Radiation Detectors
Dr Kiran Kumar BR
Scintillation Detectors
 The detection of ionising radiation by scintillation light
produced by certain materials – oldest methods on record.
 A scintillation detector consists of the following :
o A luminescent material
o An optical device – facilitate collection of light
o An optical coupling b/w luminescent material and
photomultiplier
o The photomultiplier tube
o Electrical circuit – record pulses appearing at output of
PMT
Working Principle
 Crystals such as NaI is connected to the PMT as
shown in figure
 PMT has 10 to 12 dynodes and an anode
 Each dynode is 100 volts +ve with respect to the first
 Crystal is coupled to PMT by silicon grease to prevent
loss of photons by reflection
Radiation Detectors-operation and applications.pptx
 “When a photon interacts with a crystal ,
photoelectrons are produced “
 Photoelectrons travel through crystal , excite and
ionise atoms of the crystal
 Flashes of UV or visible light are produced
 These photons are radiated in all directions
 Mno2 layer reflects the photons towards the window
side
 Low energy photons fall on photocathode of
multiplier tube and ejects no of photoelectrons
Photoelectrons are accelerated by potential
applied b/w cathode and first dynode of the tube
On striking first dynode , each photoelectron
ejects several electrons by secondary emission
Electron multiplication process repeated at
subsequent dynodes , each at higher potential
than preceding one
After multiplication of about 10^5 to 10^9 , the
avalanche arrives at collector plate
Produces a voltage pulse at output condenser ,
coupled to an exit pulse amplifying circuit
The initial energy of single ionising particle is
transformed to single volt pulse
Whole system enclosed in light tight box –
eliminate effects others than those due to
incident ionising radiation
No of electrons ejected from photocathode α
amount of light reaches it
Amount of light α energy absorbed from photon
beam by crystal
“Size of pulse from the anode α energy of
incident photon in crystal”
• ADVANTAGES: higher quantum efficiency and
compact size.
• DISADVANTAGES : Electronic noise is a major
problem due to the small signal amplitude
GM Counter
• Geiger counter instrument used for the detection of
ionizing radiation
• GM Counter is a Gas detectors that detect ionizing
radiation , alpha particles, beta particles, or gamma
rays .
• This have an easily read digital display, emit an
audible "beep´ as a constant indication of exposure
rate,
• GM counters cannot determine the type or energy
of the detected radiation.
GM Tube
GM tube is the sensing element of the GM
counter.
Uses the Townsend Avalanche phenomenon
to produce an easily detectable electronic
pulse.
Cylindrical metal envelope
Filled with Inert gas
(Helium, neon, or argon)
at a low pressure (~0.1 atm)
• Wall of the tube are metal
( Stainless steel or nickle)
inside coated with metal or
graphite to form cathode.
• Anode is a wire (tungsten) of about o.oo3 -
0.004 inch diameter stretched along the center
of the cylinder
• high PD applied between this electrodes
• The tube operates in
the "Geiger" region of
ion pair generation
WORKING
 When ionizing radiation strikes the tube some
molecules of the filled gas are ionized (Directly by
incident radiation or indirectly by means of secondary
electrons produced in the wall of the tube)
 This creates positively charged Ions (cation) and
negatively charged (electron).
 The strong electric field created by the tube’s
electrodes accelerates the positive ions toward the
cathode and electrons toward the anode.
• Due to high volt (GM region 1000v-1400v)
the ions, accelerated by the electric field, can gain enough
energy to produce ionization by collision with gas
molecules. This results in a rapid multiplication of ions and
creates a large no. of electrons avalanche which spread
along the anode.
 This is the gas multiplication effect , which is able to
produce a significant output pulse from a single ionizing
event.
• One avalanche per original ionizing event will
produce 10^6 to 10^8 excited molecules.
• The excited gas atoms produce uv photons in the
original avalanche.
Moves laterally to the axis of the anode to
instigate further ionizing event by collision with
gas molecule.
 These collision produce further avalanche and the
process continued in a chain reaction.
 This process is known as multiple avalanches
 production of multiple avalanches results in an
increased multiplication factor which can produce
109 to 1010 ion pairs
Radiation Detectors-operation and applications.pptx
Radiation Detectors-operation and applications.pptx
• The speed of avalanche is typically 2-4 cm per
microsecond.
• This short, intense pulse of current can be measured
as a count event in the form of a voltage pulse
Dead Time :
After a count has been recorded, it takes the
G-M tube a certain amount of time to reset itself to
be ready to record the next count.
• The Dead Time of Geiger tube is the Period between
the initial pulse time and time at which a second
Geiger discharge
• The tubes can produce no further pulses during the
dead time (typically 50 - 100 microseconds)
• The time interval required for GM counter to return
to its original state to deliver full amplitude pulse is
called Recovery Time.
Quenching
• If a poor diatomic gas quencher were introduced
to the tube, the positive argon ions, during their
motion toward the cathode, would have multiple
collisions with the quencher gas molecules and
transfer their charge and some energy to them.
• Neutral argon atoms would then be produced
and the quencher gas ions would reach the
cathode instead
• Disadvantage of quenching is that for a short
time after a discharge pulse has occurred (the
so-called dead time, which is typically 50 - 100
microseconds), the tube is rendered
insensitive
Detection of neutrons
 GM tubes will not detect neutrons since these do not ionize the
gas.
 To make tube neutron sensitive ,
I. Tube wall should be coated with boron. OR
II. Tube should contain Helium – 3 or boron Tri-fluoride.
 The neutron interact with Boron nuclei, producing alpha particles
or directly with the Helium – 3 nuclei producing hydrogen and
tritium ion and electrons.
 These charged particles then trigger the normal avalanche process.
Characteristics
 Geiger counters are used to detect radiation
(low level)
 It respond to commonly alpha and beta particles as well as
gamma and x-ray
 Has no energy information.
 High sensitivity (100% for each ionizing event)
 Measures low exposure rates (0.1 mR / hr)
 Large dead time (~ 100μs)
 large output signal and reasonable cost.
Proportional counters
• It works in the proportional region of voltage
response curve.
• Principle is similar to that of GM counter – gas
multiplication effect
Difference between propotionl and GM
counter
• The proportional counter is operating in linear mode,
so the energy pulse measured by the detector is
proportional to the energy of the radioactive
particle.
• Thus, the average current flow in the detector is a
function of both the activity (curies) and the energy
(kev) of the source.
• This detector is better at measuring the dose rate of
the source.
• The Geiger-Muller detector, on the other hand, operates in
avalanche mode, so the energy pulse measured by the
detector is not proportional to the energy of the
radioactive particle.
• Thus, the average current flow in the detector is a function
of only the activity (curies) of the source. This detector is
better at measuring the activity of the source, and can be
more sensitive to lower energy particles, at the loss of
discrimination of what those particles are.
Thank you

More Related Content

PPT
Gas filled detectors
Amara Usman
 
PDF
Introduction to radiation detectors for ug level
link003ab
 
PPTX
Nuclear radiation detecters
SangeethaSivakumar10
 
PPTX
Gm counter& scintillation counter
SoniyaS13
 
PPTX
Radio-active Counters
Nizam Ashraf
 
PPTX
UNIT V: MEASUREMENT OF RADIOACTIVITY,BPHARM
rikttahahmed
 
PDF
Geiger muller counting system
Gaurav Bhati
 
DOCX
Assignment radiation
Sany Ali
 
Gas filled detectors
Amara Usman
 
Introduction to radiation detectors for ug level
link003ab
 
Nuclear radiation detecters
SangeethaSivakumar10
 
Gm counter& scintillation counter
SoniyaS13
 
Radio-active Counters
Nizam Ashraf
 
UNIT V: MEASUREMENT OF RADIOACTIVITY,BPHARM
rikttahahmed
 
Geiger muller counting system
Gaurav Bhati
 
Assignment radiation
Sany Ali
 

Similar to Radiation Detectors-operation and applications.pptx (20)

PPTX
Radiation detection and measurement.pptx
RamJoshi36
 
PPTX
Radiation detection instrumentation fundamentals
Nitish Virmani
 
PPTX
Geiger muller counter
ARCHITBHARTI
 
PPTX
Radiation detectors
Jonathan Lalrinmawia
 
PPTX
Geiger–Müller Counter
ABHISEK KUMAR SINGH
 
PPTX
Nuclear detectors
zeeshan farooq
 
PPTX
Chapter 3 detection devices
ROBERT ESHUN
 
PDF
Geiger Muller counter
ShahzadRafiq6
 
PPTX
Slide share
Vasuki Arumugam
 
PPTX
Radiation detection and measurements
Vasuki Arumugam
 
PPTX
Radiopharmaceuticals.pptx
MihirDeshmukh7
 
PPTX
Analytical biochemistry
Monika Uma Shankar
 
PDF
Neutron detection in medical industry and hsopital
medicalHunter2
 
PPTX
MEASUREMENT OF RADIATION DOSE.pptx
VishnuDutt47
 
PPTX
Basic dosimetric principle and dosimeters
Vinay Desai
 
PPT
25 -radiation_detection_&_measurement_i
murty61
 
PPTX
Geiger muller counter
MTahirYounas
 
PPTX
Geiger muller counter- brief introduction
subhananthini jeyamurugan
 
PPT
Nuclear i nstrumentation
shalet kochumuttath Shaji
 
Radiation detection and measurement.pptx
RamJoshi36
 
Radiation detection instrumentation fundamentals
Nitish Virmani
 
Geiger muller counter
ARCHITBHARTI
 
Radiation detectors
Jonathan Lalrinmawia
 
Geiger–Müller Counter
ABHISEK KUMAR SINGH
 
Nuclear detectors
zeeshan farooq
 
Chapter 3 detection devices
ROBERT ESHUN
 
Geiger Muller counter
ShahzadRafiq6
 
Slide share
Vasuki Arumugam
 
Radiation detection and measurements
Vasuki Arumugam
 
Radiopharmaceuticals.pptx
MihirDeshmukh7
 
Analytical biochemistry
Monika Uma Shankar
 
Neutron detection in medical industry and hsopital
medicalHunter2
 
MEASUREMENT OF RADIATION DOSE.pptx
VishnuDutt47
 
Basic dosimetric principle and dosimeters
Vinay Desai
 
25 -radiation_detection_&_measurement_i
murty61
 
Geiger muller counter
MTahirYounas
 
Geiger muller counter- brief introduction
subhananthini jeyamurugan
 
Nuclear i nstrumentation
shalet kochumuttath Shaji
 
Ad

More from debeshidutta2 (10)

PPTX
Nanosensor-basics, application and future.pptx
debeshidutta2
 
PPTX
FUNDAMENTALS OF MECHATRONICS AND APPLICATIONS.pptx
debeshidutta2
 
PPT
WEINER FILTERING AND BASIS FUNCTIONS.ppt
debeshidutta2
 
PPTX
Thermoluminiscent Dosimetry-Principle, Charateristics, and Applications.pptx
debeshidutta2
 
PPT
Introduction to adaptive filtering and its applications.ppt
debeshidutta2
 
PPT
Introduction, concepts, and mathematics of IIR filters.ppt
debeshidutta2
 
PPT
DIGITAL FILTER SPECIFICATIONS AND MATHEMATICS.ppt
debeshidutta2
 
PPT
BIOMEDICAL SENSORS AND MEASUREMENT SYSTEMS.ppt
debeshidutta2
 
PPTX
Basic programming in Python Environment.pptx
debeshidutta2
 
PPTX
Transducers in healthcare applications.pptx
debeshidutta2
 
Nanosensor-basics, application and future.pptx
debeshidutta2
 
FUNDAMENTALS OF MECHATRONICS AND APPLICATIONS.pptx
debeshidutta2
 
WEINER FILTERING AND BASIS FUNCTIONS.ppt
debeshidutta2
 
Thermoluminiscent Dosimetry-Principle, Charateristics, and Applications.pptx
debeshidutta2
 
Introduction to adaptive filtering and its applications.ppt
debeshidutta2
 
Introduction, concepts, and mathematics of IIR filters.ppt
debeshidutta2
 
DIGITAL FILTER SPECIFICATIONS AND MATHEMATICS.ppt
debeshidutta2
 
BIOMEDICAL SENSORS AND MEASUREMENT SYSTEMS.ppt
debeshidutta2
 
Basic programming in Python Environment.pptx
debeshidutta2
 
Transducers in healthcare applications.pptx
debeshidutta2
 
Ad

Recently uploaded (20)

PDF
2010_Book_EnvironmentalBioengineering (1).pdf
EmilianoRodriguezTll
 
PDF
67243-Cooling and Heating & Calculation.pdf
DHAKA POLYTECHNIC
 
PPTX
Chapter_Seven_Construction_Reliability_Elective_III_Msc CM
SubashKumarBhattarai
 
PDF
Zero carbon Building Design Guidelines V4
BassemOsman1
 
PPTX
MT Chapter 1.pptx- Magnetic particle testing
ABCAnyBodyCanRelax
 
PPTX
easa module 3 funtamental electronics.pptx
tryanothert7
 
PDF
Advanced LangChain & RAG: Building a Financial AI Assistant with Real-Time Data
Soufiane Sejjari
 
PDF
Top 10 read articles In Managing Information Technology.pdf
IJMIT JOURNAL
 
DOCX
SAR - EEEfdfdsdasdsdasdasdasdasdasdasdasda.docx
Kanimozhi676285
 
PDF
Chad Ayach - A Versatile Aerospace Professional
Chad Ayach
 
PDF
Software Testing Tools - names and explanation
shruti533256
 
PPT
Ppt for engineering students application on field effect
lakshmi.ec
 
PPTX
Tunnel Ventilation System in Kanpur Metro
220105053
 
PPTX
Module2 Data Base Design- ER and NF.pptx
gomathisankariv2
 
PDF
The Effect of Artifact Removal from EEG Signals on the Detection of Epileptic...
Partho Prosad
 
PDF
JUAL EFIX C5 IMU GNSS GEODETIC PERFECT BASE OR ROVER
Budi Minds
 
PDF
67243-Cooling and Heating & Calculation.pdf
DHAKA POLYTECHNIC
 
PDF
Cryptography and Information :Security Fundamentals
Dr. Madhuri Jawale
 
PDF
Packaging Tips for Stainless Steel Tubes and Pipes
heavymetalsandtubes
 
PDF
Zero Carbon Building Performance standard
BassemOsman1
 
2010_Book_EnvironmentalBioengineering (1).pdf
EmilianoRodriguezTll
 
67243-Cooling and Heating & Calculation.pdf
DHAKA POLYTECHNIC
 
Chapter_Seven_Construction_Reliability_Elective_III_Msc CM
SubashKumarBhattarai
 
Zero carbon Building Design Guidelines V4
BassemOsman1
 
MT Chapter 1.pptx- Magnetic particle testing
ABCAnyBodyCanRelax
 
easa module 3 funtamental electronics.pptx
tryanothert7
 
Advanced LangChain & RAG: Building a Financial AI Assistant with Real-Time Data
Soufiane Sejjari
 
Top 10 read articles In Managing Information Technology.pdf
IJMIT JOURNAL
 
SAR - EEEfdfdsdasdsdasdasdasdasdasdasdasda.docx
Kanimozhi676285
 
Chad Ayach - A Versatile Aerospace Professional
Chad Ayach
 
Software Testing Tools - names and explanation
shruti533256
 
Ppt for engineering students application on field effect
lakshmi.ec
 
Tunnel Ventilation System in Kanpur Metro
220105053
 
Module2 Data Base Design- ER and NF.pptx
gomathisankariv2
 
The Effect of Artifact Removal from EEG Signals on the Detection of Epileptic...
Partho Prosad
 
JUAL EFIX C5 IMU GNSS GEODETIC PERFECT BASE OR ROVER
Budi Minds
 
67243-Cooling and Heating & Calculation.pdf
DHAKA POLYTECHNIC
 
Cryptography and Information :Security Fundamentals
Dr. Madhuri Jawale
 
Packaging Tips for Stainless Steel Tubes and Pipes
heavymetalsandtubes
 
Zero Carbon Building Performance standard
BassemOsman1
 

Radiation Detectors-operation and applications.pptx

  • 2. Scintillation Detectors  The detection of ionising radiation by scintillation light produced by certain materials – oldest methods on record.  A scintillation detector consists of the following : o A luminescent material o An optical device – facilitate collection of light o An optical coupling b/w luminescent material and photomultiplier o The photomultiplier tube o Electrical circuit – record pulses appearing at output of PMT
  • 3. Working Principle  Crystals such as NaI is connected to the PMT as shown in figure  PMT has 10 to 12 dynodes and an anode  Each dynode is 100 volts +ve with respect to the first  Crystal is coupled to PMT by silicon grease to prevent loss of photons by reflection
  • 5.  “When a photon interacts with a crystal , photoelectrons are produced “  Photoelectrons travel through crystal , excite and ionise atoms of the crystal  Flashes of UV or visible light are produced  These photons are radiated in all directions  Mno2 layer reflects the photons towards the window side  Low energy photons fall on photocathode of multiplier tube and ejects no of photoelectrons
  • 6. Photoelectrons are accelerated by potential applied b/w cathode and first dynode of the tube On striking first dynode , each photoelectron ejects several electrons by secondary emission Electron multiplication process repeated at subsequent dynodes , each at higher potential than preceding one After multiplication of about 10^5 to 10^9 , the avalanche arrives at collector plate Produces a voltage pulse at output condenser , coupled to an exit pulse amplifying circuit
  • 7. The initial energy of single ionising particle is transformed to single volt pulse Whole system enclosed in light tight box – eliminate effects others than those due to incident ionising radiation No of electrons ejected from photocathode α amount of light reaches it Amount of light α energy absorbed from photon beam by crystal “Size of pulse from the anode α energy of incident photon in crystal”
  • 8. • ADVANTAGES: higher quantum efficiency and compact size. • DISADVANTAGES : Electronic noise is a major problem due to the small signal amplitude
  • 9. GM Counter • Geiger counter instrument used for the detection of ionizing radiation • GM Counter is a Gas detectors that detect ionizing radiation , alpha particles, beta particles, or gamma rays . • This have an easily read digital display, emit an audible "beep´ as a constant indication of exposure rate, • GM counters cannot determine the type or energy of the detected radiation.
  • 10. GM Tube GM tube is the sensing element of the GM counter. Uses the Townsend Avalanche phenomenon to produce an easily detectable electronic pulse.
  • 11. Cylindrical metal envelope Filled with Inert gas (Helium, neon, or argon) at a low pressure (~0.1 atm) • Wall of the tube are metal ( Stainless steel or nickle) inside coated with metal or graphite to form cathode.
  • 12. • Anode is a wire (tungsten) of about o.oo3 - 0.004 inch diameter stretched along the center of the cylinder • high PD applied between this electrodes • The tube operates in the "Geiger" region of ion pair generation
  • 13. WORKING  When ionizing radiation strikes the tube some molecules of the filled gas are ionized (Directly by incident radiation or indirectly by means of secondary electrons produced in the wall of the tube)  This creates positively charged Ions (cation) and negatively charged (electron).  The strong electric field created by the tube’s electrodes accelerates the positive ions toward the cathode and electrons toward the anode.
  • 14. • Due to high volt (GM region 1000v-1400v) the ions, accelerated by the electric field, can gain enough energy to produce ionization by collision with gas molecules. This results in a rapid multiplication of ions and creates a large no. of electrons avalanche which spread along the anode.  This is the gas multiplication effect , which is able to produce a significant output pulse from a single ionizing event.
  • 15. • One avalanche per original ionizing event will produce 10^6 to 10^8 excited molecules. • The excited gas atoms produce uv photons in the original avalanche. Moves laterally to the axis of the anode to instigate further ionizing event by collision with gas molecule.
  • 16.  These collision produce further avalanche and the process continued in a chain reaction.  This process is known as multiple avalanches  production of multiple avalanches results in an increased multiplication factor which can produce 109 to 1010 ion pairs
  • 19. • The speed of avalanche is typically 2-4 cm per microsecond. • This short, intense pulse of current can be measured as a count event in the form of a voltage pulse Dead Time : After a count has been recorded, it takes the G-M tube a certain amount of time to reset itself to be ready to record the next count.
  • 20. • The Dead Time of Geiger tube is the Period between the initial pulse time and time at which a second Geiger discharge • The tubes can produce no further pulses during the dead time (typically 50 - 100 microseconds) • The time interval required for GM counter to return to its original state to deliver full amplitude pulse is called Recovery Time.
  • 21. Quenching • If a poor diatomic gas quencher were introduced to the tube, the positive argon ions, during their motion toward the cathode, would have multiple collisions with the quencher gas molecules and transfer their charge and some energy to them. • Neutral argon atoms would then be produced and the quencher gas ions would reach the cathode instead
  • 22. • Disadvantage of quenching is that for a short time after a discharge pulse has occurred (the so-called dead time, which is typically 50 - 100 microseconds), the tube is rendered insensitive
  • 23. Detection of neutrons  GM tubes will not detect neutrons since these do not ionize the gas.  To make tube neutron sensitive , I. Tube wall should be coated with boron. OR II. Tube should contain Helium – 3 or boron Tri-fluoride.  The neutron interact with Boron nuclei, producing alpha particles or directly with the Helium – 3 nuclei producing hydrogen and tritium ion and electrons.  These charged particles then trigger the normal avalanche process.
  • 24. Characteristics  Geiger counters are used to detect radiation (low level)  It respond to commonly alpha and beta particles as well as gamma and x-ray  Has no energy information.  High sensitivity (100% for each ionizing event)  Measures low exposure rates (0.1 mR / hr)  Large dead time (~ 100μs)  large output signal and reasonable cost.
  • 25. Proportional counters • It works in the proportional region of voltage response curve. • Principle is similar to that of GM counter – gas multiplication effect
  • 26. Difference between propotionl and GM counter • The proportional counter is operating in linear mode, so the energy pulse measured by the detector is proportional to the energy of the radioactive particle. • Thus, the average current flow in the detector is a function of both the activity (curies) and the energy (kev) of the source. • This detector is better at measuring the dose rate of the source.
  • 27. • The Geiger-Muller detector, on the other hand, operates in avalanche mode, so the energy pulse measured by the detector is not proportional to the energy of the radioactive particle. • Thus, the average current flow in the detector is a function of only the activity (curies) of the source. This detector is better at measuring the activity of the source, and can be more sensitive to lower energy particles, at the loss of discrimination of what those particles are.