Internal Quality Control in
coagulation lab
Dr Ankit Raiyani
Haematology Department
Sahyadri Speciality Hospital
Presentation layout
• Introduction to Quality Assurance and Quality Control
• Definitions of common terms, Accuracy vs. Precision
• Preparation/selection of reagents-
• PTT reagent PT reagent MNPT
• QC for manual coagulation tests
• Quality control procedures- schedule
• Internal Quality Control
• LJ charts
• Westgard rules
• Other measures for IQC
Introduction
In the haematology laboratory it is essential to ensure that the right test is
carried out on the right specimen and that the correct results are delivered to
the appropriate recipient without delay.
Quality assurance (QA) is defined as the overall programme for achieving these
objectives.
Quality control (QC) is defined as measures that must be included during each
assay run to verify that the test is working properly.
Quality Assurance vs. Quality Control
Quality Assurance Quality Control
An overall
management plan to
guarantee the
integrity of data
(The “system”)
A series of
analytical
measurements used
to assess the
quality of the
analytical data
(The “tools”)
The Quality Assurance Cycle
•Data and Lab
Management
•Safety
•Customer
Service
Patient/Client Prep
Sample Collection
Sample Receipt
and Accessioning
Sample TransportQuality Control
Record Keeping
Reporting
Personnel Competency
Test Evaluations
Testing
Hurdles in Quality Assurance
6
Hurdles in Quality Assurance
7
Pre-analytical
Hurdles in Quality Assurance
8
Analytical
Hurdles in Quality Assurance
9
Post-analytical
Terms relating to quality control in clinical
laboratory practice
Precision Can be controlled by replicate tests and by repeated tests on previously measured
specimens
Accuracy Can be checked only by the use of reference materials that have been assayed by
reference methods
Controls Are preparations that are used for either internal quality control or external quality
assessment. They should not be used as standards
Error The discrepancy between the result of a measurement and the true (or accepted
true) value
Random
error
An error which varies in an unpredictable manner, in magnitude and sign, when a
large number of measurements of the same quantity are made under effectively
identical conditions
Systemic
error (bias)
An error which, in the course of a number of measurements of the same value of a
given quantity, remains constant when measurements are made under the same
conditions or varies according to a definite law when conditions change
Accuracy vs. Precision
Accuracy Precision
Meaning How well a
measurement
agrees with an
accepted value
How well a
series of
measurement
s agree with
each other
Achieved
with…
Calibration Quality
control
When low… Increases
systemic errors
Increases
random
errors
Calibration
Calibration is a comparison of measuring equipment
against a standard instrument of higher accuracy to
detect, correlate, adjust, rectify and document the
accuracy of the instrument being compared.
Preparation/selection of reagents
Reagent selection for PTT
LA-sensitive reagents- (low-phospholipid reagents)
• Siemens Actin FSL
• IL HemosIL aPTT-SP
• Stago PTT-LA
Intermediate responsiveness to LA
• Stago STAPTT-LA
• IL HemosIl SynthASil
Low responsiveness to LA
• Siemens Actin FS
• Stago C.K. Prest
Reagent selection for PTT
Selection of aPTT reagent depends on indication of aPTT--
The routine laboratory aPTT
• used for evaluating the factor VIII, IX, XI deficiency and effect of heparin
• sensitive to deficiencies of factors VIII, IX and XI at concentrations of 0.35–0.4 iu/ml.
• responsive to unfractionated heparin over the therapeutic range of approximately 0.3–0.7 iu/ml.
• low responsiveness to LA
LA screening
• High responsiveness to LA (silica+ low PL content)
• Two phospholipid based clotting assays, one of which may be aPTT based, are required1
Intermediate responsiveness to both factor levels and LA
• Used in labs with mixed sample load, when only one reagent is feasible/desired
1PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M. and DE GROOT, P. G. (2009), Update of the ISTH guidelines for lupus anticoagulant detection.
Journal of Thrombosis and Haemostasis, 7: 1737–1740.
Reagent selection for PT (Thromboplastin)
Contains tissue factor and coagulant phospholipids
Should have ISI value as close to 1 as possible.
Source- human, bovine, rabbit brain/lung, human placenta, recombinant.
Animal/human source- potential hazard of viral, prion and other infections from handling
Recombinant thromboplastins-
• Manufactured using recombinant human tissue factor produced in Escherichia coli and synthetic phospholipids,
• Do not contain any other clotting factors such as prothrombin, factor VII and factor X.
• Highly sensitive to factor deficiencies and oral anticoagulant-treated patient plasma samples
What is ISI?
International Sensitivity Index
It is the quantitative measure, (in term of first
international reference preparation of thromboplastin,
human, combined, coded 67/40,) of the
responsiveness of PT system to the defect induced by
oral anticoagulants.
Calculation of ISI
Calibration of a test thromboplastin must be against a reference thromboplastin of the same species e.g.
human against human, rabbit against rabbit
Tests are performed By using reference thromboplastin and test thromboplastinon 20 normal donors not on
anticoagulants and 60 patients who have been on oral anticoagulant treatment for at least 6 weeks and has
stable INR.
Prothrombin Times are performed in duplicate for each sample and the mean for each pair of tests derived.
If there is more than a 10% difference in the clotting times between duplicate samples, the tests on that
plasma sample should be repeated
Mean of each pair of PT results are plotted on double-log paper with the reference sample on the Y axis and
the test plasma on the X-axis.
• A best fit line is drawn with
points above the highest
recorded PT and the lowest PT
(see figure above). The slope of
the line is calculated
• Slope= AC/BC
• ISItest = ISIStandard X Slope
• INR= (PT/MNPT)ISI
Preparation of pooled normal plasma and
derivation of MNPT
Citrated blood samples from at least 20 healthy volunteers (with PT, aPTT in normal range) should
be collected. Equal numbers of male and female (not pregnant, not on OC pills) volunteers.
Platelet poor plasma should be prepared. After pooling it should be immediately frozen at -80⁰ C in
aliquots.
The mean of PT of individual samples is calculated and taken as mean normal prothrombin time
(MNPT).
SD is calculated and reference range is taken as -2 SD to +2 SD.
This pooled plasma is used as control with each batch of samples.
MNPT calculation- 18/6/15
No PT PTT
1 11.5 28.6
2 12.4 27.6
3 10.5 29.2
4 10.6 26.2
5 10.6 32.1
6 11.7 29.8
7 11.1 30.7
8 12.2 31.6
9 11.9 30.9
10 11.3 31.5
11 10.6 29.1
12 12.3 29.7
13 10 26.3
14 11.9 28.7
15 11.6 33.3
16 11.2 31.4
17 10.5 28.2
18 12.5 28.7
19 10.6 27.8
20 11.1 27.7
21 10.9 29.5
22 11.3 31.9
23 11.3 29.9
24 12.5 32
total 272.1 712.4
Arithmetic Mean 11.3 29.6
SD 0.724 1.899
Range 9.8 to 12.7 25.8 to 33.4
QC for manual coagulation tests
All tests and controls should be performed in duplicates.
Operator skill
Required instruments should be properly calibrated.
Centrifuge
• Serviced and Calibrated every 6 months or after break down
• Put on sturdy platform to reduce vibrations.
• Sample load should be well balanced.
Waterbaths
• Temperature- 37⁰ C (checked in all four corners)
• Transparent walls for transillumination
• Filled with distilled water till indicated level.
Pipette
• Calibrated.
• Disposable tips.
Calibration of centrifuge
Tachometer is used for calibration of RPM.
Actual RPM is measured with tachometer at uniform intervals (minimum 5) for 5 times.
Difference between actual and set values is calculated.
If difference is more than +/- 5%, then repeat the test. If repeat reading again shows significant
difference corrective measures are required.
Timer and Temperature, when applicable, should also be calibrated.
Calibration of pipettes
Analytical balance is used to weigh pipetted volume in controlled environment.
Distilled water is used as substrate.
The pipette is checked 5 times with the maximum volume (nominal volume), the
minimum volume or 10% of the maximum volume, whichever is higher.
With multichannel pipettes, both volumes are tested with the two edge channels
Conversion of mass to volume V = (w + e) x Z
• V = Volume (µl), w = Weight (mg), e = Evaporation loss (mg),
• Z = Conversion factor for mg/µl conversion (temperature, pressure)
Acceptable accuracy and precision range varies with pipette capacity.
Quality control procedures
Quality control procedures
The procedures that should be included in a quality assurance programme vary with
the tests undertaken, the instruments used and (especially if these include a fully
automatic system) the size of the laboratory and the numbers of specimens handled.
Some control procedures should be performed daily and other performance checks
should be done at appropriate intervals.
The latter is particularly important when there is a change in staff and after
maintenance service or repair has been carried out on equipment.
Schedule for quality control procedures
1. Calibration with reference standards
Instruments, pipettes
6-month intervals or more frequently
if control chart or EQA indicates bias
or fluctuation in results and after any
repair/service
Others: thermometers, scales, etc Annually
2. Control chart with control material
Daily or more frequently with each batch of specimens
Duplicate tests on two or three patients’ samples: if control chart or delta
check shows discrepancies
3. Analysis of patients’ results
Cumulative results: following previous tests and if changes in clinical state
4. EQAS performance
Assessment monthly
Components of Quality control
Internal quality control (IQC)
• Monitoring the haematology test procedures to ensure continual
evaluation of the reliability of the daily work of the laboratory with
validation of tests before reports are released
External quality assessment (EQA)
• Evaluation by an outside agency of the between-laboratory and
between-method comparability. It can be organized nationally,
regionally or internationally
Internal Quality Control
IQC is based on monitoring the haematology test procedures that are performed in the
laboratory and includes measurements on specially prepared materials and repeated
measurements on routine specimens, together with daily statistical analysis of the data.
Primarily a demonstration of Precision.
It ensures continual checks that the established reliability of the laboratory’s work does
not fluctuate and that reports are validated before they are released
IQC includes
Control charts with tests on control materials
Duplicate tests on a proportion of the specimens
Consistency of mean values of patient data
Correlation check
Control charts
These were first applied in clinical chemistry by Levey and Jennings.
They are now widely used in haematology for both automated and manual
procedures.
Samples of the control specimen are included in every batch of patients’
specimens and the results are checked on a control chart.
To check precision, it is not necessary to know the exact value of the control
specimen.
Controls
Level 1 controls- normal controls
Level 2 controls- low abnormal controls
Level 3 controls- high abnormal controls (not used routinely)
Minimum of one level QC at least once a day.
Patient samples >25 per day  Employ 2 levels of QC at least once a day.
Patient samples >75 per day Employ 2 levels of QC at least twice a day at appropriate intervals
Controls are intended to simulate random sampling, they must be treated exactly like the patients’ specimens.
Preparation of LJ charts
Mean value and standard deviation (SD) of the control specimen should first
be established in the laboratory where the tests are performed.
Using arithmetic graph paper, a horizontal line is drawn to represent the mean
(as a base), and on an appropriate scale of quantity and unit, lines representing
+2SD and −2SD are drawn above and below the mean.
The results of successive control sample measurements are plotted. If the test
is satisfactory, sequential results oscillate about the mean value and <5% of the
results fall outside 2SD.
LJ chart PT / low abnormal May 2015
LJ chart PT / low abnormal June 2015
LJ chart PT / low abnormal June 2015
Reason ??
LJ chart aPTT / NCPNP
LJ chart F VIIIc / NCPNP
Westgard rules
“Multirule Quality Control” developed by Dr.
James O. Westgard based on statistical concepts
Uses a combination of decision criteria or
control rules
Allows determination of whether an analytical
run is “in-control” or “out-of-control”
Westgard multirule system
12S rule
13S rule
22S rule
R4S rule
41S rule
10X rule
Westgard 12S rule
“warning rule”
One of two control results falls outside ±2SD
Alerts tech to possible problems
Not cause for rejecting a run
Must then evaluate the 13S rule
12S Rule = A warning to trigger careful inspection of the control data
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
12S rule
violation
Westgard – 13S Rule
If either of the two control results falls outside of ±3SD,
rule is violated
Run must be rejected
If 13S not violated, check 22S
13S Rule = Reject the run when a single control
measurement exceeds the +3SD or -3SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
13S rule
violation
Westgard – 22S Rule
2 consecutive control values for the same level fall outside of ±2SD in the same
direction, or
Both controls in the same run exceed ±2SD
Patient results cannot be reported
Requires corrective action
22S Rule = Reject the run when 2 consecutive control
measurements exceed the same +2SD or -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
22S rule
violation
Westgard – R4S Rule
One control exceeds the mean by –2SD, and the other
control exceeds the mean by +2SD
The range between the two results will therefore exceed
4 SD
Random error has occurred, test run must be rejected
R4S Rule = Reject the run when 1 control measurement
exceed the +2SD and the other exceeds the -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
R4S rule
violation
Westgard – 41S Rule
Requires control data from previous runs
Four consecutive QC results for one level of control are
outside ±1SD, or
Both levels of control have consecutive results that are
outside ±1SD
Westgard – 10X Rule
Requires control data from previous runs
Ten consecutive QC results for one level of control are on
one side of the mean, or
Both levels of control have five consecutive results that
are on the same side of the mean
10x Rule = Reject the run when 10 consecutive control
measurements fall on one side of the mean
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
10x rule
violation
Alternate ways of checking the precision of
routine work
Duplicate tests on patients’ samples
• Test of precision can be performed by repeat analysis on samples from two or three patients
Correlation Check
• any unexpected result of a test must be checked to see whether it can be explained on clinical grounds or
whether it correlates with other tests
“Delta” check
• Blood count from the same patient checked repeatedly over 2-3 weeks
• Values should not change significantly providing the patients condition has not altered significantly
Patient data
• Need enough through put (i.e. >100 samples /day)
• MCV, MCHC, MCH should not vary, providing the patient population is stable and samples from different sets of
patients are not batched together.
• Mean of 20 successive patients can be plotted and any drift in the calibration of the analyzer will be recognized
THANK YOU!!

More Related Content

PPTX
Preparation and staining of peripheral blood smear
PPTX
Lab Diagnosis of Chronic lymphoproliferative disorders (CLPD); Flowcytometric...
PPT
Hplc interpretation
PPTX
Hemoglobinopathies - Lab diagnosis
PPTX
Calibration and Quality controls of automated hematology analyzer
PPT
QC in coagulation
PPTX
Quality assurance in haematology
PPTX
Automation histopathology
Preparation and staining of peripheral blood smear
Lab Diagnosis of Chronic lymphoproliferative disorders (CLPD); Flowcytometric...
Hplc interpretation
Hemoglobinopathies - Lab diagnosis
Calibration and Quality controls of automated hematology analyzer
QC in coagulation
Quality assurance in haematology
Automation histopathology

What's hot (20)

PPTX
Westgard's rules and LJ (Levey Jennings) Charts.
PPTX
Reticulocyte count
PPTX
quality control in clinical pathology
PPTX
Automation in hematology
PPTX
Demonstration of le cells
PPT
PPSX
Fluid cytology in CSF
PPTX
PPTX
Osmotic Fragility Test
PPTX
9. automation in urine analysis
PPTX
Cytochemical staining checked
PDF
Quality Assurance in Hematology laboratory
PPSX
Quality control in clinical biochemistry
PPTX
Quality Control In Histopathology Dr.Rami amawi.pptx
PPTX
6. platelet function_tests
PPTX
Prothrombin time and aptt
PPTX
PPT
quality control in pathology
PPTX
Platelet Function Tests
Westgard's rules and LJ (Levey Jennings) Charts.
Reticulocyte count
quality control in clinical pathology
Automation in hematology
Demonstration of le cells
Fluid cytology in CSF
Osmotic Fragility Test
9. automation in urine analysis
Cytochemical staining checked
Quality Assurance in Hematology laboratory
Quality control in clinical biochemistry
Quality Control In Histopathology Dr.Rami amawi.pptx
6. platelet function_tests
Prothrombin time and aptt
quality control in pathology
Platelet Function Tests
Ad

Viewers also liked (20)

PPTX
Quality control in clinical laboratory
PPT
Quality in microbiological lab
PPT
Laboratory Internal Quality Control presentation master revision, 2014
PPTX
Quality control in microbiology
PPT
Flow Cytometry - basics, principles and applications
PPTX
Histopathology specimen processing
PPT
Quality control in the medical laboratory
PDF
Good CBC Morphological AssessmentPDF
PDF
Itp in pregnancy
PDF
Thrombocytopenia during pregnancy
PPTX
BLOOD FILM EXAMINATION: ITS RECENT INVESTIGATIVE METHODOLOGY IN THE DIAGNOSIS...
PPT
Xt 2000i cell counter Autoanalyser
PPTX
Gene therapy
PDF
Celta task answers
ODP
Gene therapy
PPTX
Optimal Antibiotic Strategies in ICU
PPSX
Recent advances in multi drug resistant tuberculosis &rntcp
PPTX
AUTOMATION IN CLINICAL MICROBIOLOGY
PPTX
Antibiotic policy
PPT
Infection Control and Antibiotic Stewardship
Quality control in clinical laboratory
Quality in microbiological lab
Laboratory Internal Quality Control presentation master revision, 2014
Quality control in microbiology
Flow Cytometry - basics, principles and applications
Histopathology specimen processing
Quality control in the medical laboratory
Good CBC Morphological AssessmentPDF
Itp in pregnancy
Thrombocytopenia during pregnancy
BLOOD FILM EXAMINATION: ITS RECENT INVESTIGATIVE METHODOLOGY IN THE DIAGNOSIS...
Xt 2000i cell counter Autoanalyser
Gene therapy
Celta task answers
Gene therapy
Optimal Antibiotic Strategies in ICU
Recent advances in multi drug resistant tuberculosis &rntcp
AUTOMATION IN CLINICAL MICROBIOLOGY
Antibiotic policy
Infection Control and Antibiotic Stewardship
Ad

Similar to Internal quality control (IQC) in coagulation lab (20)

PPTX
(IQC) - Internal Quality Control.pptx
PPTX
Principles of Laboratory Medicine.pptx imp
PPT
2. Quality Control Notes
PPT
Hema II Chapter11_ Automation in Hematologya.ppt
PPTX
QUALITY CHECK IN A BIOCHEMISTRY LAB SYSTEM
PDF
Method Validation in Medical Laboratory..
PPTX
Quality control of tti
PDF
Hematology Chemistry Coagulation
PPT
Laboratory Quality Control .ppt
PPTX
Quality control.pptx
PPTX
new QUALITY CONTROL IN BLOOD BANKING.pptx
PPTX
QUALITY CONTROL IN HEMATOLOGY BSC. MLS 3RD
PPTX
Validation of lab instruments and quantitative test methods
PDF
Quality management system BY M .REJA
PDF
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
PPTX
validation and calibration of HPLC
PPTX
Control of analytical quality using stable control materials postgrad
PPTX
1.qualitycontrol final
PPTX
Chemistry QC, Calibration&Reportable Range REFINED.pptx
PPTX
Quality Control In Clinical Laboratory
(IQC) - Internal Quality Control.pptx
Principles of Laboratory Medicine.pptx imp
2. Quality Control Notes
Hema II Chapter11_ Automation in Hematologya.ppt
QUALITY CHECK IN A BIOCHEMISTRY LAB SYSTEM
Method Validation in Medical Laboratory..
Quality control of tti
Hematology Chemistry Coagulation
Laboratory Quality Control .ppt
Quality control.pptx
new QUALITY CONTROL IN BLOOD BANKING.pptx
QUALITY CONTROL IN HEMATOLOGY BSC. MLS 3RD
Validation of lab instruments and quantitative test methods
Quality management system BY M .REJA
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
validation and calibration of HPLC
Control of analytical quality using stable control materials postgrad
1.qualitycontrol final
Chemistry QC, Calibration&Reportable Range REFINED.pptx
Quality Control In Clinical Laboratory

More from Ankit Raiyani (14)

PPTX
Immune thrombocyopenia (ITP)
PPSX
Primary (AL) Amyloidosis
PPSX
Classical Hodgkin’s lymphoma
PDF
WHO Classification of Tumours of Haematopoietic and Lymphoid Tissues : 2016 U...
PPT
Rivaroxaban (XARELTO)
PPT
Lineage switch in Acute Leukemia
PPTX
Acute Meningoencephalitis - Thesis presentation
PPTX
Leptospirosis
PPT
Somatosensory seizures
PPTX
Motor neuron disease in HIV
PPTX
Horner's syndrome and Internuclear ophthalmoplegia
PPTX
Diffuse idiopathic skeletal hyperostosis
PPTX
Pachydermoperiostosis
PPTX
AN UNUSUAL CAUSE OF DELAYED RECOVERY FROM NEUROMUSCULAR PARALYSIS DURING GENE...
Immune thrombocyopenia (ITP)
Primary (AL) Amyloidosis
Classical Hodgkin’s lymphoma
WHO Classification of Tumours of Haematopoietic and Lymphoid Tissues : 2016 U...
Rivaroxaban (XARELTO)
Lineage switch in Acute Leukemia
Acute Meningoencephalitis - Thesis presentation
Leptospirosis
Somatosensory seizures
Motor neuron disease in HIV
Horner's syndrome and Internuclear ophthalmoplegia
Diffuse idiopathic skeletal hyperostosis
Pachydermoperiostosis
AN UNUSUAL CAUSE OF DELAYED RECOVERY FROM NEUROMUSCULAR PARALYSIS DURING GENE...

Recently uploaded (20)

PPT
INTRODUCTION TO BIOSTATISTICS BY DR. SADAKAT.ppt
PPTX
Common Bacterial infections-converted_64bcdc4f77a3b7b90bdeb611f66c6ddd.pptx
PDF
CASE PRESENTATION1.pdf bipolar disorder in which both mania and depression h...
PPTX
migraine heaEDDDDDDDADFAAAAAAFdache (1).pptx
PDF
odontologia na oncologia - carie de radiação
PPTX
Drugs used in treatment of Malaria. Antimalarial Drugs.pptx
PPTX
Case report session Apendisitis Akut people.pptx
PPT
NIPT panel discussion HHI 09.12.2017 modified.ppt
PPTX
PPTX
Biomechanical preparation in primary teeth – Instrumentation and seminar 5 (2...
PDF
WHO Global TUBERCULOSIS Report 2018-2019
PPTX
Conflict Management: Defining conflict and understanding why it occurs is t...
PPTX
Non-Variceal-Upper-GI-Bleeding_-Comprehensive-Review_121037.pptx
PPTX
Emotional Well Being & Conflict Resolution_VKV.pptx
PPTX
Symphosium Dr saurab ad Dr. Khushbu.pptx
PPTX
VITAL PULP THERAPY in pediatric dentistry
PPT
Doppler - 5.ppt .........................
PDF
Liver Cirrhosis: Causes, Symptoms, Stages & Expert Treatment in Pune
PPTX
Right Lateral Medullary Syndrome (1).pptx
PPTX
SlideEgg_100085- World Mental Health Day.pptx
INTRODUCTION TO BIOSTATISTICS BY DR. SADAKAT.ppt
Common Bacterial infections-converted_64bcdc4f77a3b7b90bdeb611f66c6ddd.pptx
CASE PRESENTATION1.pdf bipolar disorder in which both mania and depression h...
migraine heaEDDDDDDDADFAAAAAAFdache (1).pptx
odontologia na oncologia - carie de radiação
Drugs used in treatment of Malaria. Antimalarial Drugs.pptx
Case report session Apendisitis Akut people.pptx
NIPT panel discussion HHI 09.12.2017 modified.ppt
Biomechanical preparation in primary teeth – Instrumentation and seminar 5 (2...
WHO Global TUBERCULOSIS Report 2018-2019
Conflict Management: Defining conflict and understanding why it occurs is t...
Non-Variceal-Upper-GI-Bleeding_-Comprehensive-Review_121037.pptx
Emotional Well Being & Conflict Resolution_VKV.pptx
Symphosium Dr saurab ad Dr. Khushbu.pptx
VITAL PULP THERAPY in pediatric dentistry
Doppler - 5.ppt .........................
Liver Cirrhosis: Causes, Symptoms, Stages & Expert Treatment in Pune
Right Lateral Medullary Syndrome (1).pptx
SlideEgg_100085- World Mental Health Day.pptx

Internal quality control (IQC) in coagulation lab

  • 1. Internal Quality Control in coagulation lab Dr Ankit Raiyani Haematology Department Sahyadri Speciality Hospital
  • 2. Presentation layout • Introduction to Quality Assurance and Quality Control • Definitions of common terms, Accuracy vs. Precision • Preparation/selection of reagents- • PTT reagent PT reagent MNPT • QC for manual coagulation tests • Quality control procedures- schedule • Internal Quality Control • LJ charts • Westgard rules • Other measures for IQC
  • 3. Introduction In the haematology laboratory it is essential to ensure that the right test is carried out on the right specimen and that the correct results are delivered to the appropriate recipient without delay. Quality assurance (QA) is defined as the overall programme for achieving these objectives. Quality control (QC) is defined as measures that must be included during each assay run to verify that the test is working properly.
  • 4. Quality Assurance vs. Quality Control Quality Assurance Quality Control An overall management plan to guarantee the integrity of data (The “system”) A series of analytical measurements used to assess the quality of the analytical data (The “tools”)
  • 5. The Quality Assurance Cycle •Data and Lab Management •Safety •Customer Service Patient/Client Prep Sample Collection Sample Receipt and Accessioning Sample TransportQuality Control Record Keeping Reporting Personnel Competency Test Evaluations Testing
  • 6. Hurdles in Quality Assurance 6
  • 7. Hurdles in Quality Assurance 7 Pre-analytical
  • 8. Hurdles in Quality Assurance 8 Analytical
  • 9. Hurdles in Quality Assurance 9 Post-analytical
  • 10. Terms relating to quality control in clinical laboratory practice Precision Can be controlled by replicate tests and by repeated tests on previously measured specimens Accuracy Can be checked only by the use of reference materials that have been assayed by reference methods Controls Are preparations that are used for either internal quality control or external quality assessment. They should not be used as standards Error The discrepancy between the result of a measurement and the true (or accepted true) value Random error An error which varies in an unpredictable manner, in magnitude and sign, when a large number of measurements of the same quantity are made under effectively identical conditions Systemic error (bias) An error which, in the course of a number of measurements of the same value of a given quantity, remains constant when measurements are made under the same conditions or varies according to a definite law when conditions change
  • 11. Accuracy vs. Precision Accuracy Precision Meaning How well a measurement agrees with an accepted value How well a series of measurement s agree with each other Achieved with… Calibration Quality control When low… Increases systemic errors Increases random errors
  • 12. Calibration Calibration is a comparison of measuring equipment against a standard instrument of higher accuracy to detect, correlate, adjust, rectify and document the accuracy of the instrument being compared.
  • 14. Reagent selection for PTT LA-sensitive reagents- (low-phospholipid reagents) • Siemens Actin FSL • IL HemosIL aPTT-SP • Stago PTT-LA Intermediate responsiveness to LA • Stago STAPTT-LA • IL HemosIl SynthASil Low responsiveness to LA • Siemens Actin FS • Stago C.K. Prest
  • 15. Reagent selection for PTT Selection of aPTT reagent depends on indication of aPTT-- The routine laboratory aPTT • used for evaluating the factor VIII, IX, XI deficiency and effect of heparin • sensitive to deficiencies of factors VIII, IX and XI at concentrations of 0.35–0.4 iu/ml. • responsive to unfractionated heparin over the therapeutic range of approximately 0.3–0.7 iu/ml. • low responsiveness to LA LA screening • High responsiveness to LA (silica+ low PL content) • Two phospholipid based clotting assays, one of which may be aPTT based, are required1 Intermediate responsiveness to both factor levels and LA • Used in labs with mixed sample load, when only one reagent is feasible/desired 1PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M. and DE GROOT, P. G. (2009), Update of the ISTH guidelines for lupus anticoagulant detection. Journal of Thrombosis and Haemostasis, 7: 1737–1740.
  • 16. Reagent selection for PT (Thromboplastin) Contains tissue factor and coagulant phospholipids Should have ISI value as close to 1 as possible. Source- human, bovine, rabbit brain/lung, human placenta, recombinant. Animal/human source- potential hazard of viral, prion and other infections from handling Recombinant thromboplastins- • Manufactured using recombinant human tissue factor produced in Escherichia coli and synthetic phospholipids, • Do not contain any other clotting factors such as prothrombin, factor VII and factor X. • Highly sensitive to factor deficiencies and oral anticoagulant-treated patient plasma samples
  • 17. What is ISI? International Sensitivity Index It is the quantitative measure, (in term of first international reference preparation of thromboplastin, human, combined, coded 67/40,) of the responsiveness of PT system to the defect induced by oral anticoagulants.
  • 18. Calculation of ISI Calibration of a test thromboplastin must be against a reference thromboplastin of the same species e.g. human against human, rabbit against rabbit Tests are performed By using reference thromboplastin and test thromboplastinon 20 normal donors not on anticoagulants and 60 patients who have been on oral anticoagulant treatment for at least 6 weeks and has stable INR. Prothrombin Times are performed in duplicate for each sample and the mean for each pair of tests derived. If there is more than a 10% difference in the clotting times between duplicate samples, the tests on that plasma sample should be repeated Mean of each pair of PT results are plotted on double-log paper with the reference sample on the Y axis and the test plasma on the X-axis.
  • 19. • A best fit line is drawn with points above the highest recorded PT and the lowest PT (see figure above). The slope of the line is calculated • Slope= AC/BC • ISItest = ISIStandard X Slope • INR= (PT/MNPT)ISI
  • 20. Preparation of pooled normal plasma and derivation of MNPT Citrated blood samples from at least 20 healthy volunteers (with PT, aPTT in normal range) should be collected. Equal numbers of male and female (not pregnant, not on OC pills) volunteers. Platelet poor plasma should be prepared. After pooling it should be immediately frozen at -80⁰ C in aliquots. The mean of PT of individual samples is calculated and taken as mean normal prothrombin time (MNPT). SD is calculated and reference range is taken as -2 SD to +2 SD. This pooled plasma is used as control with each batch of samples.
  • 21. MNPT calculation- 18/6/15 No PT PTT 1 11.5 28.6 2 12.4 27.6 3 10.5 29.2 4 10.6 26.2 5 10.6 32.1 6 11.7 29.8 7 11.1 30.7 8 12.2 31.6 9 11.9 30.9 10 11.3 31.5 11 10.6 29.1 12 12.3 29.7 13 10 26.3 14 11.9 28.7 15 11.6 33.3 16 11.2 31.4 17 10.5 28.2 18 12.5 28.7 19 10.6 27.8 20 11.1 27.7 21 10.9 29.5 22 11.3 31.9 23 11.3 29.9 24 12.5 32 total 272.1 712.4 Arithmetic Mean 11.3 29.6 SD 0.724 1.899 Range 9.8 to 12.7 25.8 to 33.4
  • 22. QC for manual coagulation tests All tests and controls should be performed in duplicates. Operator skill Required instruments should be properly calibrated. Centrifuge • Serviced and Calibrated every 6 months or after break down • Put on sturdy platform to reduce vibrations. • Sample load should be well balanced. Waterbaths • Temperature- 37⁰ C (checked in all four corners) • Transparent walls for transillumination • Filled with distilled water till indicated level. Pipette • Calibrated. • Disposable tips.
  • 23. Calibration of centrifuge Tachometer is used for calibration of RPM. Actual RPM is measured with tachometer at uniform intervals (minimum 5) for 5 times. Difference between actual and set values is calculated. If difference is more than +/- 5%, then repeat the test. If repeat reading again shows significant difference corrective measures are required. Timer and Temperature, when applicable, should also be calibrated.
  • 24. Calibration of pipettes Analytical balance is used to weigh pipetted volume in controlled environment. Distilled water is used as substrate. The pipette is checked 5 times with the maximum volume (nominal volume), the minimum volume or 10% of the maximum volume, whichever is higher. With multichannel pipettes, both volumes are tested with the two edge channels Conversion of mass to volume V = (w + e) x Z • V = Volume (µl), w = Weight (mg), e = Evaporation loss (mg), • Z = Conversion factor for mg/µl conversion (temperature, pressure) Acceptable accuracy and precision range varies with pipette capacity.
  • 26. Quality control procedures The procedures that should be included in a quality assurance programme vary with the tests undertaken, the instruments used and (especially if these include a fully automatic system) the size of the laboratory and the numbers of specimens handled. Some control procedures should be performed daily and other performance checks should be done at appropriate intervals. The latter is particularly important when there is a change in staff and after maintenance service or repair has been carried out on equipment.
  • 27. Schedule for quality control procedures 1. Calibration with reference standards Instruments, pipettes 6-month intervals or more frequently if control chart or EQA indicates bias or fluctuation in results and after any repair/service Others: thermometers, scales, etc Annually 2. Control chart with control material Daily or more frequently with each batch of specimens Duplicate tests on two or three patients’ samples: if control chart or delta check shows discrepancies 3. Analysis of patients’ results Cumulative results: following previous tests and if changes in clinical state 4. EQAS performance Assessment monthly
  • 28. Components of Quality control Internal quality control (IQC) • Monitoring the haematology test procedures to ensure continual evaluation of the reliability of the daily work of the laboratory with validation of tests before reports are released External quality assessment (EQA) • Evaluation by an outside agency of the between-laboratory and between-method comparability. It can be organized nationally, regionally or internationally
  • 29. Internal Quality Control IQC is based on monitoring the haematology test procedures that are performed in the laboratory and includes measurements on specially prepared materials and repeated measurements on routine specimens, together with daily statistical analysis of the data. Primarily a demonstration of Precision. It ensures continual checks that the established reliability of the laboratory’s work does not fluctuate and that reports are validated before they are released
  • 30. IQC includes Control charts with tests on control materials Duplicate tests on a proportion of the specimens Consistency of mean values of patient data Correlation check
  • 31. Control charts These were first applied in clinical chemistry by Levey and Jennings. They are now widely used in haematology for both automated and manual procedures. Samples of the control specimen are included in every batch of patients’ specimens and the results are checked on a control chart. To check precision, it is not necessary to know the exact value of the control specimen.
  • 32. Controls Level 1 controls- normal controls Level 2 controls- low abnormal controls Level 3 controls- high abnormal controls (not used routinely) Minimum of one level QC at least once a day. Patient samples >25 per day  Employ 2 levels of QC at least once a day. Patient samples >75 per day Employ 2 levels of QC at least twice a day at appropriate intervals Controls are intended to simulate random sampling, they must be treated exactly like the patients’ specimens.
  • 33. Preparation of LJ charts Mean value and standard deviation (SD) of the control specimen should first be established in the laboratory where the tests are performed. Using arithmetic graph paper, a horizontal line is drawn to represent the mean (as a base), and on an appropriate scale of quantity and unit, lines representing +2SD and −2SD are drawn above and below the mean. The results of successive control sample measurements are plotted. If the test is satisfactory, sequential results oscillate about the mean value and <5% of the results fall outside 2SD.
  • 34. LJ chart PT / low abnormal May 2015
  • 35. LJ chart PT / low abnormal June 2015
  • 36. LJ chart PT / low abnormal June 2015 Reason ??
  • 37. LJ chart aPTT / NCPNP
  • 38. LJ chart F VIIIc / NCPNP
  • 39. Westgard rules “Multirule Quality Control” developed by Dr. James O. Westgard based on statistical concepts Uses a combination of decision criteria or control rules Allows determination of whether an analytical run is “in-control” or “out-of-control”
  • 40. Westgard multirule system 12S rule 13S rule 22S rule R4S rule 41S rule 10X rule
  • 41. Westgard 12S rule “warning rule” One of two control results falls outside ±2SD Alerts tech to possible problems Not cause for rejecting a run Must then evaluate the 13S rule
  • 42. 12S Rule = A warning to trigger careful inspection of the control data 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 12S rule violation
  • 43. Westgard – 13S Rule If either of the two control results falls outside of ±3SD, rule is violated Run must be rejected If 13S not violated, check 22S
  • 44. 13S Rule = Reject the run when a single control measurement exceeds the +3SD or -3SD control limit 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 13S rule violation
  • 45. Westgard – 22S Rule 2 consecutive control values for the same level fall outside of ±2SD in the same direction, or Both controls in the same run exceed ±2SD Patient results cannot be reported Requires corrective action
  • 46. 22S Rule = Reject the run when 2 consecutive control measurements exceed the same +2SD or -2SD control limit 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 22S rule violation
  • 47. Westgard – R4S Rule One control exceeds the mean by –2SD, and the other control exceeds the mean by +2SD The range between the two results will therefore exceed 4 SD Random error has occurred, test run must be rejected
  • 48. R4S Rule = Reject the run when 1 control measurement exceed the +2SD and the other exceeds the -2SD control limit 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD R4S rule violation
  • 49. Westgard – 41S Rule Requires control data from previous runs Four consecutive QC results for one level of control are outside ±1SD, or Both levels of control have consecutive results that are outside ±1SD
  • 50. Westgard – 10X Rule Requires control data from previous runs Ten consecutive QC results for one level of control are on one side of the mean, or Both levels of control have five consecutive results that are on the same side of the mean
  • 51. 10x Rule = Reject the run when 10 consecutive control measurements fall on one side of the mean 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 10x rule violation
  • 52. Alternate ways of checking the precision of routine work Duplicate tests on patients’ samples • Test of precision can be performed by repeat analysis on samples from two or three patients Correlation Check • any unexpected result of a test must be checked to see whether it can be explained on clinical grounds or whether it correlates with other tests “Delta” check • Blood count from the same patient checked repeatedly over 2-3 weeks • Values should not change significantly providing the patients condition has not altered significantly Patient data • Need enough through put (i.e. >100 samples /day) • MCV, MCHC, MCH should not vary, providing the patient population is stable and samples from different sets of patients are not batched together. • Mean of 20 successive patients can be plotted and any drift in the calibration of the analyzer will be recognized

Editor's Notes

  • #12: Target, bull’s eye
  • #16: Silica, celite, ellagic acid, kaolin Hemosil synthasil- silica
  • #17: Screening using a single aPTT assay is inadequate to rule out the presence of LA
  • #18: RecombiPlasTin 2g
  • #23: When to calculate MNPT- 1. fresh pooled plasma, 2. change in reagent lot, 3. post instrument calibration.
  • #25: If 7 sample tubes then use one tube with water to balance
  • #26: Using view port in centrifuge. Source- SOP, John Hopkins