General requirements for dosimeters
#Dosimeter is a device that measures directly
or indirectly
• Exposure
• Kerma
• Absorbed dose
• Equivalent dose
• Or other related quantities.
#The dosimeter along with its reader is
referred to as a
dosimetry system.
A useful dosimeter exhibits the following
properties:
• High accuracy and precision
• Linearity of signal with dose over a wide
range
• Small dose and dose rate dependence
• Flat Energy response
• Small directional dependence
• High spatial resolution
• Large dynamic range
 Accuracy specifies the proximity of the mean
value of a measurement to the true value.
 Precision specifies the degree of
reproducibility of a measurement.
Note:
High precision
is equivalent to small standard deviation.
Examples for use of precision and accuracy:
High precision High precision Low precision Low precision
and and and and
High accuracy Low accuracy High accuracy Low accuracy
Note: The accuracy and precision associated
with a measurement is often expressed in
terms of its uncertainty.
New Concept by the International Organization
for Standardization (ISO) "Guide to the
expression of uncertainty in measurement“
 This new guide serves as a clear procedure
for characterizing the quality of a
measurement.
 It is easily understood and generally
accepted.
 It defines uncertainty as a quantifiable
attribute.
 Standard uncertainty: is the uncertainty of a
result expressed as standard deviation.
 Type A standard uncertainty is evaluated by
statistical analysis of a series of observations.
 Type B standard uncertainty is evaluated by
means other than statistical analysis. This
classification is for convenience of discussion
only.
 It is not meant to indicate that there is a
difference in the nature of
 the uncertainty such as random or systematic.
 Combined uncertainties:
The determination of the final result is normally
based on several components.
Linearity:
 The dosimeter reading should be linearly
proportional to the dosimetric quantity.
 Beyond a certain range, usually a non-linearity sets
in.
 This effect depends on the type of dosimeter.
Two possible cases
Case A:
• linearity
• supralinearity
• saturation
Case B:
• linearity
• saturation
Dose rate dependence :
 M/D may be called the response of a dosimeter
system
 When an integrated response is measured, the
dosimetric quantity should be independent of the
dose rate dD/dt of the quantity.
 Other formulation:
The response M/D should be constant for different
dose rates (dD/dt)1 and (dD/dt)
2. M = 􀀁 (M / D)(dD / dt)dt
M = (M / D)􀀁 (dD / dt)dt
 Energy: The response of a dosimetric system
is
generally a function of the radiation energy.
 The term "radiation quality" is often used to
express a specific distribution of the energy
of radiation.
 Therefore, a dependence on energy can also
be called a dependence on radiation quality.
 Since calibration is done at a specified beam
quality, a reading should generally be
corrected if the user's beam quality is not
identical to the calibration beam quality.
A small radiation monitoring device worn by persons
entering environments that may contain radiation .
# Desirable characteristics
 Should be lightweight, durable, and reliable
 Should be inexpensive
 Healthcare or laboratory workers in non-
emergency environments that may contain
radiation
 Examples: radiology, nuclear medicine, and
radiation oncology department staff
 Workers in emergency environments that
may contain radiation
 Examples: first responders and first receivers
 Workers in industrial environments where
radiation is used
 Examples: nuclear power plant workers or
employees at radiation sterilizing facilities
 Flat badges are usually worn on the torso, at the
collar or chest level, but can be worn on the
belt, or forearm
 Ring shaped badges can be worn on the finger
when dose to the finger may exceed dose to the
badge worn elsewhere on the body
 First responders and first receivers
 Wear water-resistant personal dosimeters on the
outer layer of personal protective equipment (PPE).
 Should be able to easily see and hear a dosimeter
alarm while wearing PPE
 May wear a personal dosimeter underneath
waterproof outerwear
 Film badge
 Pocket ionization chambers
 Thermo luminescent dosimeters (TLD)
 Optically stimulated luminescence (OSL)
 Solid State
 Most widely used and most economical
 Consists of three parts:
 Plastic film holder
 Metal filters
 Film packet
 Can read x, gamma, and beta radiation
 Accurate from 10mrem - 500rem
 Developed and read by densitometer
 A certain density value equals a certain level of radiation
 Read with a control badge
 Results generally sent as a printout
 Lightweight, durable,
portable
 Cost efficient
 Permanent legal record
 Can differentiate between
scatter and primary beam
 Can discriminate between
x, gamma, and beta
radiation
 Can indicate direction from
where radiation came from
 Control badge can indicate
if exposed in transit
 Only records exposure
where it’s worn
 Not effective if not worn
 Can be affected by heat
and humidity
 Sensitivity is decreased
above and below 50 keV
 Exposure cannot be
determined on day of
exposure
 Accuracy limited to + or -
20%
 The most sensitive personnel dosimeter
 Two types
 Self-reading
 Non self-reading
 Can only be read once
 Detects gamma or x-radiation
 Small, compact,
easy to use
 Reasonably
accurate and
sensitive
 Provides
immediate reading
 Expensive
 Readings can be
lost
 Must be read each
day
 No permanent
record
 Susceptible to
false readout if
dropped or jarred
 Looks like a film badge
 Contains a lithium fluoride crystal
 Responds to radiation similarly to skin
 Measured by a TLD analyzer
 Crystal will luminescence if exposed to
radiation, then heated
 More accurate than a film badge
 Crystals contained
in TLD interact
with ionizing
radiation as tissue
does
 Determines dose
more accurately
 The initial cost is
greater than that
of a film badge
 Can only be read
once
 Records exposure
only where worn
• “Captures” information in an Aluminum
Oxide matrix
• Releases information by laser stimulation
• Can be reread after processing
• Durable
• Landauer Only
 Provides instantaneous information
regarding dose accumulation
 Simple to use
 Not a “legal” record
 Dose range device dependent
Radiation Dosimeters
Radiation Dosimeters

Radiation Dosimeters

  • 2.
    General requirements fordosimeters #Dosimeter is a device that measures directly or indirectly • Exposure • Kerma • Absorbed dose • Equivalent dose • Or other related quantities. #The dosimeter along with its reader is referred to as a dosimetry system.
  • 3.
    A useful dosimeterexhibits the following properties: • High accuracy and precision • Linearity of signal with dose over a wide range • Small dose and dose rate dependence • Flat Energy response • Small directional dependence • High spatial resolution • Large dynamic range
  • 4.
     Accuracy specifiesthe proximity of the mean value of a measurement to the true value.  Precision specifies the degree of reproducibility of a measurement. Note: High precision is equivalent to small standard deviation.
  • 5.
    Examples for useof precision and accuracy: High precision High precision Low precision Low precision and and and and High accuracy Low accuracy High accuracy Low accuracy
  • 6.
    Note: The accuracyand precision associated with a measurement is often expressed in terms of its uncertainty.
  • 7.
    New Concept bythe International Organization for Standardization (ISO) "Guide to the expression of uncertainty in measurement“  This new guide serves as a clear procedure for characterizing the quality of a measurement.  It is easily understood and generally accepted.  It defines uncertainty as a quantifiable attribute.
  • 8.
     Standard uncertainty:is the uncertainty of a result expressed as standard deviation.  Type A standard uncertainty is evaluated by statistical analysis of a series of observations.  Type B standard uncertainty is evaluated by means other than statistical analysis. This classification is for convenience of discussion only.  It is not meant to indicate that there is a difference in the nature of  the uncertainty such as random or systematic.
  • 9.
     Combined uncertainties: Thedetermination of the final result is normally based on several components. Linearity:  The dosimeter reading should be linearly proportional to the dosimetric quantity.  Beyond a certain range, usually a non-linearity sets in.  This effect depends on the type of dosimeter.
  • 10.
    Two possible cases CaseA: • linearity • supralinearity • saturation Case B: • linearity • saturation
  • 11.
    Dose rate dependence:  M/D may be called the response of a dosimeter system  When an integrated response is measured, the dosimetric quantity should be independent of the dose rate dD/dt of the quantity.  Other formulation: The response M/D should be constant for different dose rates (dD/dt)1 and (dD/dt) 2. M = 􀀁 (M / D)(dD / dt)dt M = (M / D)􀀁 (dD / dt)dt
  • 12.
     Energy: Theresponse of a dosimetric system is generally a function of the radiation energy.  The term "radiation quality" is often used to express a specific distribution of the energy of radiation.  Therefore, a dependence on energy can also be called a dependence on radiation quality.  Since calibration is done at a specified beam quality, a reading should generally be corrected if the user's beam quality is not identical to the calibration beam quality.
  • 13.
    A small radiationmonitoring device worn by persons entering environments that may contain radiation . # Desirable characteristics  Should be lightweight, durable, and reliable  Should be inexpensive
  • 14.
     Healthcare orlaboratory workers in non- emergency environments that may contain radiation  Examples: radiology, nuclear medicine, and radiation oncology department staff  Workers in emergency environments that may contain radiation  Examples: first responders and first receivers  Workers in industrial environments where radiation is used  Examples: nuclear power plant workers or employees at radiation sterilizing facilities
  • 15.
     Flat badgesare usually worn on the torso, at the collar or chest level, but can be worn on the belt, or forearm  Ring shaped badges can be worn on the finger when dose to the finger may exceed dose to the badge worn elsewhere on the body  First responders and first receivers  Wear water-resistant personal dosimeters on the outer layer of personal protective equipment (PPE).  Should be able to easily see and hear a dosimeter alarm while wearing PPE  May wear a personal dosimeter underneath waterproof outerwear
  • 16.
     Film badge Pocket ionization chambers  Thermo luminescent dosimeters (TLD)  Optically stimulated luminescence (OSL)  Solid State
  • 17.
     Most widelyused and most economical  Consists of three parts:  Plastic film holder  Metal filters  Film packet  Can read x, gamma, and beta radiation  Accurate from 10mrem - 500rem  Developed and read by densitometer  A certain density value equals a certain level of radiation  Read with a control badge  Results generally sent as a printout
  • 18.
     Lightweight, durable, portable Cost efficient  Permanent legal record  Can differentiate between scatter and primary beam  Can discriminate between x, gamma, and beta radiation  Can indicate direction from where radiation came from  Control badge can indicate if exposed in transit  Only records exposure where it’s worn  Not effective if not worn  Can be affected by heat and humidity  Sensitivity is decreased above and below 50 keV  Exposure cannot be determined on day of exposure  Accuracy limited to + or - 20%
  • 19.
     The mostsensitive personnel dosimeter  Two types  Self-reading  Non self-reading  Can only be read once  Detects gamma or x-radiation
  • 20.
     Small, compact, easyto use  Reasonably accurate and sensitive  Provides immediate reading  Expensive  Readings can be lost  Must be read each day  No permanent record  Susceptible to false readout if dropped or jarred
  • 21.
     Looks likea film badge  Contains a lithium fluoride crystal  Responds to radiation similarly to skin  Measured by a TLD analyzer  Crystal will luminescence if exposed to radiation, then heated  More accurate than a film badge
  • 22.
     Crystals contained inTLD interact with ionizing radiation as tissue does  Determines dose more accurately  The initial cost is greater than that of a film badge  Can only be read once  Records exposure only where worn
  • 23.
    • “Captures” informationin an Aluminum Oxide matrix • Releases information by laser stimulation • Can be reread after processing • Durable • Landauer Only
  • 24.
     Provides instantaneousinformation regarding dose accumulation  Simple to use  Not a “legal” record  Dose range device dependent