Product FICO2 = 0.1% FICO2 = 0.5%
Amsorb (Armstrong, yellow) FCU = -­­3.096*FGF + 15.48 FCU = -­­1.797*FGF+8.987
LoFloSorb (Intersurgical, light blue) FCU = -­­2.557*FGF + 12.78 FCU = -­­1.791*FGF+8.954
Medisorb (GE, red) FCU = -­­2.223*FGF + 11.11 FCU = -­­1.648*FGF+8.231
Medisorb EF (GE, pink) FCU = -­­2.472*FGF + 12.36 FCU = -­­2.009*FGF+10.05
Spherasorb (Intersurgical, dark blue)
Spiralith (Micropore, black)
FCU = -­­1.868*FGF + 9.342
FCU = -­­1.249*FGF + 6.243
FCU = -­­1.326*FGF+6.631
FCU = -­­1.186*FGF+5.931
Fractionalcanisteruse(%/h)
Fractionalcanisteruse(%/h)
In Vitro Canister Life of COs Absorber Prepacks with Different Fresh Gas Flows with
the Aisys Anesthesia Machine
Presenting Author: Jan FA Hendrickx, MD, PhD, OLV Hospital, Aalst Belgium
Co-­­Authors: Lemmens H, MD, PhD, Stanford University, CA; De Hert S, MD, PhD, Gent
University, Belgium; De Wolf AM, MD, Northwestern University, Chicago, IL
Introduction: Low fresh gas flows (FGF) reduce inhaled anesthetic and carrier gas
waste, but increase CO2 absorbent use. The performance of CO2 absorbents remains
poorly studied, especially in a setting that resembles clinical practice. We measured
canister life of prepacks of 6 brands over a wide FGF range. Because circuit
configuration affects the rebreathing characteristics and thus performance, only those
that fitted on 1 particular anesthesia machine were studied.
Methods: Via a CO2 flow meter (MEDEC, Aalst, Belgium; accuracy of 3.7 mL/min), 160
mL/min CO2 flowed into the tip of a 2L breathing bag that was ventilated with an Aisys®
machine (GE, Madison , WI) using controlled mechanical ventilation (tidal volume 500
mL, rate 10/min, I:E 1:1). For each brand (see Figure 1), canister life was determined for
12 canisters (all of the same lot), each of which was assigned to 1 of 12 different FGFs
that ranged from 0.25 to 4 L/min O2/air (see X-­­axis in Figure 1). For each FGF, we
measured the time until the inspired CO2 (FICO2) reached either 0.1% or 0.5%. To
facilitate comparison between the canisters and to easily calculate costs, this duration
was converted into how much of one canister was used per hour (expressed as % of 1
canister/h); this so-­­called “fractional canister usage” or FCU was calculated for both
FICO2 thresholds. E.g., if it took 8h to FICO2 = 0.1%, FCU was 12.5 %. Because this FGF-­­
FCU relationship turned out to be linear, performance of each brand (for each FICO2
threshold) could be described by a linear fit.
Results:
12
10
8
6
4
2
0
FICO2 = 0.1% FICO2 = 0.5%
20
16
12
8
4
0
0 1 2 3 4 5 FGF (L/min)
0 1 2 3 4 5 FGF (L/min)
Conclusion: The only LiOH based product, SpiraLith®
, has the longest canister life and
produces no compound A nor CO (1). Of the Ca(OH)2 based canisters, those without
NaOH (Amsorb, LoFloSorb) are less efficient but do not produce compound A nor CO.
References: (1) ASA 2010 Abstract 1693

More Related Content

PDF
Gas yield from bubble volume
DOCX
workover
PDF
Hw4 with updated answer key
PPT
Enthalpy diagrams
DOC
Midterm 2 guide
PDF
Math modelled approach to gas-exchange monitoring
PPT
Vo2max what does it really mean
PPT
Oxygen delivery-systems
Gas yield from bubble volume
workover
Hw4 with updated answer key
Enthalpy diagrams
Midterm 2 guide
Math modelled approach to gas-exchange monitoring
Vo2max what does it really mean
Oxygen delivery-systems

More from Meditech Equipment Co.,Ltd (20)

PDF
Syringe pump catalogue from Meditech -medical manufacturer
PDF
Defibrillator > AED Trainer > Defi5T from Meditech
PDF
Meditech D-Sense automated external defibrillator analyzer
PDF
meditech #medical_equipment_catalog Ver.17D
PDF
meditech products for veterinary use
PDF
Et3 Meditech Mainstream capnography
PDF
Et7 Sidestream capnography from Meditech
PDF
Sonovet ,veterinary ultrasound scanner
PDF
Meditech Mobile health care technology
PDF
معرفة تأثير تمرينات الايروبك على زيادة نسبة الأوكسجين بالدم لمرضى السكري غير ...
PDF
Cuore di bimbo BabySono per ascoltare e registrare il battito cardiaco del fe...
PDF
Oxygen consumption
PDF
Spirox pro ,Advanced spirometer with PC Software
PDF
PDF
Oxy plus LED fingertip pulse oximeter
PDF
patient monitor veterinary use
PDF
PDF
Md100 digital spo2 user's manual
PDF
Meditech SpO2 Module MD200
PDF
12 Channel ECG machine EKG1212T
Syringe pump catalogue from Meditech -medical manufacturer
Defibrillator > AED Trainer > Defi5T from Meditech
Meditech D-Sense automated external defibrillator analyzer
meditech #medical_equipment_catalog Ver.17D
meditech products for veterinary use
Et3 Meditech Mainstream capnography
Et7 Sidestream capnography from Meditech
Sonovet ,veterinary ultrasound scanner
Meditech Mobile health care technology
معرفة تأثير تمرينات الايروبك على زيادة نسبة الأوكسجين بالدم لمرضى السكري غير ...
Cuore di bimbo BabySono per ascoltare e registrare il battito cardiaco del fe...
Oxygen consumption
Spirox pro ,Advanced spirometer with PC Software
Oxy plus LED fingertip pulse oximeter
patient monitor veterinary use
Md100 digital spo2 user's manual
Meditech SpO2 Module MD200
12 Channel ECG machine EKG1212T
Ad

Recently uploaded (20)

DOCX
Advanced Nursing Procedures.....realted to advance nursing practice M.Sc. 1st...
PPTX
MONOCHORIONIC TWIN PREGNANCY detailed.pptx
PPT
FRACTURE CLASSIFICATION AND MANAGEMENT..
PDF
Indonesian Healthtech Innovation_11Sep2019_Industry_Geraldine Seow_1.pdf
PPTX
Common Bacterial infections-converted_64bcdc4f77a3b7b90bdeb611f66c6ddd.pptx
PPTX
applied physics dental materials basic principles
PPT
Doppler - 5.ppt .........................
PPTX
Case report session Apendisitis Akut people.pptx
PDF
Chapter 8. HHS Facility Design and Construction _ HHS.gov.pdf
PDF
CASE PRESENTATION1.pdf bipolar disorder in which both mania and depression h...
PPTX
(PWD) Persons with intellectual disabilities ACT - 1.pptx
PDF
Key Updates for Pulmonary Tuberculosis Multiple Drug Resistance
PPTX
health care concerns.pptx by hemant kumari
PDF
odontologia na oncologia - carie de radiação
PPTX
Seizures in paediatrics as a big cause of morbidity.pptx
PPT
NEPHROTIC SYNDROME POWER POINT PRESENTATION
PPTX
Non-Variceal-Upper-GI-Bleeding_-Comprehensive-Review_121037.pptx
PPTX
CLASS III MALOCCLUSION IN ORTHODONTICS
PPT
53afocus7fluidelectrolytesacid-basebalance-121205082904-phpapp01.ppt
PPTX
Maternal and child health. The normal new born.pptx
Advanced Nursing Procedures.....realted to advance nursing practice M.Sc. 1st...
MONOCHORIONIC TWIN PREGNANCY detailed.pptx
FRACTURE CLASSIFICATION AND MANAGEMENT..
Indonesian Healthtech Innovation_11Sep2019_Industry_Geraldine Seow_1.pdf
Common Bacterial infections-converted_64bcdc4f77a3b7b90bdeb611f66c6ddd.pptx
applied physics dental materials basic principles
Doppler - 5.ppt .........................
Case report session Apendisitis Akut people.pptx
Chapter 8. HHS Facility Design and Construction _ HHS.gov.pdf
CASE PRESENTATION1.pdf bipolar disorder in which both mania and depression h...
(PWD) Persons with intellectual disabilities ACT - 1.pptx
Key Updates for Pulmonary Tuberculosis Multiple Drug Resistance
health care concerns.pptx by hemant kumari
odontologia na oncologia - carie de radiação
Seizures in paediatrics as a big cause of morbidity.pptx
NEPHROTIC SYNDROME POWER POINT PRESENTATION
Non-Variceal-Upper-GI-Bleeding_-Comprehensive-Review_121037.pptx
CLASS III MALOCCLUSION IN ORTHODONTICS
53afocus7fluidelectrolytesacid-basebalance-121205082904-phpapp01.ppt
Maternal and child health. The normal new born.pptx
Ad

In vitro canister life of cos absorber prepacks with different fresh gas flows with the aisys anesthesia machine

  • 1. Product FICO2 = 0.1% FICO2 = 0.5% Amsorb (Armstrong, yellow) FCU = -­­3.096*FGF + 15.48 FCU = -­­1.797*FGF+8.987 LoFloSorb (Intersurgical, light blue) FCU = -­­2.557*FGF + 12.78 FCU = -­­1.791*FGF+8.954 Medisorb (GE, red) FCU = -­­2.223*FGF + 11.11 FCU = -­­1.648*FGF+8.231 Medisorb EF (GE, pink) FCU = -­­2.472*FGF + 12.36 FCU = -­­2.009*FGF+10.05 Spherasorb (Intersurgical, dark blue) Spiralith (Micropore, black) FCU = -­­1.868*FGF + 9.342 FCU = -­­1.249*FGF + 6.243 FCU = -­­1.326*FGF+6.631 FCU = -­­1.186*FGF+5.931 Fractionalcanisteruse(%/h) Fractionalcanisteruse(%/h) In Vitro Canister Life of COs Absorber Prepacks with Different Fresh Gas Flows with the Aisys Anesthesia Machine Presenting Author: Jan FA Hendrickx, MD, PhD, OLV Hospital, Aalst Belgium Co-­­Authors: Lemmens H, MD, PhD, Stanford University, CA; De Hert S, MD, PhD, Gent University, Belgium; De Wolf AM, MD, Northwestern University, Chicago, IL Introduction: Low fresh gas flows (FGF) reduce inhaled anesthetic and carrier gas waste, but increase CO2 absorbent use. The performance of CO2 absorbents remains poorly studied, especially in a setting that resembles clinical practice. We measured canister life of prepacks of 6 brands over a wide FGF range. Because circuit configuration affects the rebreathing characteristics and thus performance, only those that fitted on 1 particular anesthesia machine were studied. Methods: Via a CO2 flow meter (MEDEC, Aalst, Belgium; accuracy of 3.7 mL/min), 160 mL/min CO2 flowed into the tip of a 2L breathing bag that was ventilated with an Aisys® machine (GE, Madison , WI) using controlled mechanical ventilation (tidal volume 500 mL, rate 10/min, I:E 1:1). For each brand (see Figure 1), canister life was determined for 12 canisters (all of the same lot), each of which was assigned to 1 of 12 different FGFs that ranged from 0.25 to 4 L/min O2/air (see X-­­axis in Figure 1). For each FGF, we measured the time until the inspired CO2 (FICO2) reached either 0.1% or 0.5%. To facilitate comparison between the canisters and to easily calculate costs, this duration was converted into how much of one canister was used per hour (expressed as % of 1 canister/h); this so-­­called “fractional canister usage” or FCU was calculated for both FICO2 thresholds. E.g., if it took 8h to FICO2 = 0.1%, FCU was 12.5 %. Because this FGF-­­ FCU relationship turned out to be linear, performance of each brand (for each FICO2 threshold) could be described by a linear fit. Results: 12 10 8 6 4 2 0 FICO2 = 0.1% FICO2 = 0.5% 20 16 12 8 4 0 0 1 2 3 4 5 FGF (L/min) 0 1 2 3 4 5 FGF (L/min)
  • 2. Conclusion: The only LiOH based product, SpiraLith® , has the longest canister life and produces no compound A nor CO (1). Of the Ca(OH)2 based canisters, those without NaOH (Amsorb, LoFloSorb) are less efficient but do not produce compound A nor CO. References: (1) ASA 2010 Abstract 1693