trained athletes
#19:the action potential is sometimes strong enough to pass through the bundle into the ventricles and sometimes not strong enough to do so
#20:Progressive prolongation of the P-R interval until a ventricular beat is dropped and is then followed by resetting of the
P-R interval and repeating of the abnormal cycle
In type II block, there is usually a fixed number of non-conducted P waves for every QRS complex.
For example, a 2:1 block implies that there are two P waves for every QRS complex.
At other times, rhythms of 3:2 or 3:1 may develop.
#21:Intermittent (sec, min to weeks) or persistent.
Escape bcos it has escaped the influence of SA
Each time AV conduction ceases, the ventricles don’t start immediately
Delays for about 5 to 30 sec because of overdrive suppression.
Brain can remain active for 4 to 7 sec without blood - syncope
Once the ventricle takes over – ventricular escape
Syncope is relieved (40bpm is enough for brain perfusion)
Stokes-Adams
ventricular standstill at
the onset of complete block is so long that it becomes
detrimental to the patient’s health or even causes death
#25:Despite advances the basic components are similar
Battery source
Circuitry (output, sensing, telemetry, microprocessor and memory)
Metal shell
Ceramic feedthrough – a piece of wire surrounded by glass or sapphire (port for leads) –
The PG can take up to 3 leads
Sensors
Telemetry is the process of recording and transmitting the readings of an instrument.
#26:Ideal energy density
Predictable and gradual decrease in battery life.
Initial voltage outage is app 2.8V
When it reaches 2.0 – 2.4 V elective replacement indicator is triggered which change settings
#27:Lead is direct connection between the PG and the endocardium
It is made up of conducting cables and lead-tip electrode separated by an insulator
#30:Unipolar leads – usually has a cathode (neg) and the anode is incorporated into the PG.
Bipolar the difference in potential sensed or created is over a few mm
Unipolar lead spans the distance between the lead the PG across the chest
#31:(ICDs should not be used in conjunction with unipolar leads), and
In addition, most generators are compatible with unipolar
pacing through a bipolar lead if needed as a result of
lead damage or high pacing thresholds.
#33:Passive fixation leads acquire stability by becoming entrapped in myocardial trabeculations.
Active-fixation leads have distal screws that are deployed and penetrate the myocardial wall.
Active mechanism carry a lower risk of dislodge and provides versatility in the choice of implantation site (because their stability does not depend on presence of trabeculations)
active-fixation leads generate a more extensive inflammatory response – increases the threshold after implantation.
Electrodes implanted in myocardium elicit a strong foreign body granulomatous reaction that results in the formation of a local fibrous capsule at the lead tip
Steroid-eluting leads
Such leads have a reservoir of steroid near the electrode that flows through the porous electrode into the myocardium, thereby reducing the inflammatory
response and the rise in pacing threshold
#34:active pacing box (or generator) programmed
to pace in VVI mode
#35:Sensing occurs when filtered signal exceeds the set sensing threshold.
atrial and ventricular leads are 1.5 to 5 mV and 5 to 25 mV,
The lead has a recording electrode and indifferent electrode.
As a wave approaches the recording electrode and then proceeds aways from it - it creates an intrinsic deflection.
Intrinsic deflection is the transition from the approaching to the receding deflection.
an amplifier that
increases the signal from the recording electrode and a
band pass filter to delete signals with frequencies too high
or too low to represent an intrinsic deflection.
#36:Pulse width is the time duration over which the impulse was delivered.
These two parameters need to be optimized to enhance battery life.
They can be reprogrammed to prevent extracardiac (e.g. phrenic stim)
by lowering the pacing voltage to min far-field capture and increasing the pulse width to ensure cardiac stimulation.
Impedance increases in insulating fracture
Impedance increases in conductor fractures
50% of the current drain from the battery is used for pacing,
Whereas the other one-half is used for sensing and housekeeping functions
It is the flow of current that drains the battery and not the voltage.
The voltage in the difference in electrical potential.
Thus if you create a pathway for the flow of current to flow the potential begins to diminish as electrons move to establish an equilibrium.
The more conducting the pathway established the faster the drain of the battery.
The more resistive the pathway the slower the flow of current.
Pacing lead are designed such that the resistance is high enough to avoid complete drainage of the battery within a short period but
With good conductance enough to deliver the necessary current needed to stimulate the heart
Just like gravitational potential the high you are the higher the potential but the lower you are the lower the potential.
#37:T - a sensed event leads to delivery of a pacing stimulus
It is used to prevent inappropriate inhibition by incorrectly sensed events such as skeletal muscles myopotential.
This is life saving in patients who are pacemaker dependent.
I – inhibition of pacemaker stimulus delivery in designated chamber
(a new timing cycle is initiated)
The presence of an intrinsic dep above pacing rate inhibits pacing
D – Triggered and inhibited –
sensing of intrinsic atrial activity leads to inhibition of an atrial pacing stimulus.
An AV clock is then started (a programmed AV delay).
A sensed vent impulse – leads to inhibited pacing
No sensed vent impulse – trigger but it results in triggering of a ventricular pacing stimulus after the programmed AV delay.
Pacing in the ventricle or atrium does not occur for a programmable time
after intrinsic vent activity is sensed
#38:Simple programmable - programmability of the pacemaker is limited to certain parameters;
Multiprogrammable - able to program a large number of parameters.
The communicating function refers to the ability to exchange signals between the pacemaker and a pacemaker programming device.
Rate modulation is the capability of the pacemaker to adjust the pacing rate according to a sensor device.
A variety of sensors that attempt to match the individual’s physical activity or metabolic demands by responding to different sources,
such as motion, temperature, and chest wall impedance (as an estimate of minute ventilation).
#39:AAI – requires normal AV conduction and functioning pathway to AV node
It is seldom used because most patients require vent pacing
VVI
Used in patients with CHB or adv 2nd degree; esp if they have Atrial fibrillation.
AV synchrony is not preserved – used mainly with elderly patients with sedentary lifestyle
VOO
Old generation of pacemakers
Asynchronous pacing
Failure to sensing any electrical impulse (from the heart or interferences)
Incurrent practice, these modes are only used temporarily to prevent oversensing
DDD
Also called universal pacing because of its universal application
Prserved AV synchrony
Sinus node dysfunction and AV blocks
Dual stimulation reduces battery life
NB= If used in atrial tavhyarrhytmias – it can lead to ventricular tachycardia
Because the atrial events are tracked in the ventricle
VDD
In all pacing modes, a lower rate limit indicates the rate below which pacing occurs
(this is the slowest heart rate that should be present)
#41: presence of ≥2 consecutive, non-conducted P waves
The ventricular arrhythmias presumed to be due to AV block
Other medical conditions requiring drugs that result in symptomatic bradycardia.
AF and bradycardia with 1 or more pauses of at least 5sec.
#42:When type II second-degree AV block occurs with a wide QRS, including isolated right bundle branch block, pacing becomes a C
Pacing induced infra-His block that is not physiological
Alternating bundle branch block
Incidental prolonged HV interval ≥ 100ms on electrophysiological study.
#43:Neurocardiogenic syncope, also known as vasovagal syncope,
is a sudden loss of
rugada syndrome (BrS)
is a genetic disorder in which the electrical activity of the heart is
abnormal due to channelopathy consciousness that occurs when the body overreacts to a trigger.
Brugada syndrome despite a normal heart rhythm, because of a sudden drop in blood pressure, known as vasovagal syncope
#44:Temporary placement of a standard (screw-in) pacemaker lead is suggested.
This serves as a bridge to implantation of a permanent pacemaker.
#46:If one is to use a passive fixation – avoid site for PPI
If for active fixation – Permanent sites
#48:Injury or radiation therapy – potential abnormal venous drainage.
On table venograpgy may be done to assess for abnormal drainage.
#49:LA should be administered to the pectoral region
Oblique incision along the deltopectoral groove
Blunt dissection until a fatty pad between the pectoralis or deltoid muscles.
Dissection of the cephalic vein free from surrounding tissues.
Silk 0 sutures placed around the vein proximally and distally around the vein.
The cephalic vein is incised and lead introduced through with a vein lifter.
#51:Avoid the RV free wall due risk of ectopy and decreased lead stability,
While a lead is being placed through the sheath, the
patient should be instructed to transiently refrain from
talking or breathing to avoid complication by air embolus
through the introducer sheath
The lead and guidewire introduced gently to avoid injury to the heart or vessels
There should be some slack to the lead to avoid dislodgements
#52: Soft stylets give more leverage in manipulating a lead to the desired position (e.g. mid septal pacing) and
stiff stylets help maintain stability until fixation (e.g. ICD leads, tricky positions)
#53:AP, 40° LAO, 30° RAO and left lateral fluoroscopic images showing lead tip at apical septum in a patient with dual-chamber pacemaker
#55:Determine the impedance, sensing and pacing threshold.
An R wave of 5 mV
Pacing threshold of 1 mA or less are adequate
Impedance < 1000ohms
#57:Anchoring sleeve of the lead is plicated to the pectoralis muscle and fascia.
Final sensing and pacing threshold
Subcutaneous pocket is created medial to the delto-pectoral groove to house the PG
#58:The leads are coiled under the generator, and the entire system is implanted into the pocket
Subcutaneous tissues and skin are closed with absorbable sutures
Skin is closed in a subcuticular manner
#59:In some centres – patients are placed in a sling for 24hrs.
After 24hrs patients can use limbs but cannot lift objects that are 2.5-5kg or raise arms above the shoulder for 6weeks
X-ray to detect some of the complicatiosn
#61:Constrerictive pericarditis
It could involve an immune-mediated inflammatory response to initial cardiac injury - Dressler's syndrome.
It could involve direct irritation of the pericardium by minimally protruding electrodes
Tricuspid regurgitation
Lead-leaflet interaction: The pacemaker lead can mechanically interfere with the tricuspid valve leaflets, causing them to become scarred, restricted, or perforated
Inflammatory response: The lead can stimulate an inflammatory response that causes fibrosis and adhesions between the valve and the lead
Asynchronized activation: The pacemaker can cause the right ventricle to activate out of sync
#64:When including only a clear perforation, the sensitivity of chest radiography was 27.7
#68:In this
system, the pulse generator is placed within a pocket
created between the fourth and fifth intercostal spaces at
the midaxillary line. An electrode is positioned through
two subcutaneous tunnels, extending from the pocket to
a small xiphoid incision and from the xiphoid to a superior
sternal incision.
#72:Retrograde ventriculoatrial (VA) conduction – atrial sensing which then triggers ventricular pacing
Circuituous pacemaker-mediated tachycardia
Treatment – magnet
Algorithms that extend the post-vent atrial refractory period
Absence of rate response to physiologic need