BASIC PRINCIPLES OF UROLOGIC
ULTRASONOGRAPHY
Eko Indra P
PHYSICAL PRINCIPLES
• Interaction of soundwaves
w/ tissues and structures
within the human body 
ultrasound imaging
• The soundwaves are
longitudinal waves  the
particle motion is in the
same direction as the
propagation of the wave
• The transducer = a
receiver for the returning
sound wave reconverting
the mechanical wave to
electrical energy
RESOLUTION
• Refers to the ability to
discriminate two objects in close
proximity to one another.
• Axial resolution: the ability to
identify as separate two objects
in the direction of the traveling
sound wave – directly dependent
on the frequency of sound waves
• Lateral resolution : the ability to
identify separately objects that
are equidistant from the
transducer  The shape of the ultrasound beam is simulated in
this drawing (purple). The focal zone (A) is located
to produce the best lateral resolution of the
medial renal cortex. The location of the focal zone
is designated by the caret (B). The location of the
focal zone can be adjusted by the operator.
MECHANISMS OF ATTENUATION
• Mechanisms of
attenuation include
reflection, scattering,
interference, and
absorption.
• Reflection occurs
when ultrasound
waves strike an
object, a surface, or a
boundary (called an
interface) between
unlike tissues
MECHANISMS OF ATTENUATION
• The amount of energy reflected
from an interface is also influenced
by the impedance of the two tissues
at the interface
• Scattering occurs when sound waves
strike a small or irregular object
• Pattern of interference is partially
responsible for the echo
architecture or texture of organs.
• Absorption occurs when the
mechanical energy of the ultrasound
waves is converted to heat
Relationship between
frequency and tissue
penetration
ARTIFACTS
• A misrepresentation
produced by the
interaction of
ultrasound waves with
tissues
• Example: acoustical
shadow, increased
through-transmission,
edging artifact, and
reverberation artifact
Increased through-
transmission
Edging artifact
MODES OF ULTRASONOGRAPHY
• Gray-scale Ultrasonography: 2D
• Doppler Ultrasonography: allows for the
characterization of motion, most commonly
blood flow, but it may also useful for detecting
the flow of urine
• Harmonic Scanning: produce an image with less
artifact and greater resolution
• Spatial Compounding: reduces the amount of
artifact and noise, producing a scan of better
clarity.
MODES OF ULTRASONOGRAPHY
• Sonoelastography: an evolving ultrasound
modality that adds the ability to evaluate the
elasticity (compressibility and displacement)
of biologic tissues
• Three-Dimensional Scanning: allow the
recognition of some tissue patterns that
would otherwise be unapparent on two-
dimensional scanning
DOCUMENTATION AND IMAGE
STORAGE
• Report: should include specific
info, must be signed by the
physician, and the indications
should be displayed at the top
of the report
• By convention, the liver is
used as a benchmark for
echogenicity
• Images: should include clear
image orientation
• By convention, structures
imaged by USG should be
oriented
PATIENT SAFETY
• Biologic effects: mechanical and thermal effect
• Mechanical effect  cavitation (may cause tissue
damage in certain circumstances)  mechanical
index (MI)
• Thermal effect : result of tissue heating 
thermal index (TI)
• MI and TI estimate the potential for biologic
effects of USG based on the mode, frequency,
power output, and the time of insonation
• These indices are not safety limits
RENAL ULTRASONOGRAPHY
• Transducer used is a curved array of 3.5 – 5.0
MHz. Pediatric may use a higher frequency
• Patient in supine position
• Visualization of the left kidney often requires
the patient to be turned into a lateral
position.
• Applications: percutaneous renal biopsy
• Limitations: Renal ultrasonography has poor
sensitivity for renal masses less than 2 cm
TECHNIQUES: RIGHT KIDNEY
• Scanning of the right kidney is performed with the
patient supine.
• The kidney is located by beginning in the midclavicular
line in the right upper quadrant.
• The transducer is moved laterally until the midsagittal
plane of the kidney is imaged.
From: Campbell-Walsh Urology 10th Edition
TECHNIQUES: RIGHT KIDNEY
• The probe is rotated 90 degrees counterclockwise.
• The midtransverse plane will demonstrate the renal
hilum containing the renal vein.
• The kidney is scanned from upper pole to lower pole.
From: Campbell-Walsh Urology 10th Edition
TECHNIQUES: LEFT KIDNEY
• For left renal ultrasonography is identical to that of
the right side, but the left kidney is slightly more
cephalad than the right kidney.
• Bowel gas is more problematic on the left because
of the position of the splenic flexure of the colon.
• Visualization of the left kidney often requires the
patient to be turned into a lateral position.
From: Campbell-Walsh Urology 10th Edition
NORMAL FINDINGS
• It is helpful to understand its anatomic
position within the retroperitoneum.
• This assists identifying the midsagittal
plane, which serves as a reference point
for a complete examination
From: Campbell-Walsh Urology 10th Edition
KIDNEY ULTRASONOGRAPHIC
ANATOMY
Liver
Kidney
The lower pole of the kidney is displaced 15 degrees laterally compared with the upper pole (A).
The kidney is rotated 30 degrees posterior to the true coronal plane (B). The lower pole of the
kidney is slightly anterior compared with the upper pole.
From: Campbell-Walsh Urology 10th Edition
NORMAL FINDINGS
• The adult right kidney is usually hypoechoic with
respect to the liver.
• The central band of echoes in the kidney  a
hyperechoic area that contains the renal hilar
adipose tissue, blood vessels, and collecting
system.
• Acoustic shadowing from ribs  move the probe to
a more lateral position or into the intercostal space.
• Taking a deep breath  the kidney can be moved
inferiorly to assist complete imaging
From: Campbell-Walsh Urology 10th Edition
Midsagittal plane of the kidney. Note the relative hypoechogenicity of the renal pyramids (P)
compared with the cortex (C). The central band of echoes (B) is hyperechoic compared with the
cortex. The midsagittal plane will have the greatest length measurement pole to pole. A perfectly
sagittal plane will result in a horizontal long axis of the kidney.
From: Campbell-Walsh Urology 10th Edition
NORMAL FINDINGS:
ECHOGENICITY
• The echogenicity of the kidney varies with age.
• Infant renal cortex  more hyperechoic, smaller
and less apparent central band of echoes
• Adult renal cortex  more hypoechoic with
respect to the liver (Emamian et al, 1993a).
• Chronic medical renal diseases  renal cortex
is often thinned and isoechoic or hyperechoic
with respect to the liver (O’Neill, 2001)
From: Campbell-Walsh Urology 10th Edition
NORMAL FINDINGS: RENAL
SIZE
• Renal size changes over the lifetime of an individual.
• These are based on age, height, and weight of the
patient.
• Adult kidney  10 to 12 cm in length, 4 to 5 cm in
width.
• Measurements of renal volume may be appropriate
in cases of severe renal impairment.
• Renal measurements should be obtained in the
midsagittal plane and midtransverse plane.
From: Campbell-Walsh Urology 10th Edition
RENAL SIZE (CONT’D)
• The thickness of the parenchyma  the average
distance between the renal capsule and the
central band of echoes.
• Subjective  the midlateral renal parenchyma in
the sagittal view is a common choice.
• No universal standard
– The renal cortical thickness  greater than 7 mm (Roger et al,
1994),
– The renal parenchymal thickness  greater than 15 mm
(Emamian et al, 1993b).
From: Campbell-Walsh Urology 10th Edition
The distinction between renal cortical thickness and renal parenchymal thickness is that the renal
parenchyma is measured from the central band of echoes to the renal capsule. The renal cortex is
measured from the outer margin of the medullary pyramid to the renal capsule.
From: Campbell-Walsh Urology 10th Edition
TRANSABDOMINAL PELVIC
ULTRASONOGRAPHY
• Transducer used is a curved array of 3.5 – 5.0
MHz.
• Commonly performed with the patient supine
and the sonographer on the patient’s right side
• Bladder may also be assessed using transvaginal
and transrectal approaches.
• Applications: Transabdominal US-guided
percutaneous bladder aspiration
• Limitation: Yields limited information in patients
with an empty bladder
ULTRASONOGRAPHY OF THE
SCROTUM
• High-frequency transducers,
generally in the range of 7 to 18
MHz, may be employed.
• Patient should be supine with the
scrotum supported on a towel or on
the anterior thighs.
• Applications: percutaneous
testicular sperm aspiration,
percutaneous epididymal sperm
aspiration
• Limitations: Caution should be used
when interpreting Doppler flow
studies in the evaluation of
suspected testicular torsion
ULTRASONOGRAPHY OF THE PENIS
AND MALE URETHRA
• Best performed with a 12- to 18-MHz linear array transducer
• The more proximal aspects of the urethra and corpora cavernosa
are best assessed through a perineal approach
• Transperineal and translabial ultrasonography has also been
used for evaluation of the pelvic floor for both diagnostic
purposes and postprocedural follow-up
• Applications: evaluation of erectile dysfunction and penile
curvature
• Limitations: The complete evaluation requires a dorsal or ventral
interrogation of the exposed phallus and a perineal approach
and the evaluation of erectile dysfunction requires qualitative
and quantitative measurements of blood flow in the penile
arteries
TRANSRECTAL ULTRASONOGRAPHY
OF THE PROSTATE
• A high-frequency 7.5 - 10-MHz
transducer is usually used. This
can be a biplanar or singleplane
transducer
• It is essential to perform a digital
rectal examination before
inserting the ultrasound probe
• Applications: TRUS-guided biopy
• Limitations: Bowel preparation
is sometimes necessary for
imaging.
THANK YOU

Urologic Ultrasonography

  • 1.
    BASIC PRINCIPLES OFUROLOGIC ULTRASONOGRAPHY Eko Indra P
  • 2.
    PHYSICAL PRINCIPLES • Interactionof soundwaves w/ tissues and structures within the human body  ultrasound imaging • The soundwaves are longitudinal waves  the particle motion is in the same direction as the propagation of the wave • The transducer = a receiver for the returning sound wave reconverting the mechanical wave to electrical energy
  • 3.
    RESOLUTION • Refers tothe ability to discriminate two objects in close proximity to one another. • Axial resolution: the ability to identify as separate two objects in the direction of the traveling sound wave – directly dependent on the frequency of sound waves • Lateral resolution : the ability to identify separately objects that are equidistant from the transducer  The shape of the ultrasound beam is simulated in this drawing (purple). The focal zone (A) is located to produce the best lateral resolution of the medial renal cortex. The location of the focal zone is designated by the caret (B). The location of the focal zone can be adjusted by the operator.
  • 4.
    MECHANISMS OF ATTENUATION •Mechanisms of attenuation include reflection, scattering, interference, and absorption. • Reflection occurs when ultrasound waves strike an object, a surface, or a boundary (called an interface) between unlike tissues
  • 5.
    MECHANISMS OF ATTENUATION •The amount of energy reflected from an interface is also influenced by the impedance of the two tissues at the interface • Scattering occurs when sound waves strike a small or irregular object • Pattern of interference is partially responsible for the echo architecture or texture of organs. • Absorption occurs when the mechanical energy of the ultrasound waves is converted to heat Relationship between frequency and tissue penetration
  • 6.
    ARTIFACTS • A misrepresentation producedby the interaction of ultrasound waves with tissues • Example: acoustical shadow, increased through-transmission, edging artifact, and reverberation artifact Increased through- transmission Edging artifact
  • 7.
    MODES OF ULTRASONOGRAPHY •Gray-scale Ultrasonography: 2D • Doppler Ultrasonography: allows for the characterization of motion, most commonly blood flow, but it may also useful for detecting the flow of urine • Harmonic Scanning: produce an image with less artifact and greater resolution • Spatial Compounding: reduces the amount of artifact and noise, producing a scan of better clarity.
  • 8.
    MODES OF ULTRASONOGRAPHY •Sonoelastography: an evolving ultrasound modality that adds the ability to evaluate the elasticity (compressibility and displacement) of biologic tissues • Three-Dimensional Scanning: allow the recognition of some tissue patterns that would otherwise be unapparent on two- dimensional scanning
  • 9.
    DOCUMENTATION AND IMAGE STORAGE •Report: should include specific info, must be signed by the physician, and the indications should be displayed at the top of the report • By convention, the liver is used as a benchmark for echogenicity • Images: should include clear image orientation • By convention, structures imaged by USG should be oriented
  • 10.
    PATIENT SAFETY • Biologiceffects: mechanical and thermal effect • Mechanical effect  cavitation (may cause tissue damage in certain circumstances)  mechanical index (MI) • Thermal effect : result of tissue heating  thermal index (TI) • MI and TI estimate the potential for biologic effects of USG based on the mode, frequency, power output, and the time of insonation • These indices are not safety limits
  • 11.
    RENAL ULTRASONOGRAPHY • Transducerused is a curved array of 3.5 – 5.0 MHz. Pediatric may use a higher frequency • Patient in supine position • Visualization of the left kidney often requires the patient to be turned into a lateral position. • Applications: percutaneous renal biopsy • Limitations: Renal ultrasonography has poor sensitivity for renal masses less than 2 cm
  • 12.
    TECHNIQUES: RIGHT KIDNEY •Scanning of the right kidney is performed with the patient supine. • The kidney is located by beginning in the midclavicular line in the right upper quadrant. • The transducer is moved laterally until the midsagittal plane of the kidney is imaged. From: Campbell-Walsh Urology 10th Edition
  • 13.
    TECHNIQUES: RIGHT KIDNEY •The probe is rotated 90 degrees counterclockwise. • The midtransverse plane will demonstrate the renal hilum containing the renal vein. • The kidney is scanned from upper pole to lower pole. From: Campbell-Walsh Urology 10th Edition
  • 14.
    TECHNIQUES: LEFT KIDNEY •For left renal ultrasonography is identical to that of the right side, but the left kidney is slightly more cephalad than the right kidney. • Bowel gas is more problematic on the left because of the position of the splenic flexure of the colon. • Visualization of the left kidney often requires the patient to be turned into a lateral position. From: Campbell-Walsh Urology 10th Edition
  • 15.
    NORMAL FINDINGS • Itis helpful to understand its anatomic position within the retroperitoneum. • This assists identifying the midsagittal plane, which serves as a reference point for a complete examination From: Campbell-Walsh Urology 10th Edition
  • 16.
  • 17.
    The lower poleof the kidney is displaced 15 degrees laterally compared with the upper pole (A). The kidney is rotated 30 degrees posterior to the true coronal plane (B). The lower pole of the kidney is slightly anterior compared with the upper pole. From: Campbell-Walsh Urology 10th Edition
  • 18.
    NORMAL FINDINGS • Theadult right kidney is usually hypoechoic with respect to the liver. • The central band of echoes in the kidney  a hyperechoic area that contains the renal hilar adipose tissue, blood vessels, and collecting system. • Acoustic shadowing from ribs  move the probe to a more lateral position or into the intercostal space. • Taking a deep breath  the kidney can be moved inferiorly to assist complete imaging From: Campbell-Walsh Urology 10th Edition
  • 19.
    Midsagittal plane ofthe kidney. Note the relative hypoechogenicity of the renal pyramids (P) compared with the cortex (C). The central band of echoes (B) is hyperechoic compared with the cortex. The midsagittal plane will have the greatest length measurement pole to pole. A perfectly sagittal plane will result in a horizontal long axis of the kidney. From: Campbell-Walsh Urology 10th Edition
  • 20.
    NORMAL FINDINGS: ECHOGENICITY • Theechogenicity of the kidney varies with age. • Infant renal cortex  more hyperechoic, smaller and less apparent central band of echoes • Adult renal cortex  more hypoechoic with respect to the liver (Emamian et al, 1993a). • Chronic medical renal diseases  renal cortex is often thinned and isoechoic or hyperechoic with respect to the liver (O’Neill, 2001) From: Campbell-Walsh Urology 10th Edition
  • 21.
    NORMAL FINDINGS: RENAL SIZE •Renal size changes over the lifetime of an individual. • These are based on age, height, and weight of the patient. • Adult kidney  10 to 12 cm in length, 4 to 5 cm in width. • Measurements of renal volume may be appropriate in cases of severe renal impairment. • Renal measurements should be obtained in the midsagittal plane and midtransverse plane. From: Campbell-Walsh Urology 10th Edition
  • 22.
    RENAL SIZE (CONT’D) •The thickness of the parenchyma  the average distance between the renal capsule and the central band of echoes. • Subjective  the midlateral renal parenchyma in the sagittal view is a common choice. • No universal standard – The renal cortical thickness  greater than 7 mm (Roger et al, 1994), – The renal parenchymal thickness  greater than 15 mm (Emamian et al, 1993b). From: Campbell-Walsh Urology 10th Edition
  • 23.
    The distinction betweenrenal cortical thickness and renal parenchymal thickness is that the renal parenchyma is measured from the central band of echoes to the renal capsule. The renal cortex is measured from the outer margin of the medullary pyramid to the renal capsule. From: Campbell-Walsh Urology 10th Edition
  • 24.
    TRANSABDOMINAL PELVIC ULTRASONOGRAPHY • Transducerused is a curved array of 3.5 – 5.0 MHz. • Commonly performed with the patient supine and the sonographer on the patient’s right side • Bladder may also be assessed using transvaginal and transrectal approaches. • Applications: Transabdominal US-guided percutaneous bladder aspiration • Limitation: Yields limited information in patients with an empty bladder
  • 25.
    ULTRASONOGRAPHY OF THE SCROTUM •High-frequency transducers, generally in the range of 7 to 18 MHz, may be employed. • Patient should be supine with the scrotum supported on a towel or on the anterior thighs. • Applications: percutaneous testicular sperm aspiration, percutaneous epididymal sperm aspiration • Limitations: Caution should be used when interpreting Doppler flow studies in the evaluation of suspected testicular torsion
  • 26.
    ULTRASONOGRAPHY OF THEPENIS AND MALE URETHRA • Best performed with a 12- to 18-MHz linear array transducer • The more proximal aspects of the urethra and corpora cavernosa are best assessed through a perineal approach • Transperineal and translabial ultrasonography has also been used for evaluation of the pelvic floor for both diagnostic purposes and postprocedural follow-up • Applications: evaluation of erectile dysfunction and penile curvature • Limitations: The complete evaluation requires a dorsal or ventral interrogation of the exposed phallus and a perineal approach and the evaluation of erectile dysfunction requires qualitative and quantitative measurements of blood flow in the penile arteries
  • 27.
    TRANSRECTAL ULTRASONOGRAPHY OF THEPROSTATE • A high-frequency 7.5 - 10-MHz transducer is usually used. This can be a biplanar or singleplane transducer • It is essential to perform a digital rectal examination before inserting the ultrasound probe • Applications: TRUS-guided biopy • Limitations: Bowel preparation is sometimes necessary for imaging.
  • 28.